Result is out of range may be caused by:
● Incorrect steps taken in performing the test.
● The Control Solution temperature is not within 15°C(59°F)- 35°C(95°F) prior to testing.
● The Control Solution may be expired or contaminated or has been opened over 3 months.
● Expired or contaminated Test Strips.
● Meter malfunction.
If the problem persists, the system may not be working properly. DO NOT continue to use the system to
test your blood glucose level. If you continue to get results that fall outside the specified range, please
return the device to your local distributor or service center.
Storage and Stability
For best results, store Control Solution at room temperature; do not refrigerate. Refer to the expiration
date for stability. The solution should NOT be used after the expiration date or 3 months after opening.
Explanations of Symbols
Caution:
● Use the same model of Control Solution, Test Strip and Glucose Meter.
● DO NOT apply Control Solution to the Test Strip directly from the bottle, such operation could be
leaded to contaminate and decrease the concentration of Control Solution. You MUST place a drop
of Control Solution on a clean nonabsorbent surface first.
● Only for use outside the body (For in vitro diagnostic use).
Medical Device Safety Service GmbH
Schiffgraben 41, D-30175 Hannover, Germany
Tyson Bioresearch, Inc.
5F, No. 16, 18, 20, 22, Kedong 3rd Rd.,
Zhunan Township, Miaoli County 35053, Taiwan
Consult instructions for use
Catalogue number
Do not use if package is damaged
Batch code
Authorised Representative in the European Community
Control
2. Press and hold button for 2 seconds to change to the Control Solution test mode. " " will appear
on the display. If button is pressed again for 2 seconds, the meter will switch back to the normal
testing mode.
3. Shake the Control Solution gently. Discard the first drop of the Control Solution and squeeze a small
drop on a clean nonabsorbent surface such as a clean piece of wax paper. Do not apply the Control
Solution to the test strip directly from the bottle, as contamination may occur.
4. Introduce the tip of the test strip to the droplet of the Control Solution. The Control Solution is automatically
drawn into the strip. Hold until the meter beeps. The meter will now start to count down and the
Control Solution test result will appear on the LCD screen.
5. Compare the results with the Control Solution range printed on the Test Strip vial. The results should
fall within the specified range listed on the Test Strip vial. If the test result is out of the range, repeat
the test.
IN5HS200000000030
Rossmax Swiss GmbH, 9435 Heerbrugg, Switzerland
www.rossmax.com
OBM_OP_HS200_
ControlInsert_EN_ver2122
Temperature limitation
Manufacturer
Keep away from sunlight
Use by
In vitro diagnostic medical device