VWR Puranity TU3+ User manual

Version: 1
Issued : February, 2016
INSTRUCTION MANUAL
VWR
®
PURANITY TU+
European Catalogue Number(s)
171-1231
171-1232
171-1233
171-1234
171-1235
171-1236
171-1237
171-1238
171-1239
Puranity TU3+
Puranity TU3 UV+
Puranity TU3 UV/UF+
Puranity TU6+
Puranity TU6 UV+
Puranity TU6 UV/UF+
Puranity TU12+
Puranity TU 12 UV+
Puranity TU 12 UV/UF+

Legal Address of manufacture
VWR International bvba
Researchpark Haasrode 2020
Geldenaaksebaan 464
B - 3001 euven
+ 32 16 385011
http://be.vwr.com
Country of origin
Sweden

VWR Puranity TU+ | iii
Technical Service
Web Resources
Visit the VWR website at www.vwr.com for:
Complete technical service contact information
Access to the VWR Online Catalogue, and information about accessories and related
products
Additional product information and special offers
Contact us
For information or technical assistance contact your local VWR representative or visit
www.vwr.com.
Warranty
VWR International
warrants that this product will be free from defects in material and workmanship for a
period of two (2) years from date of delivery. If a defect is present, VWR will, at its option and cost, repair,
replace, or refund the purchase price of this product to the customer, provided it is returned during the
warranty period. This warranty does not apply if the product has been damaged by accident, abuse,
misuse, or misapplication, or from ordinary wear and tear. If the required maintenance and inspection
services are not performed according to the manuals and any local regulations, such warranty turns invalid,
except to the extent, the defect of the product is not due to such non-performance.
Items being returned must be insured by the customer against possible damage or loss. This
warranty shall be limited to the aforementioned remedies. IT IS EXPRESS Y AGREED THAT THIS
WARRANTY WI BE IN IEU OF A WARRANTIES OF FITNESS AND IN IEU OF THE WARRANTY
OF MERCHANTABI ITY.

iv
|
Puranity TU+ VWR
Explanatory notes on the operating instructions
EU Mark of Conformity
CAUTI N
Indicates a situation which, if not avoided, could result in damage to equipment or property.
WARNING
Indicates a hazardous situation which, if not avoided, could result in death or serious
injuries.
DANGER
Indicates a hazardous situation which, if not avoided, will result in death or serious injuries.
N TE!
Is used for applicational hints and useful information.
Risk of electric shock. Electrical work on the system is only to be carried out by qualified
personnel.
Protective conductor connection
Connect the power supply to an electrical socket with a protective connection.
Indicates a situation in which protective gloves or clothing must be worn.
Indicates a situation in which protective goggles must be worn.
Indicates a situation in which breathing protection must be used.

VWR Puranity TU+ | v
This information is valid for the system that is received.
For quick and correct service, please include the following information on all inquiries and
replacement parts orders which relate to your system:
The serial number (located on the back of the unit on the nameplate)
The catalog number
Compliance with local laws and regulations
The customer is responsible for applying for and obtaining the necessary regulatory approvals or
other authorisations necessary to run or use the Product in its local environment. VWR will not be
held liable for any related omission or for not obtaining the required approval or authorisation,
unless any refusal is due to a defect of the product.

