1.6 Changing Activation Distance
When you receive the Pre-Sense unit it will be set at a
predetermined distance. Should you need to increase this, it can
only be changed within the first 90 seconds from start-up:
i) Take battery out and replace after 3 seconds.
ii) Replace battery enclosure and place the unit in it’s new
position.
iii) Test the new activation distance with either a patient or
colleague.
iv) The device should be in idle mode. If not, the unit does
not need to be adjusted.
v) Hold the magnetic key over the unit reset point for 10
seconds, during which time the LED will remain Yellow.
vi) Once the yellow LED goes out, the new distance is set.
vii) To reset the distance, follow the factory reset
procedure.
1.7 Factory RESET Distance
Factory reset will return the activation distance to 300mm and
the deactivation distance to 350mm. To enter the Factory Reset
Mode:
i) Turn the unit off by taking out the battery
ii) Hold magnet over reset point
iii) Turn the unit on by replacing battery
iv) A solid, red LED will illuminate for 10 seconds
v) Remove the magnet
vi) The RED and GREEN LED’s will flash alternatively as the
Factory Reset is performed.
1.8 SLEEP Mode
After the unit has been on for more than 90 seconds, the
DISTANCE CHANGE MODE can no longer be accessed, but the
unit can be put into SLEEP MODE to preserve batteries for long
periods of inactivity.
i) Hold the magnetic key over the reset point (on face of
the Pre-Sense unit) for 10 seconds, during which time the
LED will remain Yellow.
ii) Once the Yellow LED goes out, the unit goes to sleep and
a white LED flashes every 10 seconds to indicate this.