RPX Technologies DynaVibe GX Series User manual

DynaVibe GX Series User
Manual
Software Version 1.13
Revised: May 2017
Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com


DynaVibe GX Series User Manual
Table of Contents
1Getting Started ___________________________________________________________ 2
1.1 Using this Manual ________________________________________________________2
2Background Concepts ______________________________________________________ 2
2.1 Dynamic Balancing _______________________________________________________2
2.2 Spectral Analysis _________________________________________________________2
3Applications _____________________________________________________________ 4
3.1 Propeller Balancing _______________________________________________________4
3.2 Rotorcraft Track and Balance _______________________________________________5
3.3 Vibration Spectral Analysis _________________________________________________6
3.4 Turbine Vibration Surveys __________________________________________________7
4DynaVibe GX Operation ____________________________________________________ 8
4.1 Ports and Connections_____________________________________________________8
4.2 Accelerometer Ports ______________________________________________________8
4.3 SD Card Slot _____________________________________________________________8
4.4 Optical Tach and Accessory Port _____________________________________________9
4.5 USB Port ________________________________________________________________9
4.6 Keypad _________________________________________________________________9
4.7 Turning on the System____________________________________________________10
4.8 Home Screen ___________________________________________________________10
4.9 Mode Summary _________________________________________________________12
5Mode Specifics __________________________________________________________ 14
5.1 Settings Mode __________________________________________________________14
5.2 Measure Mode _________________________________________________________15
5.3 EZ Measure Mode _______________________________________________________17
5.4 AutoBalance Mode ______________________________________________________22
5.5 Spectrum Mode _________________________________________________________32
5.6 TotalVibe Mode _________________________________________________________34
5.7 VibeSurvey Mode (GX3 Only) ______________________________________________35
5.8 Williams Mode (GX3 Only) ________________________________________________38
5.9 Rotorcraft (GX3 Only) ____________________________________________________45
6System Maintenance _____________________________________________________ 54
6.1 Cleaning _______________________________________________________________54
6.2 Battery Replacement_____________________________________________________54
7Troubleshooting _________________________________________________________ 55

DynaVibe GX Series User Manual
1
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
Important Notice
Any system that distracts a pilot while operating an aircraft is a safety hazard.
During the balancing procedure, please use extreme caution. Be very careful
with cables and components, especially on pusher-type aircraft. ALWAYS
confirm the ignition is off before rotating the propeller, main rotor, or tail
rotor.
Areas of this manual will highlight some specific safety concerns while
operating the dynamic balancing system. Before using this system, review
and follow all safety instructions in this manual, the relevant application
note(s), and the aircraft manufacturer’s documentation.
This manual is for reference only and does not modify, replace, substitute for, or supersede official
regulations or the aircraft manufacturer’s instructions. There are risks and hazards associated with
modifications to aircraft and these modifications are potentially dangerous and could cause harm. RPX
Technologies, Inc. assumes no liability, either express or implied, in the use of this procedure.

DynaVibe GX Series User Manual
2
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
1Getting Started
1.1 Using this Manual
This document describes the general operation of the DynaVibe GX series of dynamic balancers
(including the GX2 and GX3). Some modes are only available on specific model variants and will be
marked accordingly, such as “GX3 Only”.
Because of the wide range of applications for this tool, use this manual in combination with one or
more DynaVibe application notes or aircraft maintenance manuals that describe the setup and usage
for specific applications.
2Background Concepts
2.1 Dynamic Balancing
A rotating mass will always have small imbalances and tolerance variations. On a fixed wing aircraft, for
example, the engine crank, spinner, propeller, propeller extension, and starter ring mass variations will
cause vibrations that propagate through the engine, airframe, instruments, and passengers.
Vibration can cause damage throughout the aircraft. For example, vibration can damage the crank
bearings, engine mounts, firewall, instrument panel, exhaust, air intake, and more. Vibration also
causes accelerated wear and fatigue on equipment and passengers. The negative effects of vibration
can be reduced by dynamically balancing the rotating assembly as a combined unit.
Dynamic balancing is accomplished by monitoring an accelerometer and an optical tachometer
attached to the aircraft. The dynamic balancer uses these sensors to monitor assembly movement,
rotational speed, and angular position as the assembly rotates. The balancer then quantifies the
vibration magnitude and phase angle. Based on this measured vibration magnitude and phase angle,
adding weight to the assembly at critical locations improves alignment of the center of mass to the
rotational axis, thereby reducing the vibration magnitude.
While not all vibration can be eliminated, any reduction in vibration will reduce physical damage and
passenger fatigue. Note that vibration from engine combustion, gear reductions, and accessories will
not be affected by dynamic balancing. These additional vibration components can be identified and
quantified using spectral analysis.
2.2 Spectral Analysis
Spectral analysis is a data processing technique that examines a broad range of frequencies
simultaneously. Unlike dynamic balancing, where the magnitude and phase of a single vibration

