Thermo Scientific Locator Series User manual

Locator and Locator Plus
Cryogenic Storage Vessel
with or without Ultrasonic Level Monitor
Operation Manual and Parts List
LT509X9 (7008141) Rev. 5
© 2010 Thermo Fisher Scientific
All rights reserved
Printed in U.S.A.

Thermo Scientific
MANUAL NUMBER LT509X9 (7008141)
5 27087 6/11/11 Remove rate of change of validation test for level monitor (pg 6-1, 6-5) ccs
4 27323 4/13/11 Added new instructions to Section 6 ccs
3 27075/FR-2221 2/24/11 Added rack list to Replacement Parts ccs
2 -- 1/6/11 Corrected lid #CX509X10 to CV509X10 typo per Delaine/Ali (pg 5-1) ccs
1 26676/SI-10360 10/28/10 Updated per M Spence/E. Pickrell ccs
Preface
Cryogenic Storage Vessel i
Locator Catalog Number Locator Model Name
CY50925 w/o Monitor 8141 Locator Jr
CY50935 w/o Monitor 8143 Locator 4
CY50945 w/o Monitor 8145 Locator 8
CY50985 w/o Monitor 8197 Locator 6
CY509105 w/ Monitor 8201 Locator Jr.
CY509105CN w/ Monitor 8202 Locator Jr. China
CY509107 w/ Monitor 8246 Locator 4
CY509107CN w/ Monitor 8247 Locator 4 China
CY509110 w/ Monitor 8211 Locator 8
CY509110CN w/ Monitor 8212 Locator 8 China
CY509113 w/ Monitor 8198 Locator 6
Locator Plus Catalog Number Locator Plus Model Name
CY50925-70 w/o Monitor 8142 Locator Jr Plus
CY50935-70 w/o Monitor 8144 Locator 4 Plus
CY50985-70 w/o Monitor 8147 Locator 6 Plus
CY50945-70 w/o Monitor 8146 Locator 8 Plus
CY509106 w/ Monitor 8237 Locator Jr Plus
CY509106CN w/ Monitor 8238 Locator Jr Plus China
CY509108 w/ Monitor 8207 Locator 4 Plus
CY509108CN w/ Monitor 8208 Locator 4 Plus China
CY509109 w/ Monitor 8209 Locator 6 Plus
CY509109CN w/ Monitor 8210 Locator 6 Plus China
CY509111 w/ Monitor 8213 Locator 8 Plus
CY509111CN w/ Monitor 8214 Locator 8 Plus China

Thermo Scientificii Cryogenic Storage Vessel
Preface
Important Read this instruction manual. Failure to read, understand and follow the instructions in this manual
may result in damage to the unit, injury to operating personnel, and poor equipment performance. s
Caution All internal adjustments and maintenance must be performed by qualified service personnel. s
Caution It is recommended to have this vessel tested by the manufacturer or qualified cryovessel service
technician every 7-10 years, regardless of any problems (or lack thereof) you may have had in the past.
This will help insure your samples against sudden loss of liquid nitrogen due to vacuum failure. s
Material in this manual is for information purposes only. The contents and the product it describes are subject to
change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this
manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to
the use of this manual.

Thermo Scientific Cryogenic Storage Vessel iii
Preface
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this
symbol.
Extreme temperature hazards. Only qualified persons should perform procedures associated with this
symbol.
Potential biological hazards. Proper protective equipment and procedures must be used.
Marking of electrical and electronic equipment, which applies to electrical and electronic equipment
falling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the market
after 13 August 2005.
This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Fisher
Scientific has contracted with one or more recycling/disposal companies in each EU Member State
European Country, and this product should be disposed of or recycled through them. Further
information on Thermo’s compliance with this directive, the recyclers in your country and
information on Thermo products will be available at www.thermofisher.com.
4Always use the proper protective equipment (clothing, gloves, goggles, etc.)
4Always dissipate extreme cold or heat and wear protective clothing.
4Always follow good hygiene practices.
4Each individual is responsible for his or her own safety.

