Simmons 765000S User manual

765000S
CLINICALPLUSALTERNATING
PRESSUREMATTRESSSYSTEM
WITHACTIVESENSORTECHNOLOGY
USERMANUAL
Important:DonotoperatetheMattressSystem
without rst reading and understanding this manual!
Save this manual for future use.
Info: The most current version of this manual can be
found online at www.grahameld.com.
765000S-INS-LAB-RevC14
Not made with natural rubber latex.

2 765000S-INS-LAB-RevC14
CONTENTS
INTRODUCTION...................................................................................................3
INTENDED USE OF THIS DEVICE...............................................................3
CONTRAINDICATION............................................................................ 3
IMPORTANT SAFETY PRECAUTIONS ...............................................................4
DANGER........................................................................................................5
WARNING......................................................................................................6
NOTICE.......................................................................................................... 7
EQUIPMENT SYMBOLS......................................................................................8
PRODUCT DESCRIPTION...................................................................................9
PUMP AND MATTRESS SYSTEM................................................................9
PUMP: FRONT ............................................................................................10
PUMP: REAR............................................................................................... 10
PUMP: FRONT PANEL................................................................................11
INSTALLATION...................................................................................................15
UNPACKING................................................................................................ 15
PUMP AND MATTRESS SETUP ................................................................16
OPERATION .......................................................................................................18
GETTING STARTED....................................................................................18
QUICK DISCONNECT.................................................................................18
TRANSPORT CAP.......................................................................................19
CPR FUNCTION.......................................................................................... 19
MAINTENANCE..................................................................................................20
CLEANING................................................................................................... 20
GENERAL MAINTENANCE ........................................................................20
FUSE REPLACEMENT ........................................................................21
AIR FILTER REPLACEMENT............................................................... 21
STORAGE....................................................................................................21
TROUBLESHOOTING........................................................................................22
SERVICE......................................................................................................22
SPECIFICATIONS............................................................................................... 23
ONE (1) YEAR LIMITED WARRANTY ...............................................................24
APPENDIX A: EMC INFORMATION..................................................................25
SIMMONS Clinical Care products manufactured by GF Health Products, Inc.
under license from the trademark owner, Dreamwell, Ltd., a subsidiary of
Simmons Bedding Company, Atlanta, Georgia. SIMMONS is a trademark of
Dreamwell, Ltd. a subsidiary of Simmons Bedding Company.
© 2013 Dreamwell, Ltd. All rights reserved.
Graham-Field is a registered trademark of GF Health Products, Inc.
GF Health Products, Inc. is not responsible for typographical errors. All
illustrations, specications, packaging and warranties contained in this
catalog are based on the latest product information available at the time
of printing. The most current product information can be found online at
www.grahameld.com.

765000S-INS-LAB-RevC14 3
INTRODUCTION
Use this manual for set-up and operation of the
SIMMONS™ Clinical Care 765000S Clinical Plus Alternating
Pressure Mattress System with Active Sensor Technology.
Read all instructions before using the mattress system.
Save this manual for future reference.
INTENDED USE OF THIS DEVICE
The intended use of the SIMMONS™ Clinical Care 765000S
Clinical Plus Alternating Pressure Mattress System with
Active Sensor Technology is:
• Aid in the treatment of pressure ulcers while optimizing
user comfort.
• Pain management as prescribed by a physician.
• As described above, in either a homecare or long-term
care setting.
Contraindication
WARNING: DO NOT use this product in the presence of
ammable anesthesia. There is a possible re hazard
when this product is used with certain oxygen delivery
equipment.
Use nasal cannula, face mask, or 1/2 length oxygen tent
to deliver oxygen in the presence of this product.
DO NOT use a full length oxygen tent that extends past
the top surface of the mattress with this product.

