Smiths Graseby 3200 Operating manual

0473
Part Number 00SM-0130-5
August 2004
© 2004 Smiths Medical International Ltd
INFUSION RATE=
200.0 ml/h
Technical Service
Manual
Graseby 3200
In-line Pressure
Syringe Pump

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual Page i
Published by Smiths Medical International Limited.
All possible care has been taken in the preparation of this publication, but Smiths Medical International
Limited accepts no liability for any inaccuracies that may be found.
Smiths Medical reserves the right to make changes without notice both to this publication and to
the product which it describes.
© Smiths Medical International Limited 2004
No part of this publication may be reproduced, transmitted, transcribed, or stored in a retrieval
system or translated into any human or computer language in any form by any means without the
prior permission of Smiths Medical International Limited.
SMITHS MEDICAL INTERNATIONAL LIMITED,
COLONIAL WAY,
WATFORD,
HERTFORDSHIRE,
UNITED KINGDOM,
WD24 4LG
TEL: (+44) (0)1923 246434
FAX: (+44) (0)1923 231595
http/www.smiths-medical.com
Registered in England. Company number 362847.
Copyright and address
Trademarks and acknowledgements:
Graseby and Smiths are registered trademarks of Smiths Group plc.
All other trademarks are acknowledged as the property of their respective owners.

Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
Page ii
DROCEREUSSI
.oNeussIegnahcrofnosaeR segaP
detceffe etaD
2ot1seussilaitinIllA69.xorppA
3eussieRllA99ceD
4eussieRllA00yl
uJ
1.4segap2ottnemdnemA9-5dnaii00tsuguA
2.4segap2ottnemdnemA51-7,ii10.beF
5eussieRllA40tsuguA
Issue record

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual Page iii
LIST OF CONTENTS
Page
Copyright and address ...................................................................................... i
Warnings and cautions ..................................................................................... x
Abbreviations used ......................................................................................... xiii
CHAPTER 1
INTRODUCTION, FEATURES AND SPECIFICATION
Introduction ................................................................................................... 1-1
Security cover ....................................................................................... 1-1
Features .................................................................................................... 1-3
Micro-controller ............................................................................................. 1-3
Specification .................................................................................................. 1-4
AC power supply ................................................................................... 1-4
Battery type ........................................................................................... 1-4
Battery life ............................................................................................. 1-4
DC input operation ................................................................................. 1-4
Syringe brands and sizes ...................................................................... 1-4
Infusion rates and increments................................................................ 1-4
Mass unit programming range ................................................................ 1-4
Volume infused counter ......................................................................... 1-4
In-line occlusion pressure range ............................................................ 1-5
Internally adjustable occlusion pressure ................................................ 1-5
Factory set occlusion ............................................................................ 1-5
Display languages available................................................................... 1-5
Dimensions ........................................................................................... 1-5
Weight ................................................................................................... 1-5
Temperature range ................................................................................. 1-5
Drive accuracy ...................................................................................... 1-5
History ................................................................................................... 1-5
Electrical safety .................................................................................... 1-5
Design standards ................................................................................... 1-5
UK patent number.................................................................................. 1-5
Printer protocol ...................................................................................... 1-5
Brief history of the Smiths Medical bedside syringe pumps ...................... 1-6
CHAPTER 2
CONFIGURATION, DIAGNOSTICS and OCCLUSIONTHRUST
Configuration mode ....................................................................................... 2-1
Introduction ............................................................................................ 2-1
Entry into the Configuration mode ............................................................... 2-2
Changing a setting ................................................................................. 2-2
Moving to the next parameter ................................................................ 2-2
Available Configuration mode parameters and settings ............................. 2-3
Syringe brands ....................................................................................... 2-3
Syringe size ........................................................................................... 2-3
Lock progam values ............................................................................... 2-3
Contents

Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
Page iv
Contents
CHAPTER 2 (contd.)
Page
Max infusion rate ................................................................................... 2-3
Select pump modes ............................................................................... 2-4
Preset volume mode with time ............................................................... 2-4
KVO rate ................................................................................................ 2-4
Allow mass units option ......................................................................... 2-4
Infusion units ......................................................................................... 2-4
Show rate in ml/h while infusing ............................................................. 2-4
Use pressure transducer ........................................................................ 2-5
Display pressure bar graph .................................................................... 2-5
Allow bolus while running ....................................................................... 2-5
Allow rate change while running ............................................................. 2-5
Intermittent mode start delay ................................................................. 2-5
RS232 Baud rate ................................................................................... 2-5
Communication mode ............................................................................ 2-6
Pump ID ................................................................................................ 2-6
Set key beep volume ............................................................................. 2-6
Pressure units ....................................................................................... 2-6
End of menu .......................................................................................... 2-6
Diagnostic mode ............................................................................................ 2-7
Introduction ........................................................................................... 2-7
Entry into the Diagnostic mode .............................................................. 2-7
Moving to the next display ..................................................................... 2-7
Diagnostic displays ........................................................................................ 2-8
Software ................................................................................................ 2-8
Calibrate transducer ............................................................................... 2-8
Transducer alarm ................................................................................... 2-9
Battery voltage ...................................................................................... 2-9
Language ............................................................................................... 2-9
Number of faults .................................................................................... 2-9
Volume infused ...................................................................................... 2-9
Hours of use .......................................................................................... 2-9
Setting the clock .......................................................................................... 2-10
Entering the clock display ................................................................... 2-10
Disassembly and assembly of casing ......................................................... 2-11
Taking the casing apart ....................................................................... 2-11
Assembly ............................................................................................ 2-11
Occlusion measurements ............................................................................ 2-12
Thrust measurements .......................................................................... 2-12
Syringe stiction.................................................................................... 2-12
Thrust checks ..................................................................................... 2-13
Thrust adjustments .............................................................................. 2-14
(contd.)

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual Page v
CHAPTER 3
FUNCTIONAL DESCRIPTIONS
Introduction ................................................................................................... 3-1
Drive system .................................................................................................. 3-1
Introduction ........................................................................................... 3-1
Stepper motor and leadscrew ................................................................ 3-1
Microcomputer ....................................................................................... 3-1
Toggle mechanism ................................................................................. 3-1
Plunger clamp ....................................................................................... 3-2
Internal occlusion system ............................................................................. 3-2
In-line occlusion system ............................................................................... 3-2
Electro/mech control system ........................................................................ 3-3
Sensing (alarm) systems ............................................................................... 3-3
Introduction ........................................................................................... 3-3
End of infusion ...................................................................................... 3-3
Syringe nearly empty ............................................................................. 3-3
AC power failure .................................................................................... 3-3
Battery voltage low ................................................................................ 3-4
Self tests/pump malfunction .................................................................. 3-4
Drive disengaged or syringe not fitted .................................................... 3-4
Syringe sizing system ........................................................................... 3-4
Rate setting ........................................................................................... 3-4
Cannot zero extension set ..................................................................... 3-5
Cannot calibrate extension set .............................................................. 3-5
Communications (RS232) failure............................................................ 3-5
Software .................................................................................................... 3-5
Self tests ............................................................................................... 3-5
Design ................................................................................................... 3-5
CHAPTER 4
CIRCUIT DESCRIPTIONS
Introduction ................................................................................................... 4-1
Main board ................................................................................................... 4-1
Processor core ...................................................................................... 4-1
Motor interface ...................................................................................... 4-3
Power control ......................................................................................... 4-3
Sensors interface .................................................................................. 4-4
RS232 serial interface ........................................................................... 4-5
Umbilical connector ............................................................................... 4-5
Input/output serial interface ................................................................... 4-5
(contd.)
Contents

Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
Page vi
CHAPTER 4 (contd.)
Page
Status sensors ................................................................................................ 4-6
PL 11 output .......................................................................................... 4-7
PL 12 output .......................................................................................... 4-7
Setting RV1 ........................................................................................... 4-8
Syringe size sensors ..................................................................................... 4-9
Status sensors ................................................................................................ 4-9
Distribution board .......................................................................................... 4-9
Pressure sensing board ................................................................................. 4-9
Main board components ................................................................................ 4-9
Regulator board components ........................................................................ 4-9
Membrane switch panel ................................................................................. 4-9
‘D’ connector .................................................................................................. 4-9
CHAPTER 5
FAULT CODES, CLEANING, RENEWAL of FUSES AND REPAIRS
Fault codes .................................................................................................... 5-1
Cleaning .................................................................................................... 5-4
Repairs .................................................................................................... 5-4
Renewal of fuses ................................................................................... 5-4
Access to fuses .................................................................................... 5-4
Main board renewal ................................................................................ 5-5
Regulator board renewal ........................................................................ 5-5
Status sensors board renewal ................................................................ 5-5
Plunger clamp and super nut assembly renewal .................................... 5-6
Pole clamp assembly renewal ................................................................ 5-6
Leadscrew assembly renewal ................................................................ 5-6
Motor and gearbox assembly renewal .................................................... 5-7
Occlusion clutch and disc assembly renewal......................................... 5-7
Membrane switch panel renewal ............................................................ 5-8
Super nut renewal .................................................................................. 5-8
Syringe size sensors assembly renewal ................................................ 5-9
Plunger clamp repair .............................................................................. 5-9
ACAM pressure sensing assembly renewal ........................................... 5-9
Batteries. Checks and replacement ............................................................ 5-10
Checks ................................................................................................ 5-10
Replacement ....................................................................................... 5-10
Front and/or rear case repairs .................................................................... 5-10
(contd.)
Contents

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual Page vii
CHAPTER 6
Page
FUNCTIONALTESTS and MANUFACTURING SETTINGS
Functional tests ............................................................................................. 6-1
Plunger clamp alarm checks ........................................................................ 6-4
Ramp check procedures ........................................................................ 6-4
Linear accuracy .............................................................................................. 6-5
Test procedures ..................................................................................... 6-5
Plunger clamp alignment .............................................................................. 6-5
Test procedures ..................................................................................... 6-5
Manufacturing settings .................................................................................. 6-6
CHAPTER 7
ILLUSTRATED PARTS LISTS
General assembly .......................................................................................... 7-1
Plunger clamp and half nut assembly ......................................................... 7-7
Pole clamp assembly: non-rotating .............................................................. 7-8
Pole clamp assembly: Rotating .................................................................... 7-9
Leadscrew assembly ................................................................................... 7-10
Main board assembly .................................................................................. 7-11
Regulator board assembly - AC Power ....................................................... 7-13
CHAPTER 8
BRAUN PERFUSOR CONVERSION
Syringe conversion procedures ................................................................... 8-1
Introduction .......................................................................................... 8-1
Nearly empty flag conversion ............................................................... 8-1
Spacer tube fitment ............................................................................. 8-1
Plunger clamp fitment plate ................................................................. 8-1
Braun Perfusor selection ...................................................................... 8-1
‘P’ label fitment .................................................................................... 8-1
Reselecting ‘various’ syringe brands ........................................................... 8-2
Mechanical procedures ........................................................................ 8-2
Programming procedures ..................................................................... 8-2
(contd.)
Contents

Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
Page viii
Contents
CHAPTER 9
Page
DC INPUT VERSION of 3200
Introduction ........................................................................................................... 9-1
DC electrical input supply ..................................................................................... 9-1
APPENDIX
FITTING a MODIFIED SIZE SENSOR FLAG
Introduction .................................................................................................. A-1
Opening the case .......................................................................................... A-2
Removal of old SSF ...................................................................................... A-2
Reassembly ................................................................................................... A-3
Final testing .................................................................................................. A-3
Setting the Size Sensor Flag ....................................................................... A-4
LIST OF TABLES
Table Page
4.1 RS232 D connector ......................................................................................... 4-5
4.2 PL11 connections ........................................................................................... 4-7
4.3 Temperature/voltage range for setting RV ........................................................ 4-8
5.1 Main processor fault codes ............................................................................. 5-1
5.2 Fault causes and actions ................................................................................ 5-3
5.3 Front case spares kit ....................................................................................... 5-11
5.4 Size Sensor Flag spares kit ............................................................................ 5-11
5.5 Rear case spares kit ....................................................................................... 5-12
6.1 Functional tests .............................................................................................. 6-1
6.2 Manufacturing settings .................................................................................... 6-6
A.1 3200 Syringe Size Sensor Gauges (white) ...................................................... A-5
LIST OF FIGURES
Figure Page
1.1 Itemised front view of the pump ...................................................................... 1-2
2.1 Case fixing screw tightening order ................................................................. 2-11
2.2 Thrust measuring set up ................................................................................ 2-13
4.1 Overall block diagram of the 3200 system ..................................................... 4-10
4.2 Main board block diagram .............................................................................. 4-11
4.3 Processor core circuit diagram ...................................................................... 4-12
4.4 Motor interface circuit diagram ....................................................................... 4-13
4.5 Power control circuit diagram ......................................................................... 4-14
4.6 Sensors and pressure sensing interface ........................................................ 4-15
4.7 Pressure sensing interface circuit diagram .................................................... 4-16
4.8 RS232 interface circuit diagram ..................................................................... 4-17
4.9 Umbilical cable connections .......................................................................... 4-18
4.10 Input/output serial interface circuit diagram .................................................... 4-19
4.11 Overview of the regulator ............................................................................... 4-20
(contd.)

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual Page ix
Contents
Figure Page
4.12 Regulator live (primary) circuit diagram .......................................................... 4-21
4.13 Regulator isolated (secondary) circuit diagram ............................................... 4-22
4.14 Syringe size sensors circuit diagram ............................................................. 4-23
4.15 Status sensors circuit diagram ...................................................................... 4-23
4.16 Distribution board connections ....................................................................... 4-24
4.17 Pressure sensing circuit diagram ................................................................... 4-25
4.18 Membrane switch panel circuit ....................................................................... 4-26
4.19 Internal ribbon cable and ‘D’ connector connections ....................................... 4-26
5.1 Half nut (obsolete)/Super nut ........................................................................... 5-8
5.2 Strengthened front and rear case mouldings .................................................. 5-10
6.1 Outline of dual ramp gauge .............................................................................. 6-4
6.2 Ramp check - Square/Ramp style flags ........................................................... 6-4
6.3 Linear accuracy gauge ..................................................................................... 6-5
6.4 Taper gauge .................................................................................................... 6-5
7.1 General assembly ............................................................................................ 7-3
7.2 Plunger clamp and half nut assembly .............................................................. 7-6
7.3a Pole clamp assembly diagram - Non-rotating ................................................... 7-7
7.3b Pole clamp assembly diagram - Rotating ......................................................... 7-8
7.4 Leadscrew assembly ....................................................................................... 7-9
7.5 Main board assembly ..................................................................................... 7-12
7.6 Regulator board assembly - AC Power ........................................................... 7-13
8.1 Braun Perfusor conversion: parts required ....................................................... 8-2
9.1 Diagram of pump chassis connector ................................................................ 9-1
9.2 Power supply circuit diagram ........................................................................... 9-2
9.3 Layout of components on the power supply board ........................................... 9-2
A.1 New modified Syringe Size Sensor Flag .......................................................... A-1
A.2 Order for tightening the case screws................................................................ A-3
A.3 Size Sensor Flag: general details ..................................................................... A-5
LIST OF FIGURES (contd.)

Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
Page x
Warnings and cautions
AC input power connecting socket/cable
The AC input power socket that connects to the rear of the pump has three connections (live, neutral and earth)
provided by a 3-way power cable. As the casing is doubly insulated, the AC input connector situated on the pump
only utilises two connections (live and neutral), there is no third earth pin. This method of AC input enables similar AC
input sockets (if required) to that supplied by Smiths Medical to be used.
Introduction
This Technical Service Manual together with the Instruction Manual, contains all the information that is needed in order
to maintain, repair and operate the Graseby 3200 pump.The contents of this Manual are intended to be read and used
by suitably qualified personnel.
Warnings and cautions
Warnings tell you about dangerous conditions that could lead to death or serious injury to the user or patient that can
occur if you do not obey all of the instructions in this manual.
WARNINGS
1. WARNING:To avoid over- or under- infusion, always verify that the brand and size of the loaded syringe are the
same as the brand and size displayed on the screen before starting an infusion. Failure to do so may result in an
inaccurate delivery of medication, resulting in patient injury or death.
2. WARNING:To avoid incorrect or inappropriate configuration of the pump, the Configuration menu must only be
selected by qualified persons or authorised personnel. Incorrect pump configuration could lead to inappropriate
infusion resulting in patient injury or death.
3. WARNING:This equipment is not suitable for use in the presence of flammable anaesthetics, oxygen-enriched
or explosive atmospheres. The use of the device in such atmospheres may lead to explosion or fire.
4. WARNING:To avoid possible malfunction of the pump, do
not
expose the pump to X- rays, gamma rays or
ionizing radiation, or to the RF interference or strong electric/magnetic fields emitted (for example) by diathermy
equipment or mobile telephones. If the pump is used in the presence of, or in combination with Magnetic
Resonance Imaging (MRI) machines it must be protected from the magnetic field emitted by such equipment.
Malfunction of the pump can cause incorrect infusion or loss of infusion resulting in patient injury or death.
5. WARNING: Operation of the pump outside the temperature limits defined in the specification may result in
erroneous operation. Ensure that the temperature is within the specified limits. Failure to do so may result in
patient injury or user injury.
6. WARNING: In order to ensure that the intended infusion is performed, data must be entered correctly. Likewise
before confirming any displayed data the user should ensure that it is correct. Failure to do so may result in
compromised function of the product, patient injury or user injury.
7. WARNING: Failure to respond promptly to an alarm may result in patient injury or death.
8. WARNING: Failure to follow the
Service Manual’s
maintenance schedule recommendations may result in
compromised function of the product and lead to patient injury or death.
9. WARNING: It is essential that clinical staff remain within visual and audible range of the pump so that critical
alarms can be seen or heard and responded to.
10. WARNING: The user should ensure that the performance offered by the pump is fit for the intended purpose.
Failure to do so may result in compromised function of the product, patient injury or user injury.
11. WARNING: When the pump is carrying out an infusion, to ensure that electrical safety is maintained only items
of equipment that conform to EN60950 are to be connected to the RS232 connector situated at the base of the
pump, otherwise patient safety may be compromised.
12. WARNING: Correct management of battery charging is essential to ensure that the pump can operate on
batteries for the time specified. Failure to do so may lead to impaired functioning of the pump, resulting in
patient injury or death.
13. WARNING: Do
not
use a faulty pump. If the pump develops a fault then an alarm will sound; the display will
indicate a FAULT condition and the pump will fail to infuse. Incorrect performance of the pump can cause
complications resulting in patient injury or death.
If the pump develops a fault then it must be referred to a suitably qualified engineer or returned to Graseby
Medical in order to have the fault rectified.
14. WARNING: Failure to use the mains lead clamp means that the pump may be accidentally or erroneously
disconnected from the mains. Although there is battery backup in case this happens, the battery may not be
sufficiently charged. Consequently there is a risk of the pump not functioning, which could lead to patient injury
or death.
15. WARNING: If an occlusion alarm occurs, immediately clamp the line to eliminate the possibility of a bolus being
delivered to the patient. Then inspect the fluid pathway for kinks, clogged catheter, etc. in order to remove the
occlusion prior to restarting the infusion. An unintentional bolus of medication can result in patient injury or
death.

