Battery condition
The device can be used for approx. 30 continuous hours when using manganese batteries. The
battery level is shown in the following manner. Refer to “Installing the battery” if the battery needs to
be replaced. Note that the battery level display is not an equivalent amount. The level may drop
suddenly depending on the condition of the battery and the working environment.
FULL EMPTY
Battery level (4 stages)
Connecting the grounding cable
The accuracy might be affected by the static charge on the person
making the measurement if Eye-02 is ungrounded. For proper
measurement, the operator should be grounded using a wrist
strap and/or Eye-02 shall be grounded on the right side grounding
socket by the grounding lead provided with the equipment.
Connecting the grounding cable
REMEMBER
The Eye-02 case is made of conductive resin. The grounding terminal
functions as the reference for the measurement circuit. Always ground the
terminal to ensure accurate measurements. The measurement accuracy will
not be guaranteed if the terminal is not correctly grounded.
Handling the Ion Balance plate
This device is equipped with an Ion Balance plate for measuring the Ion Balance. The Ion Balance
plate is mounted on the device’s sensing section side when at the time of shipment.
Switch the measurement mode to measure the Ion Balance. Refer to “Selection of the
measurement mode” for details on switching the measurement mode.
Remove the Ion Balance plate when measuring the Static charge voltage.
Before storing the device in the soft case, attach the Ion Balance plate onto the device’s sensing
section side to provide sufficient protection.
Handling the Ion Balance plate
REMEMBER
Make certain that the surface of a white plastic plate under the Ion Balance
plate is clean before attaching it to Eye-02 for Ion Balance measurement.
REMEMBER
The Ion Balance plate should be stored or used in a place having less than
60 %RH.