Spellman MONOBLOCK Series User manual

MONOBLOCK® XRBHR/XRBD MANUAL 118184-001 Rev B
R
Instruction Manual
High Voltage X-Ray Generator
MODEL :
SERIAL# :
DATE :
SPELLMAN
HIGH VOLTAGE ELECTRONICS
CORPORATION
475 Wireless Blvd.
Hauppauge, New York, 11788
+1(631) 630-3000*FAX: +1(631) 435-1620*
E-mail: sales@spellmanhv.com
Website: www.spellmanhv.com
MONOBLOCK® SERIES
XRBHR/XRBD

118091-001 REV. B
IMPORTANT SAFETY PRECAUTIONS
SAFETY
THIS POWER SUPPLY GENERATES VOLTAGES THAT ARE DANGEROUS AND MAY BE
FATAL.
OBSERVE EXTREME CAUTION WHEN WORKING WITH THIS EQUIPMENT.
High voltage power supplies must always be grounded.
Do not touch connections unless the equipment is off and the
Capacitance of both the load and power supply is discharged.
Allow five minutes for discharge of internal capacitance of the power supply.
Do not ground yourself or work under wet or damp conditions.
SERVICING SAFETY
.
Maintenance may require removing the instrument cover with the power on.
Servicing should be done by qualified personnel aware of the electrical hazards.
WARNING note in the text call attention to hazards in operation of these units
that could lead to possible injury or death.
CAUTION notes in the text indicate procedures to be followed to avoid possible
damage to equipment.
Copyright 2000, Spellman High Voltage Electronics Corporation. All Rights Reserved.
This information contained in this publication is derived in part from proprietary and patent data. This information has
been prepared for the express purpose of assisting operating and maintenance personnel in the efficient use of the
model described herein, and publication of this information does not convey any right to reproduce it or to use it for
any purpose other than in connection with installation, operation, and maintenance of the equipment described.

118091-001 REV. B
WICHTIGE SICHERHEITSHINWEISE
SICHERHEIT
DIESES HOCHSPANNUNGSNETZTEIL ERZEUGT LEBENSGEFÄHRLICHE HOCHSPANNUNG.
SEIN SIE SEHR VORSICHTIG BEI DER ARBEIT MIT DIESEM GERÄT.
Das Hochspannungsnetzteil muß immer geerdet sein.
Berühren Sie die Stecker des Netzteiles nur, wenn das Gerät ausgeschaltet ist und die elektrischen
Kapazitäten des Netzteiles und der angeschlossenen Last entladen sind.
Die internen Kapazitäten des Hochspannungsnetzteiles benötigen ca. 5 Minuten, um sich zu entladen.
Erden Sie sich nicht, und arbeiten Sie nicht in feuchter oder nasser Umgebung.
SERVICESICHERHEIT
Notwendige Reparaturen können es erforderlich machen, den Gehäusedeckel während des Betriebes zu
entfernen.
Reparaturen dürfen nur von qualifiziertem, eingewiesenem Personal ausgeführt werden.
“WARNING” im folgenden Text weist auf gefährliche Operationen hin, die zu Verletzungen oder zum Tod
führen können.
“CAUTION” im folgenden Text weist auf Prozeduren hin, die genauestens befolgt werden müssen, um
eventuelle Beschädigungen des Gerätes zu vermeiden.

118091-001 REV. B
PRECAUTIONS IMPORTANTES POUR VOTRE SECURITE
CONSIGNES DE SÉCURITÉ
CETTE ALIMENTATION GÉNÈRE DES TENSIONS QUI SONT DANGEUREUSES ET PEUVENT ÊTRE FATALES.
SOYEZ EXTRÊMENT VIGILANTS LORSQUE VOUS UTILISEZ CET ÉQUIPEMENT.
Les alimentations haute tension doivent toujours être mises à la masse.
Ne touchez pas les connectiques sans que l’équipement soit éteint et que la capacité à la fois de la charge et de
l’alimentation soient déchargées.
Prévoyez 5 minutes pour la décharge de la capacité interne de l’alimentation.
Ne vous mettez pas à la masse, ou ne travaillez pas sous conditions mouillées ou humides.
CONSIGNES DE SÉCURITÉ EN CAS DE REPARATION
La maintenance peut nécessiter l’enlèvement du couvercle lorsque l’alimentation est encore allumée.
Les réparations doivent être effectuées par une personne qualifiée et connaissant les risques électriques.
Dans le manuel, les notes marquées « WARNING » attire l’attention sur les risques lors de la manipulation de ces
équipements, qui peuvent entrainer de possibles blessures voire la mort.
Dans le manuel, les notes marquées « CAUTION » indiquent les procédures qui doivent être suivies afin d’éviter
d’éventuels dommages sur l’équipement.

