■Connector connection diagrams
CAUTION
NC is a vacant terminal, but do not use it as a junction terminal.
CAUTION
When wiring is made, take such treatment to cover each wired
terminal with an insulation tube.
If the relay output of this meter is used to close or open an inductive
load (relay or solenoid), in order to prevent contact damage (welding,
etc.) caused by electric arc, to heighten the reliability of contacts
and also to extend a meter service life, it is recommended that a
contact protection circuit be inserted in the output of the meter.
*The above values may not always be correct depending on the
characteristic of the load used. Therefore, finally determine these
values in the actual circuit.
5.System functions
5-1 Start/Hold
If the HOLD terminal (No.7 on the lower connector) is shorted with
the COM terminal (No.8 on the lower connector), the displayed value
just after their shorting is held. In addition, measurement will
start by opening these terminals at the necessary timing. The
minimum time required for one time of measurement is about 400ms.
Perform control using mechanical contacts such as relays or
switches. When performing control using TTL or transistors, add
such a circuit as shown in Fig.5.
5-2 Decimal point position setting
First, turn off the power supply and input; insert a screwdriver
in the groove at the bottom of the front panel; and then lightly
turn the screwdriver in the groove to remove the front panel from
the case. The position of the decimal point can be set by the stud
pin on the front surface. The decimal point is not set to any
position prior to meter factory shipment. Therefore, set it to the
desired position at the site. (See Fig.6.)
5-3 Offset adjustment
Apply the minimum measured-voltage (current) to the input
terminals of the meter, then turn the ZERO adjuster from 0 to ±
1000 until the desired display is obtained. (See Fig.6.)
5-4 Fullscale adjustment
Apply the minimum measured-voltage (current) to the input
terminals of the meter, then turn the SPAN adjuster from 100 to
2000 until the desired display is obtained. (See Fig.6.)
5-5 Comparison setting
Turn the slide switch at the front to the LO position, then set
the desired value by turning the LO setting multi-rotating
variable resistor. Next, turn the slide switch to the HI position,
then set the desired value by turning the HI setting
multi-rotating variable resistor. Always set the LO setpoint to
any value smaller than the HI setpoint.
CAUTION
Do not touch any parts other than those specified.
6.Analog output
7.Maintenance
7-1 Cautions for maintenance
Store the meter at a storage temperature of -10℃to +70℃and a
humidity of less than 60%RH. As the case and panel are made of
plastics, do not use any volatile oil such as thinner for removing
stains on them.
WATANABE ELECTRIC INDUSTRY CO., LTD.
MODEL AM-212 Series (3/4)