4 765000S-INS-LAB-RevC14
IMPORTANT SAFETY PRECAUTIONS
The safety statements presented in this chapter refer to
the basic safety information that should be observed by
those using this Mattress System. There are additional
safety statements in other chapters or sections, which may
be the same as or similar to the following, or specic to the
operations.
DANGER: Indicates an imminent hazard situation that,
if not avoided, will result in death or serious injury.
WARNING: Indicates a potential hazard situation or
unsafe practice that, if not avoided, could result in
death or serious injury.
CAUTION: Indicates a potential hazard or unsafe
practice that, if not avoided, could result in minor or
moderate personal injury.
s NOTICE: Indicates a potential hazard or unsafe practice
that, if not avoided, could result in product or property
damage.
Info: Provides application recommendations or other useful
information to ensure that you get the most from your
product.

765000S-INS-LAB-RevC14 5
DANGER
To reduce the risk of electrocution:
DANGER: Always unplug this product immediately after
use.
DANGER: Do not use this product while bathing.
DANGER: Do not place or store this product where it
can fall or be pulled into a tub or sink.
DANGER: Do not place this product in or drop into
water or other liquid.
DANGER: Do not reach for a product that has fallen
into water. Unplug it immediately.

6 765000S-INS-LAB-RevC14
WARNING
To reduce the risk of burns, electrocution, re,
or personal injury:
WARNING: Use this product only as intended and
described in this manual. Do not use attachments or
accessories not recommended by Graham-Field.
WARNING: Do not leave this product unattended when
plugged in.
WARNING: Always use close supervision when this
product is used by, on, or near children or those who
require close supervision.
WARNING: Never operate this product if:
a) it has a damaged cord or plug,
b) it is not working properly,
c) it has been dropped or damaged, or
d) it has been dropped into water.
Return the product to your Graham-Field authorized
distributor.
WARNING: Keep the cord away from heated surfaces.
WARNING: Never block the air openings of this product
or place it on a soft surface, such as a bed or couch,
where the openings may be blocked. Keep the air
opening free of lint, hair, and other similar particles.
WARNING: Never drop or insert any object into any
opening or hose.
WARNING: Connect this product only to a properly
grounded power outlet.
WARNING: Maximum patient weight capacity for this
product is 400 lb (189 kg), EVENLY DISTRIBUTED.
WARNING: Notice for California Customers- California
Proposition 65 WARNING: This product contains a
chemical known to the State of California to cause
cancer and reproductive or developmental harm.

765000S-INS-LAB-RevC14 7
WARNING: Patient entrapment with bed side rails
may cause injury or death. The bed frame and its
components, including the mattress, bed side rails,
head and foot board, bedding, and any accessories
added to the bed, can all affect the risk of entrapment.
Thorough patient assessment and monitoring are
necessary to reduce the risk of entrapment, including
establishing whether the use of a bed rail is in the
best interest of the patient. Read and understand
the User Manual before using this equipment.
Graham-Field product manuals are available online at
www.grahameld.com.
Visit the FDA's Bed Safety page at www.fda.gov to
learn more about the risks of entrapment. It is the
responsibility of the facility and provider to be in
compliance with these guidelines. Refer to user
manuals for beds and rails for additional product safety
information.
After any adjustment, repair or service, and before use,
ensure all attaching hardware is securely tightened.
Bed rails with dimensions different than the original
equipment specied by the bed manufacturer may not
be interchangeable and may result in entrapment or
other injury.
NOTICE
s NOTICE: The pump can be used only with the
accompanying mattress. Do not use it for any other
purpose.

