6 Vestfrostsolutions.com EN
WARNING:
Sharp edges on cabinet, compressor compartment, evaporator,
ventilation cover and on internal equipment can occur. Please be
aware to avoid injury.
WARNING:
The condenser on the back of the appliance will in some cases
have a hot surface. Please be aware to avoid injury.
WARNING:
The appliance must be connected to power minimum 12 hours
before using it for storage of medicine.
●Always keep the keys in a separate place and out of reach of children.
●Do not step on the lower panel to reach medicine in the top of the appliance.
●Before servicing or cleaning the appliance, unplug the appliance from the mains or disconnect
the electrical power supply.
●If the supply cord is damaged, it must be replaced by the manufacturer, its service agent, or
similarly qualified persons in order to avoid a hazard.
●Relevant for Australia: Supply cord fitted with a plug complies with AS/NZS 3112.
●Frost formation on the interior evaporator wall and upper parts is a natural phenomenon. There-
fore, the appliance should be defrosted during normal cleaning or maintenance.
●Please note that changes to the appliance construction will cancel all warranty and product
liability.
●This device is intended to be used exclusively for medical products.
●If medicine is spilled in the appliance or the defrost water canal is has to be cleaned immedi-
ately to avoid the medicine to evaporate to the surroundings.
●If the instructions is lost please contact your supplier of the appliance to have a new instruction
for use.
●If service needed to this device, please be aware of only using service personnel with educa-
tion in handling medical devices.
Flammable Material Storage Freezer
Sparkfree interior to reduce the risk of internal explosion.
Interior temperature range: +1°C to 15°C
The interior comprise this safety marking:
II 3G nA IIB T6
NOTE:
●Combustible materials may be stored inside cabinet. but the cabinet is not approved for
potential explosive areas.
●Combustible materials shall be stored in approved containers.
●Danger of sparks due to electrostatic discharge. Do not use a dry cloth to clean plastic
parts. Clean plastic parts with a damp cloth only!