Scaletron 4042 Installation and operation manual

Scaletron™ Industries, Ltd.
Bedminster Industrial Park
53 Apple Tree Lane
P.O. Box 365
Plumsteadville, PA 18949 USA
Toll Free: 1-800-257-5911 (USA & Canada)
Phone: (+1) 215-766-2670 (international)
Fax: (+1) 215-766-2672
E-mail: [email protected]
Website: www.scaletronscales.com
OEM Manual
MODEL 4042™AND 4042WB™ - 3½DIGIT
SECONDARY SPILL CONTAINMENT SCALE

1
© 2007. Scaletron Industries Ltd. Plumsteadville, PA Scaletron. Scaletron Sets The Standard In Corrosion Resistance. Model 1099, Model 2305, Model 2306, Model 2307, Model
2310, Model 2330, Model 2350, Model 2352, Model 2353, Model 3001, Model 3002, Model 3001 SI, Model 3002 DI, Model 3003, Model 3004, Model 3005, Model 3006, Model 3050-1,
Model 3050-2, Model 4010, Model 4020, Model 4021, Model 4030, Model 4040, Model 4042, Model 4042WB,Model 4041, Model 4042, Model 4042 WB, Model 5305, Model 5305-2, and
ProGram Guide are trademarks of Scaletron Industries Ltd. Total Electronic System is a service mark of Scaletron Industries Ltd. All Rights Reserved. .PRINTED IN U.S.A. JULY 2007.
These instructions generally describe the installation, operation, and maintenance of subject
equipment. The manufacturer reserves the right to make engineering refinements that have not been
described herein. Should any questions arise that may not be answered specifically by these
instructions, they should be directed to Scaletron Industries Ltd., Or Our Sales Agent for a
response.
All possible precautions were taken in packaging each piece of equipment to prevent shipping
damage. Carefully inspect each item and report damages immediately. Report all damage claims to
the shipping agent involved for equipment shipped F.O.B. job site. Do not install any damaged
equipment.
All instructions given on any labels, or attached tags, should be followed. Carefully inspect all
packing material before discarding to prevent the loss of accessories, mounting hardware, spare
parts, or instructions.
TABLE OF CONTENTS PAGE
I. GENERAL DESCRIPTION…………………………………………………………………………….1
II. SPECIFICATIONS…………………………………………………………………………………..1 - 2
III. ASSEMBLY AND START-UP…………………………………………………………………………2
IV. GENERAL USE INSTRUCTIONS…………………………………………………………………….3
V. CALIBRATION PROCEEDURE………………………………………………………………..…3 - 4
VI. TROUBLESHOOTING…………………………………………………………………………………4
VII. BOARD DIAGRAM…………………………………………………………………………………5 - 6
I. General Description:
The Model 4042™ and 4042WB™ Secondary Spill Containment Scale is designed specifically
with a built-in containment basin. Off center loading or solid materials do not affect the scale
accuracy due to the four cell design. Mixers and pumps may be used with this base design.
Different base sizes are available, as well as custom bases for different size platforms and
large capacities. The maximum net weight is 199.9 lb./ or kg. with 0.1 lb/ or kg resolution, or
with a maximum net weight exceeding 199.9, the decimal is blanked at the factory, and the
display can read 1999 lb/ or kg. on the standard, 3 ½ Digit indicator.
The design of the platform includes sturdy four point support, with a polyethylene containment
basin. The base is coated with dry polyester powder coating to prevent corrosion. Scale has
four temperature compensated strain gauge transducers.
Standard Parts
1 Base assembly 1 Weigh Meter (indicator)
1 Power Cord 1 Technical Manual
Available Options
A. Remote mounting of standard enclosure
B. Load Cell Cable
C. Low Level relay contact
II. Specifications:
Display: LCD, 3 ½ Digit with minus sign liquid crystal display, 0.5 inch high, seven segment
Resolution: Tank: 199.9 lb x 0.1 lb, or 200 to 1999 lb.x 1.0 lb.

