tecan sunrise User manual

TECAN
Instructions for Use for
sunrise Absorbance Reader
+ Option Touchscreen Color
Document Part No.: 30008746
2006-03
Document Revision No.: 1.4

TECAN
2 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
Tecan Affiliates and Service Centers
Austria
Tecan Austria GmbH
Untersbergstrasse 1a
A-5082 Grödig/Salzburg
Austria
T +43 6246 8933 256
F +43 6246 72 770
Austria
Tecan Sales International GmbH
Untersbergstrasse 1a
A-5082 Grödig/Salzburg
Austria
Tel.: +43 6246 8933
Fax: +43 6246 72 770
Asia
Tecan Asia Pte Ltd.
80 Marine Parade
#10-09 Parkway Parade
Singapore 449269
Singapore
T +65 6444 1886
F +65 6444 1836
tecan@tecan.com.sg
Belgium
Tecan Benelux bvba
Vaartdijk 55
B-2800 Mechelen
Belgium
T +32 15 42 13 19
F +32 15 42 16 12
tecan-be@tecan.com
China
Tecan Group Ltd, Beijing
Representative Office
Room 2502, Building A
Jianwai SOHO
No. 39 Dongsanhuan Zhong Rd.
100022 Beijing
China
T +86 10 5869 5936
F +86 10 5869 5935
France
Tecan France SAS
26 avenue Tony Garnier
F-69007 Lyon
France
T +33 820 88 77 36
F +33 472 76 04 99
Germany
Tecan Deutschland GmbH
Theodor-Storm-Strasse 17
D-74564 Crailsheim
Germany
T +49 1805 8322 633 or
T +49 1805 TECANDE
F +49 7951 9417 92
Italy
Tecan Italia S.r.l.
Via F.lli Cervi
Palazzo Bernini
Centro Direzionale Milano 2
I-20090 Segrate (Mi)
Italy
T +39 02 215 21 28
F +39 02 215 97 441
Japan
Tecan Japan Co., Ltd.
Kawasaki Tech Center
580-16, Horikawa-cho,
Saiwai-ku
Kawasaki, Kanagawa 212-0013
Japan
T +81 44 556 7311 (Kawasaki)
F +81 44 556 7312 (Kawasaki)
T +81 66 305 8511 (Osaka)
Netherlands
Tecan Benelux bvba
Industrieweg 30
NL-4283 GZ Giessen
Netherlands
T +31 183 44 81 74
F +31 183 44 80 67
tecan-nl@tecan.com
Scandinavia
Tecan Nordic AB
Taljegårdsgatan 1
SE-431 53 Mölndal
Sweden
T +46 317 54 40 00
F +46 317 54 40 10
Spain (Barcelona)
Tecan Ibérica
Sabino de Arana 32
E-08028 Barcelona
Spain
T +34 93 490 0174
T +34 93 409 1237 (Helpdesk)
F +34 93 411 2407
helpdesk-sp@tecan.com
Spain (Madrid)
Tecan Ibérica
Gobelas 17 bjo. Izq. Urb. La Florida
E-28023 Madrid
Spain
T +34 91 151 7107
T +34 93 409 1237 (Helpdesk)
F +34 91 151 7120
helpdesk-sp@tecan.com
Portugal
Tecan Ibérica
Quinta da Fonte Edificio Pedro I
P-2780-730 Paço D'Arcos
Portugal
T +35 21 000 8216
Switzerland
Tecan Schweiz AG
Seestrasse 103
CH-8708 Männedorf
Switzerland
T +41 44 922 82 82
F +41 44 922 84 84
helpdesk-ch@tecan.com
United Kingdom
Tecan UK
Theale Court
11-13 High Street
Theale
UK-Reading RG7 5AH
United Kingdom
T +44 118 930 0300
F +44 118 930 5671
USA
Tecan US
4022 Stirrup Creek Road
Suite 310
Durham, NC 27703
USA
T +1 919 361 5200
F +1 919 361 5201
Toll Free US:
T +1 800 TECAN US or
T +1 800 832 2687
USA
Tecan Systems Inc.
2450 Zanker Road
San Jose, CA 95131
USA
T +1 408 953 3100
F +1 408 953 3101
Toll Free US/Canada:
T +1 866 798 3226