DynaVibe GX Series User Manual
3
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
frequency are determined, spectral analysis displays vibration magnitudes of a broad range of
frequencies on a graph.
For example, in a fixed-wing piston-engine-driven aircraft, vibrations from engine misfire, prop
imbalance, prop wash, imbalanced alternator, etc. will each occur at a different frequency and will be
shown as individual peaks on the spectral graph. When performing spectral analysis with the optical
tach installed, the source of each vibration peak can often be identified by comparing the ratio of the
peak’s frequency to the frequency measured by the optical tach. On a non-geared four-stroke airplane
engine with a two-blade propeller, the propeller assembly vibration will be at 1x, weak cylinders will be
at 1/2 x, prop wash will be at 2x, etc.
Refer to the sections below for more information on how to use the DynaVibe GX for spectral analysis
and how it applies to your applications. RPX application notes, such as RPX-AFA-PR-SPC-00 "Spectral
Analysis of Piston Engine Aircraft," provide additional information.

DynaVibe GX Series User Manual
4
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
3Applications
3.1 Propeller Balancing
Process Overview
Balancing is an iterative process of vibration measurement and mass adjustments.
To balance a propeller, first mount an accelerometer to the engine and an optical tach to the aircraft as
described in the application note(s) listed below, or as described by the aircraft manufacturer’s
maintenance procedure. The DynaVibe GX uses these sensors to measure the vibration magnitude and
phase angle.Based on these measurements and information from previous runs, the balancer
determines the location where dynamic balance weights need to be added or removed to balance the
assembly.
DynaVibe GX Modes to Use
The DynaVibe GX has three modes useful for balancing propellers:
AutoBalance (recommended): This is the generally recommended mode for balancing propellers. Not
only does it report vibration magnitude and phase angle in an easy-to-understand format (zero degrees
is up and angles are always measured in the clockwise direction), this mode also recommends specific
mass adjustments and “learns” from each run. This simplifies and accelerates the balancing process.
When finished, the AutoBalance mode also generates a report for documentation.
EZ Measure: EZ Measure provides less automation than AutoBalance. While it does give vibration
magnitude and phase angle in the same easy-to-understand format, it does not recommend mass
adjustments. This mode is used to manually walk through the balancing process, compare front-to-
back vibration magnitudes, and other applications where a quick reading is needed.
Measure: Measure mode gives raw vibration magnitude and phase angle measurements from one or
two accelerometer channels. In Measure mode, the phase angle is not compensated for the setup (see
the Measure mode section below for how to interpret the phase angle). Measure mode is useful for
someone who has experience with a DynaVibe classic or another balancer that gives uncompensated
vibration phase angle readings.
Spectrum (troubleshooting):Note that while propeller assembly imbalance is the most common cause
of aircraft vibration, prop balancing only balances the propeller assembly. If vibration is still present
after the propeller is balanced, then spectral analysis can help identify the source of vibration. Refer to
the spectral analysis section below for more information.