2
© 2007. Scaletron Industries Ltd. Plumsteadville, PA Scaletron. Scaletron Sets The Standard In Corrosion Resistance. Model 1099, Model 2305, Model 2306, Model 2307, Model
2310, Model 2330, Model 2350, Model 2352, Model 2353, Model 3001, Model 3002, Model 3001 SI, Model 3002 DI, Model 3003, Model 3004, Model 3005, Model 3006, Model 3050-1,
Model 3050-2, Model 4010, Model 4020, Model 4021, Model 4030, Model 4040, Model 4042, Model 4042WB,Model 4041, Model 4042, Model 4042 WB, Model 5305, Model 5305-2, and
ProGram Guide are trademarks of Scaletron Industries Ltd. Total Electronic System is a service mark of Scaletron Industries Ltd. All Rights Reserved. .PRINTED IN U.S.A. JULY 2007.
Load Cell Excitation: 12 VDC Power, 4 Load cells
Zero Adjustment: Internal potentiometer, External tare knob
Overload: Blanks with “1” displayed at the far left of the display
Accuracy: +/- 0.5% full capacity
Power: 85-265VAC ~ 50/60 Hz, single phase, 0.42 Amps Dedicated – Isolated Power With
Earth Grounding
Dimensions: Base sizes: Various sizes available
III. Assembly and Start-Up:
The Model 4042™ and 4042WB™ Secondary Spill Containment Scale is shipped partially
disassembled. Assembly instructions are as follows:
The load cells are secured to the base. All wire connections in the base have been made and
sealed at the factory. The digital weigh meter is assembled and calibrated to the base at the
factory. The customer must mount the indicator on the wall. The base should be roughly
leveled using the adjustable foot in the load cell. The wires must be run through the strain
relief and connected to the connector before scale is operated. Steps are as follows:
Step 1: Remove all parts from box and inspect for damage – bent metal, broken wires, cracks
in indicator, etc. Any shipping damage must be reported to carrier!
Step 2: Place the scale on a solid, dry, and even surface. Clear the area of all debris where
the scale base will be placed.
VERY IMPORTANT: If Model 4042WB™, make sure that expandable bladder has room
and clearance to expand without coming into contact with any other objects that may
cause a leak or obstruct proper expansion o the bladder!!!
Step 3: If conduit is to be used, we recommend at least 2 feet of FLEXIBLE conduit
horizontally, as to not interfere with free movement of the base for accurate measurement.
Feed the wire up through the conduit, and apply a fitting.
Step 4: Drill a hole in the indicator, being sure to avoid the PC board, and completely seal the
hole with the fitting. If ANY SPACE exists around the fitting allowing gas into the box, the
equipment will be damaged. You MUST SEAL any openings with a flexible caulk, such
as silicon! Fasten the indicator to the wall. Connect the wires into the connector using the
following code:
TB1 : Ret: Common, 4-20mA Return*
Out: + 4-20mA Output*
G: Green = + Signal
W: White = - Signal
Bk: Black = - Supply Voltage
R: Red = + Supply Voltage
*Connect your wires here if you are using 4-20mA output. Refer to the wiring diagram on
page 5 and 6 for 4-20mA connections. Jumper J-4 will need jumper in the top two pins for loop
powered, and the bottom two pins for scale powered (with board oriented as in drawing pg 5.)
Insert the wire, being sure not to crimp the insulation in the connector. Using a small
screwdriver, push the orange lever down to insert the wire into the proper hole and release the
lever to allow the wire to clamp into the connector. The connector holes are labeled according
to the color that is inserted there.
Step 5: Allow approximately 15 minutes warm up time before using the scale.

