General Precautions
CAUTION:
Proper use of this device de-
pends on careful reading of
all instructions and labels.
CAUTION:
Steam Sterilize (autoclave)
or ETO sterilize only. Do not
gamma ray sterilize.
WARNING:
After steam sterilization, unit
will be hot. Allow adequate
cooling time before handling.
Do not drop or mishandle
unit.
WARNING:
For use with gamma and x-
ray emitting LDR radiation
sources only.
Warnings and Cautions alert users to
dangerous conditions that can occur if
instructions in the manual are not obeyed.
Warnings are conditions that can cause
injury to the operator, while Cautions can
cause damage to the equipment.
2
!
!
!
!
Typical Sterilization Procedures
CAUTION:
Do not use dry heat or chemical
sterilization or damage to the parts
may result.
WARNING:
These items are not sterile when
shipped to the customer. Before use
they must be sterilized using steam
or ethylene oxide (EtO).
All parts are fabricated from the highest
quality aluminum, stainless steel and
plastic, and are designed to withstand
high temperatures:
•Steam sterilization (autoclave):
Use the normal cycle (250 degrees F or
121 degrees centigrade at 15 psi for 15
to 30 minutes) or the ash cycle (270
degrees F or 132 degrees centigrade
at 30 psi for 3 to 12 minutes).
•Ethylene oxide (EtO) sterilization:
Use cycle and aeration times recom-
mended by the manufacturer of the
Sterilizer or use those determined at
the hospital.
Exterior cleaning should be done with
a soft brush and a cloth. Gently brush
all surfaces to remove dirt and dust. To
remove any remaining dirt, use a cloth
slightly dampened with a solution of mild
detergent and water or a liquid disinfect-
ing agent.

Table of Contents
3
General Precautions
General Operation
Maintenance
Parts and Accessories List
Features and Specications
Service Policy
Customer Responsibility
Warranty
2
3
4
4
4
7
7
8
General Operation
PAGE
1. Attach the Seed Alignment Tray to the
Loading Cover of the Seed Sterilization
and Sorting Tray by removing the two
screws on top of the Loading Cover. Align
the captive screws on the Alignment Tray
to the nuts revealed in the Loading Cover.
Secure nger-tight.
2. Sterilize the assembly with the loaded
Sterilization Tray and spatula.
3. Remove the Sterilizaion Cover and attach
the Loading Cover.
4. Use the sterilized spatula to scoop out the
seeds from the wells.
5. Pour spacers onto the Seed Alignment
Tray.
6. Align the seeds and spacers in the slots
of the tray.
7. Use a SeedVac to pick up the seeds and
spacers.
8. Proceed with loading the needles.
The Seed Alignment Tray is used with a SeedVac™ to conveniently load needles and
reduce radiation exposure. The Seed Alignment Tray can be used alone or connected
to the Seed Sterilization and Sorting Tray.
7601 MURPHY DRIVE
MIDDLETON, WI 53562

4
As is standard practice, it is recommended that the Seed Alignment Tray be examined after
each use for deterioration. To remove any dirt, use a cloth slightly dampened with a solution
of mild detergent and water or a liquid disinfecting agent.
There are no serviceable parts on the Seed Alignment Tray.
If assistance is desired in the proper disposal of this product (including accessories and
components), after its useful life, please return to Standard Imaging.
REF
90089
80201
Description
Seed Alignment Tray
Instruction Manual
Features and Specications
Dimensions
Width
Depth
Material
4.5 in. (11.4 cm)
4.5 in. (11.4 cm)
Aluminum
Parts and Accessories
Specications are subject to change without notice.
Maintenance

