1.1. Safety and general warnings
The user of this device is obliged to become familiar with this manual before starting to use
the device and is further obliged to proceed according to this manual.
If the device is used contrary to chapter 1.1 Safety and general warnings and the user does
not follow the instructions in this chapter, it may result in injury to the operator, damage to
the device, or the loss of warranty.
Terms of use
The device must be placed on a stable base, connected together with a PC, and for normal use the
following principles must be complied with:
The device can be used in indoor areas for environmental pollution degree 2, and only for the
purposes referred to in the instructions in 1 Introduction.
Protect the device from hits, falls, vibration and shakes.
Do not place samples, agents or any liquids on the surface of the device.
The normal temperature of the environment for using the device is 15 - 40
∘
C, at a maximum
relative humidity of 80%.
The minimum working space (without PC):
–Dimensions (width, height, depth) 250 x 230 x 430 mm.
–The minimum load capacity of the table is 25 kg (the weight of the device: 17 kg).
–
The space in front of the doors must be free of any barriers that could prevent the ejection
of the tray.
–
The air vents and the fan must remain clear, otherwise the service life of the product may
be significantly reduced. The ventilation openings are located on the sides of the device
(the distance from the surrounding barriers must not be less than 1 cm). The fan is located
in the rear part of the device (its distance from surrounding barriers must not be less than
10 cm).
The device can be used at transient overvoltage up to the category II overvoltage level and during
temporary overvoltage occurring in the power feed from the grid.
sciREADER CL2
device is designed in accordance with relevant legal regulation for electromagnetic
compatibility (EMC).
Only supported formats of input samples can be used - standard SBS microtiter plate format,
slide holder: 128 x 86 mm.
Samples must be handled according to the principles of handling of biological material and with
increased caution.
Use only original accessories provided by the manufacturer.
In case of any uncertainties during the use of the device, contact the service department or
support (see contacts below).
Attention
The device must be placed on a firm base.
Scienion AG, 2019, Original user manual, sciREADER CL2 5