4.5.3 Set check weighing mode
Press F key to enter setting mode, press TARE until display show “F4 OFF”, press
ZERO key to enter, press TARE key until display show “BEEP”, press ZERO key to
enter, press TARE key to select BP 2 check mode 2), BP3 check mode 3), BP1 not
sound), press ZERO key to sure, press Print/M+ key to escape.
4.5.4 NOTE
The weight must be greater than 20 scale divisions for the check-weighing to
operate.
To disable the Check-Weighing function enter zero into both limits by pressing the
FUNC key when the current limits are shown then pressing ZERO/ENTER to store
the zero values.
4.6 Accumulated Total
The scale can be set to accumulate manually by pressing the PRINT key. See the
PARAMETERS Section for details of selecting the method using function "F5 P RT".
The accumulation function is only available when weighing. It is disabled during
parts counting.
The weight displayed will be stored in memory when the PRINT key is pressed and
the weight is stable.
The display will show "ACC 1" and then the total in memory for 2 seconds before
returning to normal. If the optional RS-232 interface is installed the weight will be
output to a printer or PC this type machine has no print function).
Remove the weight, allowing the scale to return to zero and put a second weight on.
Press the PRINT key, the display will show "ACC 2" and then the new total.
Continue until all weights have been added.
To view the totals in memory press enter the PARAMETER SECTION and use
function "F1 TOL". See below.
4.7 Animal Scales
KW can set as an animal scale when external resolution less than 1/3000, about how
to set it, see the technical manual of KW.
Let the animal on the platform, after some second, if reading data change between
the range you have set, you can hear beep sound and reading data will be locked,
after you let the animal leave the platform and reading data return to zero, KW will do
automatic accumulation operation, if you have connect with mini printer, KW will print
automatically.
To clear accumulation memory, press Print key when scale in ZERO point, this
function only available in animal scale mode.