SAE HAN Medical DCD 2IN1 User manual

1
DCD 2IN1
Self-Monitoring Blood Glucose System
User’s Manual
(Model Name: EasyPlus mini 2IN1)

2
Dear DCD 2IN1 SMBG System Owner,
Thank you for using the DCD 2IN1 Self-Monitoring Blood Glucose (SMBG) System. We designed
this system to be dependable, easy-to-use, compact, lightweight and portable to help you monitor
your blood glucose on a regular basis.
Please read this manual thoroughly before you begin testing. This manual provides you and
your diabetes care team with important information and step-by-step direction to use the DCD 2IN1
Self-Monitoring Blood Glucose System.
Thanks again for choosing the DCD 2IN1 SMBG System.

3
Intended Use
The DCD 2IN1 Self Monitoring Blood Glucose Test System is intended for the quantitative
measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is
done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter
[OTC]) by persons with diabetes, or in clinical settings by healthcare professional, as an aid to
monitor the effectiveness of diabetes control.
Although the DCD 2IN1 SMBG System is easy to use, you may need to consult your
healthcare professional (this may be your doctor, pharmacist or diabetes nurse educator)
for instructions on how to use the system. Only the correct use of the system will ensure
accu
r
ate
r
esu
l
ts
.

4
Standard Accessories
Your new DCD 2IN1 Blood Glucose meter and accessories work together to measure the amount of
glucose in your blood. The system includes:
• User’s Manual
• Quick Reference Guide
• Log Book
• Test Strip Instructions
• CR2032 Battery x 1
• Blood Glucose Meter
• Carrying Case
• Test Strip x 10 ct.
• Normal Control Solution
• Control Solution Instructions
• Lancet x 5 ct.

5
Why is it so important to test blood glucose regularly?
Testing your blood glucose regularly can make a big difference in how you manage your diabetes
everyday.We’vemadethisSMBGsystemassimpleaspossibleto helpyouto usethemeterregularly.
Your meter is easy to use, and you can adjust the lancing device for your comfort.
Do you need Help?
If you have questions or need assistance, please contact your healthcare professional. You can also
visit www.epsbio.com for diabetes management tools and product demonstrations.
Please complete your warranty card and mail it, so you receive the best customer service possible
and product update news.

6
Important Information About Your New DCD 2IN1
The DCD 2IN1 blood glucose meter is designed and approved for testing fresh capillary blood
from fingertip. The meter is for outside the body (in vitro) use. It should not be used to diagnose
diabetes.
Only use DCD 2IN1 blood glucose meter with the DCD 2IN1 Blood Glucose Test Strips. Other
test strips will give inaccurate results.
Testing is not valid for neonatal blood specimen.
Do not disassemble the meter as this may cause damage to the components resulting in
incorrect reading. Disassembling the meter will also void the warranty.
Always keep the meter clean and store it in a safe place. Protect the meter from direct sunlight
to ensure a longer lifespan.
You should not store the meter and test strips in a car, a bathroom or a refrigerator.
Keep the meter and test strips away from children and pets.
You should not test critically ill patients with home-use blood glucose meters.
Elevated levels of acetaminophen, uric acid, gentisica acid, levodopa, dopamine, ascorbic acid
and methyldopa may affect results.

7
Incorrect results may occur when performing the test. If you believe you are not feeling well,
please contact your healthcare professional.
Remove batteries if the meter will not be used for one month or more.
Warning for potential biohazard: Healthcare professionals using this system on multiple
patients should be aware that all products or objects that come in contact with human blood,
even after cleaning, should be handled as if capable of transmitting viral disease.
Do not touch the strips with wet hands.
Do not use expired strips (the expiration date is shown on the bottle.)
Do not bend, cut or twist the strips.
Altitude up to 10,000 feet above sea level has no effect on readings.

8
Health-Related Information
If you are very dehydrated, urinating frequently, low blood pressure, shock or hyperosmolar
hyperglycemic nonketotic coma (HHNKC), you may get a test result that is lower than what your
blood glucose really is. If you think you are dehydrated, call your doctor right away.
If you have followed the steps in the user’s manual, but still have symptoms that don’t seem to
match your test results, or if you have questions, please talk to your doctor.
Please read your test strip instructions carefully for additional health-related information.
Warning for potential biohazard:
Healthcare professionals using this system on multiple patients should handle all
products or objects in contact with human blood carefully to avoid transmitting viral
disease, even after cleaning.

9
Explanation of Symbols
Consult instructions for use Caution
Batch code Do not reuse
In vitro diagnostic medical
device 3V (CR2032) x 1 battery only
Use by Temperature limitation
Manufacturer Catalogue number
Serial number Control
Sufficient for Authorised representative in
the European Community
This product meets the requirements of Directive 98/79/EC in vitro diagnostic
medical devices

10
①Result amount ②Blood apply ③Hypo ④Control Solution ⑤Unit
⑥Date ⑦Time ⑨Memory ⑩Battery
⑥
②
③
⑦
⑧
①
④
⑤
⑨

11
Table of Contents
Chapter 1: Understanding Your Meter …......................................................................................14
The DCD 2IN1 Blood Glucose Meter …..............................................................14
Installing Batteries ....................................................................................................15
Setting The Time And Date ................................................................................................16
Using DCD 2IN1 Blood Glucose Test Strip ...............................................................................17
Chapter 2: Control Solution Test ..................................................................................................18
Why Run a Control Solution Test ..............................................................................................18
About The Control Solutions .....................................................................................................19
Running A Control Solution Test ...............................................................................................20
Understanding Control Solution Test Results …........................................................................22

