ViewLight Inami L-0955-SD User manual


This time, Zoom LED portable operation microscope , Thank you for purchasing.
Please use properly on before using this manual well read, and you know how to use, etc..
And stop use if you can't follow the preparation and operation, please contact the dealer or our
company.
Table of contents
1. Safety warnings and cautions ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2
2.Names of parts, accessories and product overview and structure ・・・・・・・・・・ 3
3.Assembly instructions ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 5
4.Instructions ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7
5.Replacing the cleaning of medical equipment, consumables and storage ・・・・・ 10
6.Maintenance ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 11
7.Troubleshooting ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 13
8.Spec ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 15
9.After-sale service ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 15
・Please note that there is a regarding the contents of this manual, for the improvement of future change
without notice to you.
・Measures about the contents of this manual has been created, but any errors, please contact your dealer if you
have concerns, such as omissions.
・Will be replaced if unclear, missing. Please contact the dealer.

1. Safety warnings and cautions
Before using this safety warnings and precautions well read on the right please.
Keep important this instruction manual so that you can view at any time.
Because the precautions listed in this section are provided to serious safety always keep.
The symbols and their meanings in the following.
Danger
Indicates a hazard that can cause deaths, or imminent danger of injury.
Warning
Indicates content or applied incorrectly, creates the possibility of death or injury
is expected.
Attention
Indicates the possibility of injury to people or material damage of injury.
However, it may lead to serious consequences in some situations.
Pertaining to the houses and household goods and material damage, damage.
Alerts
・During the replacement parts please be sure main power switch to the OFF (O).
Electric shock while energizing to work.
・Body 10 times over not tilt please.
Beside the body and damage the injured, houses, and other instruments.
Attention
・Please attach ground wire.Upon failure or short circuit of an electric shock.
・Unplug the power supply plug ( Plug ) Hands must be dry, with advanced plug and unplug
(Pointing ) Please. An electric shock may ignite or short.
・Please do not use firmness the power supply cord or plug, or corrupt.An electric shock may ignite
or short.
・Hurt power cord forcibly bend, pull, twist, bind, place heavy items, please do not
conductive.Power cord is damaged, can cause fire or electric shock.
・Power cord no dirt, dust, oil, so please, clean at all times.Terminals may be dirty and cause fire or
malfunction.
・Outside of clinic hours, plug-ins, unplug from the wall outlet.On the cause of the short circuit and
fire.
・Please do not use other bulbs, fuse, electrical components.Risk of fire due to heat.
・Do not use lenses other than specified.Be able to observe the proper focus and magnification.
・When placing your body between table and in table clocks-lamp system handles firmly.On the
stand next to the slide, falling damage the injured, houses, and other instruments.
・And absolutely never disassemble the body.On the cause of the failure, or fire.
2. Name parts and accessories with the product overview and structure

1) Product overview
・Zoom LED portable operation microscope Is lightweight, compact, and handling ease has been designed
for the purpose.
・You can set the position to facilitate adoption of the spring arm.
・By 45 degrees tilting mirror to get rid of ENT , Can be used, such as orthopaedic surgery.
・Microscope head and compact design for observation, and made it possible to observe for a
long time.
・You can choose from two types of models, depending on the intended use.
L-0955SD
Table clamp system
L-0955SD + K-1422
Base stand system

2)付属品
2) 各部の名称
1 dust cover 1
2 Objective lens cap 1
3 eyepiece cap 2
4 Fuse 2
5 Mouth opening rubber 2
6 Disinfection cap 7
7 Power cord 1
8 Charger 1
1 12.5 Eyepiece lens
2 PD adjustment knob
3 Microscope removal screw
4 Zoom magnification control knob
5 objective lens
6 Filter switch
7 Microscope head rotation knob
8 Second arm balance knob
9 Second arm rotation control knob
10 Power supply unit
11 Illumination control knob
12 First arm rotation control knob
13 Microscope head adjustment knob
14 Microscope head angle adjuster
15 First arm base
16 First arm
17 second arm
18 Zoom microscope head
19 Focusing knob
20 Delivery Ali
21 Table clamp (SD type)
22 Stand
23 Battery unit
24 Battery

