Solmetex Hg5 User manual

Solmetex Hg5®HV Amalgam Separator
Maximum Flow 2000 ml/min Maximum Fillable Volume 1900 ml
Installation and Maintenance Instructions
Hg5®-HV Manual – July 2017 www.solmetex.com 800-216-5505
Thoroughly read and understand instructions prior to installing,
operating and servicing the Hg5®HV Amalgam Separator. These
instructions are also available on our website at www.solmetex.com.
Table of Contents
Warnings & Warranty ........ Page 1
Installation ........................... Page 2
Installation Diagram .......... Page 3
Maintenance ...................... Page 4
Service ................................ Page 5
Solmetex Warranty
Solmetex Hg5®HV Air Water Separators are warranted against defects in material and
workmanship for a period of two (2) years from the date of purchase, established by proof
of purchase or formal warranty registration. During the warranty period Solmetex will at its
option repair or replace products that prove to be defective.
Solmetex Hg5®HV Collection Containers are warranted against defects in material and
workmanship for a period of one (1) year from the date of purchase, established by proof
of purchase or formal warranty registration. During the warranty period Solmetex will at its
option repair or replace products that prove to be defective.
Labor, transportation and service charges are not included.
Limitations of Warranty
The warranty shall not apply to defects resulting from improper installation, maintenance,
abuse, unauthorized modification, or operation outside of the environmental specifications
for the product or damages that occur due to improper repackaging of equipment for return
to Solmetex.
USE OF THIRD PARTY COLLECTION CONTAINERS OR LINE CLEANERS HAVING A pH
LESS THAN 6 OR GREATER THAN 8 WILL VOID THESE WARRANTIES. For a complete list of
recommended cleaners, visit our web site www.solmetex.com.
No other warranty is expressed or implied. Solmetex specifically disclaims the implied war-
ranties of merchantability and fitness for particular purpose.
Exclusive Remedies
The remedies provided herein are the buyer’s sole and exclusive remedies. Solmetex shall
not be liable for any direct, indirect, special, incidental or consequential damages, whether
based on contract, tort or any other legal theory.
The waste stream
treated by the
Solmetex Hg5®HV is
generated by a dental vacuum
system and as such may contain
concentrations of solid and
soluble mercury and silver.
Because of this, any spills should
be considered hazardous and
should be handled in accordance
with standard hazardous
materials (HAZMAT) handling
procedures.
Full Collection Containers
are a HAZMAT and should be
handled, stored and disposed of
according to regulations applying
to hazardous waste containing
mercury. Always wear protective
gear when handling full Solmetex
Hg5®HV collection containers
(latex gloves, safety glasses or
face shield) and dispose of per
local regulations and codes.

Hg5®HV Installation
Hg5®-HV Manual – July 2017 www.solmetex.com 800-216-5505
Instructions
When choosing the location to mount the Hg5®HV Amalgam Separator, remember that you must leave room to:
• Complete the piping.
• Access the front and sides of the unit for collection container replacement.
• Have the unit mounted higher than the pump and drain.
The wall mounted support bracket must be mounted to the wall prior to installation of the Hg5®HV Amalgam Separator. The support
bracket should be stud mounted in order to support the filled weight of approximately 100 pounds. The inlet and outlet connections are
3” Schedule 40 PVC Pipe. It is recommended that flexible pipe couplings be used to connect the unit to the existing vacuum lines. Since
pipe sizes can vary from site to site, bushings may be required to adapt the unit to the vacuum line piping at your facility.
• Remove all components from package(s) make sure you have everything, prior to proceeding.
• The unit is piped into the vacuum line between the treatment rooms and the vacuum source. The line from the treatment rooms is
piped to the inlet of the Hg5®HV and the outlet is piped to the vacuum source. DO NOT connect the unit to the outlet of the separating
tank (dry vacuum system) or water ring pump (wet vacuum system).
• Determine the location of unit with respect to existing plumbing; pump location, drain and ease of collection container change out.
• Mount the back plate to an existing wall using lag bolts. The structure and bolts should be capable of supporting 100 pounds. Mount
directly into the studs using the pre-drilled holes in the backplate. If no studs are available, molly bolts may be used providing they and
the wall are capable of supporting the load.
• After mounting the bracket, lower the Hg5®HV onto the bracket. It will sit between the two side supports. Use the four ¼-20 bolts
supplied with the unit to secure the unit to the bracket.
• After the unit is installed, the two collection containers can be installed. (See included installation instructions).
• Write the installation date on each collection container.
• Complete the registration form and return to Solmetex.
Environmental Specifications
• Overall Dimensions: 17.625” W x 28” H x 22.5” D
• One Hg5 HV will serve up to 20 chairs
• Minimize water lift height.
• Maximum Temperature = 120F (52C).
• Maximum Vacuum = 15” Hg (51 kPa)
• For Dental Use ONLY
Inlet
I
nlet
Air
Water
Separator
28” 28”
Wall
Mount
Bracket
14”
Outlet Outlet
Side View
22.5” 17 5/8”
Front View
Collecon
Containers
2

