Sensortronic scales BW-6 User manual

9007-B0N-0033-0
2006.06
BW - 1N
Battery Weighing Indicator
OWNER’S MANUAL

CONTENTS
CAUTIONS 4
INTRODUCTION 6
FEATURES & MAIN FUNCTIONS 6
SPECCIFICATION 7
PART NAME & KEY FUNCTION 8
SYSTEM MODE 11
GENERAL FUNCTION & DESCRIPTION 16
CONVERSION MODE 18
PRINT FORMAT 20
RS-232C COMMUNICATION 21
USING & BATTERY 23
HOW TO INSTALL THE BW 25
TEST MODE 26
ERROR MESSAGE DESCRIPTION & MANAGMENT 30

5
4
CAUTIONS
Do not press the key strongly Do not use the inflammable
at cleaning
Do not install in the place with
sudden temperature change
Do not install in the place with
excessive electrical noises and
shaking
Charge a battery in full to use
after negligence for a long time
Install the platform on a flat
place Never connect AC adaptor
while using the dry battery
Do not install in the place with voltage and terrible electrical noises
Do not use in the place exposed
directly to the sun and including
much dust

7
6
INTRODUCTION SPECIFICATION
We greatly appreciate for you to purchase BW of CAS.
This goods has the excellent performance and splendid properties through
strict test under severe quality management.
It is recommended to read this manual in full before using BW for good
function application.
FEATURES
Simple outdoor using in use of battery
Using general (Mn, Alkaline) battery or rechargeable battery (NiCd)
Easy operation and various function
SPACTM‚ single pass automatic span calibration
Water proof (IP65)
High precise resolution
Liquid crystal display of 6 place weight (LCD)
Using long-time battery
MAIN FUNCTIONS
Hold Function
COUNT Function (Unit weight setting, unit weight setting by sample)
Low / High Weight Comparing
Serial Communication Interface (RS-232)
Serial Printer Connection
Self Hardware Test
Independent Zero Calibration
Gravity Compensation
BW-1N
BW-6 BW-15 BW-30 BW-60 BW-150
6kg 15kg 30kg 60kg 150kg
2g 5g 10g 20g 50g
280(W) 280(D)
80(H)
405(W) 525(D) 105(H)
280(W) 350(D)
530(H)
425(W) 675(D) 755(H)
About 6 kg About 17 kg
About 6 hr
About 30 hr (BL ON),
About 60 hr (BL OFF).
ON/OFF, Zero, Net, Hold, PCS, RS232, HI, LO, Compare
Backlight
-10 ~+40
Max Capacity
6 Digit LCD(24mm Height)
Stable( ),Zero( ),Tare( ), Net ( ),Hold ( ),
Low battery ( ), Weight unit ( kg ),
High, OK, Low ( , , ),Count( )
1. DC 7.2V(6EA SIZE " C " ) 2500mAh Rechargeable battery
2. DC 9V(6EA SIZE " C " ) Dry Battery
3. DC 12V 850mA AC Adaptor
approx. 0.25W
Max Capacity
Interval
Platter Size
Product Size
Product Weigh
Charge Time
Discharge
Function
Operatint Temp
Tare
Display
Power (Option)
Power consumption
Model
Item
General
Option

9
8
PART NAME & KEY FUNCTION
LAMP SYMBOL DESCRIPTION
STABLE ON when current weight is stable
ZERO ON when current weight is 0
NET ON at displaying the pure weight
HOLD ON at weighing the moving object
TARE ON when tare is activated.
HI ON when current weight is more than high limit value.
OK ON when current weight is right weight band.
LOW ON when current weight is less than low limit value.
BAT Display the battery recharging time (rechargeable battery)
Display the battery replacing time (dry battery)
- about 6.9V remain
PCS ON when current mode is count mode.
CHARGE When charging battery, charging lamp (RED) is ON.
When it is completed, charging lamp (GREEN) is ON.
POWER ON at connecting the DC 12V adapter.
Minus ON when weight is -
Kg Weight unit
1. Part Name
2. DISPLAY
LCD DISPLAY
POWER Connector
RS-232 Connector
LOWER COVER
KEY
UPPER COVER