3. Assembly instructions
1)installation
(1) Installation conditions
·Please place it in a place not subject to water.
·Please place it in places where there is no fear of adverse effect due to air including
atmospheric pressure, temperature, humidity, ventilation, sunlight, dust, salt, ion content
etc.
·Please pay attention to stable conditions such as inclination, vibration and impact.
·Do not install in place of chemical storage or gas generation.
(2) Assembly method
·When installing, please adhere strictly to ensure that the environmental conditions at the
time of use are the following environmental conditions.
2)Assembly method
(1)for the first arm-based
①
clamp
(L-0940SD)
Stabilize the base with the handle firmly.
(Maximum table thickness:
70
mm)
②base(K-1422)
Pillar and base please tighten at the hexagon
bolt with a hole (4 screws).
(2)first arm
First arm installed Teflon washer and insert
into base.
The arm slowly insert vertically.
(3)second arm
No. 2 arm installed Teflon washer and insert
first arm.
The arm slowly insert vertically.
Environmental conditions
Temperature
+10℃~+35℃
Relative humidity
30%~75%
No.1 arm
No.2 arm
Teflon washer
No.1 arm
Teflon washer
base
pillar

(4)eyepiece with Prism box installation
Scale conversion box installed eyepiece Prism box,
clamp screws.
(5)installation of the power unit
Power supply unit to the Teflon washer Mounting
the chapter 1
Perpendicular to the arm slowly insert.
The fixed position, tighten the hex screws under the
one-arm. Screw the click in place. Please make
sure the swivel transformer.
(6)connect the power cable
Plug the power cable into the mirror cylinder of the
fix.
Plug the power cable into the power supply Jack,
please fix.
Please insert the battery cable into the power
supply unit and fix it.
Batter cable
power cable
eyepiece
Clamp screws
Scale conbersion
box No.1 arm
Teflon washer
Supply unit

4. Instructions
(How do)
(1)power switch
If you turn on the unit, press the power switch |.
(2)coaxial lighting brightness adjustment knob
Darkens the light mirror tunnel side coaxial lighting
brightness adjustment knob turns to the right and turn left.
(3)フィルター
Please select a filter using the filter switch knob on the
tube side mirror.
In each filter set you want.
Type Blue, retinal protection filters,
Color temperature change filter
(4)vertical fine adjustment knob
You can refine your focal length by
manipulating the vertical fine adjustment knob
and loosen the clamping lever. (Fine amount: 30
mm)
Vertical fine
adjustment knob rise
fall
Clamping lever
Fine amount
30mm
Filter switch knob
coaxial lighting brightness adjustment knob
Power switch

(5)turn locking knob
Tightening the lock knob rotation on postgraduate
offered by the rotation, fixed.
(6)diopter adjustment
Rotate the eyepiece and adjust the scale to a diopter
measured by.
(7)adjustment
PDcontrol knob to rotate, and adjust the scale to a
measured by eyes wide.
(Microscopic head angle adjuster)
Microscope head is the microscope head tilt locking knob loose, 0 °~ 90 °up to adjust to any position you
can.
(8)Battery
Insert the battery cable into the power supply unit and install the battery in the battery unit, you can operate
it with the battery.
When power is supplied from the AC cable, operation with the battery can not be performed. When operating
with a battery, please stop supplying power from the AC cable.
Battery cable
AC cable
Head tilt locking knob
eyepiece
PD control knob
Tightening the lock knob

(9)Battery charging
Insert the charger's power plug into the indoor outlet (100 to 240 VAC) and install the battery.
The red lamp lights up while charging. The green lamp will light up when it is fully charged.
*Lamp lighting status indicates the following meaning.
Yellow Lit:Energized Red Lit:Charging Green Blinking:Practical charging completed
Green lighting:Charging completed yellow flashing:Charging standby Red flashing:Not charging
After charging, unplug the power plug and remove the battery.