DynaVibe GX Series User Manual
5
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
Application Notes and Procedures
This manual is not intended to be a comprehensive guide to propeller balancing. For process details
please refer to the following documents:
•Application Note: RPX-AFA-PR-BAL-00 “Dynamic Propeller Balancing on Aircraft”
•Aircraft manufacturer’s maintenance manual (if a procedure is available)
3.2 Rotorcraft Track and Balance
Process Overview
Rotocraft track and balance procedures vary by rotorcraft type and manufacturer recommendations.
General overviews are provided in this document and in our application notes, but please refer to the
manufacturer’s maintenance manuals for the exact track and balance procedure for your aircraft.
In general, main rotor track and balance procedures start on the ground with an initial dynamic balance
and blade track. Once an initial track and balance is complete, the process is repeated in a hover, then
in forward flight.
Other rotating components, such as fans, shafts, and tail rotors, can also be balanced using the
DynaVibe GX.
DynaVibe GX Modes to Use
Note: When using published track and balance charts or procedures from manufacturer’s
maintenance manuals, use only Rotorcraft mode or Measure mode since these charts and procedures
require raw phase angle readings.
Rotorcraft Mode (recommended):Rotorcraft mode in the DynaVibe GX3 can be used with or without
the DynaTrack. However, Rotorcraft mode is the ONLY mode that uses the DynaTrack for optical blade
tracking.Rotorcraft mode first helps you set up the system, then allows several measurement
conditions to be taken back-to-back, such as hover and multiple forward flight speeds. Blade track and
up to two vibration measurements (i.e. vertical and/or lateral) are recorded simultaneously during each
data set acquisition. This comprehensive set of measurements can then be used to determine what
adjustments are needed between flights to track and balance the ship.
Measure:Measure mode can be used to take single vibration measurements on a rotating component
using either one accelerometer alone or two accelerometers simultaneously. Please refer to the
Measure description above in the Propeller Balancing section, as well as the Measure mode
instructions below in the mode descriptions for more information.
AutoBalance (exercise additional caution when used with rotorcraft): Although AutoBalance mode is
intended for propeller balance applications, it can be useful when balancing other rotating components
on the helicopter such as tail rotors. Keep in mind that the initial weight addition recommended by
the AutoBalance mode should NOT be used on non-propeller applications because the weight
recommendations are based on typical fixed-wing propeller/engine mass and are not appropriate for

DynaVibe GX Series User Manual
6
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
other applications. Installing excessive weight may damage the assembly and the aircraft.The user
must estimate the first trial weight addition based on the particular assembly being balanced when
using AutoBalance mode for non-propeller applications. A common initial estimate for the trial weight
is 0.1% of the overall rotating mass (0.001 x mass).
Spectrum:Spectrum mode can be used to determine the source of various vibrations in the ship. Since
the main-rotor, tail-rotor, engine fan, engine, weak cylinders, prop wash (e.g., 2-per-rev on a two-
bladed propeller), etc. all occur at unique frequencies, spectral analysis can be used to identify the
source and magnitude of each component.
Application Notes and Procedures
This manual is not intended to be a comprehensive track and balancing guide. For process details
please refer to the following:
•Refer to the rotorcraft manufacturer’s maintenance manual for specific procedures!
•Application Notes: Several application notes are available to provide guidance unique to each
type of helicopter. Some examples include:
oRPX-ARA-RB-R22-TB “Robinson R22 Track and Balance with the GX3”
oRPX-ARA-RB-R44-TB “Robinson R44 Track and Balance with the GX3”
oRPX-ARA-BE-206-TB “Bell 206 Track and Balance with the GX3”
Covers: Bell 206A, 206B3, 206 L1/L3/L4, OH-58 A/C
o(refer to www.rpxtech.com for more rotorcraft-specific application notes)
3.3 Vibration Spectral Analysis
Process Overview
Spectral analysis, often referred to as an FFT, is performed with a single accelerometer and optional
optical tach. Spectral analysis of the accelerometer signal results in a magnitude versus frequency plot
of the vibration signal. When combined with an optical tach signal, these frequencies can be
conveniently compared to the rotational frequency of a reference component, such as rotor, propeller,
or engine RPM. Since many vibration sources have frequencies that are fixed ratios to a reference RPM,
the source of each vibration peak can usually be determined.
DynaVibe GX Modes to Use
Spectrum (recommended):Spectrum mode is used for most spectral analysis with the exception of
turbine engine vibration surveys, which require a wider frequency range.
TotalVibe (related): TotalVibe mode displays a graph of total vibration versus RPM. Although not a
spectral mode, TotalVibe is often useful in applications that require spectral analysis when vibration
magnitudes change significantly with RPM.

