Turtle Tough TT-CL2 User manual

TURTLE
TOUGH
®
OPERATION MANUAL
TT-CL2 FREE CHLORINE SENSOR
Ion Specic Chlorine Analyser
turtletoughsensors.com

CONTENT
SAFETY PRECAUTIONS 3
THEORY OF OPERATION 3
SPECIFICATIONS 4
Residual Chlorine sensor probe 4
Flow cells 4
Electrochemical Probe Simulator 4
INSTALLATION 5
Site selecon 5
Unpacking 5
Mounng 6
Water connecons 6
OPERATION 6
Calibraon 6
SENSOR 7
Installaon 8
Setup 8
Maintenance 11
4.0.3.1 Roune maintenance 11
4.0.3.2 Diagnosc maintenance 11
Storage 11
Spares 11
Single owcell 11
4.0.6.1 Inseron 12
Double owcell 13
COMMISSIONING 13
TROUBLESHOOTING GUIDE 13
Further help 13
APPENDICES 14
General layout drawings 14
Cercates and approval statements 15
Returns Procedure and ‘Free of Contaminaon Sheet’ 16
Warranty 17
MSDS safety data sheets for Free Chlorine Electrolyte (ECS2.1/GEL) Electrolyte (ECS2.1/GEL 18
Calibraon informaon and sensor setup 21

T T-CL2
Free Chlorine Sensor
SAFETY PRECAUTIONS
• Before aempng to unpack, set up, or operate this instrument, please read this enre manual.
• Make certain the unit is disconnected from the power source before aempng to service or remove any
component.
Make certain the unit is disconnected from other sources of force or pressure (for example, pneumac or hydrau-
lic), before aempng to service or remove any component.
• Failure to follow these precauons could result in personal injury and damage to the equipment.
1. THEORY OF OPERATION
The free chlorine sensor has been designed to measure the concentraon of dissolved free residual chlorine.
It does this using a Potenastac 3-electrode chronoamperometric sensor insulated from the water supply
by a hydrophilic membrane. The free chlorine diuses through the membrane and is reduced at the cathode
generang a small current that is proporonal to the concentraon of free chlorine. In potable, process, or
swimming pool water this means HOCl (hypochlorous acid) and OCl- (hypochlorite ion). The relave amount of
these two species is dependent on the pH of the soluon. At low pH’s the predominant species will be HOCl.
At higher pH's the predominant species will be OCl-. Tradional amperometric measurement systems need to
be buered to an exact pH in order that changes in the pH of the sample water do not aect the Free Chlorine
measurement. The probe supplied for Free Chlorine measurement with this instrument is not aected by changes
in pH to the same degree. This means that on many plants it will not be necessary to buer the sample water at
all. If it is necessary to buer the sample water, because the pH is both high and variable it will only be necessary
to reduce the pH to below 6.5. This is best done using the CO2 Buer accessory. The probes are available for a
variety of ranges which are specied at the me of purchase.
The sensor incorporates automac temperature compensaon (ATC).
3
TURTLETOUGH
IPM19 HaloSense Free Chlorine Probe
Manual date 12/1/2011 Page 4of 29
1.0 Theory of Operation
The free chlorine sensor has been designed to measure the concentration of dissolved free residual
chlorine. It does this using a Potentiastatic 3-electrode chronoamperometric sensor insulated from the water
supply by a hydrophilic membrane. The free chlorine diffuses through the membrane and is reduced at the
cathode generating a small current that is proportional to the concentration of free chlorine. In potable,
process, or swimming pool water this means HOCl (hypochlorous acid) and OCl-(hypochlorite ion). The
relative amount of these two species is dependent on the pH of the solution. At low pH’s the predominant
species will be HOCl. At higher pH's the predominant species will be OCl-. Traditional amperometric
measurement systems need to be buffered to an exact pH in order that changes in the pH of the sample
water do not affect the Free Chlorine measurement. The probe supplied for Free Chlorine measurement
with this instrument is not affected by changes in pH to the same degree. This means that on many plants it
will not be necessary to buffer the sample water at all. If it is necessary to buffer the sample water, because
the pH is both high and variable it will only be necessary to reduce the pH to below 6.5. This is best done
using the CO2Buffer accessory. The probes are available for a variety of ranges which are specified at the
time of purchase.
The sensor incorporates automatic temperature compensation (ATC).
Typical Probe Response to pH
(Unbuffered)
0
20
40
60
80
100
120
6
6.5
7
7.5
8
8.5
9
9.5
10
Chlorine Probe Response
Typical Amperometric Probe Response