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual Page xi
16. WARNING: Use only the syringes and administration sets listed in the
Specification
(Chapter 1).
Failure to do so may result in an inaccurate delivery. Smiths Medical does not guarantee performance of the
pump if syringes other than those listed are used. Incorrect function or performance of the pump can cause
complications resulting in patient injury or death.
17. WARNING:The volume of fluid contained in the connecting tubing is a residual amount and will
not
be infused.
Hence this extra volume of fluid must be allowed for when initially filling the syringe and purging the system.
Under-delivery of medication can cause complications resulting in patient injury or death.
18. WARNING:To avoid patient embolism, ensure that the patient tubing is purged of all air bubbles before
administering any medication.The pump provides a purge facility to assist with this process. The presence of
air within the medication can result in complications leading to patient injury or death.
19. WARNING:To avoid syphoning of the syringe contents (free flow), ensure that the syringe is correctly loaded
into the pump, that the syringe plunger is properly engaged by the pump’s actuator and that the pump is placed
not more than 80cm above the infusion site. Syphoning can result in over-infusion leading to patient injury or
death.
20. WARNING:To avoid over-infusion, do not purge the infusion line when the administration set is connected to the
patient. Over-infusion of medication can result in patient injury or death.
21. WARNING: To avoid the pump becoming detached from an IV pole always make sure that the pump is securely
fixed to the pole. Always check the security and stability of the assembly with the pump mounted.
22. WARNING: Following a significant liquid spill onto the pump, it should be wiped dry and inspected by service
personnel before being returned to service. Failure to do so may result in compromised functioning of the pump,
leading to patient or user injury or death.
23. WARNING: When using a syringe smaller than 50/60 ml the occlusion pressure will increase as the diameter of
the syringe decreases, i.e. the smaller the syringe the higher the pressure.
24. WARNING: For high risk or critical infusions, the use of the dedicated 3200 Extension set incorporating the
pressure sensing disc is recommended in order that occlusion pressure can be appropriately set and
monitored during an infusion.
25. WARNING: The patient history is lost when the clock is reset.
26. WARNING: Only adequately qualified personnel should maintain and repair the pump.
27. WARNING: The pump must be disconnected from the AC power supply before the case is opened.
28.WARNING:The Configuration and Diagnostic mode must only be used by personnel who are adequately qualified and
have previous training in the use of the pump.
29.WARNING:The safety and reliability of the pump may be compromised by the use of parts other than those specified
in this Manual.
30.WARNING:When a new Syringe Size Sensor Flag has been fitted to the pump (see Appendix), then the pump must be
tested using the new syringe size sensor gauges available from Smiths Medical (part number 0131-0202). Page 3 of
the Appendix gives details of the Final Testing procedures.
31. WARNING: The pump must be set to display the brand and the size of syringe that is going to be used. Using a
different brand or size of syringe to that being displayed could lead to the incorrect amount of drug being administered,
that could result in injury or death.
32. WARNING: The internal pump batteries must be disposed of in accordance with the manufacturers instructions.
Lead acid batteries must NOT be placed in the normal waste stream.
Warnings and cautions

Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
Page xii
CAUTIONS
Cautions tell you about dangerous conditions that can occur and cause damage to the pump if you do not obey all
of the instructions in this manual.
1. CAUTION: Refer all service, repair and calibrations only to qualified technical personnel. Unauthorised
modifications to the pump must
not
be carried out.
2. CAUTION: When turning the pump on if screens similar to those illustrated are not displayed, do
not
use the
pump. Send the pump to a suitably qualified engineer or return it to Smiths Medical in order to have the fault
rectified.
3. CAUTION: Do not use cleaning and disinfecting agents other than the approved ones specified here.
4. CAUTION:The pump must
not
be immersed in any liquids or exposed to strong organic solvents. Wipe off spills
immediately, and do not allow fluid or residues to remain on the pump. Additionally, the pump is not designed to
be autoclaved, steam-sterilised, ETO-sterilised or subjected to temperatures in excess of 45° C (113° F). Failure
to observe this caution may cause serious damage to the pump.
5. CAUTION: During the removal and replacement of a pump's components, strict observance to Electro Static
Discharge (ESD) rules must be observed at all times, i.e. an earthing strap must be worn. Failure to apply ESD
protection may result in serious damage to the product and possible malfunction. Ensure that any replacement
printed circuit board or other ESD sensitive items are stored in an anti-static container.
Warnings and cautions

