Inlet
☞Connection lines at the pump inlet have to be performed conductive grounded
and gas tight (designation of the pump connections see gures).
☞The diameter of the inlet pipeline should be at the least as large as the diameter
of the pump connection pipelines.
☞Obey max. inlet pressure (see ”Technical data”).
Outlet
☞ Connection lines at the pump outlet (small ange) have to be performed conduc-
tive grounded and gas tight (designation of the pump connections see gures).
☞The diameter of the inlet pipeline should be at the least as large as the diameter
of the pump connection pipelines.
☞Obey max. inlet pressure (see ”Technical data”).
☞ Lead away exhaust gas or condensate at the outlet of the pump and the overpres-
sure safety valve under consideration of all applicable safety measures. Avoid
backow of gases or condensate denitely.
☞Escaped and/or intentional or unintentional released inammable gases or va-
pours which may lead to danger of explosion have to be lead away safely or lead
to an appropriate place. If not possible enclose the gasses, vapours or mist safely
or dispose in an other way. In case of mixtures or different gases the measures
must consider the highest risk.
Check the overpressure safety valve at the outlet regularly.
☞Attention: If the exhaust pipeline is blocked the pumped gases may escape
into the environment through the overpressure safety valve. If necessary
take appropriate measures.
If pumping dangerous gases install a second completely independent exhaust
pipeline instead of the overpressure safety valve. If necessary assemble a hose
nozzle instead of the overpressure safety valve to lead away the gases through
a second exhaust pipeline. Attention: Never combine the exhaust pipelines, i. e.
provide a sperate line for each exhaust.
Inlet and outlet of the pump
☞To avoid damage of the pump due ignition of gases or gas mixtures before or
behind the pump the user has to provide suitable safety devices against in-
coming of ames according to EN 12874. The user must ensure the suitability
concerning gas ow, chemical resistance and safeness against clogging prior to
use.
Inert gas purge
☞The user must take suitable precautions to prevent any formation of explosive
mixtures in the expansion chamber using inert gas to purge the housing. Other-
wise in case of a diaphragm crack, mechanically generated sparks, hot surfaces
or static electricity may ignite these mixtures.
☞Inert gas connection lines (inlet and outlet) have to be performed conductive
grounded and gas tight (designation of the pump connections see gures).
☞Connect the inert gas connections at the crankcase not serial but parallel.
☞ We recommend an inert gas purge of ≥ 1 l per minute.
☞ Provide an external pressure limitation to a pressure of 1.1 bar absolute if inert
gas is connected to the pump or at an air admittance valve. The inert gas has
to be dry and pure. It could be recommendable to provide an appropriate control
system to supervise the inert gas ow which switches off the pump in case of
failure (e. g. absence of inert gas) if necessary.
☞To control the faultless function of the diaphragm it could be recommendable in-
stall a gas specic detector at the outlet of the inert gas purge. A signal from
the detector indicates a diaphragm crack. Switch off the pump immediately and
check diaphragms.