T T-CL2
Free Chlorine Sensor
1.1 SPECIFICATION
1.1.1 Sensor
Type Membrane covered amperometric potenostac three-electrode system
Measurand Free Residual Chlorine
Range 0-2, 0-5, 0-10mg/l (ppm)
Resoluon 0.01mg/l (ppm)
Reproducibility ±5 %
Stability -2 % per month (without calibraon)
Working electrode Cathode made of gold
Counter electrode Anode made of stainless steel
Reference electrode Silver / silver halide
Membrane material Micro-porous hydrophilic membrane
Flow rate 0.5 l/min (0.13 gallon/minute [US])
Temperature range >5 and <45°C
Temperature compensaon Automacally by an integrated thermistor (ATC)
pH-range pH 4 up to pH 9.5
Permissible overpressure 0.5 bar
First-polarisaon me 120 min
Re-polarisaon me 30 min
Response me T90: approx. 2 min
Zero-point adjustment Not necessary
Zero running Running at zero de-polarises the probe
Calibraon Manual using DPD or Automac (Autocal)
Housing material PVC, silicone, polycarbonate, stainless steel, perspex
Dimensions Diameter approx. 25mm (1”), length 175mm (7”)
Maintenance intervals
Calibraon On control applicaons it is a week to a month & on non-control it is 1 to 3 months.
Electrolyte Quarterly
Calibraon reagent Six monthly (Autocal unit only)
Membrane Yearly
Interferences High levels of other oxidants such as Ozone and Chlorine Dioxide
1.1.2 Flow cell
Inlet ¼” or 6mm barb ng, or 8mm pipe push-ng
Outlet ½” or 12mm barb ng
No. of probes Single and double ow cells available
Service isolaon Ball valve
Flow 0-1.4 l/min (preferred ow rate 0.5 l/min)
Inlet pressure 0.5 bar
Flow cell pressure Eecvely zero, cell open to atmosphere
Weight 1kg (single), 1.5kg (double)
Filter (if ed) 100um
Flow switch available as an opon to give a low ow alarm if sample ow is lost.
1.1.3 Electrochemical probe simulator
Supplied with your instrument, if ordered. It simulates a constant output from a sensor and allows for easier
troubleshoong.
4
TURTLETOUGH

T T-CL2
Free Chlorine Sensor
2.0 INSTALLATION
As with all instrumentaon the installaon and commissioning of this instrument is crucial to its safe and accurate
funcon. This instrument must be used only for its purpose as outlined in this manual and must be installed and
commissioned in accordance with this manual and by trained and qualied personnel.
2.1 Site Selecon
Please choose a suitable locaon for the installaon of the probe and the electronics. The choice of installaon
point on any site is a compromise and is best undertaken by experienced installaon personnel. The following is
a list of the factors that need to be taken into consideraon. This list is not intended as a checklist neither is it
implied that the list is complete.
• Ensure that the mounng allows access to all serviceable parts.
• Try to mount the electronics in a posion where they are not habitually hosed down in a cleaning process.
• Consider the length of wiring runs when mounng the instrument.
• Try to keep the electronics away from substaons or other large EMI emiers.
• Consider whether the sample will be representave and well mixed.
• Consider sample return points.
• In a plasc run, with a low conducvity sample consider earthing the sample.
2.2 Unpacking
Please have a copy of your order with you when you unpack your instrument. All orders are checked when they
leave the factory. Please double check that you have all the parts that were ordered as soon as you open the box.
This may include (but not necessarily);
Probe
Probe mounng overow cell.
Electrolyte
Membranes
Abrasive paper (blue)
Manual
CO2 Buer kit - opon
Acid buer unit - opon
If anything is missing, or damaged, please contact your sales outlet immediately. If the instrument needs to be
returned for any reason please follow the instrucons given in Appendix C of this manual.
TURTLETOUGH
5