TECAN
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 3
WARNING
CAREFULLY READ AND FOLLOW THE INSTRUCTIONS PROVIDED IN
THIS PUBLICATION BEFORE OPERATING THE INSTRUMENT.
Notice
Every effort has been made to avoid errors in text and diagrams; however, Tecan
Austria GmbH assumes no responsibility for any errors, which may appear in this
publication.
It is the policy of Tecan Austria GmbH to improve products as new techniques
and components become available. Tecan Austria GmbH therefore reserves the
right to change specifications at any time with appropriate verification, validation,
and approvals.
We would appreciate any comments on this publication.
Manufacturer
Tecan Austria GmbH
Untersbergstraße 1A
A-5082 Grödig/Salzburg
AUSTRIA / EUROPE
Telephone: 0043 (0)6246/8933
FAX: 0043 (0) 6246/72770
Copyright Information
The contents of this publication are the property of Tecan Austria GmbH and are
not to be copied, reproduced or transferred to another person or persons without
prior written permission.
Copyright ©Tecan Austria GmbH
All rights reserved.
Printed in Austria.
Declaration for EU Certificate
Provided upon request where appropriate.
sunrise Absorbance Reader Intended Use
See chapter 2.2.1 Intended Use.
About the Instructions For Use
This publication is intended as Instructions for Use for the sunrise Absorbance
Reader, which is designed to measure the light absorbance (optical density) of
samples in 96 well microplates. It is intended as reference and instruction for the
user.
This document instructs how to:
• Install the instrument
• Clean and maintain the instrument
Warranty
3 Year Warranty
As an expert for microplate instrumentation Tecan proves its commitment to
quality and offers a unique 3 year warranty for your sunrise Absorbance Reader
as a standard. Warranty will be invalid if the instrument is opened or modified.

TECAN
4 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
Warnings, Cautions and Notes
The following types of notices are used throughout this publication to highlight
important information or to warn the user of potentially dangerous situations:
Note
Gives helpful information.
Caution
Indicates a possibility of instrument damage or data loss if instructions
are not followed.
WARNING
INDICATES THE POSSIBILITY OF SEVERE PERSONAL INJURY, LOSS
OF LIFE OR EQUIPMENT DAMAGE IF THE INSTRUCTIONS ARE NOT
FOLLOWED.
WARNING
INDICATES THE POSSIBLE PRESENCE OF BIOLOGICALLY
HAZARDOUS MATERIAL. PROPER LABORATORY SAFETY
PRECAUTIONS MUST BE OBSERVED.
WARNING
INDICATES LASER. DO NOT STARE INTO THE BEAM!
ATTENTION
Negative environmental impacts associated with the treatment of WEEE
• Do not treat electrical and electronic equipment as unsorted municipal
waste.
• Waste of electrical and electronic equipment must be collected
separately.

TECAN
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 5
Symbols
Manufactured by
Date of manufacture
Conformité Européenne
In vitro diagnostic medical device
Laser
Bio-hazardous
IVD

TECAN
6 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
Abbreviations
Abbreviation
A Ampere
Abs. Absorbance
ADC Analog digital converter
C Celsius
CE Conformité Européenne
cm Centimeter
F Fahrenheit
Hz Hertz
IVD In vitro diagnostics
IVD-D In vitro diagnostics Directive
kg Kilogram
l; L Liter
LED Light emitting diode
MB Megabyte
ml Milliliter
nm Nanometer
OD Optical Density
REF Reference Number/ Order Number
SN Serial Number
TYPE Name and type of instrument
USB Universal serial bus
V Volts
VA Volt ampere
VGA Video graphics array
VOLTAGE Voltage