8 765000S-INS-LAB-RevC14
EQUIPMENT SYMBOLS
Manufacturer
“BF”symbol: Indicates that this product is in accordance with the
degree of protection against electric shock for type BF equipment
Functional earth (for UL only)
Attention! Read the instructions!
Disposal of Electrical & Electronic Equipment (WEEE):
Do not treat this product as household waste. For more detailed
information with regard to returning and recycling this product,
please consult your local city ofce, household waste disposal
service, or Graham-Field authorized distributor.
Consult operating instructions
Class II (for UL only)
The system has been tested and successfully approved with the following standards:
EN 60601-1, EN 60601-1-2, EN 550011 Class B, IEC61000-3-2,
IEC61000-3-3
For U.S. and Canada only
E305034
53DG
Medical Equipment: Air Pump
with respect to electrical shock, re and
mechanical hazards only in accordance with
UL60601-1 and CAN/CSA C22.2 No. 601.1

765000S-INS-LAB-RevC14 9
PRODUCT DESCRIPTION
PUMP AND MATTRESS SYSTEM
pumpandmattresssystem
CPRvalves
quickconnectors
(mattress-pump
connectors)
mattress
footend
symbol
pump

10 765000S-INS-LAB-RevC14
PUMP: FRONT
quickconnector
front
panel
PUMP: REAR
mainpower
switch
powercord
receptacle
fuse
hanger
filter

765000S-INS-LAB-RevC14 11
PUMP: FRONT PANEL
Front panel of pump is shown below. Descriptions of
buttons and indicators follow.
Tuningbutton;
indicatorsatleftofbutton
Powerfailureindicator
Alarmmutebutton
Serviceindicator
Power(ON/OFF)button;
indicatoratrightofbutton
Panellockbutton;
indicatoratrightofbutton
Maxfirmbutton
Auto-Adjustbutton;
indicatorsatrightofbutton
Lowpressure
indicator
Therapyadjustbutton;
indicatorsatrightofbutton
Front panel, rst row
Auto-Adjust button: The auto-adjust button
adjusts the mattress pressure automatically based on
the patient's weight. To remind the user that the auto-
adjust process is ongoing, the auto-adjust indicator LEDs
will periodically ash from left to right until the Ready
LED illuminates to show the process is complete. Three
conditions trigger the auto-adjust function:
1.
Press the auto-adjust button for two seconds; the mattress
system automatically readjusts pressure setting.
2. When initial mattress ination is complete, the mattress
system automatically starts the auto-adjust function.
3. When the mattress system detects a signicant change
in patient's weight on the mattress (e.g. patient egress
or ingress), the system automatically starts the auto-
adjust function.
When auto-adjust is complete, the Ready indicator will
illuminate. To disable the auto-adjust function, press
the auto-adjust button until auto-adjust indicators de-
illuminate.
Info:
During the auto-adjust operation, it is normal for the mattress
system to cycle through a series of inflation and deflation.

12 765000S-INS-LAB-RevC14
Tuning button: The tuning button ne-tunes the
mattress system's auto-adjusted pressure. Three tuning
settings are available:
0 System auto-adjusted pressure
- One level softer than auto-adjusted pressure
+ One level rmer than auto-adjusted pressure
Info: The tuning function is only available in alternate and
static therapy modes.
Front panel, second row
Therapy adjust button: The therapy adjust
button selects the therapy mode. After power-up, the
system automatically inates the mattress, auto-adjusts
the pressure, and enters alternate therapy mode. To then
cycle among the therapy settings, press the therapy adjust
button to select static therapy mode; press again to select
alternate + seat ination therapy mode; press again to
select static + seat ination therapy mode; press again to
select alternate therapy mode, etc.
Alternate: Each tube lls for 10 minutes, then deates for
10 minutes, alternately to the next tube. When selected,
the alternate indicator illuminates.
Static:
The pressure inside all tubes is adjusted to the same
pressure. When selected, the static indicator illuminates.
Alternate + Seat Inflation: For use when the head is raised,
to increase pressure under the buttocks — in alternating
pressure therapy mode. When selected, both the alternate
and seat ination indicators illuminate.
Static + Seat Inflation: For use when the head is raised, to
increase pressure under the buttocks — in static therapy
mode. When selected, both the static and seat ination
indicators illuminate.

