Spica Polaris ll User manual

INSTRUCTIONS FOR USE
Polaris ll, Nexus, Tertius mattress systems

Contents
1. Packaging Contents...................................................................................................................................................1
2. Alternating Pressure Operation Principle .................................................................................................................1
3. Guidelines .................................................................................................................................................................1
4. Specifications ............................................................................................................................................................1
5. General Precautions..................................................................................................................................................2
5.1 Mattress Precautions..............................................................................................................................................2
5.2 Control Unit Precautions.........................................................................................................................................3
6. Storage......................................................................................................................................................................3
7. Warranty ...................................................................................................................................................................3
8. Symbols.....................................................................................................................................................................4
9. Polaris ll, Nexus, Tertius Front Panel........................................................................................................................5
10. CPR Valve Operation (Mattresses only)................................................................................................................6
11. Installation ............................................................................................................................................................6
11.1 Mattress................................................................................................................................................................6
11.2 Cushion..................................................................................................................................................................7
12. Bottoming Out Test...............................................................................................................................................8
13. Settings and Operation .........................................................................................................................................8
13.1 Factory dafult Settings ..........................................................................................................................................8
13.2 Pressure setting and Clinician override ................................................................................................................8
13.3 Dynamic and static CLP modes .............................................................................................................................8
13.4 Care Mode.............................................................................................................................................................8
13.5 Upright Boost........................................................................................................................................................8
13.6 Cushion mode.......................................................................................................................................................8
13.7 UNLOCK and Factory RE-SET ................................................................................................................................8
14. Alarms ...................................................................................................................................................................8
14.1 Low Pressure Alarm ..............................................................................................................................................9
14.2 Power Failure Alarm............................................................................................................................................10
15. Replacing the Air Filter........................................................................................................................................11
16. Cleaning...............................................................................................................................................................11
16.1 On-site cleaning ..................................................................................................................................................11
16.2 Machine washing / drying...................................................................................................................................11
16.3 Troubleshooting..................................................................................................................................................12
Appendix A: EMC Information

1. Packaging Contents
•Control Unit x 1
•Power Cable x 1
•Mattress/Cushion (depending on system) x 1
2. Alternating Pressure Operation Principle
Patients who spend a lot of time in bed will often experience vascular compression on any part of their body where
there is a bony protrusion, such as the shoulders and sacrum. This can easily cause decubitus pressure ulcers to form,
leading to further complications such as infection.
The interleaved pressure relief areas on our alternating pressure support surfaces provide periodic relief from
vascular compression to the whole body by dynamically redistributing your patient’s weight, preventing and alleviating
decubitus pressure ulcers. For your patient’s comfort, the three-cell pillow zone at the head of themattress is designed
to provide static support for your patient’s head.
3. Guidelines
Always use pressure care systems under the advice of a suitably qualified medical professional. The patient’s pressure
ulcer risk assessment scores, weight and the handling considerations for caregivers should be considered to ensure
that the appropriate system is selected.
4. Specifications
Technical Specifications
Atmospheric Specifications
Therapy modes
Dynamic & Static
Storage temperature
-25 - 70°C
Compressor air flow
Approx. 10lpm
Transport temperature
-25 - 70°C
Cycle time
10 minutes
Operating temperature
5 - 40°C
Anti-particle filter
Yes
Humidity
10 - 90%
Visual and audible alarms
Yes
Atmospheric pressure
700 - 1060hPa
Electrical power supply
220 –240V / 50Hz
Power consumption
9W
Fuse rating
F2A 250VAC
Electrical isolation
Class I 2a
Ingress protection
IP21
IEC conformity
60601-1, 60601-1-2, 60601-11
Warranty
1 year against all manufacturing
faults

5. General Precautions
✓Ensure this equipment is not setup or used near sources of high heat or excessive electromagnetic, electrostatic
or radiation fields including UV radiation, such as direct sunlight.
✓Do not use system in the presence of any flammable anaesthetic mixture with air, nitrous oxide or oxygen or in
the presence of smoking materials or open flame - risk of explosion.
✓Do not use this equipment for anything other than what it is specifically designed for. The use of accessories or
parts not recommended or specifically designed for this equipment is prohibited and be voided the warranty.
✓Do not modify or connect this equipment to other parts or equipment not specifically designed for use with this
system.
✓Do not allow young children to operate, play with or remove any part of this medical system.
✓Do not use this equipment near sources of excessive moisture such as nebulizers or steam kettles.
✓Do not use corrosive cleaning products such as industrial degreasers or acetone solvents.
✓The product can only be operated by personnel who are qualified to perform general nursing procedures and
has received adequate training in knowledge of prevention and treatment of pressure ulcer.
✓Close supervision is necessary when this product is used on or near children. The device is mains powered and
contains small parts so presents electrical and choking hazards.
✓The mattress cover has passed skin sensitization and skin irritation test. If you suspect that you may have had or
are having an allergic reaction to the mattress cover, consult a physician immediately.
5.1 Mattress Precautions
✓Do not place any layers of material between the patient and top cover of the mattress; this will compromise
the support system therapy. The following are included:
oHospital sheets - regular or fitted
oSheepskins or equivalent
oIncontinence sheets
oSlide sheets
oElectric heating blankets
✓Ensure the patient’s clothing does not cause skin damage due to ties, buttons, creases, seams, objects in
pockets and jewellery.
✓Do not place any sharp items on or near the mattress such as syringes or scalpels or any instrument that could
hole the top cover.
✓Do not place any solid items on top of the system besides the patient.
✓No natural rubber latex is used in the construction of this system.

