Safety/ Operating Instructions - Technical Data RLWE 4161
Start-up
•Power supply
•Power on → in ication
After connecting mains cable into RLWE 4161 the evice is switche on by power
switch (14) on rear si e. Refer also Figure 3. During operation of RLWE 4161 the
yellow in icator (13) is glowing.
Usage of signal channels CH1 to CH4
•There is a usage of all optical an electrical pulse signal channels at the same time in
parallel possible.
•On front panel there are channel CH1 an channel CH3 each with three non-inverting
(5) an one inverting (10, marke by “INV”) TTL signal outputs available.
•At channel CH2 an channel CH4 there are three non-inverting (8) an one inverting
(12, marke by „INV“) available to use in parallel.
•Optical signal inputs eposite on rear si e of RLWE 4161 can be use in parallel.
They can be connecte pairwise to TTL an NIM outputs resp. By means of toggle
switch (11) in position “||” it is possible to map optical inputs 1 an 2 to electrical
outputs CH1 an CH2. In position “X” of switch (11) the optical signal 1 is mappe to
output channel CH2 an the optical signal 2 is mappe to output channel 1 at the
same time (crosse -over mo e). The i entical scope of functions offer optical
channels 3 an 4 in conjunction with electrical output channels CH3 an CH4. This
way optical signal pairs 1, 2 an 3, 4 resp. can be converte optionally into TTL or NIM
signals. Activate mapping mo es are in icate by re LEDs above (6) an (7) resp.
on front panel.
•As soon as a signal is available on a channel output the yellow signal LEDs (9) an
(12) resp. are glowing. In case there is no signal available these LEDs remain
eactivate .
Usage options at signal inputs of a channel
•Via three BNC receptacles (5) on channels CH1 an CH3 resp. there are 50 Ω outputs
for TTL signals available. The same TTL signal in inverte form is provi e at each
right-han output which is marke by “INV”.
•An analogue configuration is vali for NIM output channels CH2 an CH3. This means
also at output section (8) are three non-inverting an one inverting 50 Ω outputs for
NIM signals available. NIM stan s for Nuclear Instrumentation Mo ule Stan ar
an is implemente in this pro uct accor ing to the fast negative logic stan ar
version of DOE/ER-045T.
Mounting instructions for assembling into a 19“ rack
•Four elongate holes are available on right an left en of front panel of RLWE 4161
for fastening this unit in a 19” rack. Use IEC 60297 for further instructions.
RADIOCON, owner W. Kern, Bamberger Str. 7, D-01187 Dres en, Germany
T +49 351 32 33 549, E ra iocon@ra iocon. e, www.ra iocon. e
page
Data sheet 31S0009A 5 (9)