Digital Thermometer
Model: DTR-1221BLU, DTR-1221BG,
DTR-1221P and DTR-1221BGW, DTR-1221A
This thermometer comes in four color variations
Congratulations on your purchase of this product.
Please read the instructions carefully before using
the thermometer for the rst time and keep these
instructions in a safe place. This product is intended for
measurement of human body temperature.
Operating Instructions
1. Please disinfect the probe before use.
2. Press the button to activate the thermometer.
• A short beep will sound and the thermometer runs a
check showing all the digital segments on the LCD.
• After the check, the last measurement result is
shown for 2 seconds.
• As soon as - -.- - appears in the LCD the thermome-
ter is ready.
3. Put the thermometer sensor in the desired
measurement position (oral or rectal).
4. Wait for the beep to sound and the ºC or ºF icon to
stop ashing. The measurement is now completed.
5. Read the result from the display.
This is a maximum thermometer and the displayed
temperature can slightly increase if you leave the
thermometer in the measurement position after the
beeps. Please follow the instructions described in the
“Methods of measuring temperature” section.
Fever alarm
If the detected body temperature exceeds
100.0ºF/37.8ºC the thermometer will sound an alarm:
“bi-bi-bi---bi-bi-bi---.”The alarm indicates a fever may
be approaching.
Fever indication
Over 100.0ºF/37.8ºC; elevated temperature —
fever may be approaching!
99.1-99.9ºF/37.3-37.7ºC; slightly elevated
temperature
<99.0ºF/<37.2ºC; normal temperature
Auto-o function
When the measurement is completed, switch the
thermometer o by pressing the ON/OFF button.
If not switched o, the thermometer will switch o
automatically after 10 minutes.
Memory function
When the thermometer is switched on, the last
measurement will be shown for 2 seconds. Press the
button again to turn the thermometer o. When the
thermometer will remain turned on, the result will be
overwritten by the new measurement.
Methods of measuring temperature
It is important to remember that the accuracy of
the body temperature reading depends on the
measurement position. When keeping track of the
results, the measurement position must always be
specied in order to ensure that a correct temperature
reading is recorded.
Rectal Measurement
From a medical point of view, this is the most accurate
method because the temperature at this measurement
position is closest to the core body temperature. The
thermometer tip is inserted carefully into the rectum
for a maximum of 2 cm (1”). The usual measuring time
is approximately 8 to 15 seconds.
Oral Measurement
There are dierent heat zones in the mouth. As a
general rule, the oral temperature is 0.3-0.8˚C
(0.5-1.4ºF) lower than the rectal temperature. To
ensure that the reading is as accurate as possible,
place the thermometer tip to the left or right of the
root of the tongue. The thermometer tip should
remain in constant contact with the tissue under the
tongue during the reading. The mouth should remain
closed during the reading. The usual measuring time is
approximately 8 to 15 seconds.
Important! As eating, drinking and breathing through
the mouth inuence the temperature of the mouth
tissue, we highly recommend not to eat, drink or
breathe through the mouth within 15 minutes before
taking an oral measurement.
Axillary Measurement is NOT recommended
It is NOT recommended to use this device for taking
axillary (under arm) temperature measurements.
This quick reading thermometer will not accurately
measure the axillary temperature.
Switching the temperature scale (˚C or ˚F)
First make sure the thermometer is switched o. If it is
still switched on, press the button once to switch o.
When the thermometer is o, press and hold the
button for 3 seconds to switch the measuring scale.
The new scale (˚C or ˚F) is now and saved.
Cleaning and disinfection
The best way to clean the thermometer tip is by
applying a disinfectant (e.g. 70% ethyl alcohol) with
a damp cloth. This waterproof thermometer can be
immersed in lukewarm water for thorough cleaning
and disinfection.
Safety precautions
• Do not expose the thermometer to hot water over
50ºC (122ºF).
• Do not expose the thermometer to high
temperatures or direct sunlight.
Thermometer tip
Thermometric
sensor LCD
ON/OFF
button
Battery
compartment
Fever indicator
• Do not drop the thermometer. It is neither shock-
proof nor impact-resistant.
• Do not open the device except the battery
compartment.
• Do not bend the device except than the exible tip.
Avoid bending the tip by more than 45 degrees.
• Do not clean with thinners, petrol or benzene. Only
clean the thermometer with water or disinfectant.
• Do not immerse the thermometer under water
deeper than 15cm(6”) for longer than 5 minutes.
• The thermometer contains small parts (including
a battery) which can be swallowed by children.
For this reason, do not leave the thermometer
unattended in the hands of children.
• If the ambient temperature is over 35ºC (95ºF) dip
the thermometer tip in cold water for 10 seconds
prior to taking a measurement.
• Persistent fever, in particular in children, has to
be treated by a doctor. If this occurs please get in
touch with your doctor!
Battery replacement
The battery is empty and needs replacing when the
battery symbol appears on the right of the LCD.
To replace the battery: Lay the thermometor with
the button downwards on a at surface. Remove the
screw. Pull o the battery cover. Gently slide out the
circuit board with the battery chamber, up to 1 cm
(0,5 “). Replace the battery by one of the same type
(preferably non-mercury) with the “+”polarity facing
up. After replacing the battery, reposition the circuit
board, the battery cover and the screw.
To protect the environment, please do not dispose of
batteries as normal waste. Instead, please hand them
in at a collection point or municipal recyclable material
centre as special waste.
Technical data
Type: maximum thermometer
Measurement range: 32.0~43.0ºC (89.6~109.4ºF)
Measurement accuracy :
±0.1ºC at 35.5~42.0ºC at ambient temperature 18~28ºC
±0.2ºC at 32.0~35.5ºC at other ambient temperatures.
(±0.2ºF at 95.9~107.6ºF at ambient temp. 64.4~82.4ºF
±0.4ºF at 89.6~95.9ºF at other ambient temperatures.)
Storage/transport conditions: -20~55ºC, 15%~95% RH
Ambient operating conditions: 5~35 ºC, 15%~95%RH
Min Scale: 0.1ºC /0.1ºF
Battery type: Alkaline manganese battery, type LR41, 1.5V.
Weight: Approximately 12g
This thermometer is latex free.
Explanation of LCD symbols
The battery is empty
LoºC/ LoºF The temperature is under 32 ºC or 89.6 ºF
HiºC/ HiºF The temperature is over 43 ºC or 109.4 ºF
Explanation of symbols on the label
The battery in this product complies with
the requirements stated in European
Directives 2006/66/EEC.
Type BF equipment.
Consumed electronic products must not
be mixed with conventional domestic
waste. Take this product to an appropriate
waste collection point for correct handling
and recycling.
Calibration check
This thermometer is initially calibrated at the time of
manufacture. If this thermometer is used according to
the operation instruction, periodic re-adjustment is
not required. However, if there are indications that the
product does not maintain the dened error limits or
if the calibration could have been aected by an ex-
ternal factor or by any other means a calibration check
should be carried out immediately. For calibration
please observe any national statutory or administra-
tive regulations. The calibration check can be carried
out by any competent authority or by authorised
service providers. Test instructions for a calibration
check can be provided to the relevant authorities and
authorized service providers on request.
Warranty
This Limited Warranty covers any defects in material or
workmanship under normal use during the Warranty
Period of 365 days after purchase. iProvèn will either
replace the product or repair the Product at no charge,
using new or refurbished replacement parts. A replace-
ment product or product part assumes the remaining
warranty of the original product purchase. This Limited
Warranty does not cover batteries and packaging. It
also does not include any problem that is caused by:
conditions, malfunctions or damage not resulting from
defects in material or workmanship.
Customer support
For obtaining warranty service and for any other
inquiries or suggestions please contact iProvèn
customer support at www.iproven.com
Manufactured for: iProvèn USA LLC
9450 SW Gemini Dr #48879
Beaverton - OR 97008-7105
Call us toll free if you have any questions or concerns.
+1 855-700-1668
Manual DTR-1221A, DTR-1221BLU, DTR-1221BG,
DTR-1221P and DTR-1221BGW
Manual version: 1.3