118091-001 REV. B
IMPORTANTI PRECAUZIONI DI SICUREZZA
SICUREZZA
QUESTO ALIMENTATORE GENERA TENSIONI CHE SONO PERICOLOSE E
POTREBBERO ESSERE MORTALI.
PONI ESTREMA CAUTELA QUANDO OPERI CON QUESO APPARECCHIO.
Gli alimentatori ad alta tensione devono sempre essere collegati ad un impianto di terra.
Non toccare le connessioni a meno che l’apparecchio sia stato spento e la capacità
interna del carico e dell’alimentatore stesso siano scariche.
Attendere cinque minuti per permettere la scarica della capacità interna dell’alimentatore
ad alta tensione.
Non mettere a terra il proprio corpo oppure operare in ambienti bagnati o saturi d’umidità.
SICUREZZA NELLA MANUTENZIONE.
Manutenzione potrebbe essere richiesta, rimuovendo la copertura con apparecchio
acceso.
La manutenzione deve essere svolta da personale qualificato, coscio dei rischi elettrici.
Attenzione alle AVVERTENZE contenute nel manuale, che richiamano all’attenzione ai
rischi quando si opera con tali unità e che potrebbero causare possibili ferite o morte.
Le note di CAUTELA contenute nel manuale, indicano le procedure da seguire per
evitare possibili danni all’apparecchio.

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 1 118185-001 REV B
Table of Contents PAGE
1. INTRODUCTION
1.1 Description of XRBHR/XRBD MONOBLOCKSeries................................. 2
1.2 Operating Features........................................................................................... 3
1.3 Status and Fault Diagnostic Display................................................................. 4
1.4 Interpreting the Model Number........................................................................ 5
2. INSPECTION
2.1 Initial Inspection.............................................................................................. 6
2.2 Installation....................................................................................................... 6
3. OPERATING INSTRUCTIONS
3.1 Operation......................................................................................................... 8
3.2 Signal and Power Connections......................................................................... 9
3.3 Local Mode (Analog Monitors) ....................................................................... 9
3.4 Digital Control…............................................................................................. 9
3.5 X-ray Tube Seasoning Process......................................................................... 9
4. DIAGNOSTICS
4.1 Table 1 ............................................................................................................ 10
5. FACTORY SERVICE
5.1 Warranty Repairs............................................................................................. 11
5.2 Factory Service Procedures.............................................................................. 11
5.3 Ordering Options and Modifications…............................................................ 11
5.4 Shipping Instructions....................................................................................... 11
Appendix A ..................................................................................................................... 12

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 2 118185-001 REV B
Chapter 1
INTRODUCTION
1.1 Description of the
XRBHR/XRBD MONOBLOCK
Series
he XRBHR (high reliability) and XRBD Monoblock®
X-ray source are designed for OEM applications. The
XRBHR/XRBD Monoblock® series is equipped with
an internal Bipolar X-Ray tube designed for use at voltages
up to 80k, 100kVor 160kV at power levels up to 100W,
210W, 350W or 500W. It is a complete integrated system
consisting of a high voltage power supply (HVPS) with
proprietary control and protection circuitry as well as a
filament supply, X-ray tube and oil encapsulant in one
compact enclosure. Additional advantages are the
elimination of high voltage cables and low leakage X-ray
radiation.
A Power factor corrected input regulator is utilized to
improve overall reduction in harmonic emissions and EMI.
The output provides a regulated +390VDC to downstream
switching circuitry enabling optimal performance and
efficiency.
The XRBHR/XRBD MONOBLOCKSeries
incorporates remote programming, monitoring, and fault
indicators including interlock.
The X-ray source is a sealed unit containing the HVPS, X-
ray tube and insulating oil. The insulating oil provides
electrical insulation for the high voltage components and
the X-ray tube. The oil also functions as a coolant to carry
heat away from the tube. The cooling system requirements
are outlined in Chapter 3 and in the XRBHR data sheet
SHV#128132-001 or XRBD data sheet SHV#128146-001.
A bellows internal to the X-ray source compensates for the
expansion of the oil as the oil temperature varies thereby
eliminating the need for a bulky overflow tank.
The XRBHR/XRBD is available with either a fan or cone
shape beam geometry. Proprietary emission control
circuitry provides excellent regulation of X-ray tube
current, along with outstanding stability performance. The
XRBHR is designed for long field life and comes with a
three-year warranty for standard units.
T