DynaVibe GX Series User Manual
7
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
Application Notes and Procedures
•Application Note: RPX-AFA-PR-SPC-00 “Spectral Analysis of Piston Engine Aircraft”
•Application Note: RPX-ARA-PR-SPC-00 “Spectral Analysis of Rotorcraft”
3.4 Turbine Vibration Surveys
Process Overview
Vibration surveys are used to help verify turbine engine health. By monitoring a wide range of
frequencies (up to 20 kHz) it is often possible to detect component balance issues, damaged bearings,
and other turbine engine faults.
GX Modes to Use
VibeSurvey (GX3 Only):The VibeSurvey mode measures the high frequencies required for turbine
vibration surveys. Note that the 20 kHz accelerometer (part number A20M0AS) and cable are required
for VibeSurvey mode. These parts are available in the “Vibration Survey Accessory Kit.”
Williams (GX3 Only):The Williams mode, like VibeSurvey mode, measures the high frequencies
required for turbine vibration surveys. However, Williams mode is customized for use with turbines
manufactured by Williams International and applicable procedures. Note that the 20 kHz
accelerometer (part number A20M0AS) and cable are required for Williams mode. These parts are
available in the “Vibration Survey Accessory Kit.”
Application Notes and Procedures
•Refer to the engine manufacturer’s procedure and/or maintenance program specifications for
procedures that are unique to each engine and maintenance program.

DynaVibe GX Series User Manual
8
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
4DynaVibe GX Operation
4.1 Ports and Connections
Left Right
Accelerometers SD Card Optical Tach
& Accessories
USB
(Factory Only)
The DynaVibe GX ports are shown in the diagram above.
4.2 Accelerometer Ports
The left and right accelerometer inputs are on the left top face. The accelerometer cables are labeled
“L” and “R” near the connectors. These labels indicate the proper input channel that should be used
with each accelerometer so the factory calibration is matched to the correct accelerometer. When
using the DynaVibe, the left channel is primary and used for most modes, such as AutoBalance mode.
The right channel is used for secondary measurements when two accelerometers are needed to take
simultaneous measurements. For example, when using the Rotorcraft mode, use the left channel for
lateral measurements and right channel for vertical measurements. That will ensure that the sensor
positions match what the software is expecting.
4.3 SD Card Slot
The SD memory card slot is in the center of the top face. Screen shots (using the camera button),
reports, and CSV data files are saved to the SD card. The SD card is also used when upgrading the
DynaVibe GX. The SD card slot is spring-loaded. With the DynaVibe turned off, gently press in on the SD
card to eject it. The SD card must be inserted for system operation.
Do not remove the SD card while the unit is powered on.