11
10
NET / Use to display NET weight or GROSS weight.
GROSS Enter the TEST MODE by pressing for 2 seconds
HOLD Use to weight the moving object (Manual, Automatic)
U.W. Use to display the unit weight (for approx. 5 seconds)
W. Use to display the current weight (for approx. 5 seconds)
SYMBOL DESCRIPTION
ON / OFF Use to power ON / OFF (Press for approx 3 seconds)
ZERO / Use in case of rezero (within 2% of max. capacity),
MODE To enter the system mode by pressing for 2 seconds
TARE Use at weighing in use of TARE
Enter the TIME SETTING MODE by pressing for 2 seconds
Use to turn ON / OFF Back Light (Option) or
*to print out in the case that F 2 is set to 4.
Enter the CONVERSION MODE by pressing for 2 seconds.
3. KEY Part
Hi/Low Comparing Mode key or Weighing Mode Key
Counting Mode Key
SYSTEM MODE
Simple weight mode
Display the weight after weighing.
High / Low comparing mode
Determine that the weight is high, normal or low value and perform the display.
Count mode
Display the quantity by setting the unit or sample weight.
Quantity comparing mode
Perform the display after discrimination by setting High / Low quantity.
Press Zero-Mode key for 2~3 sec. to display the following message.
Execute depending on the below mode conversion and setting method.
1. MODE CONVERSION
Input the digit in use of the below keys.
key : Use to set the input value to 0
key : Use to increase the first place of value by on.
key : Use to move the current place to the left by one place.
key : Use to move the current place to the right by one place.
key : Use to complete the input.
2. DIGIT INPUT METHOD
STAR
G/N
TARE
ENTER
MODE
H
M

13
12
In case of no resetting the High / Low value
3. SIMPLE WEIGHING MODE CONVERSION
Display KEY
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
STEP 6
Ex) For 20.5 input
In case of resetting the High / Low value
4. HIGH / LOW COMPARING MODE CONVERSION
DESCRIPTION
Press once
(Set to 0 )
Press five times.
Press once (Move to the left
by one place)
Press once
Press twice times (input 2
as the 3th place value)
Input completion
DESCRIPTION
Press once
Press once
Generally operated
Display KEY PLATFORM
STEP 1
STEP 2
Display KEY PLATFORM
STEP 1
STEP 2
DESCRIPTION
Press twice
Press once.
Generally operated
Display KEY PLATFORM
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
STEP 6
DESCRIPTION
Press twice
Press once
(Low weight input)
Refer to 'Digit input'
In page 11
High value input
Refer to 'Digit input'
In page 11
Generally operated
Empty
INPUT
INPUT
Empty
Empty
M
H
H
M
M
M
In case of setting the unit weight
(Min. range of unit weight ; Min. Division)
Step 5. Ex) Min. division is 0.1
Input the unit weight as 0.532 and round off the second place of decimal to 0.5
In case of no setting the unit / sample weight
5. COUNT MODE CONVERSION
Display KEY PLATFORM
STEP 1
STEP 2
DESCRIPTION
Press three times
Press once
Generally operated
Display KEY PLATFORM
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
DESCRIPTION
Press three times
Press once
Display the set weight
Input the unit weight
Digit input in page 11
Press 'Mode' key to round &
display unit weight.
Generally operated
HEmpty
M
M
M

15
14
In case of setting the unit weight by sample
(Samples : 10 - 200 , 10 unit)
Note : For setting the unit weight by sample count, the min. unit weight
shall be more than min. division
In case of no setting the High / Low quantity
6. COUNT COMPARING MODE CONVERSION
Display KEY PLATFORM
STEP 1
STEP 2
DESCRIPTION
Press four times
Press once
Generally operated
Display KEY PLATFORM
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
STEP 6
STEP 7
STEP 8
STEP 9
STEP 10
DESCRIPTION
Press three times
Press once
Press once
Press once
Press five times.
Simple count input(50pcs)
Press once [ENTER]
Press once with empty
[ENTER] key
Unit weight display.
Press once
Generally operated
In case of setting the High / Low quantity
Display KEY PLATFORM
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
STEP 6
DESCRIPTION
Press four times
Press once.
Low value setting
Low value input
Digit input in page 11
Press once
High value setting
High value input
Digit input in page 11
Press once
Generally operated
H
H
H
H
Empty
Empty
Empty
Empty
Sample
M
M
M
M
M