3
© 2007. Scaletron Industries Ltd. Plumsteadville, PA Scaletron. Scaletron Sets The Standard In Corrosion Resistance. Model 1099, Model 2305, Model 2306, Model 2307, Model
2310, Model 2330, Model 2350, Model 2352, Model 2353, Model 3001, Model 3002, Model 3001 SI, Model 3002 DI, Model 3003, Model 3004, Model 3005, Model 3006, Model 3050-1,
Model 3050-2, Model 4010, Model 4020, Model 4021, Model 4030, Model 4040, Model 4042, Model 4042WB,Model 4041, Model 4042, Model 4042 WB, Model 5305, Model 5305-2, and
ProGram Guide are trademarks of Scaletron Industries Ltd. Total Electronic System is a service mark of Scaletron Industries Ltd. All Rights Reserved. .PRINTED IN U.S.A. JULY 2007.
IV. General Use Instructions:
Please refer to these instructions for daily use of this scale. These instructions simulate the
procedure for every day usage.
ON AND OFF LOADING USAGE:
To start with a new, FULL tank that will be loaded on & off, please follow these steps:
Please choose a method of usage by Tare or Net Weight:
TARE: You must know the weight of your empty tank, and any other equipment mounted on
to the tank that will affect the weight. Add this weight together, and turn the black tare knob on
the front of the indicator until it reads that number in the negative. Turn the knob counter-
clockwise in order to achieve negative numbers. Once that negative tare weight is set, you
may load the tank, and apply other equipment on that tank. The weight that registers on the
scale indicator is the NET WEIGHT, or weight of the contents. At this point you will not change
the black tare knob unless you are loading on a new tank. As you use the contents, the weight
decreases. When the tank is at zero, or close to it, you will begin again by removing the empty
tank, setting the tare weight with the black knob (for the new tank + equipment), and begin
using the material again.
NET: If you do not know the tare weight (it is not stamped on the tank), you may know the Net
Weight (contents in the tank). If this is the case, you will use this procedure:
Load on your full tank and attach any equipment that affects the weight on the scale. Turn the
black tare knob on the front of the indicator until it reads the known NET WEIGHT (weight of
contents only) on the indicator. Clockwise will increase the readings and counter-clockwise
will decrease the readings. When the weight is set for the contents in the tank, do not adjust
this again until you load on a new tank. The reading will decrease until you have used all the
contents in the cylinder, where it will read 0 lb./kg. Start a new tank by removing the old empty
tank, loading on a new full tank, and attaching equipment, and set tare knob to Net weight
again.
For use when FILLING the TANK instead of on & off loading:
To start with a new, EMPTY tank that will be filled while on the scale base, please follow
these steps:
Load the new, empty tank on to the platform. Be sure to connect all equipment to the tank that
may apply weight to the scale. With this weight applied, turn the black tare knob located on
the outside, front of the indicator, until the digital reading says zero. This means that your
digital reading will show net weight, or the weight of the contents alone. As you fill the tank for
the first time, the scale shows the current weight at all times. Do not readjust the tare weight
unless you are starting with a dry, new, empty tank. Any residual weight will reflect what is left
in the tank, and should not be changed unless a new tank is put on the base.
V. Calibration Procedure:
The Model 4042™ and 4042WB™ Secondary Spill Containment Scale is pre-calibrated at
the factory to within specified accuracy and is calibrated to standards traceable to the
Bureau of Weights and Measures. No further calibration should be necessary. If the
electronics or load cells in the base are being replaced, the following procedure should
be used.
1. Plug the AC Cord into a well-grounded receptacle. There is no power on / off switch, so
power is applied as soon as it is plugged in.
2. Allow approximately 15 minutes warm-up time before calibration.