TECAN
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 7
Table of Contents
1. Safety
1.1 Instrument Safety .......................................................................... 1-1
2. General
2.1 Introduction.................................................................................... 2-1
2.2 Area of Application........................................................................ 2-3
2.2.1 Intended Use................................................................................... 2-3
2.2.2 Variations of sunrise ........................................................................ 2-4
2.2.3 Available Options for sunrise ........................................................... 2-5
2.3 Specifications ................................................................................ 2-6
2.3.1 General............................................................................................ 2-6
2.3.2 Standard Instrument (Standard Filter Carriage) .............................. 2-7
2.3.3 Touchscreen Color Instrument........................................................ 2-8
2.3.4 Tuneable Wavelength Instrument (Gradient Filter Carriage)........... 2-9
2.3.5 6 Filter Instrument (6 Filter Carriage) .............................................. 2-9
2.3.6 Temperature Control ..................................................................... 2-10
2.3.7 Barcode Laser Scanner................................................................. 2-11
2.3.8 Microplates.................................................................................... 2-11
2.3.9 Barcode Labels.............................................................................. 2-11
2.3.10 USB Stick ...................................................................................... 2-12
2.4 Instrument Description ............................................................... 2-13
2.4.1 Back Panel Connections ............................................................... 2-14
2.4.2 Handling the Microplate................................................................. 2-15
2.5 Filter Carriage Description ......................................................... 2-16
2.5.1 Standard Filter Carriage ................................................................ 2-16
2.5.2 Gradient Filter Carriage................................................................. 2-16
2.5.3 6 Filter Carriage............................................................................. 2-17
2.6 Instrument Features.................................................................... 2-18
2.6.1 Measurement Modes..................................................................... 2-18
2.6.2 Microplate Shaking........................................................................ 2-19
2.7 Instrument Accessories.............................................................. 2-19
2.8 Software for the sunrise Absorbance Reader............................ 2-20
3. Installation Procedure
3.1 Introduction.................................................................................... 3-1
3.2 Unpacking and Inspection............................................................ 3-1
3.3 Unpacking Procedure ................................................................... 3-2
3.4 Power Requirements..................................................................... 3-2
3.5 Environmental Requirements....................................................... 3-2
3.6 Instrument Installation Procedure ............................................... 3-3
3.7 Installation of Instrument Control Software................................ 3-3
3.8 Defining the Instrument Settings................................................. 3-4
3.8.1 Installation of “Sunrise Instrument Settings” Software..................... 3-4

TECAN
8 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
3.8.2 Starting the “Sunrise Instrument Settings” .......................................3-5
3.8.3 Define Instrument Mode...................................................................3-5
3.8.4 Define Filter......................................................................................3-6
3.8.5 Define Measurement Mode..............................................................3-7
4. The sunrise Touchscreen
4.1 Introduction ....................................................................................4-1
4.2 Area of Application for the Touchscreen Option........................4-2
4.3 Specifications.................................................................................4-2
4.4 Instrument Description..................................................................4-3
4.4.1 Touchscreen Back Panel Connections ............................................4-3
4.4.2 Data Handling ..................................................................................4-4
4.4.3 Printer Compatibility.........................................................................4-5
4.4.4 Handling the USB Stick....................................................................4-6
4.4.5 USB Stick Data Management...........................................................4-7
4.5 Operating the Touchscreen...........................................................4-8
4.5.1 Introduction ......................................................................................4-8
4.5.2 Starting the Software........................................................................4-8
4.5.3 Special Features of MagellanCE......................................................4-9
4.6 Change Instrument Control Mode ..............................................4-11
4.6.1 Set Instrument in Remote Mode.....................................................4-11
4.6.2 Set Instrument in Local Mode.........................................................4-12
5. Error Messages and Trouble Shooting
5.1 Introduction ....................................................................................5-1
5.1.1 Table of Error Messages and Trouble Shooting
for sunrise Mode...............................................................................5-1
5.1.2 Table of Error Messages and Trouble Shooting
for SPECTRA Mode.........................................................................5-4
5.1.3 Table of Error Messages and Trouble Shooting
for Touchscreen ...............................................................................5-5
5.2 Definition of ‘Overflow’..................................................................5-6
6. Maintenance & Cleaning
6.1 Introduction ....................................................................................6-1
6.2 Lamp Replacement ........................................................................6-2
6.3 Filter Replacement.........................................................................6-4
6.3.1 Standard Filter Carriages.................................................................6-4
6.3.2 Gradient Filter Carriages..................................................................6-5
6.3.3 6 Filter Carriages..............................................................................6-6
6.4 Fuse Replacement..........................................................................6-7
6.5 Cleaning the Instrument................................................................6-8
6.5.1 Cleaning the Device.........................................................................6-8
6.5.2 Cleaning the Touchscreen ...............................................................6-8
6.5.3 Liquid Spills......................................................................................6-8
6.6 Preventive Maintenance Plan for sunrise......................................6-9