17
16
Zero point is changed. Press 'Zero' key. Zero lamp is ON.
GENERAL FUNCTION & DESCRIPTION
Rezero band range : Within 2% of Maximum capacity
1. REZERO FUNCTIOIN (Use at change zero point)
Put the TARE on platform0
(TARE weight : 10kg) Press 'TARE' key.
(The TARE weight is saved) Put the object on platform
(object weight : 20kg)
Maximum TARE range : Maximum capacity
Press 'NET/GROSS' key
(object + TARE weight display)
In case to want know the Gross weight
Press 'NET/GROSS' key (object weight display)
Remove the TARE and object form platform to display the saved TARE weight
In case to want know the NET weight
Remove the TARE and object from load cell(up figure) and press TARE key(weight is 0)
In case to remove the Tare weight
Note : The weight including TARE weight can't exceed the maximum capacity
2. TARE FUNCTIOIN (Use at weighing in use of TARE)
Auto hold function (Hold function operated whenever weighing)
3. HOLD FUNCTIOIN (Use at weighing the moving object)
When the weight change is
stable after putting the
Object(20kg) on platform,
'Hold' message is displayed
The weight is averaged for
approx. 3~5 sec.
Press 'Hold' key
Auto hold release message is shown for 1 sec.
And then normally operated. 'Hold' lamp is OFF.
The hold weight is
displayed.
Put the object (10kg) on
platform Press 'Hold' key.
'Hold' message is displayed
for 1 sec.
The weight is averaged for
displayed.
3-5 sec.
Empty the platform Press 'Hold' key
'Auto hold' message is
displayed for 1 sec.
Hold lamp is ON
Empty the platform
To release the hold weight Empty the platform or press 'Hold' key and
return to the normal operation
Auto hold function release
Manual hold function (Executed at pressing 'Hold' key)
The hold weight is displayed. Empty the platform or press 'Hold' key
to release the hold weight and perform the normal operation.
'Hold' lamp is OFF
Hold start is more than 5.5 division and 10% of maximum capacity
after displayed hold weight.
If change weight is more than 10% of maximum capacity after displayed hold
weight, hold function is cancelled

19
18
CONVERSION MODE
1. ENTER method
If you press the *key for three seconds to normal operation state, after display "SET"
Message for one second and then enter.
2. KEY using
Move to the next menu.
Increase the set point.
Return to the normal operation at completion of setting :[ENTER] key.
3. CONVERSION menu (F01 - F04)
F - 1 : Auto power OFF setting.
F - 2 : RS-232C communication setting.
F - 3 : Select print format type.
F - 4: Stable condition
Set point Meaning
F – 1 0 Auto power OFF is not used
F – 1 1~9 Automatically power off when the weight is not change or no key is used
for 1 ~ 9 minutes
F – 1 : Auto power OFF setting
Auto power OFF function is used to spare the battery by means of power off
when this scale is not used.
Set point Meaning
F – 2 0 No transmit
F – 2 1 -
F – 2 2 Transmit at stable weight
F – 2 3 Transmit at data request. (Data request signal “D”)
F – 2 4 Use for print mode
F – 2 : RS-232C communication setting
Set point Meaning
F – 3 0 Select print format 1
F – 3 1 Select print format 2
F – 3 2 Select print format 1 + Current Time
F – 3 3 Select print format 2 + Current Time
F – 3 : Select print format type
H
Set point Meaning
F – 4 1 ~ 9 Stable level 1~ 9
F – 4 : Stable Condition

21
20
Print Format
No. 001 : 1.00 kg
< Format 1 >
2005 03. 17 11:23:46
No. 001 : 1.00 kg
< Format 1 + Time >
001 : Net 1.00 kg
Tare 0.00 kg
Gross 1.00 kg
< Format 2 >
2005 03. 17 11:23:46
001 : Net 1.00 kg
Tare 0.00 kg
Gross 1.00 kg
< Format 2 + Time >
Weight Mode
Printedweightis net weight.
No. 001 : 1.00 pcs
< Format 1 >
2005 03. 17 11:23:46 .
No. 001 : 1.00 pcs
< Format 1 + Time >
001 : Net 1.00 pcs
Tare 0.00 pcs
Gross 1.00 pcs
< Format 2 >
2005 03. 17 11:23:46
001 : Net 1.00 pcs
Tare 0.00 pcs
Gross 1.00 pcs
< Format 2 + Time >
PCS Mode
Printedcountis net count.
RS-232C communication
1. Port connection method
Connect the computer serial port with RS-232C port in the back panel
of indicator as follows to communicate with computer
2. RS-232C Output data format
Baud rate : 9600bps
Data bit : 8, Stop bit : 1, Parity bit : None
Code : ASCII
(2)TXD
(1)RXD
(5)GND
3 Receive Data
2 Transmit Data
7 Signal Ground
4 Request to Send
5 Clear to Send
6 Data Set Ready
8 Carrier Detect
20 Data Terminal Ready
CN 1 of
Indicator
( male )
Computer serial 25 pin port(female)
(2)TXD
(1)RXD
(5)GND
2 Receive Data
3 Transmit Data
5 Signal Ground
1 Carrier Detect
4 Data Terminal Ready
6 Data Set Ready
7 Request to Send
8 Clear to Send
CN 1 of
Indicator
( male )
Computer serial 9 pin port (female)

