Sirona inEos Blue User manual

MSKOMNM
áåbçë=_äìÉ
kÉï=~ë=çÑW==
båÖäáëÜ
léÉê~íáåÖ=fåëíêìÅíáçåë

62 99 205 D3446
2D3446.201.02.02.02 06.2010
Sirona Dental Systems GmbH
Table of contents Operating Instructions inEos Blue
Table of contents
1Dear Customer, ........................................................................................................ 5
2General information.................................................................................................. 6
2.1 Structure of the documents ........................................................................... 6
2.1.1 Identification of danger levels............................................................. 6
2.1.2 Formats and symbols used................................................................ 7
2.2 Legend .......................................................................................................... 7
2.3 Warranty........................................................................................................ 7
3Safety ....................................................................................................................... 8
3.1 Basic safety information ................................................................................ 8
3.1.1 Prerequisites ...................................................................................... 8
3.1.2 Connecting the unit ............................................................................ 8
3.1.2.1Connection to the CEREC 3 acquisition unit...................................... 8
3.1.2.2Connection to the CEREC AC acquisition unit................................... 9
3.1.3 Connection of external equipment ..................................................... 10
3.1.4 Maintenance and repair ..................................................................... 10
3.1.5 Accessories........................................................................................ 10
3.1.6 In case of damage.............................................................................. 10
3.2 Intended use ................................................................................................. 10
3.3 Certification ................................................................................................... 11
4Technical description................................................................................................ 12
4.1 Design and function ...................................................................................... 12
4.2 System components...................................................................................... 12
4.2.1 Components of the inEos Blue........................................................... 13
4.2.2 Accessories (included in the scope of supply) ................................... 14
4.2.3 Calibration parts (included in scope of supply) .................................. 15
4.2.3.13D acquisition system ........................................................................ 15
4.2.3.2XYZ axis alignment ............................................................................ 15
4.3 Technical data............................................................................................... 16
4.4 System requirements .................................................................................... 16
5Transport to the installation site ............................................................................... 17
5.1 Transport and unpacking .............................................................................. 17

62 99 205 D3446
D3446.201.02.02.02 06.2010 3
Sirona Dental Systems GmbH
Operating Instructions inEos Blue Table of contents
båÖäáëÜ
5.2 Installation site.............................................................................................. 17
6Initial installation of the software.............................................................................. 18
6.1 Installation of the user software and initial start-up ...................................... 18
6.2 Connecting the unit to the PC ...................................................................... 18
6.3 Switching the unit on and starting the software............................................ 19
6.4 Selecting the 3D acquisition system............................................................. 19
6.5 Adding inEos Blue to the device configuration ............................................. 20
7Connecting the unit.................................................................................................. 21
7.1 Connecting the inEos Blue to the PC........................................................... 21
7.2 Connecting the inEos Blue to the power supply........................................... 22
7.3 Connecting the rotation mouse .................................................................... 23
7.4 Connecting the foot switch ........................................................................... 23
8Controls and displays .............................................................................................. 24
8.1 Controls on the inEos Blue........................................................................... 24
8.2 Ready LED ................................................................................................... 25
9Switching the system on.......................................................................................... 26
10 Calibration ............................................................................................................... 27
10.1 General information on calibration ............................................................... 27
10.2 Calibrating the 3D camera system ............................................................... 27
10.3 Calibrating the XYZ axis alignment .............................................................. 28
11 Performing 3D acquisition ....................................................................................... 32
11.1 General......................................................................................................... 32
11.1.1 Scan models ..................................................................................... 32
11.1.2 Focusing the image........................................................................... 33
11.1.3 Overlap area during acquisition ........................................................ 33
11.1.4 Deleting images ................................................................................ 33
11.1.5 Operating the foot switch .................................................................. 34
11.2 Raster scans ................................................................................................ 34
11.2.1 Preparing the model.......................................................................... 34
11.2.2 Taking a scan.................................................................................... 34
11.3 Free scans.................................................................................................... 35
11.3.1 Preparing the model.......................................................................... 36
11.3.2 Taking a scan.................................................................................... 36