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual Page xiii
The following list shows the abbreviations that have been used at various
places throughout this Manual.
Abbreviation Full name
AC Alternating current
A-to-D Analogue-to-digital
C Capacitor or Centigrade
CMOS Complimentary metal oxide silicone
cNm Centinewton metre
COP Computer operating properly
DC Direct current
D Diode
EEPROM Electrically erasable and programmable read
only memory
EPROM Erasable program
Fig Figure
Hz Hertz (cycles per second)
IC Integrated circuit
ID IDentification
kbyte Kilo byte
kg Kilo gram
KVO Keep vein open
LC Liquid crystal
LED Light emitting diode
MCI Manually controlled infusion
MHz Mega Hertz
ml/h Millilitre per hour
ml Millilitre
mm Millimetre
PSI Pounds per square inch
PWM Pulse width modulated
R Resistance
RAM Random access memory
TCI Target controlled infusion
V Volts
VF Vacuum fluorescent
W Watts
Abbreviations
used

CHAPTER 1
INTRODUCTION, FEATURES and SPECIFICATION
3200
In-line Pressure Syringe Pump

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual 1 — 1
CHAPTER 1
INTRODUCTION, FEATURES and SPECIFICATION
The 3200 is a microcomputer controlled syringe pump that has primarily been developed for
the neonatal infusion of sterile liquids.
The pump has an In-line (wet-side) pressure sensing system for accurate occlusion detec-
tion.
The pump is capable of operating in any one of three different infusion modes, as follows:
• continuous,
• preset, or
• intermittent.
And is able to carry out an infusion using a 5, 10, 20, 30, 50/60 ml size syringe.
The pump incorporates a dot matrix vacuum fluorescent display that provides a constant
indication of the pump’s operation. A choice of any one of the following languages is avail-
able (with the appropriate software) via the diagnostic options mode:
English, Dutch, French, German, Italian, Norwegian and Spanish.
The pump can be Configured by the user to work with one of a range of brands of syringe.
It automatically senses the size of the syringe fitted to the pump. The rate of the pump can
be set to operate with either mass units or ml/hour.
The pump is able to dispense liquids at rates of between 0.05 and 200 ml/hour.
The pump keeps the running total of the volume infused, and a history of events as they
occur. About 1,500 events can be logged, complete with the date and time of each.
This compact and robust pump can be used on either a table top or mounted via its pole
clamp to an IV (Intravenous) pole. The pump’s history can either be viewed on the screen or
output to a printer.
An RS232 interface link connects the pump to either a printer or a computer. This allows the
pump’s history to be printed or externally monitored.
Safety features are built into the pump and its software. This includes a set of self test
routines which are run when the pump is switched on.
Users are warned of such incidents as occlusion or power failure by both visible and audible
alarms.
The pump can be powered from either an AC supply or internal rechargeable batteries.
The batteries can power the pump for at least two and a half hours from a fully charged
condition.
A DC version of the 3200 also exists, that requires an input supply of 10 V to 28 V DC (see
Chapter 9).
This Manual refers to pumps that have software Version 2.20 installed. Software versions
other than 2.30 may show slight deviations from certain features described in this Manual.
An optional security cover kit is available (part number 0131-0277). The kit includes all the
items required to modify the pump, including comprehensive fitting instructions. When fitted,
the kit protects the syringe from tampering only; it provides no other security. It does not
lockout the keypad or give audible or visual alarms when opened.
Introduction
Introduction
Security cover

Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
1 — 2
Figure 1.1 Itemised front view of the pump
Introduction

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual 1 — 3
Features
The key features of the 3200 are as follows:
• In-line (wet-side) pressure sensing,
• automatic syringe size sensing,
• up to 1500 History events storage,
• universal AC supply powered or battery powered,
• can be Configured to work with a range of syringe brands,
• advanced safety features,
• state-of-the-art electronics,
• clear text display,
• comprehensive range of alarms,
• RS232 interface,
• simple to use and service,
• all materials used in this product are latex free.
The pump makes use of a sophisticated micro-controller which combines microprocessor
facilities with the following:
• 0n-board non-volatile memory,
• RAM,
• ROM,
• an analogue-to-digital converter,
• communications circuitry,
• four 8-bit pulse-width modulated outputs,
• an internal watchdog (COP).
These facilities are usually provided by many separate silicon chips, the use of a single
micro-controller greatly increases the pump’s reliability.
Micro-
controller
Features

Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
1 — 4
100 V to 240 V at 50/60 Hertz, 40 VA.
The power supply uses primary switching in order to utilise the AC supplies of most
countries.
Sealed lead acid, rechargeable (Cyclon, 3 off). Smiths Medical recommend that the
batteries are checked at least annually (see page 5-10).
More than 2.5 hours of normal pump operation when the batteries are fully charged.
With the AC supply connected, up to 14 hours are required to fully recharge low voltage
batteries.
3200 DC variant pump.
10 V to 28 V DC at a maximum of 4 amps input supply (see Chapter 9).
BD Plastipak... 5, 10, 20, 30/50 and 50/60 ml.
Injectomat... 10 ml and 50 ml.
IMS Pumpjet... 30 ml, prefilled.
Monoject... 5, 10, 20, 30/35 and 50/60 ml.
Braun Omnifix... 5, 10, 20, 30 and 50/60 ml.
Terumo... 5, 10, 20, 30/35 and 50/60 ml.
BD Plastipak (A)... 5, 10, 20, 30/35 and 50/60 ml.
The BD Plastipak (A) syringe accommodates the American manufactured 10 ml syringe,
which differs slightly in length to the European version. All other syringe sizes in the BD
Plastipak selection are identical to the standard BD Plastipak syringes.
Braun Perfusor... Optional (see Chapter 8).
AC power
supply:
Battery type:
Specification
Rates Increments
0.05 to 1.0 ml/h... 0.01 ml/h
1 to 200 ml/h... 0.1 ml/h
0.01 to 999 mg/kg/h
0.01 to 999 mg/kg/min
0.01 to 999 µg/kg/h
0.01 to 999 µg/kg/min
0.01 to 999 mg/h
0.01 to 999 µg/h.
Patient weight... 0.4 to 200 kg.
Drug mass... 1 µg to 999 mg.
Drug volume... 1 ml to 60 ml.
0 to 999.9 ml in 0.1 ml increments.
Infusion rates
and increments:
Mass unit
programming
range:
Volume infused
counter:
Specification
Battery life:
Syringe brands
and sizes:
DC input
operation:

Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual 1 — 5
Specification
In-line occlusion
pressure range:
Internally adjustable
occlusion pressure:
Factory set occlusion:
Display languages
available:
Dimensions:
Weight:
Temperature
range:
Drive accuracy:
History:
Electrical safety:
Design standards:
UK patent number:
Printer protocol:
Using a 30 ml sized syringe:
0 to 7.42 kg (1000 mmHg).
0 to 1355 cmH2O.
0 to 133.3 kPa.
0 to 1316 mBar.
0 to 19.34 psi.
Using a 50 ml syringe, 0 to 4.83 kg (0 to 650 mmHg).
Using a 50 ml syringe, 1.85 kg (250 mmHg) to 4.83 kg (650 mmHg).
3.5 kg (471 mmHg) to 4.2 kg (565 mmHg).
Dutch.
English.
French.
German.
Italian.
Norwegian.
Spanish.
350 x 195 x 115 mm, with pole clamp fitted and plunger clamp closed.
Not exceeding 3.5 kg including the batteries and pole clamp.
Operating conditions:
+5°C to +40°C, 30 to 75% Rh, 700 to 1060 hPa.
Storage conditions:
-40°C to +70°C, 30 to 90% Rh, 700 to 1060 hPa.
±2%.
1500 events can be stored.
Class II
Type CF
Drip proof IPX1.
BS5724 Part 1
IEC601 Part 1
VFG1046/1984.
2229497.
Serial, 8 bits, no parity, 1 stop bit.
Specification (contd.)
Table of contents
Popular Water Pump manuals by other brands

Graco
Graco 223843 instructions

Wilo
Wilo Stratos Installation and operating instructions

Ribiland
Ribiland PRSI24JET101I User and maintenance manual

Outwell
Outwell TORNADO instruction manual

EINHELL
EINHELL GE-AW 9041 E Original operating instructions

Stokes Microvac
Stokes Microvac 412H-11 instruction manual