5
Notes

6
Notes

This product and its components will perform
properly and reliably only when operated and
maintained in accordance with the instruc-
tions contained in this manual and accompa-
nying labels. A defective device should not be
used. Parts which may be broken or missing
or are clearly worn, distorted or contaminated
should be replaced immediately with genuine
replacement parts manufactured by or made
available from Standard Imaging Inc.
CAUTION: Federal law in the U.S.A.
and Canadian law restrict the sale,
distribution, or use of this product
to, by, or on the order of a licensed
medical practitioner. The use of this
product should be restricted to the
supervision of a qualified medical
physicist. Measurement of high activ-
ity radioactive sources is potentially
hazardous and should be performed
by qualied personnel.
Should repair or replacement of this product
become necessary after the warranty pe-
riod, the customer should seek advice from
Standard Imaging Inc. prior to such repair
or replacement. If this product is in need of
repair, it should not be used until all repairs
have been made and the product is function-
ing properly and ready for use. After repair,
the product may need to be calibrated. The
owner of this product has sole responsibility
for any malfunction resulting from abuse,
improper use or maintenance, or repair by
anyone other than Standard Imaging Inc.
The information in this manual is subject to
change without notice. No part of this manual
may be copied or reproduced in any form or
by any means without prior written consent
of Standard Imaging Inc.
7
Customer Responsibility
!
If service, including recalibration, is required,
please contact Standard Imaging’s Customer
Service department by phone or email prior
to shipping the product. Standard Imaging’s
Customer Service and Technical Service
staff will attempt to address the product is-
sue via phone or email. If unable to address
the issue, a return material authorization
(RMA) number will be issued. With the
RMA number, the product can be returned
to Standard Imaging. It is the responsibility
of the customer to properly package, insure
and ship the product, with the RMA number
clearly identied on the outside of the pack-
age. The customer must immediately le
a claim with their carrier for any shipping
damage or lost shipments. Return shipping
and insurance is to be pre-paid or billed to
the customer, and the customer may request
a specic shipper. Items found to be out of
warranty are subject to a minimum service
fee of 1 hour labor (excluding recalibrations)
for diagnostic efforts and require a purchase
order (PO) before service is performed. With
concurrence from customer, the product may
be replaced if it is unserviceable or if the
required service is cost prohibitive. Products
incurring service charges may be held for
payment. Standard Imaging does not provide
loaner products. See the Standard Imaging
Warranty and Customer Responsibility for
additional information.
Service Policy

8
Warranty
Standard Imaging, Inc. sells this product under the warranty herein set forth. The war-
ranty is extended only to the buyer purchasing the product directly from Standard Imaging,
Inc. or as a new product from an authorized dealer or distributor of Standard Imaging, Inc.
For a period provided in the table below from the date of original delivery to the pur-
chaser or a distributor, this Standard Imaging, Inc. product, provided in the table is
warranted against functional defects in design, materials and workmanship, provided
it is properly operated under conditions of normal use, and that repairs and replace-
ments are made in accordance herewith. The foregoing warranty shall not apply if
the product has been altered, disassembled or repaired other than by Standard Imag-
ing, Inc. or if the product has been subject to abuse, misuse, negligence or accident.
Product Warranty Period
Standard Imaging Ionization Chambers 2 years
Standard Imaging Well Chambers 2 years
Standard Imaging Electrometers 2 years
Standard Imaging Software Products 1 year
All Other Standard Imaging Products 1 year
Standard Imaging Custom Product 90 days
Consumables 90 days
Serviced Product 90 days
Resale Products As dened by the Original Equipment Manufacturer
ADCL Product Calibration
(Standard Imaging uses the UW-ADCL for
recalibrations required under warranty)
0 - 90 days = 100% of ADCL Calibration Costs
91 - 182 days = 75% of ADCL Calibration Costs
183 – 365 days = 50% of ADCL Calibration Costs
366 – 639 days = 25% of ADCL Calibration Costs
(days from date of shipment to customer)
Standard Imaging’s sole and exclusive obligation and the purchaser’s sole and exclusive
remedy under the above warranties are, at Standard Imaging’s option, limited to repairing,
replacing free of charge or revising labeling and manual content on, a product: (1) which con-
tains a defect covered by the above warranties; (2) which are reported to Standard Imaging,
Inc. not later than seven (7) days after the expiration date of the warranty period in the table;
(3) which are returned to Standard Imaging, Inc. promptly after discovery of the defect; and (4)
which are found to be defective upon examination by Standard Imaging Inc. Transportation
related charges, (including, but not limited to shipping, customs, tariffs, taxes, and brokerage
fees) to Standard Imaging are the buyer’s responsibility. This warranty extends to every part
of the product except consumables (fuses, batteries, or glass breakage). Standard Imaging,
Inc. shall not be otherwise liable for any damages, including but not limited to, incidental
damages, consequential damages, or special damages. Repaired or replaced products are
warranted for the balance of the original warranty period, or at least 90 days.
This warranty is in lieu of all other warranties, express or implied, whether statutory or otherwise,
including any implied warranty of tness for a particular purpose. In no event shall Standard Im-
aging, Inc. be liable for any incidental or consequential damages resulting from the use, misuse
or abuse of the product or caused by any defect, failure or malfunction of the product, whether
a claim of such damages is based upon the warranty, contract, negligence, or otherwise.
This warranty represents the current standard warranty of Standard Imaging, Inc. Please refer
to the labeling or instruction manual of your Standard Imaging, Inc. product or the Standard
Imaging, Inc. web page for any warranty conditions unique to the product.
Table of contents
Other Standard Imaging Medical Equipment manuals