Proven DT-R1221 A User manual

Digital Thermometer
Model:
DT-R1221
A
and
DT-R1221
B
Thermometric
ON/OFF
sen
s
or
LCD
button
Thermome
t
er
tip
Fever
indicator
Note
:The
exterior
of
each
model
has a little difference.
Congratulations
on
your purchase of this product.
Please read the instructions carefully before using
the
thermometer
for the first time and keep these
instructions
in
a
safe
place.
This
product
is
i11tended
for
measurement
of
human body temperature.
Operating
Instructions
Before using, please disinfect the probe.
To
switch on,
press the ON/OFF button next
to
the display. A short
beep will sound indicating that the thermometer runs
a self-check test during which all the digital segments
appear on the
LCD.
This will be folowed by the last
measured temperature. After this, the letters "Lo" and
a flashing "0
C"
or
"°F" appear and the thermometer
is
ready for
use.
If
the ambient temperature
is
below
32°C
or
89.6°F , then "
Lo
°C" or "
Lo
°F"
will appear on
the
LCD.
If
the
ambient temperature
is
over 42.
9°C
or
109.2°F; "
Hi
0
("
or
"Hi
°F
"will appear on the
LCD
.
During the temperature reading, the temperature
is
displayed continuously and the "0
C"
or
"°
F"
symbol
flashes.The measurement
is
completed when a
constant temperature value
has
been reached. The
temperature value
is
considered constant when the
temperature
rises
less
than
0.1
°C
within
3 seconds.
When
the
constant temperature value
is
reached, a
beep will sound
10
times
"bi
-bi-bi-bi-
..
:;
and the "°C" or
"°F" symbol will stop flashing. The highest measured
temperature appears on the
LCD
.
Fever alarm
If
the detected
body
temperature exceeds 37.5°C/
99.5°F the
thermometer
will sound a fever alarm: "bi-
bi-bi---bi-bi-bi---
..
:
'.
Note: this a maximum thermometer and the displayed
temperature can increase slightly
if
the measurement
continues after
the
beeps. This
is
particularly the
case
with
axillary measurements which require a signifi-
cantly longer measurem~nt
time
for accurate measu-
rements.
Please
follow the instructions described in
the
"Methods
of
measuring temperature" section.
Fever indication
@ Over 37.9°C/ 100.2°F; fever!
A 37.5°C-37.9°C/ 99.5°F-100.2°F; sligh
tly
higher
\::;I temperature.
@ < 37.4°C/
99.5°F;
good temperature.
Auto
-
off
function
When the measurement
is
comp
leted, switch the
t
hermomet
er
off
by
pr
ess
ing t he
ON
/
OFF
button.
If
not
sw
it
ched off, the
thermometer
wi
ll
sh
ut
off
au
tom
ati-
ca
lly
aft
er 10 m inutes.
Memory function
When the thermometer
is
turned on, the last measu-
rement will show on
scre
en for 2 seconds.
Pr
e
ss
the
button again
to
turn the thermometer off. When the
thermometer will remain turned on, the result will be
overwritten by the new measurement.
Methods
of
measuring temperature
It
is
important
to
remember that the body tempe-
rature read ing depends on the position where
it
is
measured.
For
this reason, the measurement position
must always be specified in order
to
ensure
that
a
correct temperature reading
is
recorded.
Rectal/ Vaginal Measurement
This
is
the most accurate method from a medical poi
nt
of
view because it
is
closest
to
the core
body
tempera-
ture. The thermometer
tip
is
inserted carefully into the
rectum for a maximum
of
2 cm. The usual measuring
time
is
approximately 8
to
15
seconds.
Axillary Measurement
Pla
cing the thermometer in the armpit provides
a measurement
of
surface temperature
that
can
fluctuate by around 0.5°C/0.9°F
to
1.5°C/2.5°F from
rectal/
vaginal temperature readings in adults. The
usual measuring time for this method
is
approxi-
mately
20
to
30
seconds.
It
should be noted
that
an
exact reading won't be obtained
if,
for example, the
armpit temperature
is
lower than normal (the
armpit
temperature
has
been cooled down).
If
this
is
the
case,
extend the measuring time
to
5 minutes in order
to
obtain a reading
that
corresponds more closely
to
the
core body temperature.
Oral Measurement
There are different heat zones in the mouth.
As
a
general rule, the oral temperature
is
0.3°C/0.5°F
to
0.8°C/1.4°F lower than the
rectal/
vaginal
tempe-
rature.
To
ensure
that
the reading
is
as
accurate
as
possible, place the thermometer
tip
to
the left
or
right
of
the root
of
the tongue. The
thermometer
tip
should
remain in constant contact
with
the tissue under
the
tonque
during the reading.The
mouth
should remain
closed during the reading. Eating
or
drinking or bre-
athing
through
the
mouth
just
before measuring will
result in an inaccurate temperature reading.The usual
measuring
time
is
approximately 8
to
15
seconds.
°C
and
°F
switch
Turn the thermometer
off
.
Press
and hold the
On/Off
Button for 3 seconds to switch
the
measuring scale.