23
22
3. Transmit data format
Total 22 Bytes
Lamp status byte : Display the current state (ON/OFF)of indicator lamp (Low Enable)
Data(8 bytes) K G CR LF
US (Unstable) GS(Gross W.)
ST(Stable) NT(Net W.)
OL(Overload)
SOH em
pty
Lamp
Status
Byte Weight
Unit
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
1 STABLE HIGH LOW HOLD GROSS TARE ZERO
Data (8 bytes)
Weight data including a decimal point and _. That is, for -13.5kg :
Transmit ASCII code 8 bytes
Corresponding to _ , , , , 1 , 3 , 5 ,
Data request mode
If F 2 is set to 3, the indicator outputs the specified data when ASCII
code corresponding to D is sent to it.
Warning
Open the cover and check the battery type.
If the battery is not the rechargeable battery (this is : Mn, Alkaline),
Never connect the AC adaptor jack in the back panel
BATTERY
1. Battery replacement or charging time
Power is off in a specific time after lamp is on in the right top of display part.
Thebatterylamp is onwhentheamount of battery isabout6.9V and powerisoff
When the amount of battery is about 6.5V automatically.
When the lamp is on, replace the dry or charge the rechargeable battery.
2. General (Mn, Alkaline) battery using and replacing
Open the top cover and make as follows in the left bottom of cover at using the
general (Mn, Alkaline) battery.
Insert the battery into the battery case after checking the battery polarity.

25
24
3. Rechargeable battery using and recharging
Open the top cover and make as follows in the left bottom of cover at
using the rechargeable battery.
Open the jack protecting cap in the back panel and connect the adapter
as follows.
When the adapter is connected with indicator, RED(POWER) and
RED(CHARGE) lamp are ON.
When the charging is completed, CHARGE lamp is changed to GREEN.
It takes approximately 4-5 hours to charge in full.
Battery using time can change according to battery using period or loadcell number.
4. Battery using period
Conditions
BACK LIGHT(E.L) is not
used
BACK LIGHT(E.L) is always
used
Using Period
Recharging Battery : Approx. 70 hours
Mn Battery : Approx. 100 hours
Alkaline Battery : Approx. 250 hours
Recharging Battery : Approx. 30 hours
Mn Battery : Approx. 45 hours
Alkaline Battery : Approx. 90 hours
HOW TO INSTALL THE BW
1. Open the box with care because indicator is connected to the scale with
load cell wire.
2. Turn the post knob so as to fix indicator.
3. Pull down the wire out of post pipe and insert the post pipe to the post
bracket. (Refer to fig. 1)
4. Fasten the post pipe with two bolts. (Refer to fig. 1)
5, A long bolt should be connected to the upper side.
6. Insert the wire to the post pipe. (Refer to fig. 2, 3)
7. If the scale is not properly level, please adjust 4 leg(adjusting bolt) at the
bottom of the scale so as to center the bubble of the leveling gauge.
Note : Place the scale on a flat and stable surface.
Inside the indicated circle.