Thermo Scientificiv Cryogenic Storage Vessel
Preface
Do You Need Information or Assistance on
Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763 Direct
1-800-438-4851 Toll Free, U.S. and Canada
1-877-213-8051 FAX
http://www.thermoscientific.com Internet Worldwide Web Home Page
Service E-Mail Address
Thermo Fisher Scientific
401 Millcreek Road, Box 649
Marietta, OH 45750
Our staff can provide information on pricing and give you quotations. We can
take your order and provide delivery information on major equipment items or make
arrangements to have your local sales representative contact you. Our products are listed on the
Internet and we can be contacted through our Internet home page.
Our staff can supply technical information about proper setup, operation or
troubleshooting of your equipment. We can fill your needs for spare or replacement parts or
provide you with on-site service. We can also provide you with a quotation on our Extended
Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your
applications. If you are experiencing technical problems, working together, we will help you
locate the problem and, chances are, correct it yourself...over the telephone without a service
call.
When more extensive service is necessary, we will assist you with direct factory trained
technicians or a qualified service organization for on-the-spot repair. If your service need is
covered by the warranty, we will arrange for the unit to be repaired at our expense and to your
satisfaction.
Regardless of your needs, our professional telephone technicians are available to assist you
Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by
telephone or fax. If you wish to write, our mailing address is:
International customers, please contact your local Thermo Scientific distributor.
Sales Support
Service Support
service.led.marietta@thermofisher.com

Cryogenic Storage Vessel vThermo Scientific
Table of Contents
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Please Read . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
General Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Measuring Liquid Nitrogen Quantity . . . . . . . . . . . . . . . . . . . . . . . .4-1
General Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Storing Samples in Vapor Phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Liquid Nitrogen Level Monitor Operation . . . . . . . . . . . . . . . . . . . . . . . .6-1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Remote Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Alarm Contact Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Microprocessor Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Level Monitor Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Safety Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Audible and Visual Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
LN2 Level Monitor Troubleshooting and Parts List . . . . . . . . . . . . . . . .7-1
Ordering Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Section 1
Section 2
Section 3
Section 4
Section 5
Section 6
Section 7


Cryogenic Storage Vessel 1-1Thermo Scientific
Please Read
Section 1 Safety Information
This manual contains important operating and safety information. The
user must carefully read and understand the contents of this manual prior
to the use of this equipment.
Your Thermo Scientific cryogenic vessel has been designed with function,
reliability, and safety in mind. It is the user’s responsibility to install it in
conformance with local electrical codes. For safe operation, please pay
attention to the warnings and cautions throughout the manual.
All cryopreservation vessels, regardless of who manufactures them, use a
vacuum to provide the super insulative properties needed to keep nitrogen
in a liquid form. They are very similar, in fact, to the Thermos® vessels you
may have used to store coffee, soup or milk.
Remember how fragile they were? Since no vacuum is perfect or will last
forever, we suggest that you monitor the consumption of liquid nitrogen
used by your vessel on a regular basis, i.e. every 1-2 days. By monitoring,
you may be able to anticipate subtle changes in consumption and possibly
react to problems before they arise. You may opt to use a level monitor for
this purpose (see page 6-1 for level monitor operation.).
Also, if your samples are super critical, one-of-a-kind, or irreplaceable,
consider allocating your samples to 2 separate vessels. Remember the old
adage about placing all your eggs into one basket….
Finally, because no vacuum will last forever, have your vessel tested by the
manufacturer or a qualified cryovessel service technician every 7-10 years,
regardless of any problems (or lack thereof) you may have had in the past.
This will help insure your samples against sudden loss of nitrogen due to
vacuum failure.
Do not use this product for anything other than its intended usage.
Use of the equipment in a manner not specified by the manufacturer may
impair the protection provided by the equipment.
General Usage