5. Replacement consumables, cleaning, and storage
1) Cleaning
(1) Body
・And use a neutral detergent if the body is dirty, wipe it with a soft cloth.
(2) Operation knob Cap
・Operation knob Cap (disinfection CAP) for the disinfection of all removable has become.
・Carefully disinfected ago use.
After disinfection, the CAP is reusable.
・Operating handle, degraded operation knob CAP be removed and replace with new ones.
(3) Power cord (plug plug)
・Wipe with cloth if the power cord is dirty with dust, oil, etc.
・Water adheres to the power cord with a wet cloth, shock, becomes the cause of the fire.
(4) Cooling fan
・And wipe with dry cloth remove the cooling fan case cooling fan is dirty with dust, oil, etc..
(5) Optical components
・And use a cleaning fluid (ether 8: methanol 2) if fingerprints and dirt on the lens, mirror, gently wipe it with a
soft cloth.
・If the dust is attached to the lens, mirror, please clean with a soft brush.
・Cleaning of the lenses, mirrors, 3 x done remove from the zoom body.
・Please note 10 minutes with lenses, mirrors, do not scratch.
2)Replacement of consumables
(1) Fuse replacement
Warning
Replacing the fuse, make sure main switch
"OFF" (O)
Then has become the fuse at high
temperature which may directly, do not touch with bare hands.
For high-temperature burns.
(1)to switch the power unit "OFF" (O)
Then please.
(2)press the fuse holder to the left Rotate
And remove from the body.
(3)remove the fuse wire is cut, new fuse
Attach the fuse holder.
(4)Rotate the right press the fuse holder
Attach to the body.
3)Storage and storage
(1) How to store
・The main switch must be "OFF" (O) It please.
・Please remove from an electrical outlet power cord.
・Of the body to prevent stains, take the dust cover.
(2) Storage location
・Keep the splashed water.
・Harm caused due to air pressure, temperature, humidity, ventilation, sunlight, dust, salinity,
ion-containing keep them in unlikely places.
Fuse holder

・Tilt, shake, please note that shock and steady state.
・Please do not store flammable gas and chemicals storage place.
(3) Environmental conditions
During storage please adhere to the following conditions.
Storage conditions
Temperature
-10℃~+55℃
Relative humidity
10%~95%
6. Maintenance
Depending on operating conditions, May require parts replacement and repair.
Products normally used for the customer Please do the inspection of equipment and parts.
And also used for a long time, or long term storage after use, before using make sure
equipment successfully and safely activate when you check out.
1)Routine inspection
Check (check daily) Please be careful before using be sure any abnormalities.
Inspection items Contents of inspection How to set up and adjust
Lateral control of
the aircraft
To verify that the level horizontally at the body. Aircraft to be placed
horizontally.
Connecting the
cords
To verify that the code in all the earth connections are
accurate and complete.
Connected to ground, all of
the code.
Light coaxial
illumination
To verify that the coaxial illumination light is set to the
appropriate value.
To set the light coaxial
illumination.
Each arm To verify the movement arm is lighter and can be fixed in any
position.
To see the movement of each
arm.
2)Periodic inspection
Products normally used for the 3-6 months once the monitoring On a regular basis Inspection
(inspection) Be sure Please.
Inspection items Contents of inspection How to set up and adjust
Check coaxial
illumination
Make sure no lighting around the bluish light.
Make sure than the light can be set.
If abnormality is identified, please
contact dealer.
Check the fuse Check to see if the fuse is not blown. If abnormality is identified, refer to
the procedures for fuse
replacement, replace the fuse.
No. 2 arm
movement
To verify that no. 2 arms to move up and down
gently, and stopped at any position, you can secure
the locking knob.
Fails if the second Adjustment of
arm clamping force See, moving
the arm up and down to adjust To.
Check mirror stand
out-degree
adjustable
mounting
Mirror stand level adjuster mounting nut, sure
enough.
Fails if the mirror stand
degreesAdjustment of the clamping
forceSee Mirror stand level
adjuster Of coordinate
movements To.
No. 2 arms
Confirmation of the
mirror with a
President
The second nut arm mirror with a President before
any loosening.
If abnormality is identified, the
mirror head rotation Adjustment of
the clamping force Adjust the
mirror head rotation movement,
see To.

