9.2 Changing weighing units
9.2.1 When “units” is displayed, press [TARE], then press [SET/0~9] to
choose, then press [TARE] to confirm and exit.
Unit: H9(kg), 9(g),Lb,oz,pcs(counting). ( mperial units may not be
chosen when calibrating).
9.2.2 Pcs Counting: Press [SET] key for 3 seconds. The display will
show ‘Range’. Press [SET] key until ‘units’ is displayed. Press
the [TARE] key. The display shows ‘g’. Press the [SET] key until
‘pcs’. Press the [TARE] key. The ‘pcs’ LED will be on. Put the
sample quantity onto the scale. Press and hold the [ZERO] key.
The display shows ‘00000’, and the ‘pcs’ LED will be flashing.
Press the [TARE] key. The display shows ‘00000’ with the left hand
digit flashing. Use the [TARE] key to move right and the [ZERO]
key to increment up. Enter the sample quantity, eg, 200. Press
[TARE]. The display will now be in count mode and show the
sample quantity measured, eg, 200.
To return to normal weighing mode, press [SET] for 3 seconds until
‘units’ is displayed. Press the [TARE] key. The display shows
‘pcs’. Press the [SET] key until ‘g’ is displayed. Press the [TARE]
key.
9.3 A-oFF(
((
(choose auto power off)
))
)
When menu displays A-oFF,press [TARE] to enter,then press [SET/0~9]
to choose ‘n’ or ‘y’. ‘n’ is auto power off,’y’ is auto power off after 10
minutes of inactivity.
9.4 Change display filtering
When the display shows “F LtX” (F Lt1 means one step to display, F Lt2
means 3-4 steps to display, F Lt3 means 6-8 steps to display, F Lt4 is the
fastest filter setting). Press [SET/0~9] to change, press [TARE] to confirm
and exit.
9. Set range of auto-zero:
When ‘zero’ is displayed, press [TARE]. The display will show zero x.x(x.x
is the range:0.5d,1d,1.5d,2d,2.5d,3d,3.5d,4d,4.5d,5d. d is the division value
when n=3000). Press [SET/0~9] to choose and [TARE] to confirm.
9.6 bUZZEr (set buzzer sound on or off)
When menu displays bUZZEr,press [TARE] to enter,press [SET/0~9] to
choose on or off, and then press [TARE] to confirm.