
mini GAC®2500 Grain Calibrations
500XXXX Rev A 3
NOTE: Uploading new grain
calibrations can only be done
one at a time.
4. Once the correct product slot is selected, click the “UPLOAD New
Calibration” button. Navigate to where the calibration file you wish to
upload is stored on your computer. Click “Open” once the file is selected.
A warning will appear to confirm what product will be uploaded and which
product will be replaced. Click “Yes”.
5. To remove a product calibration, select the desired language and product
in the drop-down lists (or right-click the designed product in the Product
view) and click the “DELETE Calibration” button.
6. A confirmation will display on the computer verifying a valid calibration
upload or deletion.
7. To view the current calibrations stored in the mini GAC®2500, look on the
left side of the PC Application Main Page. The calibrations list shown in
the Installed Products window can be updated at any time by clicking the
“Refresh Product List” button.
8. Close PCT and remove cable from computer. New calibrations can be
found on product screen.
cannot be changed on the
mini 2500 GAC®unit using
the PCT.
MOISTURE AND TEST WEIGHT BIAS
In situations where moisture and test weight differences occur with a local
grain elevator, the Bias function allows for entering a correction factor in
moisture and test weight to compensate for those differences.
This Bias adjustment can be done from the PCT software tool or from the
mini GAC®2500.
To adjust Bias from the PCT Tool:
1. In the “Product List” on the left-side of the main window, right-click the
desired product.
2. Select “Edit Bias” (Refer to Figure 6).
3. Allow time for the PCT tool to collect and display the bias data.
4. Enter new values and click the “Set Bias Values” button.
5. To re-verify the Bias values, click the “Get Bias Values” button to read
®
Figure 4To Adjust Bias from the PCT Tool