62 99 205 D3446
4D3446.201.02.02.02 06.2010
Sirona Dental Systems GmbH
Table of contents Operating Instructions inEos Blue
11.3.3 Performing detail scans...................................................................... 37
11.4 Rotational scans............................................................................................ 38
11.4.1 Preparing the model........................................................................... 38
11.4.2 Taking a scan..................................................................................... 38
11.5 Acquiring impressions ................................................................................... 38
11.6 Examples ...................................................................................................... 40
11.6.1 Scan of a 3-unit bridge....................................................................... 40
11.6.1.1Acquiring a preparation .................................................................... 40
11.6.1.2Rotational scans............................................................................... 42
11.6.1.3Acquiring an antagonist.................................................................... 43
11.6.1.4Bite registration ................................................................................ 44
11.6.1.5Buccal scan...................................................................................... 44
11.6.1.6Terminating acquisition .................................................................... 45
11.6.2 Acquisition of a veneer....................................................................... 46
11.6.3 Acquisition of a crown cap ................................................................. 47
11.6.4 Exposure of an inlay........................................................................... 48
11.6.4.1Acquiring a preparation .................................................................... 48
11.6.4.2Considering the antagonist .............................................................. 50
11.6.4.3Terminating acquisition .................................................................... 50
12 Cleaning and care .................................................................................................... 51
12.1 Care of the 3D camera system ..................................................................... 51
12.2 Cleaning the outer surface ............................................................................ 51
12.3 Protection against medicaments ................................................................... 51

62 99 205 D3446
D3446.201.02.02.02 06.2010 5
Sirona Dental Systems GmbH Dear Customer,
Operating Instructions inEos Blue
båÖäáëÜ
1Dear Customer,
Thank you for purchasing your inEos Blue® from Sirona.
Improper use and handling can create hazards and cause damage. Ple-
ase read and follow these operating instructions carefully and always
keep them within easy reach.
To prevent personal injury or material damage, it is important to observe
all safety information.
To safeguard your warranty claims, please complete the attached Instal-
lation Report / Warranty Passport when the system is handed over and
send it to the indicated fax number.
Your
inEos Blue Team

62 99 205 D3446
6D3446.201.02.02.02 06.2010
General information Sirona Dental Systems GmbH
Structure of the documents Operating Instructions inEos Blue
2General information
Please read this document completely and follow the instructions exactly.
You should always keep it within reach.
Original language of the present document: German.
2.1 Structure of the documents
2.1.1 Identification of danger levels
To prevent personal injury and material damage, please observe the war-
ning and safety information provided in this document. Such information
is highlighted as follows:
NOTE: Application information and other useful information
Tip: Work simplification information
CAUTION
Be sure to observe all warnings!
Please observe the warning and safety information provided to prevent
personal injury and material damage. Any such information is highligh-
ted by a signal word, i.e. DANGER, WARNING, CAUTION or NOTICE.
DANGER
In the event of imminent danger, which results in severe injuries or even
death.
WARNING
In the event of a potentially dangerous situation, which could result in se-
vere injuries or even death.
CAUTION
In the event of a potentially dangerous situation, which could result in mi-
nor personal injuries.
NOTICE
In the event of a potentially harmful situation, which could lead to dama-
ge of the product or an object in its environment.

62 99 205 D3446
D3446.201.02.02.02 06.2010 7
Sirona Dental Systems GmbH General information
Operating Instructions inEos Blue Legend
båÖäáëÜ
2.1.2 Formats and symbols used
The symbols and character formats used in the present manual have the
following meaning:
2.2 Legend
Year of manufactu re
Year of manufacture
2.3 Warranty
To safeguard your warranty claims, please complete the attached Instal-
lation Report / Warranty Passport when the unit is handed over. Then fax
it to the specified fax no.
✔ Prerequisite
1. First action step
CAUTION! Action step war-
ning
2. Second action step
or
➢ Alternative action
Result
Requests you to do something.
see "Formats and symbols used" Identifies a reference to another
text passage and specifies its page
number.
●List Identifies a list.
"Command/menu item" Identifies commands, menu items
or quotations.
20XX