vi
|
Puranity TU+ VWR

Table of Contents
Chapter 1 Transport and packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Examination on receipt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Complaints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Packaging for return shipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Chapter 2 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Chapter 3 Extend of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Extend of delivery Puranity TU3+/6+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Available Puranity TU 3+/6+ systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Extend of Delivery Puranity TU 12+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Available storage tanks for Puranity TU 12+ system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Available Puranity TU 12+ systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Chapter 4 Intended use of the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Chapter 5 Technical specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Chapter 6 The installation area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Chapter 7 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Connections of the Puranity TU 3+/6+ system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-21
Connections of the Puranity TU 12+ system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-23
Connections of the pure water tank Puranity TU 12+. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-25
Bring your Puranity TU+ system into operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-26
Connecting an external tank to the Puranity TU 12+ system . . . . . . . . . . . . . . . . . . . . . . . .7-29
Illustration of drain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-30
Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-31
Mounting the power pack (voltage supply) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-32
Mounting the ball valve to the Puranity TU+ system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-34
Chapter 8 Flow chart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Chapter 9 How the Puranity TU+ system functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Chapter 10 Initial start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Putting the system into operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-39
Venting the 0.2 µm final filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-40
Chapter 11 perating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Chapter 12 System control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-43
Mono/dual measurement mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-43
Setting the limiting value for conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-45

Setting the limiting value for temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-46
Error message and acoustic error signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-47
Communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-48
Potential-free contact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-48
Chapter 13 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Maintenance Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-49
Replacing the ultrapure cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-50
Replacing the RO/pretreatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-52
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-53
Changing the ultrafilter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-56
UV-reactor assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-58
Replacing the UV lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-59
Replacing the 0.2 µm final filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-61
Autoclaving the 0.2 µm final filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-62
Chapter 14 Equipment disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Chapter 15 Trouble shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Chapter 16 Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-67
Chapter 17 Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Chapter 18 Maintenance record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Chapter 19 Your Distributor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Chapter 20 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

VWR Puranity TU+ | 1
Transport and packaging
Ultrapure water systems are carefully inspected and packed prior to shipping, but damage could
still possibly occur during transport. ifting and carrying the VWR Ultrapure Water System, e.g. to
the installation location, should be carried out by two people.
Examination on receipt
Check the completeness of the goods received against the packing list.
N TE!
Does the packaging show signs of damage? Inspect the system for damage.
Packaging
System

| Transport and packaging
2
|
Puranity TU+ VWR
Chapter 1
Complaints
Should damage have occurred to the goods during transport:
Immediately contact your delivery transport agency.
Save the complete packaging, including the cardboard box, for a possible inspection of them
and/or return shipment of the system.
Packaging for return shipment
If possible, use the original box and packaging material.
When these are no longer available, then:
Protect the system from shock by packing it in bubble wrap and/or packaging foam and a
strong cardboard box.
N TE!
nly a trained person should take the system out of operation.
Prior to send back a operated device, empty the water and dry the system and take out
the cartridges.
Pack the ultrapure cartridges into a bubble wrap and/or packaging foam and take it with
in into the package of the Ultrapure Water System.

VWR Puranity TU+ | 3
Safety precautions
N TE!
bserve these safety precautions for your own safety!
CAUTI N
The VWR Ultrapure Water Systems are modern water purification system intended solely for
the treatment of potable water. The water it produces is not fit for drinking.
DANGER
Work may only be performed on the system electronics when the system has been switched
off and when ESD protection is in place. nly specially trained personnel may work on the
system’s electronics.
Do not install or operate the system until you have carefully read through these operating
instructions and the notes and notices contained therein.
ifting and carrying the ultrapure water system, e.g. to the installation location, should be
carried out by two people. To do this, lift the system in tandem at the two corner points
beneath the bottom plate.
The CE mark is nullified if you make any structural changes to the system or install products
from other manufacturers in/on the system.
Protect the system from frost. The temperature at the installation area must be between
+2 °C and +40 °C.
Always observe the applicable, pertinent codes and regulations valid at the installation
location of the system and follow all applicable accident prevention regulations.
The feedwater pressure must be at least 0.1 bar and at max. 6 bar or 1.45 to 87 PSI. When
the feedwater pressure is higher, install an external pressure reducer.
A low pressure check valve is recommended to prevent back flow of feedwater from water
system.
A grounded 100-240V, 50/60Hz electrical outlet must be available (see “Electrical
connections” on page 16).