Cleaning and disinfection
The best way
to
clean the
thermometer
tip
is
by
applying a disinfectant
(e.g.
70% ethyl alcohol)
with
a
damp
cloth.
Th
is waterpr
oof
thermometer
can be
immersed in lukewarm water
for
thorough
cleaning
and disinfection.
Safety precautions
Do
not
expose the
thermometer
to
hot
water
over 50°C/122°F.
Do
not
expose the
thermometer
to high tempe
ratur
es
or
direct sunlight.

Do
not drop the thermometer. It
is
neither
shock-proof nor impact-resistant.
Do
not
bend or open the device (except the
battery compartment).
Do
not
clean with thinners, petrol or benzene.
Only clean the thermometer with water or
disinfectant.
Do
not
immerse the thermometer under water
deeper than 15cm for longer than 5 minutes.
The thermometer contains small parts (inclu-
ding a battery) which can be swallowed by
children.
For
this reason,
do
not
leave the ther-
mometer unattended in the hands
of
children.
Avoid bending the flexible tip by more than
45
degrees.
If
the ambient temperature
is
over 35°C/95°F,
dip
the thermometer
tip
in cold water for
10
seconds prior
to
measuring the temperature.
Persistent fever, in particular in children,
has
to
be treated by a doctor. Pl
ease
get in touch
with
your doctor!
Battery replacement
The battery
is
empty
and needs replacing when the
battery symbol D appears on the right
of
the
LCD
.
Re-
move the philips screw from the back
of
the thermo-
meter. Pull
off
the battery cover. Gently slide
out
the
ci
rcu it board
with
the battery chamber approximately
up
to
1 cm and replace the battery by one
of
the
same
type (preferably non-mercury)
with
the"+"
polarity
fa
cing up. After replacing the battery, reposition the
circuit board, the battery cover and the philips screw.
To
protect the environment, please
do
not
dispose
of
batteries
as
normal waste. Instead, please hand them
in at a collection
point
or
municipal recyclable material
centre
as
special waste.
Technical data
Type: maximum thermometer
Measurement range:
(32.00oC~43
.00°C)l(89.60
°F
~109.40°F)
Meas
urement accuracy:
+/-0. 1
°C/0.2°F
(35.5°C
~42.0°C/
95
.
9°F~
107
.
6°F)
at
ambi-
ent temperature
18°C
~28°C(64.4°F~82.4°F).
+I-0.2°CI0.4°F
(32.0°C
~35.5°CIB9.6°F~95.9°F)
at
other
ambient temperatures.
Storage/transportation conditions:
(-20~55°C),
15%~95%RH
Ambient operating conditions: (5~35
°C),
15%~95%RH
Min
Scale:
0.
7°C
/0.
7°F
Battery type: Alkaline manganese battery,
type
LR41,
1.SV.
Weight: Approx.
12g
Ex
planation
of
symbols
Battery
is
empty
The
battery in this product complies with
the requirements stated in European
Directives 2006/66/EEC.
Lo°C
or
Lo°F
Temperature
is
under
32
°C
or
89
.6
°F
Hi°C
or
Hi°F
Temperature
is
over
43
°C
or 109.4
°F
Type
BF
equipment
Consumed electronic products must
::f(
not
be mixed
with
conventional
-domestic waste.
Tak
e thi s product to
the appropriate wasted collection
point
s for corr
ec
t
handling and recycling.
Legal requirements and guidelines
This product complies
with
the European Directive for
Medical Devices 93/42/EEC and carries the
CE
mark.
The device also complies
with
the specifications
of
the
European Standard for Clinical Thermometers
Part
3:
Performance
of
compact electrical thermometers
(non-predictive and predictive)
with
maximum device.
The
CE
marking confirms
that
this
is
a medical device
with
a measuring function
as
defined by the Medical
Devices Act, which
has
undergone a conformity
assessment procedure. Notified Body 0197 confirms
that this product fulfils all the appropriate statutory
regulations.
This thermometer conforms
to
all
of
the
requirements
established in
ASTM
Standard E 1112. Full responsibili-
ty
for conformance
of
this
product
to
the
specification
is
assumed by iProven, 9450
SW
Gemini Dr #48879,
Beaverton
OR,
97008-71
OS
.
Calibration check
This thermometer
is
initially calibrated at the time
of
manufacture.
If
this thermometer
is
used according to
the operation instruction, periodic re-adjustment
is
not
required. The calibration check should be carried
out
immediately,
if
there
are
indications
that
the
product does
not
maintain the defined error limits
or
if
the calibration could have been affected by
an
external factor or by any
other
means. For callibration
please observe any national statutory or administra-
tive regulations. The calibration check can be carried
out
by any competent authority
or
by
authorised
service providers. Test instructions for a calibration
check
can
be provided
to
the relevant authorities and
authorized service providers on request.
Warranty
For
1 year extended warranty please register your
product at warranty.iproven.com
l~
Proven
l
Manufactured for: iProven
USA
LL
C
Address: 9450
SW
Gemini Dr #48879
Beaverton -
OR
97008-71
OS
Vers
ion: 20151222 DT-Rl
221
A and B
This manual suits for next models
1
Other Proven Thermometer manuals