1. General .............................................................................................. 3
1.1. Warranty ........................................................................................ 3
1.2. Symbols used.................................................................................. 4
1.3. Instrument Label ............................................................................. 4
1.4. Scope of this manual........................................................................ 4
2. INTRODUCTION................................................................................... 5
2.1. Heating performance........................................................................ 5
2.2. Control Performance ........................................................................ 5
2.3. Functions and connections ................................................................ 6
2.4. Operating instructions ...................................................................... 7
2.5. Temperature switch ......................................................................... 8
2.6. Specifications .................................................................................. 9
2.7. Maintenance.................................................................................... 9
2.8. Storage .........................................................................................10
2.9. Recalibration ..................................................................................10
2.10. Optional accessories........................................................................10

User Manual
TRAQC-3 FM
TRAQC-3 FM 09/2016 (R8) page : 3
1.General
1.1. Warranty
During the design and manufacturing of this instrument the utmost attention has
been given to quality and durability.
This manual contains information needed for the safe and effective use of
the capabilities of the instrument.
Please read the manual carefully before operating the instrument. By doing
so possible damage to the instrument or damage caused by the incorrect
use of the instrument can be avoided.
TRADINCO INSTRUMENTS warrants the instrument in accordance with the Standard
Terms and Conditions of the Instrument Trade as issued by the Association bearing
the name "Federation Het Instrument" ( The Instrument Federation(, filed with the
Clerk of Utrecht District Court on 13 January 1993 under number 16/93 and with the
Chamber of Commerce and Industry in Amerfoort on 18 January 1993. A copy is
available on request
TRADINCO INSTRUMENTS warrants that this product will be free from defects in materials
and workmanship for a period of 12 months from the date of shipment. If any such
product proves defective during this warranty period, TRADINCO INSTRUMENTS, at its
option, will either repair the defective product without charge for parts or labor, or will
provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify TRADINCO
INSTRUMENTS of the defect before the expiration of the warranty period and make
suitable arrangements for the performance of the service.
Customer shall be responsible for packaging and shipping of the defective product to the
service centre designated by TRADINCO INSTRUMENTS, with shipping charges prepaid.
If no defect can be found Customer may be charged for costs of the investigation.
This warranty does not apply to any defect, failure or damage caused by:
a. Improper use of the instrument.
b. Normal wear of the product.
c. Modification or repair carried out by or on behalf of the owner or by a third party.
d. Implementation of modifications to the product that are not supplied or implemented
by TRADINCO INSTRUMENTS.
TRADINCO INSTRUMENTS and its vendors will not be liable for any indirect, special,
incidental or consequential damages irrespective of whether TRADINCO INSTRUMENTS or
the vendor has advance notice of the possibility of such damages. The type number of the
product, as listed on the instrument tag plate, should always be mentioned in any
correspondence concerning the product.
Tradinco Instrumenten-Apparaten b.v.
Radonstraat 250
2718 TB Zoetermeer
The Netherlands
+31 79 2033133
www.tradinco.com
Email i[email protected]

User Manual
TRAQC-3 FM
TRAQC-3 FM 09/2016 (R8) page : 4
1.2. Symbols used
Warning for conditions or practices that could result in personal injury or loss of life
or in damage to the product or other property.
Danger of electrical shock
Attention signal or remark
1.3. Instrument Label
1.4. Scope of this manual
This manual contains information for instruments built since September 2008.
This manual is a users manual for operating the instrument, it is not a calibration
instruction manual.
Type No. : TRAQC 3FM
Serial No. : xxxx-x
Range : Amb+5 - 375°C
Power : 110 / 230 Vac
50 –60Hz, 250 VA
Fuse : 250VA 3.15 AT
Date : 30/09/2008
Zoetermeer, The Netherlands