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 3 118185-001 REV B
1.2 Operating Features
The XRBHR/XRBD MONOBLOCKSeries
incorporates several standard features designed to optimize
user satisfaction and safety.
•INTERNAL FAULT PROTECTION: The
XRBHR/XRBD MONOBLOCKSeries continually
monitors internal circuits critical to the proper
operation of the power supply. In the event that one
of these circuits does not function correctly, the fault
detection circuit latches the appropriate fault on the
front cover display and turns off the X-ray output.
•LED FAULT/STATUS INDICATORS:
Indicators are located on the cover of the Control
/Inverter Assembly and are provided to give the user a
complete indication of system operation and fault
conditions.
POWER OVERVOLTAGE (OV)
ARC FLT UNDERVOLTAGE (UV)
OVERTEMP (OT) OVER CURRENT (OC)
UNDER CURRENT (UC) X-RAY ON
•WATCHDOG OPERATION: The HVPS employs
an internal watchdog timer to discontinue X-ray
operation in instances where the host computer has
lost communication and control of the Monoblock®
for a period greater than three seconds. This feature is
may be enabled or disabled via RS-232 or Ethernet
host command and willdefault to enable at system turn
on.
•LOCAL MONITORS: Provides local analog
monitors for the X-ray tube voltage and current via the
analog interface connector on the front cover of the
control assembly. Reference XRBHR data sheet
SHV#128132-001or XRBD data sheet SHV#128146-
001.
•REMOTE PROGRAMMING: Allows remote
adjustment of the output voltage and current via RS-
232 or Ethernet interface. All program values default
to zero at system turn on.
•REMOTE MONITORS: Provides remote monitors
such as output voltage, output current and operating
temperature via RS-232 or Ethernet interface. See the
Communication Interface document SHV#101501-
695 for a full description of features and commands.
•EXTERNAL INTERLOCK: The X-ray cannot be
enabled unless the 24VDC has been applied to pins 1
and 2 of the analog interface connector as outlined in
the XRBHR data sheet SHV#128132-001 or XRBD
data sheet SHV#128146-001. During high voltage
operation, opening the interlock circuit will cause the
high voltage to be disabled.
IMPORTANT
This control signal is not a safety
interlock and should not be used for
protection from X-ray generation for
safety purposes.
•HV ON RELAY: When X-ray ON is enabled a N.O.
and a N.C. relay contact closure will occur. Refer to
the XRBHR datasheet SHV#128132-001 or XRBD
data sheet SHV#128146-001 for pin assignments and
contact ratings.
•OIL TEMPERATURE MONITOR: Provides
remote oil temperature monitoring via RS-232 or
Ethernet interface.
•FAULT SIGNAL: The Fault Signal is an open
collector output that indicates that a fault has occurred.
The connection can be accessed at pin 7 of the analog
interface connector. High = no faults.