T T-CL2
Free Chlorine Sensor
2.3 Mounng
For mounng of the ow cell (single or double), with its backplate (9mm thick), please refer to the drawings below.
2.4 Water connecons
The incoming water sample connecon is to the owcell lower barb or push ng.
In all cases the incoming water sample should be regulated to 0.5 l/min (0.13 gallon/minute [US]).
The ow cell is open to atmosphere, large changes in water pressure can cause water leaks.
If the pressure within the owcell exceeds 0.5 bar, there is a possibility of unstable readings.
The incoming ng is a ¼” or 6mm barb ng, or an 8mm pipe push-ng.
The outlet ng is a ½” or 12mm barb ng.
TURTLETOUGH
IPM19 HaloSense Free Chlorine Probe
Manual date 12/1/2011 Page 7of 29
2.3 Mounting
For mounting of the flow cell (single or double), with its backplate (9mm thick), please
refer to the drawings below.
2.4 Water connections
The incoming water sample connection is to the flowcell lower barb or push fitting.
In all cases the incoming water sample should be regulated to 0.5 l/min (0.13 gallon/minute [US]).
The flow cell is open to atmosphere, large changes in water pressure can cause water leaks. If the pressure
within the flowcell exceeds 0.5 bar, there is a possibility of unstable readings.
The incoming fitting is a ¼” or 6mm barb fitting, or an 8mm pipe push-fitting.
The outlet fitting is a ½” or 12mm barb fitting.
3.0 OPERATION
Refer to the instrument menu structure.
3.1 Calibraon
In order to perform a span of the sensor it is necessary to have an independent method of determining the level
of the measurand, e.g. a DPD test kit to measure free chlorine in the sample water. Refer to the appendix of this
manual for further calibraon informaon. Zeroing of this sensor is not required.
6

T T-CL2
Free Chlorine Sensor
4.0 SENSOR
WARNING!WARNING!WARNING!WARNING!
PLEASE READ THE INSTRUCTIONS BELOW BEFORE DOING ANYTHING WITH THIS
PROBE. FAILURE TO FOLLOW THESE INSTRUCTIONS MAY RESULT IN DAMAGE TO
THE MEMBRANE CAP AND WARRANTIES FOR THIS DEVICE MAY BE VOIDED.
You will nd a membrane cap that is loosely screwed onto the end of your probe.
DO NOT REMOVE OR TIGHTEN THIS MEMBRANE CAP UNTIL YOU HAVE READ THESE INSTRUCTIONS!
There is an elasc seal around the membrane cap which is used to seal the vent hole on the membrane cap during
normal operaon. Prior to servicing the probe the elasc seal must be slid down to expose this vent hole. Do not
loosen the membrane cap without having the vent hole exposed and unobstructed. If the vent hole is blocked
o when the membrane cap is removed, then the membrane may rupture (i.e. develop a ny tear) and the probe
may not produce accurate readings.
When the membrane cap is lled with electrolyte and fastened onto the probe (in accordance with
instrucons in the operaons manual), you will noce some electrolyte bleed out from the vent hole. This is normal.
NOTE: DO NOT INSERT THE PROBE INTO THE FLOW CELL, OR GET IT WET, WITHOUT HAVING THE
VENT HOLE COVERED WITH THE ELASTIC SEAL!
The purpose of the vent hole is to protect the membrane from being exposed to internal pressure or vacuum
when the membrane cap is screwed onto or o of the probe p.
For further assistance please contact turtetough.com.au
TURTLETOUGH
IPBI08 – Probe Warning
Elasc Seal
This is the posion
the seal should be
in when the probe
is inserted into the
ow cell.
Vent Hole
This vent hole must
be exposed prior
to unscrewing the
membrane cap from
the probe p.
7