TECAN
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 9
6.6.1 Daily................................................................................................. 6-9
6.6.2 Weekly............................................................................................. 6-9
6.6.3 Every Six Months............................................................................. 6-9
6.6.4 Yearly (Customer or Service Technician)........................................ 6-9
6.6.5 Every Four Years (Service Technician Required) ........................... 6-9
6.7 Instrument Disinfection .............................................................. 6-10
6.7.1 Disinfection Solutions .................................................................... 6-10
6.7.2 Disinfection Procedure .................................................................. 6-10
6.8 Disinfection Declaration ............................................................. 6-12
6.9 Disposal of Instrument................................................................ 6-13
6.9.1 Introduction.................................................................................... 6-13
6.9.2 Disposal of Packing Material......................................................... 6-13
6.9.3 Disposal of Operating Material...................................................... 6-13
6.9.4 Disposal of the Instrument............................................................. 6-14
7. Quality Control
7.1 Introduction.................................................................................... 7-1
7.2 Optimizing for Maximum Performance........................................ 7-2
7.2.1 Instrument Location......................................................................... 7-2
7.2.2 Operating Procedure....................................................................... 7-2
7.2.3 Self Check Procedure...................................................................... 7-3
7.3 Operational Qualification (OQ)..................................................... 7-4
7.3.1 QC PAC 2........................................................................................ 7-4
7.3.2 Microplate Test................................................................................ 7-4
7.3.3 High Meniscus Liquids..................................................................... 7-6
7.3.4 Instrument Precision with Liquid Samples....................................... 7-8
7.3.5 Instrument Linearity with Liquid Samples........................................ 7-9
Index


TECAN 1. Safety
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 1-1
1. Safety
1.1 Instrument Safety
1. Always follow basic safety precautions when using this product to reduce
the risk of injury, fire or electrical shock.
2. Read and understand all information in this document. Not reading,
understanding or not observing the instructions may result in damage to the
product, injury to operating personnel or poor instrument performance.
Tecan is not responsible for damage or injuries resulting from incorrect
handling of the device.
3. Observe all WARINGS and CAUTION statements in this
publication.
4. Always disconnect the device from the main power supply prior to
maintenance, cleaning, and disinfection.
5. Never open the sunrise while the instrument is plugged into a power source.
6. Observe suitable safety precautions for laboratories such as wearing of
protective clothing and the application of appropriate and tested laboratory
safety procedures.
WARNING
IF THE sunrise ABSORBANCE READER OR THE MAGELLAN
SOFTWARE ARE MODIFIED IN ANY WAY, THE PERFORMANCE OF THE
READER MAY BE NEGATIVELY AFFECTED,
THE WARRANTY WILL NO LONGER BE VALID AND THE INSTRUMENT
WILL NOT BE CE CONFORM.


TECAN 2. General
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2-1
2. General
2.1 Introduction
Note
sunrise Absorbance Reader is intended for use with external software only.
Caution
Before starting measurements, make sure that the microplate
position A1 is inserted correctly.
The sunrise instruments are fully automatic, microprocessor controlled readers
designed for professional use, enabling the user to measure the light absorbance
(optical density) of samples in 96 well microplates according to the specifications
described in this publication.
Note
Results obtained using the sunrise are influenced by the proper use of the
instrument and microplates, according to the instructions given in this
document, as well as the liquid compounds used (reagents, chemistry). The
instructions for use, for storage and for other manipulations in connection
with samples or reagents must be strictly followed. Taking this fact into
consideration, results must be interpreted carefully.
By reading twelve wells simultaneously, the instrument is able to measure a
microplate in approximately six seconds.
The obtained transmission values are converted into OD values according to the
following formula:
oII
TonTransmissi =
Io = incident light
I = detected light (after the sample)
The absorption (Optical Density) is the logarithm of the reciprocal transmission.
T
1
LogOD =
With an innovative range of options, this versatile microplate reader gives
diagnostic and research laboratories all the features for numerous purposes.
Based on a new design concept EPAC, the excellent optical performance and
high quality of the sunrise will guarantee fast, reproducible and accurate
measurements.
The sunrise is designed to be integrated into Tecan robotic systems.
The sunrise Absorbance Reader is an optical 96-well microplate reader for the
measurement of absorbance and turbidimetry of samples from the biological or
non-biological origin according to the specifications described in this document,
intended for professional use only.

