Model VC-2 Voltmeter Verifier
1. VOLTMETER VERIFIER
The Model VC-2 Voltmeter Verifier provides highly accurate output voltage to verify voltmeters used for Cathodic
Protection testing. The VC-2 will work with any voltmeter or multimeter on DC volts setting. The VC-2 offers the
following voltage outputs: 0.001v, 0.01v, 0.1v, 1.0v, 2.5v, 5.0v
The VC-2 is a portable, battery operated instrument. With the VC-2, you can test voltmeters on a scheduled basis
or at each use, and this performance check allows you to send voltmeters that are out of calibration on an
exception basis rather than a time based schedule.
The VC-2 is a very accurate instrument and should be put on an annual calibration cycle. Tinker & Rasor
performs calibration services in its Southern California factory with a 24-hour turn around.
2. INSTRUCTIONS FOR UNPACKING & INSPECTION
A. Open the package and ensure you have received all items stated on the packing list, usually in a plastic
pouch on the outside of the packing box.
B. INSPECTION should be made upon receipt. If damage has occurred during shipment, file a claim with
the carrier immediately.
C. If it is necessary to contact your supplier or the Tinker & Rasor concerning damaged or missing items, be
sure to include the serial number, purchase order number, and invoice number of the instrument in
question.
3. USE INSTRUCTIONS
A. Make sure that the instrument is turned OFF. No Red Power LED.
B. Connect the voltmeter to be verified to the VC-2. Connection to the VC-2 should be made using the leads
from the voltmeter being tested. (The voltmeters leads are an important part of verifying the performance
of the instrument. For this reason, the VC-2 does not come with leads of its own.)
C. When connecting the leads polarity must be observed. The RED terminal is positive and the BLACK
terminal is negative.
D. Prior to turning the VC-2 to On, move the voltage
selector knob all the way counter clockwise to the 0.001v
setting.
E. Adjust the voltmeter to the proper scale for the selected
output voltage of the VC-2 instrument.
F. Turn the VC-2 on by pressing the power button. The Red
LED lights.
G. The voltage output will begin immediately as soon as the
VC-2 is turned On.
H. Observe the reading on the voltmeter and compare the
knob position on the VC-2.
I. Move the knob clockwise and at each knob setting,
observe the voltmeter display reading against the VC-2
knob position.
J. It is recommended to always turn the VC-2 OFF when adjusting, connecting or disconnecting lead wires.