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 4 118185-001 REV B
1.3 System Status and Fault
Diagnostic Display
If a fault occurs, the power supply will react as described
and the fault will be reported via RS-232 or Ethernet. To
reset a fault, a clear command must be sent via RS-232 or
Ethernet.
•POWER ON LED: When input power is applied to
the unit it is indicated by the POWER LED on the
control cover being illuminated.
•X-ray ON LED: When the high voltage status is
“On” state it is indicated by X-RAY ON LED status
on the control cover.
•EXTERNAL INTERLOCK FAULT: Indicates that
the high voltage supply is inhibited by external
interlock protection circuitry. This fault is reported
solely via RS-232 or Ethernet communication as
(ARG 9) Interlock.
•OVER VOLTAGE: Indicates the over voltage
protection circuitry has caused the high voltage power
supply to turn off. The over voltage protection is
internally set to turn off the high voltage to the X-ray
tube if the X-ray tube voltage is 10% greater than the
set point or at 10% greater than the maximum rated
voltage. This fault is indicated by OV LED on the
control cover and via RS-232 or Ethernet as (ARG 6),
High Voltage.
•UNDER VOLTAGE: Indicates the under voltage
protection circuitry has caused the high voltage power
supply to turn off. The under voltage protection is
internally set to turn off the high voltage to the X-ray
tube if the X-ray tube voltage is 10% less than the set
point. This fault is indicated by illumination of UV
LED on the control cover and via RS-232 or Ethernet
as (ARG 5) Low Voltage.
•OVER CURRENT: Indicates the over current
protection circuitry has caused the high voltage power
supply to turn off. The over current protection is
internally set to turn off the high voltage to the X-ray
tube if the X-ray tube emission current is 18% greater
than the set point or at 10% greater than the maximum
rated emission current. This fault is indicated by
illumination of OC LED status on the control cover
and via RS-232 or Ethernet as (ARG 3) High Current.
•UNDER CURRENT: Indicates the X-ray tube
emission current is less the setpoint. This fault occurs
when the emission current is 18% less than the
programmed emission current. This fault does not
shutdown the high voltage power supply, it is
indicated by illumination of UC LED on the control
cover and via RS-232 or Ethernet as (ARG 4) Low
Current.
•ARC FAULT: Indicates that an arc has occurred.
When a single arc occurs, the event will be indicated
by illumination of the ARC LED on the control cover
and via RS-232or Ethernet as (ARG 2) ARC. Both the
ARC LED and the digital flag will be reported for 30
seconds.
Occurrences of four arcs within a period of ten seconds
will turn off the high voltage. This fault is indicated by
illumination of ARC FLT LED on control cover and
via RS-232 or Ethernet as (ARG 2) Arc.
•OVER TEMPERATURE: Indicates the over
temperature protection circuitry has caused the high
voltage power supply to turn off. The over temperature
protection is internally set to turn off the high voltage
to the X-ray tube if the internal oil temperature has
exceeded 65 degrees C. This fault is indicated by the
OT LED status on the control cover and via RS-232 or
Ethernet as (ARG 1) Temperature.
•TEMPERATURE WARNING: Indicates the oil
temperature inside the Monoblock is greater than 58
degrees C. This fault does not shutdown the high
voltage power supply. The warning is indicated by
flashing the OT LED status on the control cover and
via RS-232 or Ethernet as (ARG 11) Temperature
Warning.
•POWER LIMITFAULT: Indicatesthe output power
has exceeded maximum rated power of the
Monoblock® causing the high voltage power supply
to turn off. This fault is reported solely via RS-232 or
Ethernet communication as (ARG 8) Power.
•WATCHDOG TIMER: Indicates the host computer
has lost communication and with the Monoblock® for
a period greater than 3 seconds. This feature is enabled
on by default. This fault is reported solely via RS-232
or Ethernet communication as (ARG 7) Watchdog.

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 5 118185-001 REV B
1.4 Interpreting the Model Number
The model number of the power supply describes its
capabilities. After the series name is:
1. Maximum voltage (in kV).
2. Polarity of the unit: (PN) Bipolar
3. The maximum output (in Watts).
4. Options
5. Custom model number
XRB 100 PN 500 HR / XX / X (#)
XRB100PN500HR
Series
Name
Maximum
Voltage
Bipolar
Output
Polarity
Maximum
Power
Custom
“X”
Number
Options
See datasheet

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 6 118185-001 REV B
Chapter 2
INSPECTION &INSTALLATION
nitial inspection and preliminary checkout procedures
are recommended. For safe operation, please follow the
step-by-step procedures described in Chapter 3,
Operating Systems.
2.1 Initial Inspection
Inspect the package exterior for evidence of damage due to
handling in transit. Notify the carrier and Spellman
immediately if damage is evident. Do not destroy or
remove any of the packing material used in a damaged
shipment.
After unpacking, inspect the Monoblock and control
assembly for any visible signs of damage.
The Monoblock and control assembly are paired together
as a system. Verify that the serial number located on the
Monoblock label is also reference on the control assembly
label and vice versa. It is important that the serial number
pair of Monoblock and control assembly are installed and
operated as a pair for normal operation. If for any reason
it becomes necessary to return the system both the
Monoblock and control assembly should be returned
together.
The Spellman XRBHR/XRBD MONOBLOCKSeries
of X-ray source and components are covered by warranty.
2.2 Installation
Reference the XRBHR data sheet SHV#128132-001 or
XRBD data sheet SHV#128146-001and outline drawing
SHV# 408644-001for dimensions.
The unit can be mounted in any position. Four, M8x1.25
threaded mounting holes are provided on the top and
bottom of the Monoblock®. Threaded M6x1 holes are
also provided for exit port collimation mounting.
The Control assemble has four, M5 threaded holes 10mm
deep.
Do not block the vent holes in the top cover as indicated
on the XRBHR data sheet SHV#128132-001 or XRBD
data sheet SHV#128146-001.
I