62 99 205 D3446
8D3446.201.02.02.02 06.2010
Safety Sirona Dental Systems GmbH
Basic safety information Operating Instructions inEos Blue
3Safety
3.1 Basic safety information
3.1.1 Prerequisites
3.1.2 Connecting the unit
Description
Perform connection by following the directions given in the present ope-
rating instructions.
3.1.2.1 Connection to the CEREC 3 acquisition unit
Warning: inEos connected to CEREC 3 acquisition unit
NOTICE
Important information on the building installation
The building installation must be performed by a qualified expert in com-
pliance with the national regulations. DIN VDE 0100-710 applies in Ger-
many.
NOTICE
Restrictions regarding installation site
The system is not intended for operation in areas subject to explosion
hazards.
NOTICE
Do not damage the unit!
The unit can be damaged if opened improperly.
It is expressly prohibited to open the unit with tools!
WARNING
No acquisitions in the patient's mouth
As long as an inEos Blue is connected to a CEREC 3 acquisition unit,
no acquisitions may be made in the patient's mouth using the connected
CEREC 3D camera.

62 99 205 D3446
D3446.201.02.02.02 06.2010 9
Sirona Dental Systems GmbH Safety
Operating Instructions inEos Blue Basic safety information
båÖäáëÜ
3.1.2.2 Connection to the CEREC AC acquisition unit
Patient environment
Patient environment
Within the patient environment (A), direct contact is only permissible with
devices or system parts that are approved for use in the patient environ-
ment (A).
This applies to all possible patient positions (B) during the examination or
treatment.
inEos outside of patient environment
WARNING
No acquisitions in the patient's mouth within the patient environment
As long as an inEos Blue is connected to a CEREC AC acquisition unit,
no acquisitions may be made in the patient's mouth using the connected
CEREC Bluecam if the inEos Blue is located outside of the patient envi-
ronment.
1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59” 1500mm
59”
1500mm
59”
1500mm
59”
1500mm
59”

62 99 205 D3446
10 D3446.201.02.02.02 06.2010
Safety Sirona Dental Systems GmbH
Intended use Operating Instructions inEos Blue
3.1.3 Connection of external equipment
If any equipment not approved by Sirona is connected, it must comply
with the applicable standards:
● EN 60 950 for information technology equipment, and
● EN 61 010-1 for laboratory equipment.
3.1.4 Maintenance and repair
As manufacturers of dental instruments and laboratory equipment, we
can assume responsibility for the safety properties of the unit only if the
following points are observed:
● The Maintenance and repair of this unit may be performed only by Si-
rona or by agencies authorized by Sirona.
● Components which have failed and influence the safety of the unit
must be replaced with original (OEM) spare parts.
Please request a certificate whenever you have such work performed. It
should include:
● The type and scope of work.
● Any changes made in the rated parameters or working range.
● Date, name of company and signature.
Changes to the product
Modifications to this unit which may affect the safety of the operator, pa-
tients or third parties are prohibited by law!
3.1.5 Accessories
In order to ensure product safety, this product may be operated only with
original Sirona accessories or third-party accessories expressly approved
by Sirona. The user assumes the risk of using non-approved accessories.
3.1.6 In case of damage
In case of noticeable malfunctions or damage, stop using the instrument
immediately and notify your dental depot or the manufacturer.
3.2 Intended use
This system is used for 3D digitization of single tooth or complete jaw
saw-cut models and impressions.
Follow the operating instru ctions
If the unit is used for any purpose other than the one mentioned above, it
may be damaged.
Proper use also includes observing the present operating instructions and
the relevant maintenance instructions.
NOTICE
Follow the instructions
If the instructions for operating the unit described in this document are
not observed, the intended protection of the user may be impaired.