User Manual
TRAQC-3 FM
TRAQC-3 LH 10/2014 EN Rev.2 page : 5
2. INTRODUCTION
Accurate temperature measurements can be assured only when it is possible physically
to produce a legitimate basis for establishing traceability to temperature standards.
By setting a series of different temperatures, the inserted sensor/ display/recorder can
be calibrated accordingly.
The model TRAQC-3 may be used for all temperature sensors, incl. RTD'S and
thermocouples and temperature switches.
The use of model TRAQC-3 offers a number of great advantages above alternative
methods of temperature calibration, but requires that any user reads the instruction
manual very carefully and is aware of the fact that high temperature has to be
handled with care and only in areas without hazardous fumes or fire risks and get
acquainted with all the facilities offered by this field temperature calibrator.
2.1. Heating performance
The heating block is made from specially produced metal to achieve optimum heat
conductivity and is heated by resistance heating elements. The instrument is
electronically controlled and regulated with 2 push-buttons on the front panel.
The temperature is controlled by a precision platinum resistance thermometer (RTD
sensor) and an intelligent closed loop circuit.
2.2. Control Performance
In order to provide optimum use of the TRAQC-3 all control functions are available
through a RS232 interface.
Tradinco AutoCal Software for using the TRAQC-3 with a PC is also available.
The MS Windows based software is completely 'MENU' controlled and allows the user to
program pre-fixed calibration routines and run calibration routines fully automatically.

User Manual
TRAQC-3 FM
TRAQC-3 LH 10/2014 EN Rev.2 page : 6
2.3. Functions and connections
6
7
8
9
10
11
3
4
5
1
2
1) Heat indicator
2) Up and down push buttons
3) Hole for inserts
4) Switching indicator Led
5) Temp. switch terminals 4mm
6) Voltage selector
7) ON/OF Switch
8) Fuse
9) 9 Pin RS232 port
10) Power cord connector
11) Cooling fan

User Manual
TRAQC-3 FM
TRAQC-3 LH 10/2014 EN Rev.2 page : 7
2.4. Operating instructions
1) Select the correct voltage by means of the voltage selector on the back of the
instrument.
Note : Although the instrument is well protected against incorrect setting of
voltage, this may result in permanent damage of electronics or incorrect
measurements.
Note : Please observe proper grounding of the instrument. Use supplied
power cord if possible.
2) Place instrument in a way that the ventilation hole on the back of the instrument is
not obstructed for proper measurement results.
3) Connect the power cord to the unit and turn it on
Note:A series if different numbers will be shown in the display. These will show
resp: 8888 : check all segments of the display.
Prog, then a number: this shows the software revision number,
XXXX : a code number for manufacturers use only,
XX: a code number for manufacturers use only,
Hr XX: The amount of hours the unist has been powered, the six digits after the “Hr”
represent the hours up to 99999.9 hrs
XXXX: see above.
4) Place unit under test in the Insert hole.
Note : It is recommended to use the centre hole for all measurements.
At special request from customers, the pattern of insert holes may differ, i.e. a
temperature reference hole may be ordered as a extra option.
With the up and down pushbuttons to any desired temperature between ambient + 5
°C and 375 °C can be set
5) Set the first calibration/test point.
6) After 1 second the display indicates the actual temperature.
7) Observe at least 5 minutes stabilisation time when the set temperature is reached.
Note 1 : While heating a led (O1) in the left upper corner of the display will
flash. Flashing will be intermittent and regular as the unit reaches the set
temperature.
Note 2. While cooling two leds (O2 and A1) on the left are switched on to
show the fan is running at high speed.
8) Follow steps 5 thru 7 for every calibration point.

User Manual
TRAQC-3 FM
TRAQC-3 LH 10/2014 EN Rev.2 page : 8
9) For decreasing the temperature the same procedure as described above must be
followed.
Note : Please do not disconnect or switch off instrument until the indicated
temperature of the block is below 50 °C.
Note : The environment (temperature and draught), length and the mass of
the probe may influence the measurement considerably.
Note: If the manufacturer has enabled the most left button, one has access
to a part of a menu structure. Use of this feature is not required, Any change
may seriously change the behaviour of the instrument.
2.5. Temperature switch
Connect the temp-switch to the terminals and insert the switch in the hole
Set the TRAQC-3 for the desired temperature and let it stabilize.
Adjust the temp-switch so that the indicator-light switches at the desired temperature

