Reed Instruments R2800 User manual

Temperature
Simulator
Instruction
Manual
R2800
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

2
Table of Contents
Introduction ................................................................................................ 3
Product Quality........................................................................................... 3
Safety ......................................................................................................... 3
Features...................................................................................................... 4
Included...................................................................................................... 4
Specications..........................................................................................4-6
Output Functions ................................................................................4-5
General Specifications ........................................................................... 6
Instrument Description ............................................................................... 7
Display Description .................................................................................... 8
Operating Instructions...........................................................................8-12
Power ON/OFF ...................................................................................... 8
Auto Power-Off ...................................................................................... 9
Output Function..................................................................................... 9
Simulating Output from DC Voltage..................................................9-10
Simulating Output from Thermocouple (TC)...................................10-11
Simulating output from thermal resistance or RTD.........................11-12
Battery Replacement................................................................................ 12
Applications.............................................................................................. 13
Accessories and Replacement Parts ....................................................... 13
Product Care ............................................................................................ 13
Product Warranty ..................................................................................... 13
Product Disposal and Recycling .............................................................. 14
Product Support....................................................................................... 14
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3
Introduction
Thank you for purchasing your REED R2800 Temperature Simulator. Please
read the following instructions carefully before using your instrument. By
following the steps outlined in this manual your meter will provide years of
reliable service.
Product Quality
This product has been manufactured to meet the stated product specications.
If a certicate of calibration is required please contact the nearest authorized
REED distributor or authorized Service Center. Please note an additional fee for
this service will apply.
Safety
Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center. To
avoid injury to the user or damage to the instrument, please read the
safety information below before initial use:
• Donotoperatetheinstrumentaroundammableorexplosivegas,
vapor or dust.
• Neverapplymorethan30Vbetweenanytwoterminals,orbetweenany
terminal and ground terminal.
Note: For optimal accuracy, allow the instrument to warm up 5 minutes
before operating. If the automatic reference-junction temperature
compensation of the instrument deviates from its designed accuracy,
contact an Authorized REED Service Center.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

4
Features
• Source8thermocoupletypes,includingR,S,B,E,K,J,T,N;and7
RTD types, including Pt 10 (385), Pt100 (385), Pt200 (385), Pt500 (385),
Pt1000(385),Cu10,Cu50;plusvoltsandohms
• Basicaccuracyof0.05%
• Internalcoldjunctioncompensation
• Easy-to-read6-digitLCDdisplay
• Zeroadjustmentbutton
• Userselectableunitofmeasure(°F/°C)
• Lowbatteryindicatorandautoshutoff
Included
• TestLeads
• AlligatorClips
• ThermocoupleAdapter
• ProtectiveHolster
• Batteries
Specifications
Output Functions
Accuracyspeciedat23°C±5°C&75%RHforaperiodofoneyear
after calibration.
Output Range Output Range Resolution Accuracy
DCV 100mV -10.00to110.00mV 0.01mV 0.05%rdg.+30mV
1000mV -100.00to1100.00mV 0.1mV 0.05%rdg.+0.3mV
OHM 400Ω 0.0 to 400.0Ω 0.1Ω ±0.05%rdg.+0.2Ω
400Ω 0 to 4000Ω 1Ω ±0.05%rdg.+2Ω
continued...
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

5
Output Range Output Range Resolution Accuracy
TC
R-40to1760°C 1°C ±0.05%rdg.+3°C
(≤100°C)±0.05%rdg.
+2°C(>100°C)
S-20to1760°C 1°C
B400to1800°C 1°C
±0.05%rdg.+3°C
(400to600°C);
±0.05%rdg.+2°C
(>600°C)
E-200.0to1000.0°C 0.1°C
±0.05%rdg.+2°C
(≤-100°C)±0.05%
rdg.+1°C(>-100°C)
K -200.0to1370.0°C 0.1°C
J -200.0to1200.0°C 0.1°C
T-200.0to400.0°C 0.1°C
N-200.0to1300.0°C 0.1°C
RTD
Cu10 -10.0to250.0°C 0.1°C ±0.05%rdg.+0.6°C
Cu50 -50.0to150°C 0.1°C
Pt10 385 -200.0to850°C 0.1°C ±0.05%rdg.+0.6°C
Pt100 385 -200.0to850°C 0.1°C
Pt200 385 -200.0to630°C 0.1°C
±0.05%rdg.+0.6°C
Pt500 385 -200.0to630°C 0.1°C
Pt1000 385
-200.0to630°C 0.1°C
Notes
• Excludesaccessoryleadresistance
• Rangeofexcitationcurrent:0.5mAto3mA;Max.outputvoltage:≤2V
•
The accuracies indicated do not include the error of internal temperature
compensation caused by a sensor. The range of the internal temperature
compensationsensorisfrom-10to50°Cwithitscompensatingerrorup
to0.5°C.
continued...
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