3)How to adjust the
(1)
the second arm balance
Remove the locking knob and shaft, please open the
arm.
Adjust the balance arm, arm locking nut tightened with
aclasp (bar, etc.).
(2)microscope angle tightening force adjustment
Fixation nut side nut screws, loosen the tightening
force adjustment tool, tightening the locking nut,
adjust the microscope angle tightening force.
After the adjustment, flat nut screws, sure.
(3 )
mirror heads rotating clamping force
adjustment
Fixation nut side nut screws, loosen the fastening
and use the power tool, tightening the locking nut
and adjust the mirror head tightening torque.
After the adjustment, flat nut screws, sure.
Power tool
Side nut
locking nut
locking nut
Side nut
Power tool
shaft
strong
weak
Locking nut

7. Troubleshooting
If faulty, and refer to the items listed below, please check.
If, if still not resolved, please contact dealer or us.
Symptoms Cause How to deal with
Light does not
light
Does not contain a main power
switch
Put the main power switch from the power cables are
properly connected.
A faulty connection of main power
connector
Please connect the power supply connector is connected to the
power supply unit correctly.
Fuse is disconnected Replace new fuses.
Disconnection of the power cable Replace the power cable.
Battery will drain Please charge in a dedicated charger.
The battery is damaged Please contact the dealer.
Lighting is Dim Not been adjusted brightness adjustment
knobs are properly positioned
Brightness adjustment knob to adjust to the proper position.
Dirty optical parts Please clean the optical parts.
Do not set in the correct position Please set the filters in the correct position.
Binocular microscope
does not work
Eyepiece diopter adjustment, adjustment is
incorrect
Please fit the correct diopter the diopter adjustment eyepiece and
adjustment.
Optical axis of the microscope tele part one Please fix the optical axis of the microscope tele Division.
No. 2 arm natural fall (or
rise) to
Not in balance and coordination Go no. 2 arm balance.
Tightening knob is loose Case No. 2 arm tightening knob.
Fuses and light bulb
filaments to break in a
short period of time
Other than specified fuse or light
bulb is being used
Replace the specified rated fuse or light bulb.
Or input voltage exceeding the rated
voltage voltage enters from the primary
side is unstable.
Input voltage rated voltage ±10 Must stabilize the primary side
input voltage, if the percent change, so please contact us.
Noise on power-up The cooling fan is dirty with dust,
etc.
To remove stains, and performs the cleaning of the
cooling fan.
Each moving part is not
working properly
Each rotation knob is loose Parts and set in the use position, tightening in the fixed knob
securely.
Rotating knob parts are too tight Loosen the rotary knob parts, and set the position using clamping
with fixed knob securely.
Each moving part is dirty with dust, etc. Conducts parts cleaning, removing stains.
Each movement of the shaft surface is dirty Each moving and cleaning of shaft surface and to remove the rust.
Won't hold a charge The fall indoor circuit breaker Put the circuit breaker.
Missing power plug of the charger from the
electrical outlet
Make sure the power plug is properly applied.
Has flashing red light charger Due to deterioration of the battery is no longer charging, replace
new battery.
Experiencing an outage Please wait for the power can be restored.
The battery is damaged Please contact the dealer.