DynaVibe GX Series User Manual
9
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
4.4 Optical Tach and Accessory Port
The large circular port on the top face is for the optical tach and accessories, such as the DynaTrack.
When not using the DynaTrack accessory, use this port for the optical tach. When using the DynaTrack,
plug the DynaTrack into the GX using this accessory port and then plug the optical tach into the
DynaTrack’s optical tach connector.
4.5 USB Port
The USB port is for factory use only.
4.6 Keypad
The DynaVibe GX has several features to allow rapid access to desired functions. There are the four
function keys along the top of the keypad that allow access to the modes available on the GX. Below
the function keys is a keypad that allows you to enter numbers and text.Some keys also have arrow
symbols indicating their use as directional and cursor keys. The <5> key has a check symbol <>
indicating its secondary function as the Select key.
Keypad
The three keys along the bottom of the keypad provide access to screen image capture, the Home
screen, and power functions.
The first key in the lower left position of the keypad has a camera icon and is used for screen capture; it
allows a user to take a snapshot of the current screen and save that image to the SD card. This is useful
in circumstances where you wish to capture data on the screen. (The Camera key also functions as the
decimal key when entering numeric data and as the period key when editing text.)
The Home key is in the center of the row and the graphic on the key resembles a small house. Home is
a shortcut to the Home screen from most screens and is useful for quickly switching modes. (The Home
key also functions as the zero key, <0>, when entering numeric data.)

DynaVibe GX Series User Manual
10
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
The third key on the right of the bottom row is the Power key. Press and hold Power for a few seconds
to turn the GX off. Press Power when the system is off to turn the system on.
4.7 Turning on the System
Turn on the DynaVibe GX by pressing the Power key in the lower right corner of the keypad; the
DynaVibe logo screen will appear for a few seconds. On initial use and when the system is not resuming
a mode, the Start-Up screen will appear after the logo screen. The Start-Up screen shows the DynaVibe
model type, software version, calibration due date, and serial number of the system.
Start-Up Screen
4.8 Home Screen
From the Start-Up screen, press the <F4> function key (under the “Next” soft key label) to open the
Home screen. Use the Home screen to start the system’s modes and view system status information. To
return to this screen from most other screens, press the Home key.
Tachometer
Alignment Indicator
Soft Menu
with Mode Icons
Battery
Indicator
Icon
Additional
Modes to
Right
Additional
Modes to
Left
Real-time
Clock

DynaVibe GX Series User Manual
11
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
Home Screen Components
Starting Modes
Press the function key directly below a “soft menu” mode icon to activate that mode. Most operations
on the DynaVibe GX will utilize the soft menus.
Real-Time Clock
The system’s real-time clock keeps track of the time to provide time-stamped reports. This helps the
operator keep a history of when the work was performed. Before use, verify that the real-time clock is
set to the local date and time. To change the date or time, see 5.1 “Settings Mode” on page 14.
Tachometer Alignment
The tachometer alignment indicator is useful when the back of the optical tach is not visible and the
tape needs to be aligned with the optical tach. The indicator shows “Tach: Aligned” when the optical
tach is aligned with the reflective tape or “Tach: --------” when not aligned.
Battery Icon
The battery indicator icon shows a graphical representation of the amount of battery life remaining.
Change the batteries if the battery icon shows less than two bars or when the “Low Battery” warning
appears during use.
Soft Menu
The soft menu along the bottom of the Home screen contains mode icons that match up with the
function keys (<F1> through <F4>). Press the corresponding function key to start a mode.
The “more” arrows indicates that there are additional modes available that are not currently shown.
Press <3> (scroll right arrow) to access any additional modes to the right of the current screen. Press
<1> (scroll left arrow) to access any additional modes to the left of the current screen. As shown below,
the additional modes will appear as the menu scrolls to the right. VibeSurvey, Williams, and Rotorcraft
modes only appear on the GX3.
Home Screen – Illustrated Across Multiple Screens to Show All Mode Icons