4
© 2007. Scaletron Industries Ltd. Plumsteadville, PA Scaletron. Scaletron Sets The Standard In Corrosion Resistance. Model 1099, Model 2305, Model 2306, Model 2307, Model
2310, Model 2330, Model 2350, Model 2352, Model 2353, Model 3001, Model 3002, Model 3001 SI, Model 3002 DI, Model 3003, Model 3004, Model 3005, Model 3006, Model 3050-1,
Model 3050-2, Model 4010, Model 4020, Model 4021, Model 4030, Model 4040, Model 4042, Model 4042WB,Model 4041, Model 4042, Model 4042 WB, Model 5305, Model 5305-2, and
ProGram Guide are trademarks of Scaletron Industries Ltd. Total Electronic System is a service mark of Scaletron Industries Ltd. All Rights Reserved. .PRINTED IN U.S.A. JULY 2007.
3. While indicator is warming up, proceed to check under cylinder support disc to make sure
there is no debris.
4. WITH NO WEIGHT APPLIED TO THE SCALE, turn the black tare knob on the outside of
the door of the indicator in the clockwise direction to make sure that there is at least +10 lb. of
weight in the tare knob. If not, adjust R-6 to see a positive reading. Once that is established,
turn the black tare knob on the door of the indicator counter-clockwise until it stops (DO NOT
OVER TURN KNOB OR YOU WILL DAMAGE THE TARE POT!) Once it reaches the most
negative reading, make sure you have at least the amount of tare weight that you require for
the tanks/drums that you will use on this platform. If the tare span is very large, you may get to
a –1 reading and this is a weight under-load condition. Turn R-6 until you see a number a
number that is not “below” –199.9 lb or 1999 lb.
5. Adjust tare knob on front of weigh meter until scale reads zero with no weight on platform.
Adjust R-22 to set the 4mA reading to 0lbs.
6. Place a known weight on the disc, as close to center as possible. Weight should be equal
to ½ of capacity, or greater.
7. Adjust R-15 until display reads weight of test weight on scale disc.
8. To complete calibration of 4-20mA output, adjust R-19 for proper 20 mA adjustment
reading. The customer specifies the factory setting for 20 mA at the time the order is placed. It
should have been assigned to the net weight of your intended product. Consult white “4-20mA
settings” sticker on inside of indicator!
9. Calibration is complete if specified accuracy is obtained.
VI. Troubleshooting:
1. If display reads: “-1”:
A. Gross under-load. Turn knob clockwise. Numbers should reappear and can be
adjusted to tare weight or zero as required.
B. Check load cell connections at indicator, and connect wires as per diagram on Page
5.
2. If display reads: “1”:
A. Gross over-load. If starting point was zero, the cylinder weight is more than 199.9
lb., or kg. with the decimal, or 1999 lb./kg. without the decimal.
B. Check load cell connections according to page 2.
3. If no display (not lit):
A. Check power to weigh meter.
B. Check green light on power supply board. It should be lit. If not lit, it is shorted out.
4. Other Problems:
A. Consult the Factory!
USA & Canada Toll-Free: (800) 257-5911
Tel: (+1) 215-766-2670 ♦Fax: (+1) 215-766-2672
Notice: Do not return any equipment without first contacting the factory. A return
authorization number will be issued and it must be marked on all materials returned to the
factory, accompanying a letter that explains the problem specifically. A Serial Number will
also be required. It is located inside the indicator box.

5
© 2007. Scaletron Industries Ltd. Plumsteadville, PA Scaletron. Scaletron Sets The Standard In Corrosion Resistance. Model 1099, Model 2305, Model 2306, Model 2307, Model
2310, Model 2330, Model 2350, Model 2352, Model 2353, Model 3001, Model 3002, Model 3001 SI, Model 3002 DI, Model 3003, Model 3004, Model 3005, Model 3006, Model 3050-1,
Model 3050-2, Model 4010, Model 4020, Model 4021, Model 4030, Model 4040, Model 4042, Model 4042WB,Model 4041, Model 4042, Model 4042 WB, Model 5305, Model 5305-2, and
ProGram Guide are trademarks of Scaletron Industries Ltd. Total Electronic System is a service mark of Scaletron Industries Ltd. All Rights Reserved. .PRINTED IN U.S.A. JULY 2007.
VII. Board Diagram:

6
© 2007. Scaletron Industries Ltd. Plumsteadville, PA Scaletron. Scaletron Sets The Standard In Corrosion Resistance. Model 1099, Model 2305, Model 2306, Model 2307, Model
2310, Model 2330, Model 2350, Model 2352, Model 2353, Model 3001, Model 3002, Model 3001 SI, Model 3002 DI, Model 3003, Model 3004, Model 3005, Model 3006, Model 3050-1,
Model 3050-2, Model 4010, Model 4020, Model 4021, Model 4030, Model 4040, Model 4042, Model 4042WB,Model 4041, Model 4042, Model 4042 WB, Model 5305, Model 5305-2, and
ProGram Guide are trademarks of Scaletron Industries Ltd. Total Electronic System is a service mark of Scaletron Industries Ltd. All Rights Reserved. .PRINTED IN U.S.A. JULY 2007.
4-20mA Diagram:
This manual suits for next models
2
Table of contents
Other Scaletron Scale manuals

Scaletron
Scaletron 4021 Installation and operation manual

Scaletron
Scaletron 2308B Installation and operation manual

Scaletron
Scaletron 4041 Installation and operation manual

Scaletron
Scaletron 2350 User manual

Scaletron
Scaletron 4020 Installation and operation manual

Scaletron
Scaletron 2305 Installation and operation manual

Scaletron
Scaletron 4042-ECO Installation and operation manual

Scaletron
Scaletron 4020 Installation and operation manual

Scaletron
Scaletron 2353 Installation and operation manual

Scaletron
Scaletron 4042 Installation and operation manual