User Manual
TRAQC-3 FM
TRAQC-3 LH 10/2014 EN Rev.2 page : 9
2.6. Specifications
Temperature range : Ambient +5°C to 375°C.
Resolution : 0,1 °C/°F.
Accuracy : Over full range better than +/- 0,4 °C, typ. +/- 0.2°C
Stability : +/- 0,05 °C.
Heating time : Ambient to 300 °C 5.75 minutes.
Cool down time : 300 °C to 60 °C 14 minutes.
Stabilization : 5 minutes.
Well depth : Max 100 mm.
Switch test : Powerless switch, LED indicator, 5 VDC, 16 mA max.
Communication : RS232 for use with e.g. AutoCal
Supply voltage : 230Volt, 50/60 Hz or 110 Volt, 50/60 Hz. Selectable.
Power consumption : 250 Watts.
Fault protection : Internal sensor break detection.
Fuse : 3.15 A .
Thermal Protection switch : 75 °C
Temp switch test : 5 VDC, 16 mA max.
Safety : EN 61010-1:2010/C1:2011, EN61010-2-010:2014.
Conformity EMC : EN-61326-1:2013
Weight : approx 1740 grams, without powercord, incl. 6.2 mm insert.
Dimensions : w x d x h : 149 x 159 x 74 mm.
All specifications at 20°C +/- 2°C.
2.7. Maintenance
Allow the TRAQC-3 to cool down as described in chapter 2.4.
Switch the TRAQC-3 off and disconnect the mains plug.
Cleaning TRAQC-3 with inserts:
A small amount of dust or pollution can cause jams when the instrument is operated
with inserts. To prevent this, remove the insert from the heating block at regular
intervals and if the TRAQC3 is not going to be operated for a longer period. Flush the
heating block bore with compressed air and clean the bore and inserts with a dry cloth.
Cleaning the fan grille:
Each TRAQC-3 is fitted with a small air grille via which cooling air enters the TRAQC-3.
Clean the grille at regular intervals (vacuuming or brushing) depending on the level of air
pollution.
External cleaning:
Clean the outside of the TRAQC-3 with a damp cloth and some water or with a
solvent-free mild cleaning agent. Disconnect the instrument from mains before cleaning.
Fuse:
Two Fuses of 3.15Amp (slow) need to be used.

User Manual
TRAQC-3 FM
TRAQC-3 LH 10/2014 EN Rev.2 page : 10
2.8. Storage
Prior to transport or contact with the metal block ensure that it has cooled down to a
temperature below 50°C otherwise there is a risk of severe damage at the instrument.
In order to cool down the TRAQC-3 quickly, set the set temperature to a low
temperature, e.g. room temperature. Do not switch the instrument of or disconnect the
mains lead until the instrument has reached a temperature below 50°C
2.9. Recalibration
The TRAQC-3 is adjusted and tested with measuring equipment in accordance with
recognized national standards prior to delivery.
The TRAQC-3 should, depending on the application situation, be inspected at appropriate
intervals. We recommend you to return the calibrator to Tradinco at intervals
of max. 12 months for recalibration and service.
2.10.Optional accessories
Carrying case:
A rugged carrying case which offers maximum
protection to the TRAQC-3 during
transportation. This case is watertight and
is secured by two safety latches. It stores
the TRAQC-3, 3 pockets to store inserts
and a pocket for the power cord.
Inserts:
The following inserts are
available:
Undrilled
Drilled:
3.5 mm
6.5 mm
4.5 mm + 4.5 mm
3.5 mm + 4.5 mm
Custom drilled

User Manual
TRAQC-3 FM
TRAQC-3 LH 10/2014 EN Rev.2 page : 11
Tradinco Instrumenten-Apparaten b.v.
Radonstraat 250
2718 TB Zoetermeer
The Netherlands
+31 79 2033133
www.tradinco.com
Email i[email protected]