27
26
How to enter test mode
1. Open the top cover of indicator and power on with pressing CAL switch in
the left top. At that time, LCD screen shows ZEro = tESt, tArE = CAL
message and pressing Zero key moves to tESt 1.
2.Press G/N key for approx 3 seconds.
Test menu (tESt 1~7)
TEST 1 : Liquid Crystal Display test
TEST 2 : Load cell and A/D conversion test
TEST 3 : Key test
TEST 4 : RS-232C communication test
TEST 5 : Temperature test
TEST 6 : Battery test
TEST 7 : RTC test
TEST 1(Enter ' TEST ' mode to start TEST 1)
FUNCTION : LCD TEST
KEY
[ ENTER Key]
Press once :
Test running
One more time :
Next menu
[ * Key] : Back light test
LCD SCREEN DESCRIPTION
Display TEST 1 state
TEST MODE
TEST 2
FUNCTION : A/D Conversion TEST
KEY
[ ENTER Key]
Press once :
Test running
One more time :
Next menu
LCD SCREEN DESCRIPTION
Display TEST 2 state
Display the digital value
Corresponding to current weight
TEST 3
FUNCTION : Key test
KEY
[ ENTER Key]
Press once :
Test running
One more time :
Next menu
TEST 4
FUNCTION : RS-232C communication test (Computer link test)
KEY
[ ENTER Key]
Press once :
Test running
One more time :
Next menu
Key name code Key name code
1 ZERO 2
TARE 3 GROSS /NET 4
LCD SCREEN DESCRIPTION
Display TEST 3 state
Pressing the key to test shows
the key code.
LCD SCREEN DESCRIPTION
Display TEST 4 state
Receive : " 1 " Transmit : " 2 "
Reference 1. Connect the computer COM port with RS-232C connector of indicator
before running this test. (Possible only at installing RS-232C option)
Reference 2. Press " 1 " in computer keyboard and check that " 1 " is shown in LCD
screen of indicator and Check that the next ASCII code value, " 2 " is
inputted in the computer.

29
28
TEST 5
FUNCTION : Temperature test
KEY
[ ENTER Key]
Press once :
Test running
One more time :
Next menu
LCD SCREEN DESCRIPTION
Display TEST 5 state
for temperature
Current temperature is shown
TEST 6
FUNCTION : Battery test
KEY
[ ENTER Key]
Press once :
Test running
LCD SCREEN DESCRIPTION
Display TEST 6 state
for battery voltage value
Current battery voltage is shown.
TEST 7
FUNCTION : RTC test
KEY
[ ENTER Key]
Press once :
Can see the current
date & time
Press twice :
Exit Test Mode
[ * Key ]
Can see date & time
Alternately
LCD SCREEN DESCRIPTION
Display TEST 7 state
for current date & time
Current date & time setting method
In the normal mode,
Press the TARE Key long and can see the message and .
After this step, You can input the current date and time as follows
How to set the current date and time
KEY
[TARE] Key :
Shift one digit
to the left
[STAR] Key :
Shift one digit
to the right
[ZERO/MODE]Key :
Reset the input
value
[ N / G] Key :
Increase input
value
[ENTER] Key :
Exitthisinputmode&
Go to normal mode
LCD SCREEN DESCRIPTION
Date & time setting mode
Initial value
& : increase 5 [Year : 2005]
& : increase 2 [Month : 02]
& : increase 1
& : increase 7 [Day : 17]
& : increase 1
& : increase 8 [Hour : 18]
& : increase 1
& : increase 3 [Minute]
2005/02/17
18:13
H
If you input wrong date, this scale display "out" message and doesn't save the date.

31
30
ErrorsinWeighingMode
Err 01
Cause
The Indicator can't be initialized due to platform shaking.
Management
Put the platform on a flat place without shaking and power on.
Err 02
Cause
The load cell is connected wrongly or A/D conversion part has a fault.
Management
Check that the load cell is well connected with main unit.
Err 03
Cause
The calibration isn't executed or the data of internal memory is lost due to
a electrical shock.
Management
Perform the required specification again in setting mode.
Err 09 (Zero Key)
Cause
The current weight exceeds zero range.
Err 10
Cause
The specified TARE weight exceeds the maximum capacity.
Management
Set the TARE weight lower than maximum capacity.
Or set the maximum capacity higher than the desired TARE weight in
calibration menu and perform the calibration again.
ERROR MESSAGE DESCRIPTION & MANAGEMENT Err 13 (Zero Range with power on)
Cause
The initial zero range exceeds above 10% of maximum capacity.
Management
Check the load cell state.
Over
Cause
Too heavy object is put on platform so that it may exceed the
maximum capacity.
Management
Never put the weight exceeding the maximum capacity.
If the load cell is damaged, replace the load cell
ErrorsinCountMode
Err 61
Cause
The count unit weight and sample weight are set too high.
Management
Set the count unit weight low or the count sample quantity high.
Err 62
Cause
The count unit weight and sample weight are set too low.
Management
Set the count unit weight high or the count sample quantity low.

33
32
MEMO MEMO

34
MEMO