SUPMEA SUP-C703S User manual

User's Manual
Signal generator
U-S4.1-S4.2-MYEN1
Supmea Automation Co.,Ltd.
www.supmea.com
Headquarters
5th floor,Building 4,Singapore Hangzhou Science Technology Park,
No. 6 street,Hangzhou Economic Development Area,
Hangzhou 310018,China
Singapore / Philippines

I
Preface
Thank you for purchasing Signal calibrator. Please read this
manual carefully before operating and using it correctly to avoid
unnecessary losses caused by false operation.
Note
Modification of this manual’s contents will not be
notified as a result of some factors, such as function
upgrading.
We try our best to guarantee that the manual content
is accurate, if you find something wrong or incorrect,
please contact us.
This product is forbidden to use in explosion-proof
occasions.
Version
U-S4.1-S4.2-MYEN1

II
Safety Precautions
In order to use this product safely, be sure to follow the safety
precautions described.
About this manual
Please submit this manual to the operator for reading.
Please read the operation manual carefully before
applying the instrument. On the precondition of full
understanding.
This manual only describes the functions of the
product. The company does not guarantee that the
product will be suitable for a particular use by the
user.
Precautions for protection, safety and modification of this
product
To ensure safe use of this product and the systems it
controls, Please read carefully the operation manual
and understand the correct application methods
before putting into operation, to avoid unnecessary
losses due to operation mistakes. If the instrument is
operated in other ways not described in the manual,
the protections that the instrument give may be
destroyed, and the failures and accidents incurred

III
due to violation of precautions shall not be borne by
our company.
When installing lightning protection devices for this
product and its control system, or designing and
installing separate safety protection circuits for this
product and its control system, it needs to be
implemented by other devices.
If you need to replace parts of the product, please
use the model specifications specified by the
company.
This product is not intended for use in systems that
are directly related to personal safety.Such as
nuclear power equipment, equipment using
radioactivity, railway systems, aviation equipment,
marine equipment, aviation equipment and medical
equipment.If applied, it is the responsibility of the
user to use additional equipment or systems to
ensure personal safety.
Do not modify this product.
The following safety signs are used in this manual:
Hazard, if not taken with appropriate
precautions, will result in serious personal
injury, product damage or major property
damage.

IV
Warning:Pay special attention to the
important information linked to product or
particular part in the operation manual.
Confirm if the supply voltage is in consistent with the
rated voltage before operation.
Don’t use the instrument in a flammable and
combustible or steam area.
To prevent from electric shock, operation mistake, a
good grounding protection must be made.
Thunder prevention engineering facilities must be
well managed: the shared grounding network shall
be grounded at is-electric level, shielded, wires shall
be located rationally, SPD surge protector shall be
applied properly.
Some inner parts may carry high voltage. Do not
open the square panel in the front except our
company personnel or maintenance personnel
acknowledged by our company, to avoid electric
shock.
Cut off electric powers before making any checks, to
avoid electric shock.
Check the condition of the terminal screws regularly.

V
If it is loose, please tighten it before use.
It is not allowed to disassemble, process, modify or
repair the product without authorization, otherwise it
may cause abnormal operation, electric shock or fire
accident.
Wipe the product with a dry cotton cloth. Do not use
alcohol, benzine or other organic solvents. Prevent
all kinds of liquid from splashing on the product. If
the product falls into the water, please cut off the
power immediately, otherwise there will be leakage,
electric shock or even a fire accident.
Please check the grounding protection status
regularly. Do not operate if you think that the
protection measures such as grounding protection
and fuses are not perfect.
Ventilation holes on the product housing must be
kept clear to avoid malfunctions due to high
temperatures, abnormal operation, shortened life
and fire.
Please strictly follow the instructions in this manual,
otherwise the product's protective device may be
damaged.

VI
Don’t use the instrument if it is found damaged or
deformed at opening of package.
Prevent dust, wire end, iron fines or other objects
from entering the instrument during installation,
otherwise, it will cause abnormal movement or
failure.
During operation, to modify configuration, signal
output, startup, stop, operation safety shall be fully
considered. Operation mistakes may lead to failure
and even destruction of the instrument and
controlled equipment.
Each part of the instrument has a certain lifetime,
which must be maintained and repaired on a
regular basis for long-time use.
The product shall be scrapped as industrial wastes,
to prevent environment pollution.
When not using this product, be sure to turn off the
power switch.
If you find smoke from the product, smell odor,
abnormal noise, etc., please turn off the power
switch immediately and contact the company in
time.