5.2 Control Unit Precautions
✓Do not open the control unit, as there is risk of electric shock and it will void your warranty.
✓Do not block the air inlet at the rear of the control unit.
✓Do not spill any liquids onto the control unit. If a spillage occurs:
oTurn off power to the control unit at the wall and disconnect the power cable from the control unit.
oWipe dry any excess moisture on the external casing.
oCheck that the interior of the IEC socket, rocker switch and power plug are dry.
✓Ensure that the power lead is undamaged, safely run and properly connected so that it does not pose an
electrocution or trip hazard.
✓Ensure that the power cable is positioned away from any moving bed part, which may kink or damage the
power cord.
✓Ensure that the power cable does not interfere with cleaning around the bed or the use of cleaning liquids
especially on the floor.
6. Storage
It is recommended that the control unit is stored in a sealed plastic bag in a cool, dry area. Ensure that no heavy items
are placed on top of the system during storage as this may damage it.
7. Warranty
The control unit is guaranteed for 1 year from the date of purchase against all manufacturing faults, on the condition
that it is stored and operated in the conditions recommended in this instruction document. Proof of purchase is
required in order to claim the guarantee. The warranty will be void if the control unit is dismantled by the user as
shown on the warning sticker on the side of the unit.
This guarantee is not a substitute for any legal guarantees. Contact your local sales agent for technical support.
Fig: 1

8. Symbols
Attention: See Instructions for use
CE Marking indicating conformance to EC Directive No. 2007/47/EC concerning medical devices
Type BF Applied Part (patient isolation from electrical shock)
Class II Product
Operation Instructions
Indicates separate collection for electrical and electronic equipment (WEEE)
IP21
Protection against finger and dripping water
Manufacturer
Date of manufacture
Note: Follow the requirements of your local authority regarding disposal of the unit.

9. Polaris ll, Nexus, Tertius Front Panel
1
Unlock
2.
Static CLP therapy
3.
Dynamic pressure therapy
4.
Auto & override pressure setting
5.
Cushion mode
6.
Care/firm mode
7
Seating boost
8
Leaking/power outage alarm mute
Fig: 2
1.
2.
4.
8.
3.
7.
5.
6.

10. CPR Valve Operation (Mattresses only)
The CPR valve allows the mattress to deflate quickly when CPR is to be performed on the patient. The CPR valve is
located at the head end of the mattress. The system will operate normally when the CPR pull tag (Fig. 3) is securely
located in the CPR air outlet and will deflate when the CPR pull tag is pulled from the CPR air outlet.
Fig: 3
11. Installation
11.1 Mattresses
12.Unroll the mattress on the bed with the connector hoses at the foot end.
13.Secure the mattress to the bed base with the attachment straps.
14.Hang the control unit at the foot of the bed using the hooks
15.Connect the mattress hoses (Fig: 4) and power cable (Fig: 5) to the control unit then guide it along the bed to the
nearest plug socket and plug it in, taking care not to create a trip hazard.
16.Check that CPR tag is securely pressed down into CPR air outlet.
17.Switch on the rocker switch on the control unit. (Fig: 5)
18.Pump will flash all LEDS before defaulting to AS (Autoset) and dynamic mode
19.The pump setting is set to default to automatic –you do not need to enter the patient weight.
20.The mattress will take up to 45 minutes minutes to inflate. Cushions will take 10 minutes to inflate

Fig: 4 Fig: 5
WARNING:
The power cable must be installed to avoid interference with the articulated parts of the bed or the wheels and to
prevent personnel from tripping up. Failure to install the power cable correctly may result in risk of bodily injury
and/or material damage.
11.2Cushions
1. Place the cushion on the seat with hoses protruding from the back corner.
2. Hang the control unit on the back of the seat.
3. Connect the cushion hoses and power cable to the control unit then guide it to the nearest plug socket and
plug it in, taking care not to create a trip hazard.
4. Switch on the rocker switch on the control unit.
1. The cushion will take up to 5 minutes to inflate. Select the button with a cushion symbol if you want to pair
the pump up with an air cushion. This will ONLY work with the Andway Saiph cushion so please ensure you
use the correct cushion. WARNING> Do NOT use this cushion mode with the pump connected to a mattress –
the pressures will be incorrect and it will cause the pump to alarm