765000S-INS-LAB-RevC14 13
Maxfirm button:
The pump will automatically enter
maxrm mode every time the power is turned on. Maxrm
mode ensures the pump is able to reach its maximum
operating pressure. Maxrm mode will last for 20 minutes,
and then the system will automatically enter auto-adjust
mode. To disable maxrm mode, press the maxrm button.
When enabled, the maxrm indicator, at left of button,
illuminates.
Front panel, third row
Panel lock button: The panel lock button sets the
panel lock to protect the panel settings from accidental
change. Press for two seconds to lock; press FIRMLY for
two seconds to unlock. After the panel has been untouched
for ve minutes, the panel lock locks automatically. When
the panel is locked, the panel lock indicator will illuminate.
Power (ON/OFF) button: The power button turns
power to the mattress system ON or OFF. Press to turn ON;
press to turn OFF. When power is on, the power (ON/OFF)
indicator will illuminate. When power is turned OFF, the
mattress system will slowly deate.
Info: Main power switch on side of pump, previously shown
in PUMP: REAR, must be ON in order for the front panel
power button to function.
Alarm Mute button: The alarm mute button
temporarily suspends the LED indicator and buzzer when
either the low-pressure alarm or power failure alarm is
activated. Should the problem not be resolved within three
minutes, the alarm will resume.

14 765000S-INS-LAB-RevC14
Low pressure indicator: The low pressure
indicator (yellow LED) illuminates, and the alarm sounds,
when an abnormally low pressure level occurs. If the low
pressure indicator persistently illuminates inappropriately:
1. Check to ensure all connections are properly and
securely connected per installation instructions.
2. Check for any leakage (tubes or connecting hoses).
If necessary, contact your Graham-Field authorized
distributor to replace any damaged tubes or hoses.
3. If problem persists, contact your Graham-Field
authorized distributor.
Info: Sometimes the low pressure indicator will illuminate
when a patient gets up to leave the bed. This is normal. The
indicator will turn off automatically after a short time.
The low pressure indicator will illuminate until the low-
pressure condition is resolved.
Power failure indicator: The power failure
indicator illuminates, and the alarm sounds, when there is
a power failure. Depress the alarm mute button to disable
both alarm and LED.
Service indicator: Mechanical failure, service
required; contact your Graham-Field authorized
distributor.

765000S-INS-LAB-RevC14 15
INSTALLATION
UNPACKING
1. Carefully remove all components from the carton.
2. Inspect all components.
3. If damage is evident to the contents, please notify the
carrier and your Graham-Field authorized distributor.

16 765000S-INS-LAB-RevC14
PUMP AND MATTRESS SETUP
1. Place the mattress on top of the
bed frame; ensure that the foot
end symbol faces up at foot end as
shown at right. Secure mattress to
bed with straps.
2. Fold open the wire hangers on
the back of the pump as shown at
right.
3. Either hang the pump on the bed's
foot end, as shown at right, or
place the pump on a at surface
easily accessible to the caregiver
and/or doctor.
s NOTICE: Place the device in a
position where the caregiver
and/or doctor can access it easily.
4. Connect the mattress-pump air
hose quick connector; ensure that
the connector has clicked securely
into place before continuing.
s NOTICE: Ensure the air hoses are
not kinked or tucked under the mattress.
WARNING: Locate all cords so that they will not be
stepped on, tripped over, or otherwise subjected to
damage or stress.
WARNING:
Do not use a household extension cord if the
electrical cord does not reach the power outlet. Use of an
improper extension cord could result in re and electric
shock. If an extension cord must be used, use a three
conductor cord with ground, properly wired, in good
electrical condition, and keep it as short as possible.
hanger

765000S-INS-LAB-RevC14 17
WARNING: Ensure that the local power voltage is
appropriate for the pump unit.
5. Plug the power cord into a properly grounded electrical
outlet.

