Scan Optics SO-5800 User manual

SCAN OPTICS
SO-5800
MICROSCOPE
ASSEMBLY MANUAL

CONTENTS
INTRODUCTION .................................................................................................................... 3
PACKING LIST .................................................................................................................... 5
MECHANICAL ASSEMBLY ............................................................................................ 6
Floor stand base and pillar ................................................................................ 6
Pantograph arm ............................................................................................ 9
X-Y unit ........................................................................................................ 11
Microscope head and suspension bracket ........................................................ 13
Eyepieces, camera, handles and spare lamp module ............................................ 15
Monitor arm and bracket ................................................................................ 17
Binocular assistant microscope ................................................................................ 19
CABLE AND POWER ASSEMBLY ................................................................................ 22
MECHANICAL SETTINGS ........................................................................................................ 24
FIGURES
Figure A1 SO-5800 Microscope ............................................................................... 4
Figure A2 Assembling the floor stand base ................................................................... 7
Figure A3 Assembled floor stand base ................................................................... 7
Figure A4 Inserting the pillar in the floor stand base....................................................... 8
Figure A5 Fixing the pillar to the floor stand base ....................................................... 8
Figure A6 Placing the friction washer and arm bearing mounts ............................... 10
Figure A7 Attaching the pantograph arm ................................................................... 10
Figure A8 Attaching the X-Y unit ............................................................................... 12
Figure A9 Attaching the microscope head and suspension bracket ............................... 14
Figure A10 Attaching eyepieces, handles and camera ........................................... 16
Figure A11 Attaching the monitor arm and bracket ....................................................... 18
Figure A12 Binocular assistant microscope in RHS orientation ............................... 21
Figure A13 Binocular assistant microscope in LHS orientation ............................... 21
SO-5800 Ophthalmic Microscope Assembly Manual Page 2of 25
Issue Number 2.0 Reference SO-5800-AM-2

INTRODUCTION
Please read the following information carefully before installing and using the Scan Optics SO-5800
Ophthalmic microscope. Scan Optics is responsible for the safety, reliability and performance of the
equipment only if it is used in accordance with these instructions.
This microscope is designed for use by a certified practitioner, for magnified observation of patients,
and for use in an operating theatre as an observation aid during surgery.
Environmental storage and packing conditions of 60-95% relative humidity and 10-40 °C, are
recommended for this product.
No parts or accessories supplied with this microscope are supplied in a sterile condition.
Apart from those instructions within this manual, there are no user-serviceable parts in this
microscope. Scan Optics will retain the discretion to advise whether any repairs may be taken out by
external qualified technical personnel, or whether part(s) of the microscope must be returned to the
manufacturer’s premises for service or repairs to be carried out under warranty or otherwise. Where
appropriately qualified technical personnel are identified by a user, and ratified by Scan Optics, then
Scan Optics will make available on request any information which will assist in repairing the
equipment.
SO-5800 Ophthalmic Microscope Assembly Manual Page 3of 25
Issue Number 2.0 Reference SO-5800-AM-2

Figure A-1: SO-5800 Microscope
SO-5800 Ophthalmic Microscope Assembly Manual Page 4of 25
Issue Number 2.0 Reference SO-5800-AM-2
MICROSCOPE
HEAD ASSEMBLY
FLOOR
STAND
MONITOR
FOOT CONTROL
X-Y POSITIONING
UNIT
POWER
SUPPLY
PANTOGRAPH
ARM
HORIZONTAL
ARM
LOWER PANEL
PILLAR

PACKING LIST
Assembly Manual (this manual)
User Manual
Pillar and horizontal arm assembly, with mounting screws (M10x25) x 4
(includes power supply)
Pantograph arm assembly
(includes main power cable and video signal cable)
Suspension bracket (with mounting screws) with attached
spare lamp module
Monitor arm assembly
Floor stand legs (2 interlocking)
Floor stand joiners (top and bottom), with 4 mounting screws
Box 1: Microscope head assembly
Box 2: X-Y unit with top bearing and knob
Box 3: Monitor with mounting bracket in original packaging
including: Box Video Accessories including:
Monitor power cable
Monitor remote control
Video camera assembly
Video camera power cable
Box 4:Binocular assistant microscope assembly
Box 5:Accessories:
Mains cable
Set sterilisable knob covers
(6x ‘zoom’, 2 x ‘focus’, 6 x ‘handle’)
Microscope cover
Tool box including:
eyepieces x 2, lens cloth
Set socket keys: 9 total; 1.5 - 10 mm
Arm bearing mounts x 2 with 8 mounting screws
and friction washer
Elbow lock knob
Microscope side handles with locking pins x2
Microscope front handle with mounting screw
Box 6: Foot control
SO-5800 Ophthalmic Microscope Assembly Manual Page 5of 25
Issue Number 2.0 Reference SO-5800-AM-2