62 99 205 D3446
D3446.201.02.02.02 06.2010 11
Sirona Dental Systems GmbH Safety
Operating Instructions inEos Blue Certification
båÖäáëÜ
3.3 Certification
CE mark SU + in Eos
empty
CE mark
This product bears the CE mark in accordance with the provisions of Di-
rectives 2006/95/EC (Low-voltage Directive) and 2004/108/EC (EMC Di-
rective).
Examples E N60950 + CAN/ CSA
Examples of CE mark for connected products:
● EN 60950:1992 + A1:1993 + A2:1993 + A3:1995 + A4:1997 based on
IEC 60950
● CAN/CSA-C22.2 No.61010-1-04.
inLab+Eos GOST mark
empty
GOST mark
CSA mark
empty
CSA mark
VDE mark
empty
VDE mark
CAUTION
CE mark for connected products
Further products which are connected to this unit must also bear the CE
mark. These products must be tested according to the applicable stan-
dards.
CUS

62 99 205 D3446
12 D3446.201.02.02.02 06.2010
Technical description Sirona Dental Systems GmbH
Design and function Operating Instructions inEos Blue
4Technical description
4.1 Design and function
The inEos Blue is designed as a desktop unit and consists of several
components (see Chapter "System components" [ ➙ 12]). It is powered
from the standard local power network via an external power supply unit.
The unit is connected to a PC via a USB port. The PC must fulfill the mi-
nimum requirements (see Chapter on "System requirements" [ ➙ 16]).
The PC serves both for controlling inEos Blue and for displaying the cap-
tured images by means of the user software.
4.2 System components
"inEos Blue"
consists of a base plate and a vertical unit. In addition, diffe-
rent accessories as well as the required calibration kits are included in the
scope of supply of the system.
A
B
C
D
AExternal power supply unit
BSupply board
CRotation mouse
D3D acquisition system

62 99 205 D3446
D3446.201.02.02.02 06.2010 13
Sirona Dental Systems GmbH Technical description
Operating Instructions inEos Blue System components
båÖäáëÜ
4.2.1 Components of the inEos Blue
empty
Vertical unit and front side of base plate
AVertical unit EBase plate
BRotary knobs for height
adjustment
FReady LED (blue)
CHeight adjustment unit
with 3D camera system
GShifting plate
DCone with glass protec-
tion cover
HStart button
IRotation mouse

62 99 205 D3446
14 D3446.201.02.02.02 06.2010
Technical description Sirona Dental Systems GmbH
System components Operating Instructions inEos Blue
empty
Rear side of base plate
4.2.2 Accessories (included in the scope of supply)
AMain switch
BSupply voltage input
CUSB port
DRotation mouse
AModel holder (with ball-and-socket joint)
BSliding block
CClamps
DModel pot 31 (3x)
EModel pot 52 (2x)
FPlate
E
D
F
A
B
C

62 99 205 D3446
D3446.201.02.02.02 06.2010 15
Sirona Dental Systems GmbH Technical description
Operating Instructions inEos Blue System components
båÖäáëÜ
4.2.3 Calibration parts (included in scope of supply)
4.2.3.1 3D acquisition system
Calibration part for calibrating the 3D acquisition system.
4.2.3.2 XYZ axis alignment
Calibration parts for calibrating the XYZ axis alignment
Foot control
ACalibration part
A
ACalibration plate
BSpacer
A B

62 99 205 D3446
16 D3446.201.02.02.02 06.2010
Technical description Sirona Dental Systems GmbH
Technical data Operating Instructions inEos Blue
4.3 Technical data
4.4 System requirements
An inLab PC 64bit is required to run this software. The hardware version
must be PC Hardware L or higher.
Model designation inEos Blue
Power connection via power supply unit:
INPUT: 100-240VAC / 47- 63Hz / 1.0 A max.
permissible line voltage fluctuations:
± 10% of nominal voltage
Overvoltage category II
OUTPUT: +12VDC / 2.08A / 25W max.
Power consumption 15 W
Ambient conditions For indoor use
Pollution degree 2
Temperature: 10°C – 35°C (50°F ~ 95°F)
Relative humidity: 30% – 75%
Barometric pressure: 700hPa – 1060hPa
Transport and
storage conditions
Temperature: -40°C – +70°C (-40°F ~ 158°F)
Relative humidity: 10% – 95%
Barometric pressure: 500hPa – 1060hPa
Protection class Class I device
Degree of protection against
ingress of water
Ordinary equipment (not protected)
Mode of operation Continuous operation with intermittent loading corresponding to the la-
boratory mode of working.
Tabletop unit
Tests/approvals This device complies with the requirements of
EN 61 010-1: 2001