18 765000S-INS-LAB-RevC14
OPERATION
GETTING STARTED
1. Turn the main power switch
to the ON position.
2. Press the front panel power
button. The power indicator
will illuminate.
Info: The mattress system will
automatically enter maxfirm mode for the quickest inflation
at set-up; when inflated, it will then automatically enter auto-
adjust mode to select the appropriate pressure.
3. The pump will begin delivering air into the mattress.
The low pressure indicator and the maxrm indicator
will illuminate for approximately twenty minutes until
the mattress is fully inated.
4. Upon ination, the system will automatically enter auto-
adjust mode to automatically select the appropriate
pressure according to the patient's weight. The auto-
adjust indicator LEDs will ash from left to right, and
the system will cycle through ination and deation,
until the Ready LED illuminates to show the process
is complete, and the mattress system automatically
enters alternate therapy mode.
5. Replace all sheets, blankets, and pillows on the bed.
6.
You can now adjust therapy modes or tuning. Please see
earlier section, PUMP: FRONT PANEL, for a complete
description of available adjustments, buttons and
indicators.
QUICK DISCONNECT
Pull the power plug from the wall connector to disconnect
the device quickly.
main
power
switch
powerbutton

765000S-INS-LAB-RevC14 19
TRANSPORT CAP
In case of power failure or
transport: Disconnect the quick
connector and cap the mattress
with the transport cap to slow
deation.
Info: This system is equipped
with cell-on-cell technology to
provide added support and prevent
bottoming out in the event of a
power failure. Lower cells will hold
air for approximately eight hours.
CPR FUNCTION
If CPR must be performed on the patient while the mattress
is in use, to immediately deate the mattress, either:
1. Pull the CPR valves from the
mattress in the location of the
arrow shown at right, or
2. Disconnect the quick connector
from the pump in the location of
the arrow shown at right.
quick connector
transport cap

20 765000S-INS-LAB-RevC14
MAINTENANCE
CLEANING
WARNING: To reduce the risk of increased bacterial
growth, infection, illness, or injury from contamination,
thoroughly clean and dry the mattress system before
use and as necessary during use as follows:
Pump: Wipe the pump with a clean, damp cloth and mild
detergent, and keep it away from dust. Air dry.
s NOTICE: Do not use phenolic products or corrosive or
powdered cleansers to clean the pump.
s NOTICE: Do not immerse or soak the pump.
Mattress: Wipe the mattress unit with a clean, damp cloth
and mild detergent. The mattress may also be cleaned
using a 10% solution of sodium hypochlorite (bleach)
diluted in water. Air dry all parts thoroughly before use.
s NOTICE: Do not use phenolic products to clean the
mattress.
s NOTICE: After cleaning, air dry the mattress without
direct exposure to sunlight.
GENERAL MAINTENANCE
1. Check main power cord for abrasion or excessive wear.
2. Check mattress cover for signs of wear or damage.
Ensure mattress cover and tubes are connected
correctly.
3. Check airow from the air hose connector. The airow
should alternate between each connector every half-
cycle time while in alternating mode.
4. Check the air hoses for any kink or break. For
replacement, please contact your Graham-Field
authorized distributor.
Table of contents
Other Simmons Indoor Furnishing manuals
Popular Indoor Furnishing manuals by other brands

American Standard
American Standard 3636.LCW Specifications

Mantion
Mantion SLID'UP 2500 Instructions for assembly

Wren Kitchens
Wren Kitchens Vogue WALL UNIT 700 Tall 2 Door Assembly guide

Hesperide
Hesperide 159197 manual

Edsal
Edsal EBC30 Quick assembly

Bob's Discount Furniture
Bob's Discount Furniture Montana Counter Storage Bench 20058856001 Assembly instructions

OSP Home Furnishings
OSP Home Furnishings MLL52 Assembly instructions

Bouclair
Bouclair 9399425 Assembly instructions

Ameriwood HOME
Ameriwood HOME Carver 5987196COM Instruction booklet

Safavieh
Safavieh Jayana SFV7510A quick start guide

De Toekomst
De Toekomst Twice user manual

BLACK RED WHITE
BLACK RED WHITE KENTAKI S320-LOZ/160 Assembly