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 7 118185-001 REV B
Chapter 3
OPERATING INSTRUCTIONS
3.1 Operation
THIS EQUIPMENT GENERATES DANGEROUS
VOLTAGES THAT MAY BE FATAL.
PROPER GROUNDING OF ALL HIGH VOLTAGE
EQUIPMENT IS ESSENTIAL.
WARNING X-RAY RADIATION EXPOSURE IS
HAZARDOUS
Failure to follow these procedures may void the warranty.
Check the input voltage rating on the nameplate of the
supply and make certain that this is the rating of the
available power source. The Spellman XRBHR/XRBD
MONOBLOCKoperates on 90-264VAC, single phase
50 or 60Hz.
Bellow Vent hole:
Do not block or insert anything into the vent hole located
on top cover as indicated on the XRBHR data sheet
SHV#128132-001 or XRBD data sheet SHV#128146-001.
This hole provides venting for the internal bellows that
allows for the expansion and contraction of the insulating
oil as temperature varies.
Cooling:
For convection/forced air cooled models:
The Monoblock® is designed to operate in an ambient
temperature of 0-40 degrees C. For maximum performance
and reliability, it is important that sufficient cooling is
provided to maintain the oil temperature of the
Monoblock® at or below 55 degrees C. Convection
cooling may be sufficient. However, in some applications
an external cooling fan (to be provided by the user) may be
necessary to maintain the oil temperature of the
Monoblock® at or below 55 degrees C. Please see below
for suggest fan placement.
For models with an integrated cooling system:
Models equipped with an integrated cooling system are
designed to operate in an ambient temperature of 0-40
degrees C. For maximum performance and reliability, it is
important that the ambient operating temperature of the
Monoblock® is maintained at or below 40 degrees C.
Preventative maintenance is recommended in order to
keep the fans and heat exchanger free from dust and
debris.
WARNING

