Smitson JM-LED01 User manual

LED Shadowless Lamp
JM-LED01
USER MANUAL
Smitson Jupiter Ltd
Gillingham, United Kingdom
www.smitsonjupiter.com
info@smitsonjupiter.com

1
1. Expected use
This product is used in various surgical lighting in hospital operating room. It can
also be use in medical lighting situations in emergency room.
2. Conditions of use
2.1 Normal working conditions
Temperature: 10℃~30℃;
Relative humidity: ≤70%;
Atmospheric pressure range: 860hPa~1060hPa;
This equipment can not be worked in flammable and explosive gas environment.
The type of device security: I type equipment;
Working system: continuous operation;
Liquid injection protection level: IPXO.
2.2 Power parameters
Power supply voltage: AC220(±10%)V;
Frequency: 50Hz;
Input power: 70VA;
Main lamp power: 70W.
2.3 suitable mounting height: 3200~3300 mm
Low-storey may bring inconvenience to users. please refer to the product diagram.
in special circumstances, you should contact with the manufacturer when purchasing.
3.Technical specifications
Type
JM-LED01 Main lamp
Diameter of Light Head (mm)
500
Quantity of LEDS
48
Illumination (10 Steps Adjustable)
40~160klx
Color rendering index
Ra>95
Color temperature (5 Steps Adjustable)
3500-5000K
Depth of illumination (mm)
1000
Spot diameter
120-250mm
Endoscopy Mode
Available
Endoscopy Illumination
40,000Lux
LED Life
>50,000 Hours
The model of LED
OSRAM GWCS8PM1.CM
UPS
Available
Control Panel
LCD
www.smitsonjupiter.com

2
4. Simplified Diagram
AC220V
RT1-20 2A
RT1-20 2A
power switch of
operating room
transform er of
the m ain lam p
transform er of
the assistance
lam p
constant-
current
source
constant-
current
source
drive control
drive control
m ain
lam p
assistant
lam p
user interface
user interface
5. Installation
The equipment is fixedly mounted on the roof. The floor must bear the weight of
the device, but also bear the force moment because of partial load and the operation.
In order to ensure the safety, the floor should have enough bearing capacity and
strength.
Users can make construction according to the following pictures.
www.smitsonjupiter.com

3
Floor installation program
φ360
L
pressure plate
floor plate
on the roof
roof of
decorative layer
m ounting
bracket
m ounting base
picture 1: the roof w ith decorative layer
pressure plate
floor plate
on the roof
m ounting disk
m ounting base
picture 2: the roof w ithout decorative layer
m o un ting
d isk
m o un ting b a se
screw M 1 2
nut M 12
sp ring w asher 1 2
flat w a sher 1 2
p ressure p la te
p icture 3 : enla rg ed view o f screw a ssem b ly
www.smitsonjupiter.com

4
φ360
φ240
φ120
8-φ14
material: GB700: Q235
ASTM : A283GRC A306GR55
JIS : G3113-87 SAPH38
φ300
φ360
φ120
φ300
thickness: 8m m
sup erficia l trea tm ent:a ntirust lacq uer
thickness: 12m m
sup erficia l trea tm ent:a ntirust lacq uer
p icture 4 : p ressure p late p icture 5 : m o unting d isk
All fasteners of the installed components must be reliable and fastening to prevent
the harm due to loose. when mounting, the bottom of mounting base must be adjusted
to level condition, otherwise it would cause the drift of lamp.
Ground wire(color: yellow or green) must be grounded reliably, in order to ensure the
use of safety.
With installed problems, please contact with the manufacturer before construction.
Mounted steps
1. according to picture 1 and picture 2 in the operation instruction, drill four Ф16mm
holes on the floor plate with impact drill in determined place. pressure plate and
mounting disk must be tightened with screw M12 in the appropriate length, according
to picture 3.
2. According to mounted diagram, connect the rotating arm to the mounting plate,
regulate it to horizontal place, and tighten it according to picture 3.
3. Connect two balance arms to rotating arms respectively, and connect the
conductive plug. then tighten the two parts with the distributive screw.
4. According to the instruction for the balance arm of Ondal, link the conductive plug,
connect two lamps to the balance arms respectively, and make sure that it is reliable.
5. After careful examination and confirmation, debug the equipment with energizing
according to the operation instruction.
www.smitsonjupiter.com