62 99 205 D3446
D3446.201.02.02.02 06.2010 17
Sirona Dental Systems GmbH Transport to the installation site
Operating Instructions inEos Blue Transport and unpacking
båÖäáëÜ
5Transport to the installation site
5.1 Transport and unpacking
Transport and unpacking
All Sirona products are carefully checked prior to shipment. Please per-
form an incoming inspection immediately after delivery.
1. Check the delivery note to ensure that the consignment is complete.
2. Check whether the product shows any visible signs of damage.
If return shipment is required, please use the original packaging for ship-
ment.
5.2 Installation site
The unit is designed for desktop use and requires a level footprint of ap-
prox. 35 x 45 cm for installation.
Make sure that suitable operating conditions are given (see section on
“Technical data“ [ ➙ 16]).
NOTICE
Damage during transport
If the product was damaged during transport, please contact your car-
rying agent.
NOTICE
Install inEos Blue in a protected area
3D image acquisition may be adversely affected by bright light.
Set the inEos Blue up so that it is not located directly in the beam path
of an extreme light source and not exposed to direct sunlight.

62 99 205 D3446
18 D3446.201.02.02.02 06.2010
Initial installation of the software Sirona Dental Systems GmbH
Installation of the user software and initial start-up Operating Instructions inEos Blue
6Initial installation of the software
6.1 Installation of the user software and initial start-up
1. Start your PC.
2. Insert the DVD into the DVD drive of your PC.
ªThe installation wizard opens.
3. Click the
"next"
button.
4. Read the license agreement.
5. Confirm your acceptance by clicking the
"Yes"
button.
6. Click
"Standard Installation"
.
ªThe standard installation is then executed.
7. Follow the instructions of the installation wizard to complete the in-
stallation.
8. Restart the system.
ªThe software and drivers are installed.
6.2 Connecting the unit to the PC
➢Connect the inEos Blue to the PC (see also section “Connecting the
unit“ [ ➙ 21]).
NOTICE
Initial installation without inEos Blue
Perform the initial installation of the software without connecting inEos
Blue.
NOTICE
Installation does not start
The installation will start automatically. If it does not start, the CD au-
torun feature is disabled on your PC. In this case start the setup.exe
file on the CD-ROM by double-clicking it.
NOTICE
User-defined installation
The user-defined installation option is recommended only for experien-
ced users.

62 99 205 D3446
D3446.201.02.02.02 06.2010 19
Sirona Dental Systems GmbH Initial installation of the software
Operating Instructions inEos Blue Switching the unit on and starting the software
båÖäáëÜ
6.3 Switching the unit on and starting the software
Let unit dry of f after storage
1. Switch the unit on at the main switch.
2. Start the software at the PC via the Windows start menu.
6.4 Selecting the 3D acquisition system
1. Click
"Settings"
/
"Configuration"
/
"Acquisition system"
in the user soft-
ware.
ªThe menu for selection of the 3D acquisition system opens.
2. Select
"inEos"
.
3. Click the
"OK"
button.
NOTICE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment, con-
densation may form and result in a short circuit.
✔Install the unit at room temperature.
➢Wait until the unit has reached room temperature and is absolutely
dry.
ªThe unit is dry and can be put into operation.

62 99 205 D3446
20 D3446.201.02.02.02 06.2010
Initial installation of the software Sirona Dental Systems GmbH
Adding inEos Blue to the device configuration Operating Instructions inEos Blue
6.5 Adding inEos Blue to the device configuration
Calibrating the system
1. Click
"Settings"
/
"Configuration"
/
"Devices..."
in the user software.
ªThe
"Configure Devices"
dialog box opens.
2. Click the
"Add automatically"
button and enter a name for this device
(
"inEos Blue"
is suggested).
3. Click the
"Add Device"
button.
ªThe installation is now completed.
Calibrating the system
NOTICE
inEos Blue is not displayed
If inEos Blue does not appear in the
"Configure Devices"
dialog, check
the configuration of the interface (see the section entitled “Checking and
configuring the serial interface“).
NOTICE
Calibrating the system
After installing the inEos Blue, you must calibrate the system (see the
chapter entitled "Calibration" [ ➙ 27]).
Other manuals for inEos Blue
2
Table of contents
Other Sirona Scanner manuals