| Safety precautions
4
|
Puranity TU+ VWR
Chapter 2
Access to the power supply cord and plug may never be restricted or obstructed.
Unplug the system from the power outlet for all maintenance work on the system.
An atmospherically vented floor drain with a nominal diameter of at least 63 mm (2.48 inch)
(DN50 tube) must be present at the installation location. If no drain is provided it is
recommended that a water detector be installed for safety reasons (for European
specifications only). Failure to provide this will release the manufacturer from liability for any
water-induced damage that may result.
Proceed as follows if the system is not to be operated for an extended period, e.g., over
extended weekend, or during a vacation period:
Switch the system off (unplug the mains plug).
Close the feedwater inlet (close the feedwater tap).
The pump would be damaged if the system were to run without any supply of feedwater. The
manufacturer will not accept any liability should this occur.
The system must be disinfected or rinsed after an extended down time. The cleaning
procedure is described under “Cleaning” on page 53.
The surface or wall on which the system is to be installed or mounted must have an adequate
load-carrying capacity (check the capacity and stability of the wall). The dry weight of the
system is given under “Dimensions and weight Puranity TU 3 +/6+” on page 14 and
“Dimensions and weight of Puranity TU 12+” on page 15.
The surface on which the system is installed must be level and stable not to exceed a
maximum of 2% deviation from evenness is recommended.
When installing the water purification system, always ensure that there is adequate space all
around the system (see “Accessibility to Puranity TU+ systems and pure water tank” on
page 17.) to ensure that ease of use or easy replacement of materials (e.g., filter change,
connection) is possible at all times.
Visually inspect the system at regular intervals. Clean up any water or spills found around the
system immediately.

VWR Puranity TU+ | 5
Safety precautions | Chapter 2
WARNING
Never look directly into a switched-on UV-lamp, as UV-light endangers eyesight!
CAUTI N
To avoid the risk of pinching, crushing, cutting or electrical shock, never perform maintenance
on the system without its protective housing, or while it is in operation. Maintenance work on
the system may only be performed by trained, authorized specialists.
Wear safety gloves when working with cleaning solutions.
If your skin should come into contact with a chlorine product, rinse it immediately with ample,
fresh water.
The system, or system components, may heat up as a result of a defect. It is recommended to
always wear appropriate safety gloves to prevent skin damage or burns.
Wear safety gloves when changing the UV-lamp, in order to prevent that your skin comes in
contact with the UV-lamp glass.
Wear safety glasses when working with cleaning solutions.
If your eyes come into contact with a chlorine product, rinse them immediately with ample,
fresh water and contact a physician at once.
Check the UV-lamp before initial start.
If the UV-lamp is broken:
wear a breathing protector, filter category FFP3 and replace the UV-lamp. For
disposal the UV-lamp refer to “Equipment disposal” on page 63.
ventilate the room well.
To avoid tripping, ensure that the tubings do not lay over the floor.
Apply the general rules of hygiene for laboratories when working with the system.
Do not use any oxidative cleaning agents for cleaning the system. These can damage the
system.

| Safety precautions
6
|
Puranity TU+ VWR
Chapter 2
Proceed as follows when the system has a defect:
Switch the system off and unplug the system from power outlet.
Shut off the feedwater supply.
Contact your local service organization.

VWR Puranity TU+ | 7
Extend of delivery
Extend of delivery Puranity TU +/6+
The following items are included with the Puranity TU 3+/6+ ultrapure water system: Cat. No.:
1x Puranity TU 3+/6+ ultrapure water system:
Puranity TU3+
Puranity TU3 UV+
Puranity TU3 UV/UF+
Puranity TU6+
Puranity TU6 UV+
Puranity TU6 UV/UF+
171-1231
171-1232
171-1233
171-1234
171-1235
171-1236
1 x Ultrapure cartridge 171-1141
1x RO/pretreatment cartridge for a 3 / h system or 171-1142
for a 6 / h system 171-1143
including 1 x assembly kit 171-1140
consisting of:
1 x Final final filter (sterile filter 0.2 μm) 171-1105
1 x Sterile vent air filter 171-1166
1 x Feedwater connecting kit 171-1145
1 x Connectiong tube AD 1/4“ 6 m/8.22 inch 171-1146
1 x Ball valve 1/4“ 171-1147
1 x Cleaning adapter 171-1148
1 x Table top power pack 171-1121
1 x Universal adapter 171-1129
1 x Universal holder 171-1130
1 x Rubber connector to nema plug connector 171-1131
1x Rubber connector to british ST plug connector 171-1132
1x Rubber connector to euro plug connector 171-1133
1 x Plug angle connector 1/4“ 171-1149
1 x T-connector 1/4“ 171-1150

