OPERATION
Rotorinstallation.Openthelidoftherotorchamber.Visuallyalignthepinonthedriveshaftwiththegrooveonthe
bottomoftherotor.Thencarefullylowertherdtorontothedriveshaft.Rotatetherotorbyhandtobesurethepinis
linedupwiththegroove.Securetheassemblybyscrewingtheretainingknobintothedriveshaftabovetherotor.
Tightenitfirmlybutnotexcessively.Closethelid.
Alwaysbalancerotorloads.Anunbalancedrotorcausesvibrationthatwillwearoutthebearings.Loadthe
rotoruniformly.Thereneednotbea tubeineachholder,butthetubesmustbeevenlyspacedaroundtherotor.
Whenusinga rotorwithaluminumtubeholders,inserttubeholderssymmetrically.Whendryingdifferenttypesof
sampleinthesamedryingrun,distributeeachtypesymmetrically.Otherwise,theirdifferentdryingratesmight
introduceprecessionintotherotor'srotation.Alwaysselecta rotorintowhichthesampletubesfitsnugly.
Pre-heating.Ifdesired,operatetheDRYINGRATEswitchatleast15minutesbeforeyoustartthedryingoperation.
•LOWleavesthechamberatambienttemperature
•MEDIUMheatsthechamberto43°C
•HIGHheatsthechamberto60°C
SC110operation.Afterinstallingthesamples,closethecoverandswitchtheROTORswitchtoON.Donotapply
vacuumuntiltherotorreachesfullspeed(allow20seconds).Prematureapplicationofvacuumcancause
bumpingandfoaming;insomecases,samplecouldboiloutofthetubes.
*Youcancontinuetherununtila pre-determinedamountoftimehaselapsedoruntilthevacuumgaugeindicatesa
*pre-determinedvaporpressureinthesamplechamber.Whenrepeatabilityofresultsisimportant,conductoneor
moretestrunswithexpendablebatchesofsubstance.
Toendtherun,firstbleedthesamplestoatmosphere.(Allow10seconds.)Whenthesamplechamberis
pressurized,stoptherunbymovingtheROTORswitchtotheOFF/BRAKEposition.Therotorbrakesuntilthe
rotationdecreasesto6-RPM.Donotopenthecoveruntilrotationiscompletelystopped.All220-voltunitshave
acoverlockthatpreventsliftingofthecoverwhentherotorisspinningfasterthan6 RPM.
SC110A/SC210AoperationisthesameasgivenabovefortheSC110,exceptthatyouneednotmanuallyoperate
thevacuumvalve.A speedsensorinsidetheconcentratoractivatesvacuumonlywhentherotorreaches1000
RPM.
WhenyoustoptherunbymovingtheROTORswitchtotheOFF/BRAKEposition,theunitimmediately
bleedstherotorchamber,butwaits10secondsbeforebrakingtherotor.
OperationwiththeSC1021SystemController.
•SC1021SystemController.Usevacuumtubingtoconnectthehosefittingattherearoftheconcentratortothe
otherinstrumentsasdescribedintheSC1021InstructionManual.
MovetheROTORswitchtotheONpositionandleaveitthere.TheSC1021operatesthemotor.TheROTOR
switchontheconcentratorcanbeswitchedOFFasanemergencystopfortherotor.TheDRYINGRATEswitch
controlstheconcentrator'schamberheater.
Removethesmallmetalplateattherearoftheconcentratorcabinet,exposingthe9-pinconnector.Attachthe
controlcablefromthe
SC1021.
Plugthelinecordintothe
SC1021.
ThetextdisplayoftheSC1021givesyoustep-by-stepguidance.TheSC1021operatestheconcentrator,
sequencesrotationandapplicationofvacuum,automaticallyendstherunaccordingtoyourspecifications,and
providesadvancedfeaturessuchasstoredprogramsandrunstatistics.RefertotheSC1021InstructionManual
fordetails.
ThechamberheateriscontrolledbytheDRYINGRATEswitchontheconcentratorfrontpanel.
MAINTENANCE
Maintaininga cleaninstrumentiscrucial.Cleanspillsimmediately.Driedsolventscanbuildupandimpairrotor
rotation.
Periodiccleaningofthesamplechamberpreventsproblems.Usea detergentsolutionona spongeor
gauze,
thenthoroughlywipethechamberdry.ExcessmoistureinthechamberisremovedonthenextSpeedVac
runbutcanvarytheperformanceofthisrun.
AsdescribedunderInstallation,keepthedriveshaftandrotorcenterlightlycoatedwithvacuumgrease.Donot
lubricatethecoverO-ring.