DynaVibe GX Series User Manual
12
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
4.9 Mode Summary
The DynaVibe GX family supports task-specific modes. Some modes are model-specific. The major
function of each DynaVibe mode is summarized in the following table:
Mode Name
Model
Purpose and Features
ALL
•Automated weight solutions supporting multiple runs (up to 8)
•Vibration magnitude and phase angle (with polar charting)
•Phase IS compensated for setup
•Save HTML report with weight additions, vibration phase angle and
magnitude, final spectrum, graphical summary, and run log
•Single accelerometer input (left channel)
ALL
•Vibration magnitude and phase angle (with polar charting)
•Phase IS compensated for setup
•No automated solutions
•Single or dual-channel accelerometer input
ALL
•Vibration magnitude and phase angle (with polar charting)
•Phase is NOT compensated for setup
•No automated solutions
•Single or dual-channel accelerometer input
•Phase angle representation compatible with most third-party procedures
ALL
•Spectrum analysis up to 24,000 RPM
•Velocity (IPS) or Acceleration (G) views (plotted vs. frequency)
•Harmonic markers
•Switchable between left and right accelerometer inputs
•Identify vibration sources based on frequency
ALL
•Acceleration (G) versus frequency with harmonic markers
•Sweep throttle to find high vibration RPM ranges
•Switchable between left and right accelerometer inputs
GX3
•Analyze high speed vibration up to 20 kHz
•Velocity and Acceleration views (plotted versus frequency)
•Configure up to three limit ranges (Velocity view)
•For use only with 20 kHz Accelerometer (available separately)
•Save spectra to CSV files on SD card

DynaVibe GX Series User Manual
13
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
Mode Name
Model
Purpose and Features
GX3
•Designed for use by Williams International Authorized Service Centers
•Analyze high speed vibration up to 20 kHz
•Velocity versus frequency with up to three limit ranges
•For use only with 20 kHz Accelerometer (available separately)
•Acquire multiple conditions and generate HTML reports
GX3
•Methods specific to rotorcraft track and balance procedures
•Supports measurements while on the ground, in a hover, and at varying
forward flight speeds
•Use with dual accelerometers and/or DynaTrack Blade Tracker
•Supports Push-to-Talk (PTT) input for single-user operation
ALL
•Browse screenshots on the SD Memory Card
ALL
•Change Date and Time
•Disable or Enable Auto Power Off (APO)

DynaVibe GX Series User Manual
14
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
5Mode Specifics
5.1 Settings Mode
Use the Settings mode to set the real-time clock’s date/time or adjust the Auto Power Off (APO)
function.
Press <3> (scroll right arrow) several times to scroll to the right-most mode icon on the Home screen.
Press the function key corresponding to the “Settings” entry to activate the Settings mode.
Settings
Press <F2> (“Set Clock”) to set the real-time clock to your local date and time. This will initiate a series
of input screens to set the year, month, day, a.m./p.m., hour, and minute.
If a value already has the correct setting, press <F4> (“Accept”) to leave it unchanged and move on to
the next setting. To change a value, press <F1> (“clear”) to erase the current value or press <F2> (“”)
to delete the rightmost character. Use the numeric keys to enter the new value. When “Accept” is
pressed for the minutes’ entry, the clock will be set to the time and date as specified. The seconds’ field
will be reset to 00.
APO is enabled by default each time the system is turned on and will allow the system to shut down
and conserve battery power when the system is idle. APO[X] indicates Auto Power Off is active and
the unit will shut down after a few minutes of inactivity (no button presses). Press <F3> to deactivate
APO and change the label to APO[ ] . However, if the unit is accidentally left on while APO is disabled
then the batteries will drain prematurely.
NOTE: VibeSurvey and Williams modes override the APO setting.
VibeSurvey increases the APO timeout to 15 minutes. Williams
mode disables APO.