5
6. Use
NANTONG MEDICAL APPARATUS CO.,LTD
handhold
push button
handle
picture 6:lamp head
Position adjustment
Pull the handhold on the lamp in order to the position and the point which meet
operators’ requirement.
Illumination adjustment
Press the power switch, the lamp lights. each time you press the button
“ increasing brightness”, the illumination will rise a gear. When up to the maximum
10 gear, the maximum illumination symbols will lights; each time you press the
button “weakening brightness”, the illumination will decline a gear, up to the lowest
illumination. The electronic control system of the equipment has memory function. So
it can adjust to the last used gear illumination when rebooting.
Spot adjustment
It is not required usually adjustment, because that the product has been set on
www.smitsonjupiter.com

6
appropriate position when it is out of the factory. When using,place the lamp
Head at 1m away from the surgical site. If it needs to adjust slightly, rotate the handle
in order to the site which meet the surgical needs.
Handle’s remove
When disinfecting, hold the lower part of the handle with hand, and press the
button with your thumb, pull down. Then you can remove it. When installation, keep
the holes on the direction of the button, and push it gently until the button pops out.
Handle’s disinfection
Do not put the handle in high pressure steam sterilizer, in order to avoid
deforming. It can use peroxyacetic acid solution, glutaruc dialdehyde solution and
medical alcohol to disinfect in normal temperature.
7. Maintain
Damping adjustable rotating arm
The center seat of 2 rotating arms has 2 damping adjustment screws. If it drifts
after using for some time, you can tighten the damp with wrench according to picture
8.
picture 9
Lamp head lamp adjustment
Lamp head and two brackets should have appropriate damping,to avoid
transposition. If appearing the phenomenon, tighten the screws properly according to
picture 9.
Balance arm adjustment
Please operate it according to the instruction of Ondal.
www.smitsonjupiter.com

7
Replace the fuse
If the lamp can not light ,in the case of normal electricity network, unscrew the
fuse holder , check whether the fuse is damaged and identify the reasons, according to
the following picture. After confirming, replace the fuse.
Precautions when maintaining
(A) the product is not anti-inlet device. It can not be cleaned with water when
cleaning. You can wipe it with wet cloth containing neutral detergent solution, then
wipe it with wet cloth containing clear water for several times. It also can be wiped
with wet cloth containing disinfectant.
(B) After surgery, cut off power supply of shadowless operating lamps, rise the lamp
head to the highest place to be convenient to clean the operating room and to avoid
that the cleaning equipments collide the shadowless operating lamps.
(C) Non-professionals do not dismount the device at option. In addition to the LED
module inside the device, there is no device what users can replace.
8. Warranty
The warranty period of our shadowless lamp is one year. We carry out the
national 3 guarantees policy. Please save the relevant receipt of purchasing and
maintenance card.
Do not let person without the authorization by the manufacturer dismantle the
equipment. If there is any needs on the fault or use of the equipment, please contact
with customer service of the manufacturer.
www.smitsonjupiter.com

8
9. Tips for environmental protection
The equipment does not contain mercury vapor and explosive devices. deal with
the packaging materials and the equipment after stopping working according to local
environmental regulations. Metallic article may become renewable resources ,and
e-waste should be treated with harmless.
10. Packing List
(1) Rotating arm 1
(2) Balance arm Ondal 1
(3) Main lamp 1
(4) Handle 4
(5) Product certificate 1
(6) Manual 1
(7) Warranty sheet 1
(8) Instruction for balance arm Ondal 1
(9) Inner hexagon spanner (L/M/S) 1
(10) Fuse (F2AL250V) 4
Smitson Jupiter Ltd
Gillingham, United Kingdom
www.smitsonjupiter.com
info@smitsonjupiter.com
Table of contents
Other Smitson Lighting Equipment manuals
Popular Lighting Equipment manuals by other brands

Goobay
Goobay 49867 user manual

Multikomplex
Multikomplex SLT973 Installation, Use and Maintenance Instruction

Home Decorators Collection
Home Decorators Collection HD7838E Use and care guide

GTV
GTV LD-KONML-RGBW-PIL Assembly instruction

Clas Ohlson
Clas Ohlson WX-LED44-260WW-10-2 quick start guide

Alien
Alien RG STAR user manual