6
General Specifications
Display: 6-DigitLCD
ZeroAdjustmentButton: Yes
KickStand: Yes
PowerSupply: 2xAABatteries
ColdJunctionCompensation: Yes
AutoShutOff: Yes(after15minutes/off)
LowBatteryIndicator: Yes
ReplaceableTestLeads: Yes
Product Certications: CE
OperatingTemperature: 32to122°F(0to50°C)
OperatingHumidityRange: 0to85%
StorageTemperature: 14to122°F(-10to50°C)
Dimensions: 7.1x3.5x1.9"(180x90x47mm)
Weight: 8.2oz (500g)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

7
Instrument Description
R2800
SIMULATOR
Ω°C°F
mV
Pt1000 Pt200 Pt500 Cu50 Cu10
OUTPUT
RJ-ON
1
2
3
4
5
6
7
8
9
10
1. LCDDisplay
2. ON/OFFButton
3. °C/°FSelectionButton
4. RJ-ONButton(T/CReferenceJunctionCompensationButton)
5. Function Selection Output Button
6. Range Selection Button
7. ZeroResetButton
8. OutputValueSettingButton
9. Output Digit Selection Button
10. Output Terminals
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

8
Display Description
3
B
Ω°C°F
mV
Pt1000 Pt200 Pt500 Cu50 Cu10
OUTPUT RJ-ON
4
567
1. Thermocouple Type Indicator
2. Indicates which output value is being set
3. Current output value (unit of measure)
4. RTD Type Indicator
5. LowBatteryIndicator
6. Instrument is performing its reference-cold junction compensation
7. Output State Indicator
Operating Instructions
Power ON/OFF
Press the POWER button to turn the instrument on. Press and hold the
POWER button to turn the instrument off. When the instrument is on, it
will begin an internal self-diagnosis at which time the full screen will be
displayed. When complete, the instrument is ready to take measurements.
continued...
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

9
Auto Power-Off
As a default the instrument will automatically turn off after 15 minutes of
inactivity. To turn off this feature:
1. Turn the meter off.
2. Press the POWER button (to display the full screen).
3. Quickly press the RANG button when the instrument is in the
maintenancestateasindicatedby"1.8.8.888".
4. "AP-XX"willnowappearonthedisplay.
5. Press the buttontotogglebetween"AP-ON"&"AP-OF"."AP-OF"
indicates that the no automatic power-off function is disabled, while
"AP-ON"indicatestheautomaticpower-offfunctionisenabled.
6. Press the ZERO button to save the required setting.
7. PressandholdthePOWERbuttontoexitthemaintenancestateand
turn the instrument off.
Output Function
• TheoutputterminaloftheinstrumentcanproduceDCvoltagessetby
the user or can simulate resistance.
Do not apply any voltage to the output terminal during the operation.
If any improper voltage is applied to the output terminal, it will cause
damage to the internal circuit.
Simulating Output from DC Voltage
1. Insert one end of the test lead into
theoutput(TC/mV)jackofthe
meter and connect the other end
to the input of the instrument under
test, as shown in the diagram to
the right.
2. Press the FUNC button to select
"DCVoltageOutput"."V"will
appear on the display.
continued...
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

10
3. Press the RANGbuttontoselecttherangeof1.0000Vor100.00mV.
"V"or"mV"willappearonthedisplay.
4. Press the and buttons to move the on-screen cursor in order to
select the desired digit on the display.
5. Press the and buttons to change the numerical value of each
digit.(Numericalvaluecannotbechangedbeyondrangemaximum.)
6. Press the ZERObuttonandtheoutputwillbesetto00.00mV
or0.0000V.
Simulating Output from Thermocouple (TC)
1. Insertoneendofthetestleadintotheoutput(TC/mV)jackofthe
meter and connect the other end to the input of the instrument under
test, as shown in the above diagram.
2.
Press the FUNCbuttontoselect"ThermocoupleOutput"."R"and
"°C"willappearonthedisplay.
3. Press the RANG button to select the type of thermocouple.
4. Press the and buttons to move the on-screen cursor in order to
select the desired digit on the display.
5. Press the and buttons to change the numerical value of each
digit.(Numericalvaluecannotbechangedbeyondrangemaximum.)
6. Automatic compensation for reference-junction temperature.
7. Press the ZERObuttonandtheoutputwillbedirectlysetto0000°C
(RorStype),400°C(Btype)or0000.0°C(othertype).
8. Press the °C/°Fbuttontoselecttheunit°Cor°F.
Note: During the direct calibration of an instrument with reference-junction
temperature compensation, it is recommended to press the RJ-ON button
so that the instrument can start the function of automatic reference-
junction compensation. Thus providing the required thermo-electromotive
forceforoutputfollowedbydisplayingthesymbol"RJ-ON"where:
Output thermoelectric force = corresponding emf of set
temperature – emf of room temperature
•
It takes two seconds for the instrument to start its internal referencejunction
temperature. After this, each automatic compensation occurs at
10 second intervals.
continued...
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