MECHANICAL ASSEMBLY
NOTE: MINIMUM 2 PERSONS REQUIRED
Floor stand base and pillar: refer figures A-2 to A-5
1. Remove the floor stand base from the crate
2. Place the top joiner upside-down on the floor and insert the upper and lower
interlocking legs as shown
3. Place the bottom joiner over the crossed legs and fix the assembly together
using the four M8x70 socket head cap screws and 6 mm socket key
provided. Tighten the screws alternately to ensure even pressure is applied
and that the assembly is seated square and true.
4. The floor stand base is now assembled but upside-down. Place the floor
stand base the right way up on the floor and lock the two wheel stops by
applying foot pressure.
5. Unpack the pillar/horizontal arm assembly from the crate and remove the four
M10 x 25 socket head cap screws and washers from the pillar legs.
6. Insert the pillar pins into the floor stand base holes so that the front panel
faces out in between the legs with wheel locks.Unlock the horizontal arm
and rotate it until it is in line with the front panel. Then lock it in to place
again.
7. Take care with the next step:
Use the side handles to carefully lean the pillar and base back, then carefully
chock up the elevated end. While one person holds the pillar flush against
the base, the other can insert the four M10 x 25 socket head cap screws and
washers from underneath the floor stand base assembly into the threaded
holes in the end of the pillar pins. Tighten alternately with the 8 mm socket
key until secure. The entire assembly may now be placed back upright on the
floor.
SO-5800 Ophthalmic Microscope Assembly Manual Page 6of 25
Issue Number 2.0 Reference SO-5800-AM-2

Figure A-2: Assembling the floor stand base
Figure A-3: Assembled floor stand base
SO-5800 Ophthalmic Microscope Assembly Manual Page 7of 25
Issue Number 2.0 Reference SO-5800-AM-2
TOP JOINER
BOTTOM JOINER
MOUNTING SCREWS
LOWER LEG
UPPER LEG

Figure A-4: Inserting the pillar in the floor stand base
Figure A-5: Fixing the pillar to the floor stand base
SO-5800 Ophthalmic Microscope Assembly Manual Page 8of 25
Issue Number 2.0 Reference SO-5800-AM-2

Pantograph arm: refer figures A-6 and A-7
1. Identify and remove the arm bearing mounts from the box.
Remove the pantograph arm assembly from the crate. Place the friction
washer over the top exposed shaft at one end of the pantograph arm. The
top side may be identified as the plain side; the bottom side features the
sliding knob safety stop assembly.
Place the arm bearing mounts in their correct orientation over the ends of the
shaft on the pantograph arm. Note that the lower arm bearing mount must be
supported by hand from underneath to prevent it dropping down.
With the arm bearing mounts in place insert the pantograph arm in to the end
of the horizontal arm assembly. Note that some shuffling movement will help
to move the arm in to place. When the arm bearing mounts are flush against
the horizontal arm, insert the 8 M6 x 16 countersunk screws and tighten them
alternately using the 4 mm socket key.
2. Identify and remove the elbow lock knob from the box. Screw the knob in
underneath the end of the horizontal arm.
SO-5800 Ophthalmic Microscope Assembly Manual Page 9of 25
Issue Number 2.0 Reference SO-5800-AM-2

Figure A-6: Placing the friction washer and arm bearing mounts
Figure A-7: Attaching the pantograph arm
SO-5800 Ophthalmic Microscope Assembly Manual Page 10 of 25
Issue Number 2.0 Reference SO-5800-AM-2
Other manuals for SO-5800
1
Table of contents
Other Scan Optics Microscope manuals
Popular Microscope manuals by other brands

VWR
VWR VisiScope 384 Series instruction manual

Nikon
Nikon ECLIPSE E200 POL instructions

Leica
Leica DI C800 User's manual & installation instructions

ThermoFisher Scientific
ThermoFisher Scientific Continuµm manual

ThermoFisher Scientific
ThermoFisher Scientific Continuµm manual

Olympus
Olympus SZ61 instructions