1-2 Cryogenic Storage Vessel Thermo Scientific
Section 1
Safety Information
Cautions
• Liquid nitrogen is extremely cold; it boils at -196°C.
• To avoid injury due to frostbite or ruptured vials use extreme care
whenever handling liquid nitrogen, liquid nitrogen storage or transfer
vessels, or any objects which have come in contact with liquid
nitrogen.
• Leave no areas of skin exposed.
• Always wear proper safety attire over clothing: face shield, cryogenic
gloves, cryogenic apron.
• Do not tightly seal liquid nitrogen containers or prevent nitrogen gas
from escaping.
• Always handle ampules with tongs. They may explode when removed
from the vessel.
• Use extreme care to prevent spilling and splashing liquid nitrogen
during transfer and removal of storage contents and holders.
• Immediately remove any clothing or safety attire on which liquid
nitrogen has been spilled or splashed.
• Get immediate medical attention for any frostbite injuries due to
liquid nitrogen.
• Never overfill liquid nitrogen vessels. Liquid nitrogen level should
never be above the tops of the racks inside the vessel. The excess
weight may crack the neck and result in vacuum failure.
• When inserting or removing racks, be careful not to come in contact
with the neck tube area of the vessel. Remove or insert racks slowly
in a vertical manner. Scratches on the neck tube area can cause
premature vacuum failure.
• Do not tamper with or remove vacuum port (covered by a black
plastic cap on side of vessel); this will void warranty.
• Do not spill liquid nitrogen on vacuum port - this can cause vacuum
failure.
• Always keep liquid nitrogen vessel in an upright position.
• Do not use this Locator vessel for transport.
Warnings

Cryogenic Storage Vessel 2-1Thermo Scientific
Section 2 General Specifications
Operating: 17°C - 27°C; 20% to 80% relative humidity, non-condensing.
Installation Category II (overvoltage) in accordance with IEC 664.
Pollution Degree 2 in accordance with IEC 664.
Altitude limit: 2,000 meters.
Storage: -25°C to 65°C; 10% to 85% relative humidity.
Locator and Locator Plus units are intended for indoor use only.
Locator 4
(#CY50935)
Locator 4 Plus
(#CY50935-70)
Locator 6
(#CY50985)
Locator 6 Plus
(#CY50985-70)
Locator 8
(#CY50945)
Locator 8 Plus
(#CY50945-70)
Locator JR
(#CY50925)
Locator JR Plus
(#CY50925-70)
Height
(including lid)
37.5”
(95.2cm) 39.5” (100.3cm) 39.5”
(100.3cm) 39.5” (100.3cm) 37.5”
(95.2cm) 39.5” (100.3cm) 26.5”
(67.3cm) 28.9” (73.4cm)
Diameter 22”
(55.9cm) 22” (55.9cm) 26” (66cm) 26” (66cm) 22”
(55.9cm) 22” (55.9cm) 22” (55.9cm) 22” (55.9cm)
LN2 Capacity,
Liquid Phase
Storage
111L 121L 184L 184L 111L 121L 60L 71L
Static LN2
Evaporation
rate - L/Day
.99 .99 .99 .99 .60 .60 .85 .85
Liquid Phase
Capacity (2mL
Vial2actually
submerged in
liquid nitrogen)
3600 4000 6000 6000 1800 2000 1600 2000
Power Supply/
100VSupply/
100-240VAC
TN509X2 TN509X2 TN509X2 TN509X2 TN509X2 TN509X2 TN509X2 TN509X2
Power Supply/
100VSupply/
100-240VAC
TN509X2
1The liquid nitrogen level monitor will add 1 inch to these heights. The transportation cart will add 4 inches to these heights.
2Using Nalgene System 100 cryoboxes with a 10 x 10 vial configuration (Nalgene System 100 vial or equivalent are needed).
Environmental
Conditions

