MAN 13.463 PT02 Pocket Scale Manual
1. Features:
•Auto shut off
•Sim le calibration
•Auto zero tracking
•Low battery indication
•Backlit dis lay
2. Key Functions:
[ON/OFF] Key Turns the scale on and off
[UNIT] Key Switches the scale between the available units
[TARE] Key Zeros the scale and o erates the Tare function
3. Pre aration for weighing:
•Please make sure the ower is turned on (remove the lastic under
the battery first)
•The scale will turn off automatically after 2 minutes if it is not used.
4. Weighing:
•Without a container
Press [TARE] to zero the scale and dis lay “0” & “ ”. Place the
article to be weighed on the an, wait for the reading to become stable
and take the reading.
•With a container
Place a container on the an, ress [TARE] to dis lay “0” when the
reading is stable. Then lace the article to be weighed in the
container, wait for the reading to become stable and take the reading.
5. Calibration.
•With the scale off ress both the [ON/OFF] and [UNIT] key for a few
seconds, the dis lay will show the A/D value.
•Press the [ON/OFF] key again. The dis lay will show the desired
calibration weight, and then enter into the weight calibration mode.
•Put the standard weight as dis layed (use a 200g weight on the an if
the value dis layed is 200g) on the latform calibration, and ress the
[ON/OFF]key. After 2 seconds the dis lay will Show "PASS” If the
calibration is successful and then return to weighing mode.
•If the calibration is unsuccessful (this can ha en when the weight on
the latform is unstable), the dis lay will show“O_ld”. Re-try the
calibration ensuring the weights are stable and there is no vibration or
air flow to disturb the readings.
PT Limited www. t-global.com
Product Manual