T T-CL2
Free Chlorine Sensor
TURTLETOUGH
The chlorine sensor is a membrane covered potenostac 3-electrode system. The measuring electrode is
membrane covered and is in the electrolyte area together with the reference electrode. This electrolyte area
contains a special electrolyte and is separated by a second membrane from the sample water.
In this measuring method chlorine diuses out of the sample water, through the membranes and causes an
electrical signal at the measuring electrode. The electrical signal at the measuring electrode is proporonal to the
chlorine concentraon and is amplied by the electronics in the sensor. The measuring signal is independent from
the temperature of the measuring water due to an integrated temperature compensator.
4.0.1 Installaon
• Please refer to Site Selecon, secon 2.1
• Ensure the probe housing is mounted vercally, to ensure correct water ow over the sensor
• Ensure the installaon is stable, secure and away from vibraon
• Ensure the sample is free from bubbles and is well mixed
4.0.2 Setup
CAUTION:
When you unscrew the lled (or unlled) membrane cap, ensure
the vent band has been removed from the vent hole, to allow
air to enter, otherwise the membrane will be destroyed by the
vacuum.
• Unscrew the membrane cap from the electrode sha.
• Place the membrane cap and the G-holder onto a clean and
non-absorpve bench.
CAUTION: Never polish this area of the probe.
8

T T-CL2
Free Chlorine Sensor
9
TURTLETOUGH
• Fill the membrane cap up to the edge with the enclosed electrolyte.
• Be careful, so that there are hardly any bubbles.
• Then replace it onto the bench.
NOTE: Read the electrolyte MSDS sheet.
• Fill up the G-holder with electrolyte.
• Be careful that there are no bubbles.
• Hold the electrode sha upright and push the electrode nger carefully
into the lled G-holder.

T T-CL2
Free Chlorine Sensor
10
TURTLETOUGH
• Hold the electrode sha (with the G-holder) upright and put it on the
lled membrane cap.
• Then screw the membrane cap onto the electrode sha.
• Turn it anclockwise unl the thread engages, then screw slowly the
electrode sha clockwise (by hand) onto the membrane cap.
• Excess electrolyte will escape through the vent hole in the membrane cap.
• Do not close this vent with your nger.
WARNING: Electrolyte may spurt from the vent hole.
Excess electrolyte, or electrolyte on your skin or in your eye,
should be washed o immediately with water.
The electrolyte contains potassium halide.
• Check that the membrane cap is completely screwed in up to the stop.
• There is initial resistance from the o-ring, continue past this to the
electrode body.
• Ensure the vent band is seated correctly.
• Wash o the excess electrolyte with water.
• Inial run-in is 1 hour prior to calibraon.
• Calibraon should be repeated aer 24 hours.
Roune calibraon should be weekly.
CAUTION: When you unscrew the lled
membrane cap, ensure the vent band has
been removed from the vent hole, to allow
air to enter, otherwise the membrane will be
destroyed by the vacuum.

T T-CL2
Free Chlorine Sensor
11
TURTLETOUGH
4.0.3 Maintenance
Cauon: The coang of the electrode nger must not be cleaned with emery cloth or other polishing agents!
Do not unscrew the metal end ring of the membrane cap, as this will destroy the membrane.
4.0.3.1 Roune maintenance
The electrolyte should be changed every three months, although it may be possible to extend the period if low
levels of chlorine are measured. It may be necessary to use a new membrane cap when changing the electrolyte.
In any event, the membrane cap should be changed every 12 months. If your unit has been ordered with the auto
calibrator then the calibraon reagent should be changed every six months.
4.0.3.2 Diagnosc maintenance
4.0.4 Storage
To store the probe the membrane cap is unscrewed (note previous warning). The membrane G-holder must
be removed and discarded. The membrane cap and electrode nger should be rinsed in clean water and dried
in a place free of dust with a so paper towel (do not damage the electrode areas). The dry membrane cap is
then loosely screwed onto the electrode sha. The membrane must not rest against the measuring electrode.
Do not fully ghten the membrane cap.
When pung the probe back into use aer storage, the electrode p must be cleaned with the special abrasive
paper supplied (see 4.0.3.2) and a new membrane cap and G-holder must be used (follow 4.0.2).
4.0.5 Spares
Free Chlorine Membrane cap M48 G
Free Chlorine Electrolyte ECS2.1/GEL
Please contact us regarding any other spares.
4.0.6 Single owcell
The o-ring is rst inserted in the 1” opening of the adaptor followed by the PVC shim. Then the 1” PVC threaded
end cap is screwed in loosely. The probe is inserted into this prepared probe housing. Then the probe is xed
in place by ghtening (hand ght only) the threaded end cap, so the probe cannot move away from its posion.
If calibraon is impossible due to unstable or low output,
remove the sensor, displace the vent seal so that the vent hole
opening is free. The membrane cap can then be unscrewed. The
electrode nger is cleaned with a clean, dry, so paper towel.
With the special abrasive paper supplied (blue) just the p of
the dry electrode (the working electrode) is cleaned. Place the
special abrasive paper on a paper towel (on a at, dry surface), hold
it at the corners and rub the electrode p of the perpendicularly
held probe two or three mes across the abrasive paper. Then
rell with electrolyte (see secon 4.0.2). If necessary, use a new
membrane cap.
Recommendaon: Change the electrolyte every 3 months.