(CE models only) We hereby declare under our sole responsibility that this
product conforms with the technical requirements of the following
standards:
EMC:
EN 61000-3-2: Limits for Harmonic Current Emissions
EN 61000-3-3: Limits for Voltage Fluctuations and Flicker
EN 61326-1 Electrical Equipment for Measurement, Control, and
Laboratory Use; Part I: General Requirements
Safety:
EN 61010-1 Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use; Part I: General
Requirements
per the provisions of the Electromagnetic Compatibility Directive
2004/108/EC.
Low Voltage Directive: 2006/95/EC
The authorized representative located within the European Community is:
Thermo Electron LED GmbH
Robert-Bosch-Strasse 1
D-63505 Langenselbold, Germany
Sales Ph: 0800 1 536376
Cryogenics Dewars (liquid nitrogen containers) meet the requirements of
Annex V and VII Medical Devices Directive 93/42/EEC
Copies of the Declaration of Conformity are available upon request.
2-2 Cryogenic Storage Vessel Thermo Scientific
Section 2
General Specifications
Declaration of
Conformity
Electrical Inputs . . .100-240VDC, .08 - .04 amps, single phase
Frequency . . . . . . . . . . . . . . . . . . . . . . . .50-60 Hz
Output . . . . . .15VDC (±10%), 0.2 amps, 3.0 watts

Cryogenic Storage Vessel 3-1Thermo Scientific
Section 3 Unpacking
Before using your new Locator or Locator Plus vessel, carefully inspect the
vessel prior to use. Check for signs of damage which may have occurred in
shipment. It is advisable to fill (see filling instructions) all new units with
liquid nitrogen and watch liquid nitrogen loss rate for a few days. If there
are any problems, call Customer Service as soon as possible.
The warranty registration card enclosed with the vessel must be completed
and returned to the factory within 30 days to make warranty effective. This
information must include the serial number, which is located on the label
on the vessel.
Note The most prevalent cause of failure of liquid nitrogen storage vessels
is mechanical. The vessel necktube supports the full weight of the inner
shell and all the liquid nitrogen it contains. A side blow to the vessel causes
the inner shell to swing in a pendulum-like motion causing the necktube
to be damaged. Any storage vessel which has been in an accident, has been
dropped, or lowered to hit on its side, will tend to fail more rapidly than
one that has not. s
Caution Exercise caution when moving your Locator or Locator Plus
vessel. Locator and Locator Plus cryobiological storage systems are not
transportation vessels. Transport carts are designed for mobility within the
labor lab to lab only. Moving full vessels long distances, over cracks in
floor, thresholds, on inclined ramps or in elevators can cause premature
vacuum failure. s
Note If you must transport samples under cryogenic conditions, consider
the Arctic Express line of Dry Shippers available from Thermo Scientific. s


Cryogenic Storage Vessel 4-1Thermo Scientific
Section 4 Operation
Caution Never overfill your Locator or Locator Plus vessel with liquid
nitrogen. The liquid nitrogen level in your tank (with racks inserted)
should never be above 20 inches (50cm) for Locator 4 and Locator 8, 10
inches (25cm) for Locator Jr., 22 inches (55.8cm) for Locator 4, 6, and 8
Plus and 12 inches for Locator Jr. Plus. Filling the tank up to or above the
bottom of the necktube may cause immediate or premature vacuum failure
to occur. s
Caution When inserting or removing racks, be careful not to come in
contact with the necktube area of the vessel. Remove or insert racks slowly
in a vertical manner. Scratches on the necktube area can cause premature
vacuum failure. s
Caution Do not spill liquid nitrogen on vacuum port (covered by a black
plastic cap on side of vessel). This can cause vacuum failure. s
To avoid damage to your Locator or Locator Plus cryogenic storage vessel
which may result in premature vacuum loss, it is important that the
following procedure be used when adding liquid nitrogen to a warm vessel.
1. Add only a small amount of liquid nitrogen (5-10 liters) to a new or
warm vessel.
2. Allow this small amount of liquid nitrogen to sit in the covered vessel
for a minimum of 2 hours. This will limit stress caused by the sudden
temperature change associated with adding liquid nitrogen to a warm
vessel.
3. Add an additional 15 liters of liquid nitrogen to the vessel.
4. Allow vessel to sit for 48 hours and monitor liquid nitrogen
consumption with a wooden yarsdstick, optional dip measurer (part
#180143) or Level Monitor.
5. Fill Locator or Locator Plus as indicated (see Caution in this section).
Allow for displacement of liquid nitrogen when racks and boxes are
inserted.
6. Insert and remove racks slowly. Allow liquid nitrogen to run out of
boxes and off racks.
Filling Instructions