Notes on the use of medical electrical equipment
1)Does not use equipment other than skilled.
2)To note the following when installing the equipment.
(1) To install water-free place.
(2) Of adverse effects caused by air pressure, temperature, humidity, ventilation, sunlight, dust, salinity, Ion,
including placing in unlikely places.
(3) To note stable, such as tilt, vibration, shock (including transportation).
(4) Be no storage of chemicals or flammable gases.
(5) Power frequency voltage and current ( Or power ) To note that.
(6) Battery power status ( , Such as discharge, polar ) To make sure.
(7) Successfully connecting to the Earth.
3) To note the following before using the equipment.
(1) Sure to check switch contact State, polarity, dial settings and parameters of the device to work properly.
(2) To make sure that the Earth is connected permanently.
(3) Sure to connect all of the code is accurate and complete.
(4) To note or loosening equipment combined with an accurate diagnosis, and is likely to cause danger, so 10
minutes.
(5) Direct contact with patients with external circuit to review it.
(6) To make sure the battery power.
4)That the equipment is in use, to note the following.
(1) Be careful not exceeding the time required for the diagnosis, treatment and the amount.
(2) Constantly monitoring equipment in General and patients without anomaly.
(3) Appropriate measures found abnormalities in patients and equipment, patients stop operation of
the equipment in a safe condition and that.
(4) Sure that no patient is touching the equipment.
5)After using the equipment, the following note.
(1) By prescribed procedure switches, turning after the dial was restored to use.
(2) Have the code when detaching cables, without excessive force, such as pulling out.
(3) Note the following about the location.
iTo keep protected against moisture and water.
iiCaused adverse effects due to atmospheric pressure, temperature, humidity, ventilation, sunlight, dust, salinity,
ion-containing air to keep non-place.
iii To note stable, such as tilt, vibration, shock (including transportation).
ivThat do not store flammable gas and chemicals storage place.
(4) After cleaning accessories, codes, guide, organize and summarize.
(5) Equipment not to use it next time so that you always clean.
6)That without breaking anything when you failed to show proper repair is left to the professionals.
7)Equipment modification should not be.
8) Maintenance and inspection
(1) Always done maintenance equipment and parts.
(2) To ensure that reuse equipment not used for a while, before using always equipment properly and safely
operate.
9)Do not use equipment prescribed purposes.

8. Specification
L-0955SD
L-0955SD+K-1422
Electrical
Departme
nt
Use power
AC100 - 240 V 50 / 60 Hz
Power
consumption
70 VA
Use LED
13.5 V 16 W
Filter
Blue filter
Retinal protection filter
Color temperature change filter
Of the
microscope
Objective lens
Focal length f = 175
Eyepiece
12.5
Total
magnification 5X - 15 x (zoom ratio 1:3)
Field of view
diameter
(mm)
Φ35 - 10
Light
60000 Lux (Max.)
Battery 25.2 V 3Ah
Dimensions (mm)
210 (W) x 1150 (D) x 580 (H) 540 (W) x 1200 (D) x 1230 (H)
Weight
17 kg
28 kg
9. After-sale service
1) Service (repair and maintenance)
・If you believe failure to refer "maintenance and troubleshooting" section first, and then, check again.
・If despite correct use status if the fault should occur, please contact dealer or us.
・Please contact me if you please inform specific product name, model, serial number, purchase date and
failure of.
2)Warranty
Warranty on this product. Since we issued the dealer purchased a warranty purchased, make sure it is
filled dealer, taken by mark, please contact.
・Please keep this because the repair free of charge within the warranty period, if the fault should occur,
even though the correct usage.
・Free repair except as described in the certificate of guarantee will be paid.
3)For the retention period of the aftermarket performance parts
Performance parts of the instrument (to maintain the features of the product parts) of the retention
period
Manufacturing after cessation, 7 years Is.
4)Duration of use (useful life)
The instrument, usually with environmental conditions and normal use period (life) Has been
established for 7 years and from the year of manufacture.
Depending on operating conditions, May require replacement parts and repairs, must be Go to
regular maintenance and inspection.
Age ( Life ) Order to maintain the quality, efficacy and safety when used beyond the replacement
parts, repair, and overhaul is needed.
Duration of use ( Life ) Please contact the company if you use beyond.