T T-CL2
Free Chlorine Sensor
12
TURTLETOUGH
4.0.6.1 Inseron
There should be 25mm (1”) of clear space between the p of the sensor and the conical base of the owcell tube.

T T-CL2
Free Chlorine Sensor
13
TURTLETOUGH
4.0.7 Double owcell
The sensor ts into the right-hand entry (if used with a pH or ORP probe). The o-ring is rst inserted in the 1”
opening of the adaptor followed by the PVC shim. Then the 1” PVC threaded end cap is screwed in loosely. The
probe is inserted into this prepared probe housing, so the ow passes directly over the probe p. Then the probe
is xed in place by ghtening (hand ght only) the threaded end cap, so the probe cannot move away from its
posion.
As with the single owcell, there should be 25mm (1”) of clear space between the p of the sensor and the conical
base of the owcell tube (refer to 4.0.6.1) if using the le-hand entry.
5.0 Commissioning
It is recommended that the instrument is commissioned by a trained commissioning engineer. During
the commissioning the engineer can train the operators in the correct operaon of the instrument.
The installaon and commissioning should proceed in the following manner:
1) Install the unit
2) Check the sensor for damage
3) Make up the sensor
4) Ensure that water ows through the owcell and over each sensor’s measuring end.
5) Calibraon
Aer the Cl2 sensor has completely polarized (from 2-12 hours of water ow with the electronics turned on
and chlorine in the water), calibrate the sensor using a suitable test kit (eg DPD for Cl2).
6.0 Troubleshoong Guide
Sensor faults will usually show themselves as unexpected readings. Unexpected readings can also be caused
by the electronics, so when troubleshoong it is important to determine the cause of the variable readings.
Cl2 sensors can be supplied with an electrochemical probe simulator. If the instrument displays unexpected
readings, unplug the probe and plug in the probe simulator. This supplies a xed constant reading to the analyser.
If the reading is not constant, please contact your supplier.
Symptom Possible cause Possible soluon
Chlorine
incorrect
Sensor disconnected Check sensor connected
Probe depolarised aer
extended zero period
Probe will re-polarise aer
2-12 hours
Sensor connected
incorrectly
Check sensor connected to
correct channel
6.1 Further help
For more troubleshoong, please visitt www.turtletoughsensors.com

T T-CL2
Free Chlorine Sensor
14
TURTLETOUGH
Appendix A
General Layout Drawing - Flowcell