12. Bottoming Out Test
It is important to check that your patient has not bottomed out on the support system. This is where the patient’s
sacrum sinks too far into the mattress and contact pressure from the bed frame is exerted on the patient. To check
for this, slide two fingers between the air cells underneath the patient’s sacrum. The patient should be clear of your
fingers. If the patient has bottomed out, increase the pressure setting on the control unit then wait for 10 minutes
before repeating the bottoming out test. This test does not refer to cushion systems.
13. Settings & operation
13.1 Factory default settings
The default setting for the Polaris ll, Nexus, Tertius control unit is dynamic mode. The control panel on your Polaris
ll, Nexus, Tertius control unit will lock after 30 seconds of inactivity. You can unlock with a single press of the unlock
button and re-lock with a single press of the button (or it will lock automatically after 30 seconds)
The pump setting for patient weight and pressure is automatic. When you switch on, it will default to AS (AutoSet)
so you do not need to enter the patient weight. It will also automatically select Dynamic mode .
13.2 Clinician auto override. Item 1, Figure 2, Section 9
The default setting for the patients pressure setting is automatic. However if a trained clinician decides that his or hers
patients weight or body shape does not suit the automatic selection then they can override the AutoSet setting. Press
and hold the unlock button (Item 1 Figure 2 section 9) for 3 seconds until the lock button LED light goes out then use
the arrow button (Item 4 Figure 2 Section 9) to scroll through the choices available. The first choice is +l which is 10%
firmer than AS, if you press the up arrow again then +ll appears in the LED screen (item 8, Figure 2,Section 9) this is
20% firmer than AS. If you press the up arrow again then is goes to -ll which is 20% softer than AS, and if you press the
up arrow again then -l will appear in the LED screen which is 10% softer than AS. If you press the up arrow again you
return to AS the Automatic pressure setting for your patients weight. Please REMEMBER this setting is based on weight
ONLY and it cannot determine body shape. Short wide patients can require a different setting to say tall thin patients
to double amputees –even though they may weigh the same. A clinician should perform a hand test (section 12) to
determine if the patient is correctly supported.
13.3 Dynamic & Static (CLP) modes. Items 2 & 3, Figure 2, Section 9
The factory default setting for this pump is dynamic (alternating) which means that cells A will inflate for 5 minutes
followed by cells B inflating for 5 minutes whilst cells A are deflated. This is a 10 minute cycle. However a trained
clinician MAY choose static CLP therapy mode. This mode means that ALL the cells remain inflated with any dynamic
or alternating at much lower internal pressure. Press and hold unlock for 3 seconds until the LED light goes out and
then press the CLP therapy mode.
13.4 Care Mode (also known as FIRM or Nursing mode)Item 6, figure 2 section 9
The factory default setting for care mode is OFF. However this mode maybe selected for use when changing dressings,
toileting a patient and other general assistance. This mode will inflate the mattress to maximum and is very firm. The
mattress will NOT alternate Because of this, this mode will automatically switch OFF after 30 minutes use to protect
the patient and should NOT be reused straightaway unless by a trained clinician. To select this mode press and hold
UNLOCK for 3 seconds until the LED light goes out and then select Care. The Dymanic mode LED light will then goes
out. After 30 minutes the Care mode will automatically switch OFF and the Dymanic mode will come back resume by
default.