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 8 118185-001 REV B
BEFORE CONNECTING THE MONOBLOCK® TO
THE AC LINE, FOLLOW THIS PROCEDURE.
1) PROPER GROUNDING TECHNIQUES: The
Monoblock® must be grounded, by making a ground
connection to the M5 threaded hole on the tank flange. The
control assembly must also be grounded using the M5
ground stud located on the cover of the control assembly.
Note: For optimal EMC results the Tank and Controller
should be mounted to the same ground plane, if not a
ground wire (preferably a braided wire) should be
connected between the tank and the controller.
2)Connect the controlcable to the Monoblock®and secure
the cable ground to the Monoblock® using the M5
threaded hole on the tank flange.
3) ENSURE THAT THE EXIT PORT IS PROPERLY
MATED TO THE COLLIMATOR OR
SATISFACTORILY SHIELDED WITH LEAD PLUG
TO LIMIT EXPOSURE TO LEAKAGE RADIATION.
X-ray Safety Procedures must be followed when
testing this unit. The XRBHR/XRBD is capable of
producing Lethal Voltages and X-ray Radiation.
Only proceed with operation of the HVPS after
a) consulting with the Manufacturer and
verification of X-ray setup for the proper
precautions.
b) reading this entire document.
NEVER OPERATE THIS UNIT WITH AN OPEN
X-RAY EXIT PORT.
It is recommended not to allow leakage radiation
exceeding 0.5mR/hr at 5cm from any surface of
the XRBHR/XRBD MONOBLOCK.
3.2 Signal and Power Interface
Connections
1) Connect Ground wires as suggested above.
2) Connect digital signal interface RS-232 or Ethernet
communication cable from the host computer to the
connector on Control Assembly.
3) Connect 90 - 264VAC line to the AC input connector.
Reference power and interface connections located on the
XRBHR data sheet SHV#128132-001 or XRBD data sheet
SHV#128146-001. Make sure AC line voltage is OFF
when connecting to the unit.
4) Connect 24VDC to pins 1 and 2 of the heat dissipation
unit. It is critical that power is supplied to the heat
dissipation unit to ensure proper operation of the
Monoblock. Reference DC input for heat dissipation unit
located on the XRBHR data sheet SHV#128132-001 or
XRBD data sheet SHV#128146-001.
5) Apply the 24VDC external Interlock by connecting
24VDC to pins 1 and 2 of the analog interface connector.
Reference analog interface located on the XRBHR data
sheet SHV#128132-001 or XRBD data sheet
SHV#128146-001.
6) Determine unit’s idle time and refer to Appendix A for
the recommended seasoning schedule. Also reference
paragraph 4.5.21 and 4.5.28 of the Communications
Interface document SHV#101501-695.
3.3 Analog monitors
1) The X-ray tube voltage can be monitored by measuring
the voltage at pins 4 and 5 of the analog interface
connector. Reference analog interface located on the
XRBHR data sheet SHV#128132-001 or XRBD data sheet
SHV#128146-001.
2) The emission current can be monitored by measuring the
voltage at pins 6 and 5 of the analog interface connector.
Reference analog interface located on the XRBHR data
sheet SHV#128132-001 or XRBD data sheet
SHV#128146-001.
3.4Digital control
G.U.I and operation instruction
1) Install the G.U.I software to the host computer (See CD
with GUI).
2) Start the G.U.I.
3) Read agreement then click AGREE.
4) Operate the unit using the GUI. For instruction on GUI
operation, see the “Monoblock® X User Manual.
3.5 X-ray Tube Seasoning Process
Caution: To prevent premature degradation of the X-ray
tube, it is highly recommended to season the X-ray tube if
the unit has been idle for greater than 24hours. Please refer
to Appendix A for the recommended seasoning schedule.
Also reference paragraph 4.5.21 and 4.5.28 of the
Communications Interface document SHV#101501-695.
WARNING

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 9 118185-001 REV B
Chapter 4
Diagnostics
4.1
Fault/Symptom
Possible Cause
Over Voltage fault resulting in
shutdown of the high voltage
power supply.
kV programming set greater than maximum specified
kV.
Under Voltage fault resulting in
shutdown of the high voltage
power supply.
AC input voltage lower than minimum specified.
Over Current fault resulting in
shutdown of the high voltage
power supply.
mA programming greater than maximum specified mA.
Under Current fault
kV programming is less than 35kV or AC input voltage
below spec.
Over Temperature fault
resulting in shutdown of the
high voltage power supply.
Internal oil temperature is above 65 degrees C. May be
caused by operation at ambient temperature greater than
40°C or failure of a pump/cooling fan. Check that the
fans and pump are operational and that there are no
obstructions.
Temperature Warning fault
Indicates the oil temperature is greater than 58 degrees C.
Verify that the ambient temperature is less than 40°C.
Check that the fans and pump are operational and that
there are no obstructions.
ARC fault resulting in the ARC
fault being reported for 30
seconds while the high voltage
power supply remains on.
Occurrence of tube arc. If the arcing becomes regular the
X-ray tube may require additional warm up or seasoning.
ARC fault resulting in
shutdown of the high voltage
power supply.
At least four arcs occurred in a ten second period. The X-
ray tube may require additional seasoning. Season the X-
ray tube following the 14-30 days seasoning protocol. If
problem continues contact Spellman service department.
Overpower fault
The unit has exceeded the maximum rated power.
Interlock fault resulting in
shutdown or failure to turn on
the high voltage power supply.
Verify the 24VDC interlock enable signal is present at
pins 1 and 2 of the analog interface connector.
TABLE 1