VII
Disclaimer
The company does not make any guarantees for the
terms outside the scope of this product warranty.
This company is not responsible for damage to the
instrument or loss of parts or unpredictable damage
caused directly or indirectly by improper operation of
the user.
Package contents
No.
Item Name
Quantity
Remarks
1
Signal calibrator
1
2
Test lead
3
One for each color:
red, black and
yellow
3
Manual
1
4
Certificate
1
After opening the box, please confirm the package contents
before starting the operation.If you find that the model and
quantity are incorrect or there is physical damage in
appearance, please contact us.

VIII
Contents
Chapter 1 Overview........................................................................ 1
1.1 Introduction..................................................................1
1.2 Functions..................................................................... 1
1.3 Parameter....................................................................2
1.4 Specifications.............................................................. 4
Chapter 2 Component and functions............................................ 5
2.1 Terminal....................................................................... 6
2.2 Navigation button........................................................6
2.3 LCD display................................................................. 8
Chapter 3 Signal output.................................................................. 9
3.1 Voltage and Active Current........................................9
3.2 4-20 mA....................................................................... 9
3.3 Thermocouple...........................................................10
3.4 Resistance/Thermal Resistance............................ 11
3.5 Passive Current........................................................ 11
3.6 Output or Measurement of Voltage and Current
Signals based on Display Range (Free from Range
Conversion)..................................................................... 12

Contents
IX
Chapter 4 Signal Measurement...................................................14
4.1 Voltage and Active Current..................................... 14
4.2 Passive Current........................................................ 15
4.3 Resistance/Thermal Resistance............................ 17
4.4 Thermocouple...........................................................17
4.5 Regulating Valve...................................................... 19
Chapter 5 Programmable Output................................................ 20
5.1 Split Output Function (n/m)..................................... 20
5.2 Linear Output Function............................................ 21
5.3 Automatic Stepping Function.................................. 22
Chapter 6 Troubleshooting and maintenance........................... 24
6.1 Troubleshooting........................................................ 24
6.2 Maintenance............................................................. 25


Chapter 1 Overview
www.supmea.com
-1-
Chapter 1 Overview
1.1 Introduction
This signal calibrator can measure and output various signals,
including voltage, current, thermocouple, resistance, and
thermal resistance signals. An HD LCD and silicone button with
distinct functions are used, providing clear display and easy
operation. The calibrator has a long standby time, a high
precision level and programmable output. It is widely used for
commissioning for laboratories, industrial site PLC and process
instruments and electric valves.
1.2 Functions
Voltage signal: output and measurement of 0-30 V,
0-24 mV, and 0-100 mV
(Maximum output is 15 V)
Current signal: output and measurement of active
and passive 0-24 mA and 4-20 mA
Thermocouple: output and measurement of K, E, J, T,
R, B, S and N
Resistance/thermal resistance: output and
measurement of 0-400 Ω and -199°C-650°C (the
minimum resistance output is 15Ω)
* Note: Resistance/thermal resistance is only for SUP-C703S

Chapter 1 Overview
-www.supmea.com
-2-
1.3 Parameter
Table 1 Table for measurement and output functions
Item
Signal
Range
Accuracy
Resolution
Remarks
DC voltage
20 mV
0.00-
24.00mV
±0.2%
0.01 mV
100 mV
0.0-100.0mV
±0.2%
0.1 mV
V
Output 0.00-
15.00 V
±0.2%
0.01 V
Output:
maximum
current 30 mA
Measurement:
input
impedance1.2
MΩ
Output 0.00-
30.00 V
±0.2%
0.01 V
DC current
mA
0.00- 24.00
mA
±0.2%
0.01 mA
Output:
maximum load
750 Ω
Measurement:
Output
impedance 100
Ω
4-20
mA
4/8/12/16/20
mA
±0.2%
0.01 mA
Passive
current
mA
0.00- 24.00
mA
±0.2%
0.01 mA
Output:
external power
supply 16-30 V

Chapter 1 Overview
www.supmea.com
-3-
Electric
power
distribution
24 V
LOOP
24 V/16 V
10%
/
Drive current 24
mA
Thermo-
couple
K
0-1372°C
±1%
1°C
The output or
measurement
temperature
cannot be
lower than cold
conjunction
compensation
temperature.
Note: The
minimum
excitation
current is 0.5
mA and the
maximum
excitation
current is 3
mA.
E
0-1000°C
±1%
1°C
J
0-1200°C
±1%
1°C
T
0-400°C
±1%
1°C
R
0-1768°C
±1%
1°C
B
250-1820°C
±1%
1°C
S
0-1768°C
±1%
1°C
N
0-1300°C
±1%
1°C
Resistance
Ω
Output
15.0-400.0Ω
±0.2%
0.1 Ω
Measuremen
t 0.0-400.0 Ω
±0.2%
0.1 Ω
Thermal
resistance
PT100
-199.0°C-
650.0°C
±0.2%
0.1°C

