
Volumetric Infusion Pumps
Smiths Medical
Warnings and Cautions
Warnings tell you about dangerous
conditions, that could lead to death or
serious injury to the user or patient, that
can occur if you do not obey all of the
instructions in this manual.
1. WARNING:You should ensure that the
performance offered by the pump is fit for the
intended purpose. Failure to do so may result
in compromised function of the product,
patient injury or user injury.
2. WARNING: Do not use a faulty pump. If the
pump detects a fault when it is first turned on,
or if it develops a fault during operation then a
continuous system alarm sounds.The pump
must be referred to a suitably qualified
technician or returned to Smiths Medical in
order to have the fault rectified.
3. WARNING: Before using the pump, it should be
inspected for physical damage.The pump
should not be used if damage is evident, and
should be returned to service personnel for
repair before being returned to use. Failure to
do so may result in compromised function of
the product, patient injury or user injury.
4. WARNING: Do not use the pump if you detect
any cracks, chips and loose or bent parts, or
if the buttons do not move in and out freely
when they are pressed. Failure to do so could
cause inadvertent disconnection of the pumps.
5. WARNING:To avoid possible malfunction of the
pump, do not expose the pump to X- rays,
gamma rays or ionizing radiation, or to the RF
interference or strong electric/magnetic fields
emitted (for example) by diathermy equipment
or mobile telephones. If the pump is used in
the presence of, or in combination with
Magnetic Resonance Imaging (MRI) machines
it must be protected from the magnetic field
emitted by such equipment. Malfunction of the
pump can cause incorrect infusion or loss of
infusion resulting in patient injury or death.
6. WARNING: Do not push or pull on the pumps,
or the IV pole may tip over or the pumps fall
to the floor. Do not try to remove modular
connected pumps from the IV pole whilst they
are joined together. Either of these could
cause the administration set to separate from
the fluid container thus spilling the
medication, or the pumps themselves could be
damaged.
7. WARNING: Correct entry of data is essential in
order to ensure that the intended infusion is
performed. Before confirming any displayed
data when setting up an infusion, you should
ensure that it is correct. Failure to do so may
result in compromised function of the product,
patient injury or user injury.
8. WARNING: Dose-rate calculation requires
care in entering data. Refer to specific product
drug labelling for information on appropriate
administration techniques and dosages.
Entering incorrect data may result in patient
injury or death.
9. WARNING:When delivering drugs in the
epidural space, use only those medications
specifically indicated for epidural use.
Epidural administration of other drugs could
result in serious patient injury or death.
10. WARNING:The use of administration sets
incorporating injection sites could lead to an
improper or inappropriate infusion resulting in
serious patient injury or death.
11. WARNING: Failure to clearly identify the pump
and administration sets could lead to an
improper or inappropriate infusion resulting in
serious patient injury or death.
12. WARNING: Remove any air to prevent air
embolism.The presence of air within the
infusion can result in complications resulting
in patient injury or death.
13. WARNING: To avoid over infusion, do not
prime the infusion line when the admini-
stration set is connected to the patient. Over
infusion can result in patient injury or death.
14. WARNING:The Occlusion alarm level must be
checked before starting an infusion to ensure
that it is appropriate for the infusion. Failure
to do so may result in an unacceptably slow
time to Occlusion alarm, resulting in patient
injury or death.
15. WARNING: Prior to starting an infusion,
inspect the fluid path for a closed clamp or
any other obstructions or restriction. Failure
to do so may result in the infusion not being
delivered correctly, resulting in patient injury
or death.
16. WARNING: If using a blood pressure cuff above
the patient’s venipuncture site take extra care
in setting the Occlusion alarm pressures.
Failure to do so may result in unnecessary
Occlusion alarms, resulting in patient injury
or death.
Warnings