T T-CL2
Free Chlorine Sensor
15
TURTLETOUGH
Appendix B
CERTIFICATES AND APPROVALS
Declaraon of conformity
The product meets the legal requirements of the harmonised European standards.
Process Instruments (UK) conrms compliance with the standards by axing the CE
symbol.
Waste electrical and electronic equipment (WEEE) Important informaon.
Disposal of old electrical & electronic equipment
This symbol on the product or on its packaging indicates that this product
shall not be treated as household waste. Instead, it shall be handed over to the
applicable collecon point for the recycling of electrical and electronic equipment. By
ensuring this product is disposed of correctly, you will help prevent potenal negave
consequences for the environment and human health, which could
otherwise be caused by inappropriate waste handling of this product. The recycling of
materials will help to conserve natural resources. For more detailed informaon about
recycling of this product, please contact your local council or city oce, your waste
disposal service or the organisaon from which you purchased the product.
Appendix C
RETURNS PROCEDURE AND ‘FREE OF CONTAMINATION SHEET’
If the analyser has to be repaired, please return it cleaned to your local sale organisaon or Turtle Tough.
Please use the original packaging, if possible. Please enclose the completed Contaminant sheet
(copy the Contaminants page, on the following page) with the packaging and shipping documents.
No repair will be undertaken without a completed contaminant sheet!

T T-CL2
Free Chlorine Sensor
16
TURTLETOUGH
CONTAMINATION SHEET
Dear customer
Because of legal requirements and for the safety of our employees, we need this “Contaminaon sheet” lled
in (with your signature) before your instrument can be handled. Please put the completely lled in declaraon
with the instrument and the shipping documents. Add any safety data sheets (MSDS) and/or specic handling
instrucons if necessary.
Type of instrument / sensor:
Serial number:
Medium / concentraon:
Temperature: Pressure: Conducvity: Viscosity:
Cleaned with:
Please mark the appropriate warnings.
Radioacve
Explosive
Causc
Poisonous
Harmful to health
Biologically hazardous
Inammable
Safe
Company:
Contact name:
Address:
Telephone/fax no:
Email:
Order no.:
I hereby cerfy that the returned equipment has been cleaned and decontaminated according to good pracces
and is in compliance with all regulaons. This equipment poses no health or safety risks due to contaminaon.
Company stamp (if available) and
legally binding signature
Date: / /
For more informaon about service and repair go to www.turtletoughsensors.com

T T-CL2
Free Chlorine Sensor
17
TURTLETOUGH
Appendix D
WARRANTY
Turtle Tough warrants its product to be free of defects in material and workmanship for a period of one (1) year
from date of shipment to the original customer. Upon receipt of wrien noce from the customer, Turtle Tough
shall repair or replace (at the discreon of Turtle Tough) the defecve equipment or components. Turtle Tough
assumes no responsibility for equipment damage or failure caused by:
A. Improper installaon, operaon, or maintenance of the equipment.
B. Abnormal wear and tear on moving parts caused by some processes.
C. Acts of nature (i.e. lightning, ooding, etc.)
This warranty represents the exclusive remedy of damage or failure of the equipment.
Under no circumstances shall Turtle Tough be liable for any special, incidental, or consequenal damage, such as
loss of producon, prots or product quality. The warranty cannot be guaranteed if the customer fails to service
and maintain the equipment in accordance with Turtle Tough wrien instrucons and policies, as stated in the
Operaons Manual.

T T-CL2
Free Chlorine Sensor
18
Appendix E
Safety Data Sheet
Electrolyte ECS2.1/GEL/Electrolyte FC1 for Membrane Free Cl2- Sensor
IPM19 HaloSense Free Chlorine Probe
Manual date 12/1/2011 Page 25 of 29
Appendix E

T T-CL2
Free Chlorine Sensor
19
Appendix E
IPM19 HaloSense Free Chlorine Probe
Manual date 12/1/2011 Page 26 of 29

T T-CL2
Free Chlorine Sensor
IPM19 HaloSense Free Chlorine Probe
Manual date 12/1/2011 Page 27 of 29
20
Table of contents
Other Turtle Tough Accessories manuals
Popular Accessories manuals by other brands

Texas Instruments
Texas Instruments IWRL6432BOOST user guide

Aeon Labs
Aeon Labs MultiSensor 6 user manual

Nexmosphere
Nexmosphere X-EYE 200 Series product manual

Hathaway
Hathaway MAGNUM NG1042D owner's manual

Bowflex
Bowflex SelectTech 5.1 Bench Assembly & owners manual

LEGRAND
LEGRAND 0 489 36 quick start guide