Treatment
Rooms
Hg5-HV
Inlet
Tank Inlet Air
Water
Separang
Ta
nk
Any
Dry
Vacuum Unit
DO NOT connect
Hg5-HV
to outlet of
tank.
Hg5-HV Outlet OR
Pump
Inlet
Any Water
Ring Pump
DO NOT connect Hg5 HV to
outlet of the water ring pump.
NOTE: The Hg5 HV should be installed so that the outlet of the Hg5 HV
is higher than the tank or pump inlet.
Hg5®-HV Manual – July 2017 www.solmetex.com 800-216-5505
3
Hg5®HV Installation Diagram

Hg5®HV Maintenance
Hg5®-HV Manual – July 2017 www.solmetex.com 800-216-5505
Maintenance
• Collection Container must be replaced/changed once every twelve (12) months or when the sediment level reaches the full line;
whichever occurs first.
• Check the sedimentation level of the collection container weekly.
• Line Rinsing & Vacuum Line Cleaners - Cleaners should have these qualities: Non-Foaming, De-Odorizing, Sanitizing & pH between 6
and 8. USE OF LINE CLEANERS OUTSIDE OF THIS pH RANGE WILL VOID THE WARRANTY.
For a complete list of recommended cleaners, visit our web site www.solmetex.com.
WITH DENTAL VACUUM SYSTEMS: Follow manufacturers instructions.
Note: Plan to rinse no more than 5 chairs every 10 minutes. Limit the total maximum flow to the Hg5® during rinsing to 1 liter per minute.
Rinsing too much or too rapidly can overfill the Hg5® and could aect the unit’s eciency.
4

Hg5®HV Service
Hg5®-HV Manual – July 2017 www.solmetex.com 800-216-5505
Service
The Hg5®HV is designed to provide years of trouble-free service, with minimal attention. In the unlikely event of system related problems,
please consult the troubleshooting and maintenance chart below.
Problem: Little or No vacuum at the hand piece
1. Check sediment level of collection container, if full change the collection container.
2. If the vacuum gauge reads normally but there is little or no vacuum to the hand piece, there is probably a clog or a leak between
the hand piece and the Hg5®HV.
3 .If the vacuum gauge reads lower than normal these are the possible causes:
Did you just replace the collection container?
Yes:
Check installation of the new collection container. If the vacuum is low there is a possibility that the o-rings on the collection
container did not seal into the receivers. Remove the collection container, check the o-rings and re-install per the collection
container installation procedure.
No:
If you have a solids collection cup (installed by others) - Check & Clean or replace element/screen if necessary.
Check operation of the vacuum pump.
• Listen for vacuum leaks.
• Check all connections for breaks or cracks.
Check all flexible hose for kinks, breaks, or loose clamps.
Problem: Water in the Upper Chamber (Air Water Separator)
1. Check sediment level of collection container, if full change collection container.
2. Check the pH of vacuum line cleaner, if below 6 or above 8 change collection container and line cleaner. Visit www.solmetex.com for
a recommended list of cleaners.
3. Call your dental dealer.
• If the problem cannot be solved easily, call or e-mail your dealer for support.
5
Problems
Soluon
Solids reach full line of collecon container
Change the collecon container
Leave the vacuum running during process
Solids above full line of collecon container
Change the collecon container
Inspect the top chamber for solids
Top chamber has some solids
System is backed up
•Turn on vacuum
•Remove pins
•Tilt container towards manifold to allow air into top
chamber
•
Place container back onand insert pins
Top chamber is full with solids
System is in bypass
•Reducon in sucon
•Solids released into waste stream and environment
•Top chamber needs to be replaced
•
Full top chamber needs to be recycled
Bring a flashlight
•Most equipment/ulity rooms have poor lighng. Using
a flashlight from the backside of the system and shining
it forward will help determine the level of
sedimentaon. Also can be used to inspect the top
chamber using the same procedure.
Table of contents
Other Solmetex Water Filtration System manuals
Popular Water Filtration System manuals by other brands

Safeway water
Safeway water 4 Stage Ro Installation & maintenance guide

Commodity Axis
Commodity Axis Poly-Reactor ViaAqua PR 20/20 Booklet

Thermal Care
Thermal Care SF Series owner's manual

Spectra Watermakers
Spectra Watermakers AQUIFER 360 EXPEDITION owner's manual

Axi
Axi STS 6030 Instruction, Operation and Maintenance Manual

Moen
Moen Automobile Parts Replacement instructions

ATB
ATB AQUAmax BASIC series Operation and maintenance manual

Eureka Forbes
Eureka Forbes Aquaguard Aquasure Smart Plus RO+UV+MTDS user manual

Ecosoft
Ecosoft FK Series instruction manual

Dell
Dell OptiPlex 5090 install guide

Puricom
Puricom Haktiva user manual

Pentair
Pentair THS Series Installation and user guide