4-2 Cryogenic Storage Vessel Thermo Scientific
Section 4
Operation
1. Use a wooden yardstick or optional dip measurer (part #180143) to
measure liquid nitrogen level. NEVER use a hollow tube or plastic
dipstick to measure the liquid nitrogen level.
2. Level will be indicated by the frostline which develops when dipstick is
removed.
3. Level Monitor can be used for constant measurements. See Section 6.
1. Remove the 2 bottom-most cryoboxes from each rack.
2. Measure the height of 2 cryoboxes stacked on top of each other.
3. Fill vessel with liquid nitrogen to the height obtained from Step 2,
using a yardstick to measure liquid level in vessel.
4. Place racks into vessel without the 2 bottom-most cryoboxes. Samples
are now above the liquid nitrogen and are stored in the vapor phase.
Important A level monitor should always be used when you store your
samples in the vapor phase because of the already low levels of liquid
nitrogen. Refer to Section 5 for correct level monitor for your system. s
Wipe exterior surfaces with lightly dampened cloth containing mild soap
solution.
Storing Samples in
Vapor Phase
General Cleaning
Instructions
Measuring Liquid
Nitrogen Quantity

Cryogenic Storage Vessel 5-1Thermo Scientific
Section 5 Replacement Parts
Note Cork and glue cannot be replaced on covers with level monitor or
level monitor will malfunction. Contact Technical Services for return
authorization to send cover in for repair. s
Note Locator and Locator Plus vessels include racks and covers. However,
these items may also be ordered separately. s
Catalog # Description - For Standard Covers Only
1950355 Replacement cork and glue for Locator 4, Locator 4 Plus, Jr. and Jr. Plus
1950356 Replacement cork and glue for Locator 8 and Locator 8 Plus
1950354 Replacement cork and glue for Locator 6 and Locator 6 Plus
317035 Glue for corks
1950343 Vent Port Cap
Locator and Locator Plus Vessels and Covers
Japan, North/SouthAmerica, Europe 100-240VAC
Type Tank Only
No Racks or Covers Standard Covers
Locator Jr. CS509X11A (317030) CV509X10
Locator Jr. Plus CS509X20A-70 (317146) CV509X10
Locator 4 CS509X12A (317159) CV509X10
Locator 4 Plus CS509X21A-70 (317173) CV509X10
Locator 6 CS509X22A-70 (317185) CV509X40-70
Locator 6 Plus CS509X22A-70 (317185) CV509X40-70
Locator 8 CS509X13A (317199) CV509X9
Locator 8 Plus CS509X23A-70 (317215) CV509X9
All level monitors listed operate on 15VDC (±10%), 0.2 amps, 3.0 watts.

5-2 Cryogenic Storage Vessel Thermo Scientific
Section 5
Replacement Parts
Catalog # Rack Description
HR509X15A (317050) For Locator Jr (8141, 8201, 8202) with 4 - 2 inch boxes
HR509X2A (317161) For Locator 4 (8143, 8246, 8247) with 9 - 2 inch boxes
HR509X19A (315903) For Locator Jr (8141, 8201, 8202) with 2 - 4 inch boxes
HR509X20A (315905) For Locator 4 (8143, 8246, 8247) with 5 - 4 inch boxes
HR509X15A-70 (317148) For Locator Jr Plus (8142, 8237, 8238) with 5 - 2 inch boxes
HR509X2A-70 (317175) For Locator 4 Plus (8144, 8207, 8208) with 10 - 2 inch boxes
HR509X31A-70 (317192) For Locator 6 and 6 Plus (8147, 8197, 8198, 8209, 8210) with 10 - 2 inch boxes
HR509X1A (317205) For Locator 8 (8211, 8212) with 9 - 2 inch boxes (25 cell)
HR509X1A-70 (317217) For Locator 8 Plus (8146, 8213, 8214) with 10 - 2 inch boxes (25 cell)
HR509X19A-70 (315904) For Locator Jr. Plus (8142, 8237, 8238) with 2 - 4 inch boxes + 1 - 2 inch box
HR509X20A-70 (315906) For Locator 4 Plus (8144, 8207, 8208) with 5 - 4 inch boxes + 1 - 2 inch box
HR509X32A-70 (315907) For Locator 6 and 6 Plus (8147, 8197, 8198, 8209, 8210) with 5 - 4 inch boxes + 1 - 2 inch box