2. General TECAN
2-2 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
The sunrise Absorbance Reader is available with Touchscreen, tuneable
wavelength, temperature control and barcode scanner options.
Magellan is a reader control and data reduction software package, and also can
be used for concentration and titer determination.
Note
It is important to note that the proper installation of the instrument and
Magellan software alone will not ensure regulatory compliance . A range
of policies and standard operating procedures according to local
regulations must also be established.
Make sure the instrument and the software are suitable for use with the reagents,
chemicals and microplates that are being used (see 7.2 Optimizing for Maximum
Performance and 7.3 Operational Qualification).
Caution
If the instructions given in this publication are not correctly followed,
the instrument may become damaged or procedures may not be
correctly performed and the safety of the instrument cannot be
guaranteed.
Note
For more information about the operating instructions,
see Magellan or XFluor4 Instructions for Use.
WARNING
CAREFULLY READ AND UNDERSTAND ALL INFORMATION IN THIS
DOCUMENT. FAILURE TO READ, UNDERSTAND AND FOLLOW THE
INSTRUCTIONS IN THIS PUBLICATION MAY RESULT IN DAMAGE TO
THE PRODUCT, INJURY TO OPERATING PERSONNEL OR POOR
INSTRUMENT PERFORMANCE.

TECAN 2. General
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2-3
2.2 Area of Application
2.2.1 Intended Use
The sunrise Absorbance Reader is an optical 96-well microplate reader for the
measurement of absorbance, including turbidimetry (optical density) of samples
from biological or non-biological origin. According to the specifications described
in this manual, the instrument is intended for professional use in:
Diagnostic laboratories, Pharmaceutical, biotechnological and life science
industry, for research use, for food analysis, and veterinary applications.
Note
If the sunrise Absorbance Reader or Magellan software
are modified in any way the instrument loses its guarantee
and regulatory compliance.
Note
The operating authority must use only CE-labeled test kits for clinical
diagnostic applications. The operating authority must assure that the
combination of a particular CE-labeled test kit used with the CE-labeled
sunrise Absorbance Reader has been designed to meet the IVD directive
98/79/EC or other relevant national or local regulations.
If the sunrise Absorbance Reader is used with other software other than
Magellan, additional validation must be carried out by the operating authority to
ensure the safety of the results. The operating authority is fully responsible for
any results obtained using software other than Magellan with the sunrise
Absorbance Reader.

2. General TECAN
2-4 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
2.2.2 Variations of sunrise
All variations of the sunrise Absorbance Reader and its options are designed to
meet the IVD directive 98/79/EC of European Union.
sunrise Hardware Variations
The following table shows the compatibility of the filter slides with each sunrise
hardware variation:
sunrise Variations 4 Filterslide 6 Filterslide Gradient
filterslide
Standard X
6 Filter X X
Tuneable wavelength X X X
sunrise Options
The sunrise options work with all hardware variations and accessories and can
be used in any combination:
sunrise Options
Temperature Control
Touchscreen
Barcode
sunrise Accessories
The sunrise accessories work with all hardware variations and options and can
be used in any combination:
IVD compliant for Europe
sunrise
Accessories Included in CE1 Separate CE2 Research Use
Only None/
Unrelated3
4 Filterslide X
6 Filterslide X
Gradient Filter X
sunrise Instrument
Settings X
XFluor 4 X
Magellan X
Magellan TS X
QC Pac 1 X
QC Pac 2 X
1 IVD compliant for Europe. Included in the CE declaration for the instrument.
2IVD compliant for Europe. CE declaration is separate from that of the instrument.
3 Can be used in an IVD environment. CE declaration does not exist or it is unrelated to IVD compliance.