12
Chapter 3: Testing Your Blood Glucose .......................................................................................24
Using the Bulit-in Lancing Device .................................................................................24
Inserting A Lancet into the Lancing Device ...............................................................................25
Running A Blood Glucose Test With Blood From Your Fingertip .............................................28
Discarding Used Lancets ….......................................................................................................30
Understanding Your Test Results …........................................................................................31
Unusual Test Results …........................................................................................32
Symptoms of High or Low Blood Glucose..................................................................................33
Comparing Your Meter Result To A Lab Result …....................................................................35
Chapter 4: Meter Memory, Setup …...............................................................................................36
Memory, Storing Test Result ….................................................................................................36
Viewing & Deleting Test Results …............................................................................................37

13
Chapter 5: Maintenance And Troubleshooting ….......................................................................39
Installing Batteries ….................................................................................................................39
Cleaning Your Meter ……..........................................................................................................40
Maintenance And Testing .........................................................................................................41
Cleaning Your Lancing Device ….............................................................................................42
Screen Messages And Troubleshooting …...............................................................................43
Chapter 6: Technical Information …..............................................................................................47
Specifications …........................................................................................................................47
Limitation …...............................................................................................................................48
Device Information …................................................................................................................51
Warranty ....................................................................................................................................52

14
Chapter 1: Understanding Your Meter
DCD 2IN1 Blood Glucose Meter
Test Strip Slot-
Insert test stri
p
here.
Button-
Press to enter
memory,
adjust setting,
and scroll
through
results.
Button-
Press to turn
the meter
on, confirm
setting, and
scroll through
results.
Depth
Indicato
r
Cap
Cocking
Button
Batter
y
Doo
r
Display
Release
Button
Carrie
r
-
Insert lancing
here.

15
Installing Batteries
1. Flip open the battery door by pushing
the tab in the direction of the arrow and
pulling the door up.
2. Insert a battery.
3. Put the battery door back in place and
snap it closed. The meter beeps and
turns on automatically.

16
Setting The Time and Date—First Time Use
Setting the current time and date in your meter is important if you intend to use the meter memory or
if you want to download your results to a computer.
1. If your meter does not turn on automatically, press for 3
seconds to turn on the meter.
3. Repeat step 2 to set the date and time. The flashing field is
the one you are currently setting. When finishing the setting,
the meter beeps.
2. The last 2-digits of the year flash at the center of the
display. Press to adjust the year and press to
confirm the settin
g
.

17
Using DCD 2IN1 Blood Glucose Test Strip
Use only with DCD 2IN1 Blood Glucose Meter.
Run a control solution test every time you open a new box of test strips (See Chapter 2
"Control Solution Testing.")
Keep the test strips in their original bottle.
Close the bottle tightly right away, after you take a test strip out. This keeps the test strip dry.
Use the test strip within three minutes after you take it out of the bottle.
The strip is for single use only. Do not reuse it.
Record the date you open the test strip bottle. Be sure to check the “Expiration date” on the test
strip bottle. The test strip is good for three months from the date the bottle is opened or until the
expiration date on the bottle, whichever comes first.
Store the test strip bottle and your meter in a cool dry place.
Store the test strips between 2°C - 30°C (35.6°F- 86°F). Do not freeze.
Do not apply blood or control solution to the test strip until you insert it into the meter.
Do not touch the test strip with wet hands. Do not bend, cut, or twist the test strips.

18
Chapter 2: Control Solution Testing
Why Run A Control Solution Test
We recommend that you run the DCD 2IN1 Normal control test because it lets you know that your
meter and test strips are working properly to give you reliable results. You should run the control
solution test when:
You use the DCD 2IN1 Blood Glucose Meter for the first time.
You open a new bottle of test strips.
You think the meter or test strips may be working incorrectly.
You drop the meter.
You have repeated a test and the test results are still lower or higher than expected.
You are practicing the test procedure.
Professional users are instructed to follow federal, state, and local guidelines.

19
About The Control Solution
Use only with DCD 2IN1 test strips.
Write the date you opened the bottle on the bottle label. The control solution is good for three
months from the date the bottle is opened or until the expiration date on the bottle, whichever
comes first.
Do not use a control solution that is past the expiration date.
The control solution can stain clothing. If you spill it, wash your clothes with soap and water.
Close the bottle tightly after every use.
Left over control solution should not be added back into the control bottle.
Store the bottle of control solution at room temperature, between 2°C- 30°C (35.6°F - 86°F). Do
not freeze.

20
Running A Control Solution Test
You need the meter, a test strip, and control solution.
1. Put a test strip into the meter in the direction of the arrow.
4. Place the meter on a flat surface, like a table.
5. Remove the control solution bottle cap and wipe the tip of
the bottle with a tissue.
2. The meter beeps and turns on automatically. The icon
shows itself.
3. Press the , the icon shows and the meter is ready
to run a Control Solution Test.
Or press to cancel the Control Solution Test Mode.
Table of contents
Popular Blood Glucose Meter manuals by other brands

Accu-Chek
Accu-Chek Compact Plus owner's booklet

Trividia Health
Trividia Health TRUE METRIX GO user manual

Roche
Roche ACCU-CHEK Instant S quick start guide

Roche
Roche Accu-Chek FastClix quick start guide

Lifescan
Lifescan SureStep Blood Glucose Monitor Operator's guide

OKmeter Cameo
OKmeter Cameo OK-P user guide