Chapter 1 Overview
-www.supmea.com
-4-
1.4 Specifications
Table 2 Specifications
Operating Temperature
and Humidity
-10°C-55°C,20%-80% RH (non-condensing)
Storage temperature
-20°C - 70°C
Dimension
115x71x30 (mm)
Weight
143g
Power supply
4*AAA batteries or external 5 V/1A adapter
Power consumption
About 200 mA; with power supplied by 4*AAA
batteries ( each nominal capacity of 1100
mAh ), can be used for 4 hours with a full load
and 17 hours standing by.
Reverse connection and
Over current protection
30 V

Chapter 2 Component Names and Functions
www.supmea.com
-5-
Chapter 2 Component and functions
Figure 1 Front view of the signal calibrator

Chapter 2 Component Names and Functions
-www.supmea.com
-6-
2.1 Terminal
The port for the test pen is as shown in Figure 1.
Please see Table 3 to its function .
If the measurement end is used for signal
measurement, insert the test pen into ports ①and ③
in Figure 1.
If the output end is used for signal output, insert the
test pen into ports ①and ②in Figure 1.
Table 3 Terminals function description
No.
Name
Description
①
Black terminal
Common
②
Yellow terminal
Output signal
③
Red terminal
Measurement signal
2.2 Navigation button
Table 4 Button and switch description
No.
Button
Description
④
Move the decimal point/switch between positive
sign and negative sign
⑤
Turn on or off the measurement function.
⑥
Display/change cold conjunction compensation
(only works for measuring the thermocouple).

Chapter 2 Component Names and Functions
www.supmea.com
-7-
⑦
Turn on or off the power.
⑧
Turn on or off the signal output.
⑨
Switch
(OFF by
default)
1. Automatic power-off: the device will
automatically be turned off if there is no operation
for 10 minutes.
2. Manual cold conjunction compensation: set
cold conjunction compensation value manually
for thermocouple measurement.
3. Passive output: output passive current signals
for analog transmitters.
4. Low-load mode: during passive current input,
output 16 V voltage to the transmitter to lower the
power dissipation and prolong the service life.
⑩
Switch the type of measurement/output signals.
Switch the measurement/output range.
Turn on the programming function.
Switch the programmable output wave form
(linear output/stepping output))
Increase or decrease the value.

Chapter 2 Component Names and Functions
-www.supmea.com
-8-
2.3 LCD display
A Measurement value display: four digits with a unit
B. Source signal value: four digits with a unit
20mV/100mV/4-20mA/K/E/J/T/R/B/S/N。
C. Signal and cold conjunction compensation mode:
20mV/100mV/4-20mA/K/E/J/T/R/B/S/N
RJA is the automatic cold end and M is manual.
D. Programming function:
n/m is split output. The output value = (main set value)
x (n/m)
“SWEEP” is linear output that outputs signals in a
linear manner based on the time you set.
“STEP” is stepping output that outputs signals in a
stepping manner based on the time you set.
“TIME” is the output time for each step, configurable
from 0 to 999 s.
“COUNT” is the number of output cycle, configurable
from 0 t0 999. 0 means an infinite loop.
E. Unit: mA/mV/V/°C
F. Range switch function: RL is the lower limit of the
display range, while RH is the upper limit. SL is the
lower limit of the signal, while SH is the upper limit.
G. Battery level: the real-time battery level is
displayed. Three bars are the full level.

Chapter 3 Signal Output
www.supmea.com
-9-
Chapter 3 Signal output
The signal calibrator can source voltage, active current,
passive current, thermocouple signal, and PT100 signal.
* Note: PT100 signal is available only for SUP-C703S
3.1 Voltage and Active Current
Connect the black signal cable to the public end and
the yellow signal cable to the output end.
Press the yellow signal button to switch the
signal type.
Press or to adjust the output value.
Press the yellow source button , and the value of
on the LCD will turn from to
and the output is started.
3.2 4-20 mA
The 4-20 mA signal increases/decreases by 4 mA each time.
Select 4-20 mA to be the signal type.
Use the corresponding button of 4-20 mA to
set the output value in a stepping manner, with an
increase/decrease of 4 mA each time in the order of
4→8→12→16→20. You can also use the regular
Table of contents