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 10 118185-001 REV B
Chapter 5
FACTORY SERVICE
5.1 Warranty Repairs
During the Warranty period, Spellman will repair all units
free of charge. The Warranty is void if the unit is worked
on by other than Spellman personnel. See the Warranty in
the rear of this manual for more information. Follow the
return procedures described in Section 5.2. The customer
shall pay for shipping to and from Spellman.
5.2 Factory Service Procedures
Spellman has a well-equipped factory repair department. If
a unit is returned to the factory for calibration or repair, a
detailed description of the specific problem should be
attached.
For all units returned for repair, please obtain an
authorization to ship from the Customer Service
Department, either by phone or mail prior to shipping.
When you call, please state the model and serial numbers,
which are on the plate on the rear of the power supply, and
the purchase order number for the repair. A Service
Validation Order number is needed for all returns. This
SVO number should be marked clearly on the outside of
the shipping container. Packages received without an SVO
number will be returned to the customer. The Customer
shall pay for shipping to and from Spellman.
A preliminary estimate for repairs will be given by phone
by Customer Service. A purchase order for this amount is
requested upon issuance of the SVO number. A more
detailed estimate will be made when the power supply is
received at the Spellman Repair Center. In the event that
repair work is extensive, Spellman will call to seek
additional authorization from your company before
completing the repairs.
5.3 Ordering Options and
Modifications
Many of the options can be retrofitted into Spellman’s
power supplies by our factory. For prices and
arrangements, contact the Spellman Sales Department.
5.4 Shipping Instructions
All power supplies returned to Spellman must be sent
shipping prepaid. Pack the units carefully and securely in
a suitable container, preferably in the original container, if
available. The power supply should be surrounded by at
least four inches of shock absorbing material. Please return
all associated materials, i.e. high voltage output cables,
interconnection cables, etc., so that we can examine and
test the entire system.
All correspondence and phone calls should be directed to:
Spellman High Voltage Electronics Corp.
475 Wireless Boulevard
Hauppauge, New York 11788
TEL: (631) 630-3000 FAX: (631) 435-1620
E-Mail: sales@Spellmanhv.com

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 11 118185-001 REV B
Appendix A
X-ray Tube Seasoning Tables

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 12 118185-001 REV B
X-RAY TUBE RE-SEASONING PROCESS
for
XRB80PN210HR
Parameters
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Step 9
Step 10
Step11
Step12
Voltage (kV)
50
60
70
80
85
80
80
80
70
60
50
40
Current (mA)
0.35
0.65
0.90
1.2
1.5
1.8
2.2
2.6
3.0
3.5
4.2
5.2
For units with idle
time greater than
30 days
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
For units with idle
time of 14-30 days
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
For units with idle
time of 7-14 days
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
For units with idle
time of 1-7 days
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 13 118185-001 REV B
X-RAY TUBE RE-SEASONING PROCESS
for
XRB80PN350HR
Parameters
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Step 9
Step 10
Step11
Step12
Voltage (kV)
50
60
70
80
80
80
80
85
70
55
50
40
Current (mA)
0.35
0.75
1.5
2.25
3.0
3.5
4.4
4.0
5.0
6.0
7.0
8.0
For units with idle
time greater than
30 days
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
For units with idle
time of 14-30 days
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
For units with idle
time of 7-14 days
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
For units with idle
time of 1-7 days
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 14 118185-001 REV B
X-RAY TUBE RE-SEASONING PROCESS
for
XRB80PN500HR
Parameters
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Step 9
Step 10
Step11
Step12
Voltage (kV)
50
60
70
80
80
80
80
80
85
80
70
60
Current (mA)
0.35
0.75
1.5
2.25
3.0
3.5
4.0
4.5
5.5
6.25
7.0
8.0
For units with idle
time greater than
30 days
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
For units with idle
time of 14-30 days
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
For units with idle
time of 7-14 days
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
For units with idle
time of 1-7 days
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec

XRBHR/XRBD SERIES MONOBLOCK® MANUAL 15 118185-001 REV B
X-RAY TUBE RE-SEASONING PROCESS
for
XRB100PN100HR
Parameters
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Step 9
Step 10
Step11
Step12
Voltage (kV)
50
60
70
80
90
100
105
80
75
65
55
50
Current (mA)
0.4
0.6
0.7
0.8
0.9
1.0
2.0
1.25
1.3
1.5
1.8
2.0
For units with idle
time greater than
30 days
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
20 min
For units with idle
time of 14-30 days
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
For units with idle
time of 7-14 days
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
60 sec
For units with idle
time of 1-7 days
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
30 sec
Other manuals for MONOBLOCK Series
1
This manual suits for next models
2
Table of contents
Other Spellman Portable Generator manuals