| Extend of delivery
8
|
Puranity TU+ VWR
Chapter 3
Available Puranity TU +/6+ systems
Extend of Delivery Puranity TU 12+
Cat. No.: System
171-1231 Puranity TU 3+
171-1232 Puranity TU 3+ UV
171-1233 Puranity TU 3+ UV/UF
171-1234 Puranity TU 6+
171-1235 Puranity TU 6+ UV
171-1236 Puranity TU 6+ UV/UF
The following items are included with the Puranity TU 12+ pure water system: Cat. No.:
1 x Puranity TU 12+ ultrapure water system:
Puranity TU12+
Puranity TU12 UV+
Puranity TU12 UV/UF+
171-1237
171-1238
171-1239
1 x Pure water tank with 30 capacity or 171-1170
60 capacity 171-1171
1 x Ultrapure cartridge 171-1141
1 x RO/pretreatment cartridge 171-1165
including 1 x assembly kit 171-1185
consisting of:
1 x Final final filter (sterile filter 0.2 μm) 171-1105
1 x Feedwater connecting kit 171-1145
1 x Connectiong tube outer diameter 1/4“ 6 m/8.22 inch 171-1146
1 x Ball valve 1/4“ 171-1147
1 x Cleaning adapter 171-1148
1 x Table top power pack 171-1121
1 x Universal adapter 171-1129
1 x Universal holder 171-1130
1 x Rubber connector to nema plug connector 171-1131
1x Rubber connector to british ST plug connector 171-1132
1x Rubber connector to euro plug connector 171-1133
1 x Plug angle connector 1/4“ 171-1149
1 x T-connector 1/4“ 171-1150

VWR Puranity TU+ | 9
Extend of delivery | Chapter 3
Available storage tanks for Puranity TU 12+ system
Available Puranity TU 12+ systems
Cat. No.: System Capacity
171-1170 Pure water tank 30 liter
171-1171 Pure water tank 60 liter
Cat. No.: System
171-1237 Puranity TU 12+
171-1238 Puranity TU 12+ UV
171-1239 Puranity TU 12+ UV/UF

| Extend of delivery
10
|
Puranity TU+ VWR
Chapter 3

VWR Puranity TU+ | 11
Intended use of the device
The VWR Ultrapure Water Systems are laboratory system and is used for treatment of water. The
system allows the purification of water into the water categories mentioned in the standards of
ASTM 11.01 and ASTM 11.02.
The system must not be operated outside of the specifications as described in the operating
manual. In particular, the system may not be used for production of drinking water and drugs
manufacturing. The system must not be used as a medical device and outside of laboratories.

| Intended use of the device
12
|
Puranity TU+ VWR
Chapter 4
This manual suits for next models
8
Table of contents
Other VWR Water System manuals
Popular Water System manuals by other brands

Watts
Watts Dead Level Trench Drains Instructions for installing

August
August AT-6 manual

Ecodyne Water Systems
Ecodyne Water Systems TB-RO Installation and operation manual

Albion
Albion ART 26 Installation & operating manual

HYDRO GUARD
HYDRO GUARD LongNeck Standard HG-31 User's operation manual

salmson
salmson THERMO'CLOCK Installation and starting instructions