13.5 Upright boost. Item 7, Figure 9,Section 9
The factory default setting for upright boost is OFF. However this mode maybe selected for heavier patients, patients
who due to their body shape tend to sit lower in the mattress and double/single amputees to give them a boost when
the bed is profiled in sitting/eating/watching TV position. This mode gives about an extra 20% more internal pressure
to support the patient when they are in an upright position. To select this mode press the UNLOCK button for 3 seconds
until the LED light goes out. Then select UPRIGHT BOOST. The UPRIGHT BOOST LED light will light up. This mode has
NO time limit and will remain on until the clinician turns it off. To turn it off, press the uNLOCK button and press the
UPRIGHT BOOST once –and the LED light will go out.
13.6 Cushion mode Item 5, Figure 2, Section 9
The factory setting for this mode is OFF. However if you want select this mode Please read section 11.2 FIRST. You
must NOT use this mode unless the pump is connected to a Andway Saiph cushion . You can cause injury to the patient
if you use CUSHION mode whilst the pump is operating the mattress. To select this mode connect the pump to the
cushion as in 11.2 and then press the UNLOCK button for 3 seconds until the LED light goes out and then select
CUSHION mode. The CUSHION mode LED light will come on. There is NO time limit to this mode and it will remain in
CUSHION mode until the clinician changes it.
13.7 Unlock and factory re-set –Item 1, Figure 2, Section 9
The pump will automatically lock after 30 seconds of inactivity. The LED light will come on and none of the pump
functions will work when locked apart from the alarm mode. To unlock the pump press and hold the UNLOCK button
for 3 seconds. Then you make your function selection before the pump automatically locks after 30 seconds. The
UNLOCK buttons also acts as a factory re-set should you become confused about what is in the pumps memory mode
especially when you swap the pump to a new patient. To reset the pump back to factory settings switch the pump off
at the rocker switch on the side. Now press and HOLD the UNLOCK button and keeping holding it whilst you switch
the rocker switch back on. You will hear three rapid beeps –you may now release the UNLOCK button and the pump
is now reset to factory settings.
14. Alarms
14.1 Low Pressure Alarm Item 8, Figure 2, Section 9
When there is low pressure in the system the control unit will beep continuously and the corresponding red LED
will flash.
Press and hold the unlock button for two seconds to unlock the control panel. (Fig: 2)
Press the low pressure alarm mute button (Fig: 2) to mute the alarm, and then check the following connections
for leakage:
•The connection between the mattress and the control unit
•The connection of the CPR valve to the mattress
•The connection between the cells and the connecting hose
If the fault is not solved the check the following for leakage:
•The CPR valve
•The connecting hose
•The air cells

If the fault persists, contact your Andway Spica distributor for service repair.
14.2 Power Failure Alarm Item 8, Figure 2, Section 9
When the power to the control unit fails, it will beep continuously and the corresponding red LED will flash.
Press and hold the unlock button for two seconds to unlock the control panel. (Fig: 2)
Press the power failure alarm mute button (Fig: 2) and check the following;
•That the power cable is connected to the power outlet and that the outlet is switched on
•That the power cable is connected to the control unit
•That the fuse in the power cable has not blown
•That the fuse in the IEC socket on the control unit has not blown
If one, or both, of the fuses has blown check the power cable for damage and replace it immediately if necessary.
If the fault persists, contact your Andway Spica distributor for service repair.

15. Replacing the Air Filter
The air filter must be replaced annually. To replace the air filter, remove the air filter cover (Fig: 8) from the back of
the control unit and replace the cotton filter inside after cleaning any residual dust from the air filter cavity and cover.
Fig: 8
16. Cleaning
16.1 On-site cleaning
When necessary the top cover and base can be cleaned and disinfected on site once the patient has been removed
from the mattress. The following on-site cleaning procedure is recommended for top cover and control unit. Note
that a summary of the below is printed at the foot end flap of the top cover. Do not immerse the control unit in
water.
1. Ensure gloves are worn and all disinfection and occupational health and safety protocols of the facility are
adhered to.
2. Wipe down with a clean cloth using a disinfectant solution comprising of warm water and a neutral detergent
or with a sodium hypochlorite solution (Chlorine 1000 ppm)
3. Proprietary disinfectants may be used provided manufacturer’s instructions are followed.
4. All cleaning agents and disinfectants must be thoroughly rinsed off and the surface dried before storage or
re-use. Failure to do this may result in the accumulation of reagent that could damage the polyurethane
coating, react with the bed frame or negate the bio-compatibility results of the fabric.
16.2 Machine washing / drying
The mattress cover and base can be machine washed. To machine wash the base - all cells and hoses must be
removed. Wash at a temperature up to 71°C (160°F), using normal detergents.
Dry the cover by air drying, spinning or tumbling at temperatures up to 130°C (266°F); do not mangle.
In washing machines, it may be difficult to fully soak the mattress cover. Also, spinning and tumbling may not
remove water trapped between layers. It may be helpful to interrupt the washing or drying cycles to alleviate this.

16.3 Troubleshooting
The instructions below are for end-user maintenance of the pressure care system. If these instructions do not
resolve the fault, contact your Andway Spica distributor for service repair.
Fault
Solution
Control unit does not operate
1. Check that the power cable is connected to the wall socket and
the IEC socket on the control unit correctly.
2. Check the fuses in the power plug and in the IEC socket. Be sure
to use the correct fuse specification.
Low pressure in mattress
1. Check that the mattress is connected to the control unit and that
the CPR tag is securely pressed into the CPR air outlet.
2. Check the air cells and connecting hose for punctures, tears or
leaking connections.
Air cells at uneven heights
1. Remove the top cover and check for air cell press studs that are
disconnected.



This manual suits for next models
2
Table of contents
Other Spica Network Hardware manuals