11
• Ifthereisachangeintheoperatingambienttemperature,donotstart
the operation until the built-in compensating sensor has become stable
(approx.10minutes).
• Ifthereisnoneedforthesimulatortoperformthefunctionofautomatic
reference-junction compensation, press the RJ-ON button and the
symbol"RJ-ON"willnolongerappearinthedisplay.
Simulating output from thermal resistance or RTD
1. Insert one end of the test lead
intotheoutput(RTD/Ω)jack
of the meter and connect
the other end to the input of
the instrument under test as
shown in the diagram to the
right. The dedicated test leads
supplied with the simulator are
intended for a 2-wire system
for testing output according to
user's requirement.
Note: An additional set of test leads are required for 3-wire or
4-wire systems.
2. Thedisplayofthesymbol"OUTPUT"denotestheinstrumentisinan
output state.
3. Press the FUNC button to select thermal resistance output or RTD
outputasindicatedby"Ω"or"°C"and"Cu10"(Copper10ohmRTD)
ontheLCDdisplay.
4. While RTD function is selected, press the RANG button to select the
type of RTD.
5. Press the and buttons to move the on-screen cursor in order to
select the desired digit on the display.
6. Press the and buttons to change the numerical value of each
digit.(Numericalvaluecannotbechangedbeyondrangemaximum.)
7. Press the ZERObuttonandtheoutputwillbedirectlysetto000.0°C.
8. Press the °C/°Fbuttontoselecttheunit°Cor°F.
2w
RTD 4w3w
continued...
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

12
Note: The instrument produces a simulation resistance up to 400Ω at
itsoutputterminal(RTD/Ω).Themethodofsimulatingresistanceoutput
istosendoutanappropriatevoltage'Vx'accordingtotheexcitation
current'Ix'producedbythecalibratedinstrument.BecauseR(set
resistance)=Vx(outputvoltage)/Ix(excitationcurrent),thecalibrated
devicemustprovideanexcitationcurrenttothesimulator.Theexcitation
current should range from 0.5mA to 2mA.
Note:
When testing the resistance output of a 4-wire system while
employing a 2-wire connection, you should take into consideration the
error(approx.0.1Ω)arisingfromtheleadresistanceofthetestleads.If
the capacitance between the resistance output terminal of the simulator
and the tested instrument is more than 0.1μf, the simulator will produce
improper resistance.
Battery Replacement
When the battery symbol, , appears on the display, the battery needs to
be replaced. In order to replace the battery, proceed with the following steps.
1. Ensure that the meter is turned OFF
and remove any test leads from the
meter's terminals.
2. Liftthetiltstandonthebackoftheunit
to access the battery compartment
door which can be removed using a
Phillips head screwdriver.
3. Replacethe2xAAbatteriesinthelid
of the battery compartment.
4. Reinstall the compartment lid by ensuring that battery terminals
touch the unit's contact points and snap into place.
5. Tighten the screw to secure the battery compartment cover.
Note: Ensure battery door is closed and secured in place before using
the meter. To ensure proper operation, please wait 5 seconds before
turning meter on after changing batteries.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

13
Applications
• Vericationandcalibrationoftemperatureinstrumentsincluding
thermocouple and RTD thermometers.
Accessories and Replacement Parts
CA-05A Soft Carrying Case
C-820 Soft Carrying Case
AD-1 Thermocouple Adapter
R1000SafetyTestLead
FC-300FusedTestLeadSet
Don't see your part listed here? For a complete list of all accessories and
replacement parts visit your product page on www.reedinstruments.com.
Product Care
To keep your instrument in good working order we recommend the following:
• Storeyourproductinaclean,dryplace.
• Changethebatteryasneeded.
• Ifyourinstrumentisn'tbeingusedforaperiodofonemonthorlonger
please remove the battery.
• Cleanyourproductandaccessorieswithbiodegradablecleaner.Donot
spraythecleanerdirectlyontheinstrument.Useonexternalpartsonly.
Product Warranty
REED Instruments guarantees this instrument to be free of defects in
material or workmanship for a period of one (1) year from date of shipment.
During the warranty period, REED Instruments will repair or replace, at no
charge, products or parts of a product that proves to be defective because
of improper material or workmanship, under normal use and maintenance.
REED Instruments total liability is limited to repair or replacement of the
product. REED Instruments shall not be liable for damages to goods,
property, or persons due to improper use or through attempts to utilize the
instrumentunderconditionswhichexceedthedesignedcapabilities.In
order to begin the warranty service process, please contact us by phone at
1-877-849-2127orbyemailat[email protected]todiscussthe
claim and determine the appropriate steps to process the warranty.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

14
Product Disposal and Recycling
Please follow local laws and regulations when disposing or recycling your
instrument.Yourproductcontainselectroniccomponentsandmustbe
disposed of separately from standard waste products.
Product Support
If you have any questions on your product, please contact your authorized
REED distributor or REED Instruments Customer Service by phone at
1-877-849-2127orbyemailat[email protected].
Please visit www.REEDINSTRUMENTS.com for the most
up-to-date manuals, datasheets, product guides and software.
Product specifications subject to change without notice.
All rights reserved. Any unauthorized copying or reproduction of
this manual is strictly prohibited without prior written permission
from REED Instruments.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

TEST & MEASURE
WITH CONFIDENCE
CHECK OUT OUR LATEST PRODUCTS!
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Table of contents