Section 6 Liquid Nitrogen Level
Monitor Operation
The liquid nitrogen (LN2) level monitor provides a constant indication of
the LN2 level and notifies you to low level conditions through audible and
visual alarms. It alerts you when your vessel needs filling. The level
monitor is shipped mounted directly to the appropriate Locator or Locator
Plus cover. It can be wired to a remote alarm system to alert you to
problems when you are not in your laboratory. This equipment was
intended to be operated with an external power supply provided by the
manufacturer. It is designed for use with Thermo Scientific LN2 vessels
only.
Caution Use this product for its intended usage only. s
Warning Potential electrical hazards. Only qualified persons should
perform procedures associated with this symbol. s
Note If the level monitor was purchased as part of a Locator or Locator
Plus system, skip to Step 3. If it is a retrofit for an existing Locator or
Locator Plus, begin with Step 1. s
1. Remove the level monitor and Locator or Locator Plus cover from box.
2. Remove present cover from your Locator or Locator Plus and replace
with your new level monitor and cover.
3. Check the electrical specifications on the power supply for electrical
data and plug in to appropriately grounded receptacle.
Cryogenic Storage Vessel 6-1Thermo Scientific
Installation

1. Disconnect the liquid nitrogen level monitor from the power supply.
Disconnect power to your alarm system.
2. Remove the three screws securing the liquid nitrogen level monitor’s
cover. Pull the cover up and out of the way (see Figure 6-1).
When connected to your remote alarm system, the LN2 level monitor will
alert you to a problem even when your laboratory is unoccupied. The unit’s
alarm contacts are activated at the same time the audible alarm sounds (see
Microprocessor Functions).
3. Remove the black plug from the side of the cover. Insert the wires for
the remote alarm through this hole.
4. Connect the remote alarm wires as shown in Figures 6-1, 2 and 3.
5. Replace the LN2 level monitor’s cover. Reconnect power to the LN2
level monitor and to your remote alarm system.
Note The LN2 level monitor can be configured with either a normally
open contact or a normally closed contact. Which mode is currently
operative, depends on the placement of a single jumper on the circuit
board. For a normally open contact, the jumper should be placed on J6.
For a normally closed contact, the jumper should be placed on J4 (see
Figure 6-2). Verify that the jumper is set appropriately for your remote
alarm system. s
The LN2 level monitor’s alarm contact is a Photomos type relay.
Contact Form: Type 1 form B
Load Voltage Maximum: 24VDC Max
Load Current Maximum: 120mA
Typical On Resistance: Maximum 26 Ω
6-2 Cryogenic Storage Vessel Thermo Scientific
Section 6
LN2 Level Monitor Operation
Alarm Contact
Specifications
Level Monitor
Remote Alarm
Connection to Remote
Alarms

Cryogenic Storage Vessel 6-3Thermo Scientific
Section 6
LN2 Level Monitor Operation
1
1
1
For normally closed contact, place jumper here.
For normally open contact, place jumper here.
Remote Alarm wires connect here
(Note polarity.)
-+
J3
J4 J6
Figure 6-1. Connect to Remote Alarm
Figure 6-2. Jumper Location
Other manuals for Locator Series
1
This manual suits for next models
39
Table of contents
Languages:
Other Thermo Scientific Storage manuals
Popular Storage manuals by other brands

Fulterer
Fulterer FR 1535 Mounting instructions

Stack-On
Stack-On GCB-2SF-BB Assembly instructions

Sun Microsystems
Sun Microsystems StorageTek StreamLine SL8500 Best practices guide

LapSafe
LapSafe ClassBuddy Desk 10 user manual

HP
HP Pavilion p6000 - Desktop PC Replacement instructions

Seagate Technology
Seagate Technology ST2000VT000 product manual