
IonPac AS3, AS4, AS5, CS2 Document No. 032226-14 Page 14 of 14
one of the proportioned eluents is less than 5%.
C. Columncontamination canleadto alossof columncapacity.Thisisbecausealloftheionexchangesiteswillnolonger
beavailableforthesampleions.Polyvalentanionsmightbeconcentratingonthecolumn.RefertoSection6.3,“Column
Cleanup,” for recommended column cleanup procedures.
Possiblesourcesofcolumncontaminationareimpuritiesinchemicalsandinthedeionizedwaterbeingused.Beespecially
careful to make sure that the recommended chemicals are used. The deionized water should have a specific resistance
ofatleast18.2 megohm-cm.
D. Dilutingthe eluentwillimprovepeak resolution, butwillalsoincrease theanalytes’retentiontimes.Ifa10%dilution
of the eluent is not sufficient to obtain the desired peak resolution, or if the resulting increase in retention times is
unacceptable, clean the column (see Section 6.3, “Column Cleanup”).
Aftercleaning the column, reinstall it in the system and let it equilibrate with eluent for about 30 minutes. Thecolumn
is equilibrated when consecutive injections of the standard give reproducible retention times. The original column
capacity should be restored by this treatment, since the contaminants should be eluted from the column. If you need
assistance in solving resolution problems, contact the nearest DIONEX Office (see, “DIONEX Worldwide Offices”).
5.3.3 Loss of Front End Resolution
Ifpoorresolutionsandefficienciesareobservedfortheveryearlyelutingpeaksnearthesystemvoidvolumecomparedtothelater
eluting peaks, check the following:
A. Impropereluent concentrationmaybethe problem.Remaketheeluentasrequiredforyourapplication.Ensurethatthe
water and chemicals used are of the required purity.
B. Columnoverloading maybe theproblem.Reducetheamountofsampleionsbeinginjectedontotheanalyticalcolumn
by either diluting the sample or injecting a smaller volume onto the column.
C. Sluggish operation of the injection valve may be the problem. Checktheairpressureandmakesurethereareno gas
leaks or partially plugged port faces. Refer to the valve manual for instructions.
D. Improperlysweptoutvolumesanywhereinthesystempriortotheguard andanalyticalcolumnsmaybetheproblem.
Swapcomponents,oneatatime,inthesystempriortotheanalyticalcolumnandtestforfront-endresolutionafterevery
system change.
5.4 Spurious Peaks
A. Thecolumns maybecontaminated. Ifthesamplescontainanappreciablelevelofpolyvalentionsandthecolumnisused
withaweakeluentsystem,polyvalentanionsmaycontaminatetheanalyticalcolumn.Theretentiontimesfortheanalytes
willthendecreaseandspurious,inefficient(broad)peakscanshowupatunexpectedtimes.Cleanthecolumnasindicated
inSection6.3,“ColumnCleanup”.
Ifyouneedassistanceindeterminingthebestwaytocleanstronglyretainedsolutesinyour specific sample matrix off
ofDIONEXcolumns,contactthenearestDIONEXOffice(see,“DIONEXWorldwideOffices”).
B. The injection valve may need maintenance. Whenaninjectionvalveisactuated,thepossibilityofcreatingabaseline
disturbanceexists.Thisbaselineupsetcanshowupasapeakofvaryingsizeandshape.Itwillhappenwhentheinjection
valve needs to be cleaned or retorqued (see Valve Manual). Check to see that there are no restrictions in the tubing
connected to the valve. Also check the valve port faces for blockage and replace them if necessary. Refer to the Valve
Manual for troubleshooting and service procedures. Small baseline disturbances at the beginning or at the end of the
chromatogram can be overlooked as long as they do not interfere with the quantification of the peaks of interest.
Ifcleaningandretorquingthevalvedoesnothelp,replacethevalve.ConsultthenearestDIONEXOffice(see,“DIONEX
Worldwide Offices”) for help in services or replacing a suspect valve.