This manual suits for next models

4

Other Proven Thermometer manuals

Proven DT-R1221 A User manual

Proven

Proven DT-R1221 A User manual

Proven DMT-489 User manual

Proven

Proven DMT-489 User manual

Proven ET-828-BT User manual

Proven

Proven ET-828-BT User manual

Proven DMT-489 User manual

Proven

Proven DMT-489 User manual

Proven DMT-489 User manual

Proven

Proven DMT-489 User manual

Popular Thermometer manuals by other brands

innovu health INH-GB-01 user manual

innovu health

innovu health INH-GB-01 user manual

AEMC 1822 user manual

AEMC

AEMC 1822 user manual

THERM PRO TP-16 quick start guide

THERM PRO

THERM PRO TP-16 quick start guide

Fluke 574 user manual

Fluke

Fluke 574 user manual

ThermoWorks Pro-Surface Thermapen operating instructions

ThermoWorks

ThermoWorks Pro-Surface Thermapen operating instructions

Omega Engineering HH501BJ manual

Omega Engineering

Omega Engineering HH501BJ manual

ThermoWorks Precision operating instructions

ThermoWorks

ThermoWorks Precision operating instructions

Radio Shack 63-1034 owner's manual

Radio Shack

Radio Shack 63-1034 owner's manual

TMS Mini First-time setup

TMS

TMS Mini First-time setup

Craftmade Accolade TH003 Assembling instructions

Craftmade

Craftmade Accolade TH003 Assembling instructions

Fluke Series II user manual

Fluke

Fluke Series II user manual

TESTO TEO104 instruction manual

TESTO

TESTO TEO104 instruction manual

Hanna Instruments HI 93551R instruction manual

Hanna Instruments

Hanna Instruments HI 93551R instruction manual

SP tools SP62022 user manual

SP tools

SP tools SP62022 user manual

Ametek LAND IQ Series user guide

Ametek

Ametek LAND IQ Series user guide

Lutron Electronics TM-946 manual

Lutron Electronics

Lutron Electronics TM-946 manual

Braun WelchAllyn ThermoScan PRO 4000 user manual

Braun

Braun WelchAllyn ThermoScan PRO 4000 user manual

TFA 10.4001 instruction manual

TFA

TFA 10.4001 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.