TECAN 2. General
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2-5
2.2.3 Available Options for sunrise
The sunrise is a modular system, so you can create your own tailor-made
instrument that meets exactly your needs. Options such as a Touchscreen
combined with a WindowsCE based on-board software, free wavelength
selection, temperature control and a barcode scanner can be added onto the
basic system which is dedicated to remote controlled computer operation.
Factory Installed Options
Touchscreen
10037302
sunrise Absorbance Reader
Tuneable Wavelength
from 400 - 700 nm
10037306
Temperature Control
10037303
6 Filter
30004221
Upgrade in field possible
Barcode Scanner
10037304
Barcode Upgrade Kit
30000322

2. General TECAN
2-6 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
2.3 Specifications
The tables below list the specifications for the instruments.
2.3.1 General
For all instrument options:
PARAMETERS CHARACTERISTICS
Main power input 100 - 120 & 220 - 240 V ± 10% 50/60 Hz
(Auto sensing)
Consumption Stand by mode: approx.50 VA
Operational mode: max.110 VA
Fuse rating 2 x F 2.0 A / 250 V (Fast Blow)
Outside dimensions Width: 25.5 cm (11.22 inch)
Depth: 34.0 cm (13.39 inch)
Height: 14.5 cm (5.71 inch)
Height with touchscreen 21.0 cm (8.27 inch)
Weight 6 kg
Min. (Standard RC) 6.2 kg
Max. (incl. all options) 8.6 kg
Ambient temperature:
Operation 15°C to 35°C (59°F to 95°F)
Storage -20°C to 60°C (-4°F to 140°F)
Relative humidity:
Operation 20 % to 90 %
Storage 5 % to 95 %
Over voltage category II
Pollution degree 2
Method of disposal Contaminated waste
Environment See 3.5 Environmental Requirements for more
information.
Stability:
Normal measurements After 15 minutes warm up time
max. +/- 0.001 OD

TECAN 2. General
2006-03 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2-7
2.3.2 Standard Instrument (Standard Filter Carriage)
PARAMETERS CHARACTERISTICS
Measurement time:
dual wavelength
single wavelength
8 seconds
6 seconds
Wavelength range:
Standard
340 - 750 nm
Measurement range:
340 - 399 nm
400 - 750 nm
0 - 3.000 OD
0 - 4.000 OD
Resolution: 0.001 OD
Accuracy:
(492 nm) 0.000 - 2.000 OD
(492 nm) 2.000 - 3.000 OD
Better than +/- (1.0 % + 0.010 OD)
Better than +/- (1.5 % + 0.010 OD)
Precision:
(492 nm) 0.000 - 2.000 OD
(492 nm) 2.000 - 3.000 OD
Better than +/- (0.5 % + 0.005 OD)
Better than +/- (1.0 % + 0.005 OD)
Linearity:
(340 - 399 nm) 0.000 - 2.000 OD
(400 - 750 nm) 0.000 - 2.000 OD
2.000 - 3.000 OD
Better than +/- 2 %
Better than +/- 1 %
Better than +/- 1.5 %
Wavelength selection:
Standard filter
Narrow band interference filters. Up to
four filters can be mounted in a filter
carriage. The instrument can use up to
eight different filter carriages.
Filter wavelength accuracy Central wavelength +/- 2 nm
Filter bandwidth
At 50% transmission:
10 +/- 2 nm
Light source: Halogen lamp 20 W
All connected devices must be approved and listed as per EN 60950, UL 1950
or CSA C22.2 No. 950 for Data Processing Devices.
Computer interface:
Serial RS 232 C
Printer
300 - 38,400 baud
Only with the Touchscreen option.

2. General TECAN
2-8 Instructions for Use for sunrise Absorbance Reader No. 30008746 Rev. No. 1.4 2006-03
2.3.3 Touchscreen Color Instrument
PARAMETERS CHARACTERISTICS
Screen size 640 x 480 pixel, app. 19 cm (7.5´´) diagonal
Color VGA
Operating system WindowsCE
Operating software MagellanCE
Memory 64 MB (RAM)
Storage >= 32 MB (USB stick)
Computer interface:
Serial RS 232 C
Printer
USB master ‘USB stick’
USB master ‘Printer’
USB slave ‘ACS’
MagellanCE software
300 - 38,400 baud
Parallel connection to the printer.
Port for USB stick.
USB connection to the printer.
Only for service.
See Magellan Manual
Personal computer:
Interface
Operating system
For USB stick data exchange and/or backup.
USB port
Windows 98/98SE, ME, 2000, XP
Other manuals for sunrise
2
Table of contents
Other tecan Card Reader manuals