DynaVibe GX Series User Manual
15
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
5.2 Measure Mode
Left
Accel.
Input
Right
Accel.
Input
Optical
Tach
Resume
After
Power Off
DynaTrack
Accessory
PTT
Switch
APO
(Auto Power Off)
Setting
Required:
Standard
(1 kHz)
Accel.
Optional:
Standard
(1 kHz)
Accel.
Required
Yes
Not
Supported
Optional
Yes:
Supports Global
APO Setting
Measure mode supports one or two channels of vibration analysis. It is a quick way to plot vibration
measurements from one or two channels simultaneously on a polar chart.
Measure mode generates raw vibration readings that are not compensated for sensor positions or the
direction of assembly rotation. Therefore, readings from Measure mode are compatible with many
adjustment charts and processes from other sources, such as maintenance manuals. This mode will
generate data equivalent to the DynaVibe Classic, showing magnitude and phase relative to the
accelerometer position when the assembly is aligned with the optical tach.
This mode requires that a standard DynaVibe accelerometer (1 kHz) is connected to the left input prior
to activating the mode. A second standard accelerometer may optionally be connected to the right
accelerometer input.
When using the Measure mode, the phase angle of maximum vibration is reported relative to the
accelerometer position. In other words, the accelerometer position defines 0°. The phase angle is
physically located, with the engine and ignition off, by aligning the reflective tape to the optical tach
then using the accelerometer as the zero-degree reference and measuring the reported angle in the
direction the assembly would rotate during operation. The phase angle typically indicates the heavy
location.
Measure Mode Description
Measure mode starts with a description screen as shown above. Press <F4> “Continue” to continue to
the live polar chart.
The Measure mode polar chart screen plots live accelerometer reading(s) in a polar chart on the left
and displays the tachometer and accelerometer readings on the right. In the following example screen,
the motor is rotating at 2232 RPM and two accelerometers are mounted at different angular positions.

DynaVibe GX Series User Manual
16
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
On the chart, the Left Channel accelerometer is labeled "1" and the Right Channel accelerometer is
labeled "2".
Measure Polar Chart
F1: Help Describes basic functions of the polar chart.
F3: Average Begins an averaging cycle before proceeding to the “Measurement Results”.
Since the averaging is a set number of rotations, the length of time to complete
the average depends on the RPM, but is typically 5 to 10 seconds. Averaging can
improve the accuracy of the magnitude and phase readings, especially at lower
magnitude vibration levels. “Average” can also be started by holding in the
optional PTT switch.
F4: Accept Proceeds to the “Measurements Results” screen using the readings.“Average”
(<F3>) should typically be used instead of “Accept” (<F4> or a short press of the
optional PTT switch).
The measurement results screen summarizes the final readings on the polar chart screen. Press <F1>
“Back” to return to the live polar chart or <F4> “Menu” to return to the Home screen.
Configuration Examples
•Propeller balancing is typically performed by using a single accelerometer at the front of the
engine.
•After balancing the propeller, Measure mode can be used to do a front-back survey of the
engine using both accelerometers simultaneously. When propeller balancing is complete, it is
possible that the back of the engine has residual imbalance. This can be caused by a number of
complex interactions and further troubleshooting may be necessary.
•Helicopter track and balance procedures typically use one or two accelerometers.
NOTE: In Measure mode, phase is NOT corrected for setup: the
accelerometer position defines 0° and angular degrees increase
in direction of rotation.

DynaVibe GX Series User Manual
17
Copyright © 2017 RPX Technologies, Inc. Doc: RPX-AXU-RP-GX2-113-A
www.rpxtech.com Software Version: 1.13
5.3 EZ Measure Mode
Left
Accel.
Input
Right
Accel.
Input
Optical
Tach
Resume
After
Power Off
DynaTrack
Accessory
PTT
Switch
APO
(Auto Power Off)
Setting
Required:
Standard
(1 kHz)
Accel.
Optional:
Standard
(1 kHz)
Accel.
Required
Yes
Not
Supported
Optional
Yes:
Supports Global
APO Setting
EZ Measure mode can simplify interpretation of the phase angle by compensating the measurement for
sensor mounting locations.
EZ Measure Mode Description
Once configured, EZ Measure corrects for the tachometer position, accelerometer position, and
direction of rotation to provide vibration magnitude and location relative to 0° being at the 12 o'clock
(12:00) position.
EZ Measure Mode Setup - # of blades
The first step is to configure the number of blades on the propeller. Press the right arrow (<6>) to
increase the number of blades. Press the left arrow (<4>) to decrease the number of blades. Press
“Next” (<F3>) when the number of blades is correct.
This manual suits for next models
2
Table of contents