Weber Ultrasonics SONIC DIGITAL HS MD User manual

A
rt. Nr. 90003403GB
Software: 3.10
A
s per: 29.01.2009
V
ersion: 01
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Operating Instructions
for Ultrasonic Generators of the
SONIC DIGITAL HS MD
MULTI PURPOSE

- 2 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Contents
Introduction 3
Safety instructions 3
Assembly 5
Operation 10
Functions in the Settings menu 18
Functions in the Administrator Settings menu 20
Error messages and troubleshooting 27
Maintenance 28
Specifications 28
Installation variants of the generator 28
Warranty 29
CE Declaration of Conformity 29
Index 29
Service hotline 30
Spare parts 30
Production, sales and service:
Weber Ultrasonics GmbH Phone: +49 (0)72 48 / 92 07-0
Im Hinteracker 7 Fax: +49 (0)72 48 / 92 07-11
Germany www.weber-ultrasonics.de

- 3 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Introduction
Dear Customer, Thank you for purchasing this Weber Ultrasonics product. You have chosen a first-class product that was
developed using the latest technology.
It is vital that you read these operating instructions and follow them carefully before installing or commissioning
your product. Failure to observe these instructions can present a risk to life.
Units may be operated by trained personnel only. Failure to comply with this stipulation will result in a loss of
warranty rights.
The device may only be operated and maintained by personnel who have read and understood this operating
manual and are familiar with the applicable legal regulations for accident prevention and workplace safety.
Safety instructions
Before starting up your device, please read through the following instructions carefully, both for your own safety and for the
safety of the device.
Keep the operating manual in a place that is accessible to all users at any time.
The electrical connections may only be established by trained personnel.
The ultrasonic generator may only be operated by trained personnel.
Due to its principle of operation, additional safety measures must taken if the device is to be used in potentially explosive
atmospheres.
The electromagnetic compatibility complies with the standards and regulations listed in the specifications.
All necessary settings were either made at the factory or are described in this manual. However, should problems occur during
start-up, please do not make any unauthorised adjustments to the device, as this would endanger your warranty rights.
If in doubt, please contact our technical service staff.
Work inside the device may only be carried out to the scope described and, as with the electrical connection, should only be
performed by skilled personnel. When performing such work, the ultrasonic generator must be completely disconnected from
the mains (unplug the mains connection). (Disconnect mains plug)
Inputs or outputs that are used for controlling or monitoring purposes should be twisted and shielded. The device must not be in
close proximity to electrically charged components or cables.
The shielding should be connected to the generator's earth on one side of the generator.
Important: All connections for the signal or control lines are galvanically connected to the generator.
Always observe any warnings or instructions given on the device itself.
The device must always be disconnected from the mains before cleaning or when installing/uninstalling an option. Do not use
liquid cleaners or sprays. Only use a damp cloth.

- 4 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Never operate the device in areas where moisture could penetrate the device.
The platform for the device must be sufficiently stable, as severe damage could result if the device is jolted or dropped.
Ensure that the power supply specifications given on the device are met.
Only converters with the correct frequency, power output and dimensions may be used with this generator.
HF lines between the generator and the transducer and mains cables to the generator should not be rolled up if they are too
long, but must be shortened to the necessary length due to the potential risk of overheating.
With the exception of the permitted tasks listed in the manual, you should never attempt to repair or modify the device yourself.
In the following cases you should disconnect the device from the mains and contact a qualified service engineer:
•If the mains cable or connector is damaged
•If liquid has penetrated into the device
•If the device has been dropped or the housing is damaged
•If the device displays noticeably different behaviour than during standard operation
IMPORTANT: Repairs and modifications may only be carried out by competent, skilled personnel.

- 5 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Assembly
Notes on the installation location 6
Power supply 6
Connections on the back of the generator 7
Assignment of the 15-pole interface socket 8
Description of the interface 8

- 6 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Notes on the installation location
During operation the generator will become very hot.
To ensure that the generator can dissipate this heat well, the generator must be fitted with the fan facing
downward. If the generator is to be fitted horizontally, it must be assembled with the cooling profile at the top. Ensure that there
is sufficient space for air circulation.
If the generator cannot dissipate this heat sufficiently, after a short time it will display an error message signalling that the
temperature is too high (also refer to the "TEMPERATURE LIMIT REACHED" error description).
In addition, ambient temperatures of over 30°C should be avoided at the installation site.
Choose a suitable location that will protect the device from moisture, water, excessive sunlight and heat.
IMPORTANT: • Choose a location that will prevent steam or any other aggressive vapours from penetrating the
device.
• Long-term exposure to chemically contaminated ambient air can result in irreparable damage to
the device.
Power supply
The ultrasonic generator is supplied with alternating voltage of 230 V / 50 / 60 Hz via an external cable, which is connected
using a plug.
It is equipped with an internal main fuse (10 AF).
If you need to change the fuses, unscrew the cover of the housing.
IMPORTANT: • For safety reasons always disconnect the mains cable before replacing the fuse.
• Only connect the generator to earthed mains sockets.
• Only replace blown fuses with new fuses of the same type.
• This should only be performed by qualified and skilled personnel.

- 7 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Connections of the generator
Assignment of the mains connection
Connection PIN 1: 230 VAC (L)
Connection PIN 2: 230 VAC (N)
Connection PIN 3: PE conductor (PE)
BNC-socket:
Converter negative connection: Shield of the BNC socket
Converter positive connection: Inner conductor of the BNC socket
Use only the specially provided connection cable from Weber Ultrasonics to connect the converter!
ATTENTION: • Only use cables specified by the manufacturer.
• Use only shielded transducer connection cables!
• Connent the shielding to the PE conductor on the generator side.
• Only use cables with sufficient cross-section.
• Minimum cross-section 1,5 mm2.
DANGER: RISK TO LIFE!
During operation, i.e. while ultrasound is being emitted, there is a life-threatening voltage on the HF socket!

- 8 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Assignment of the 15-pole interface socket
PIN no. on DSUB socket
interface
Signal name Description
1 +15 VOLT OUT 15 Volt for external use
2 POUT Output 0 – 10 Volt = Power output 0 – 100%
3 P-EXT.-IN Input 5 – 10 Volt for power control
4 GND Shared reference point = Ground
5/9 HF-DA-ERROR Relay root (shared) for "HF-DA" and "ERROR"
6 HF-DA Relay output "HF-DA"
7 ERROR Relay output "ERROR"
8 NC Not assigned! Do not connect!
9/5 HF-DA-ERROR Relay root (shared) for "HF-DA" and "ERROR"
10 NC Not assigned! Do not connect!
11 <> Nominal Open collector output for nominal value
12 FAN-ON Monitoring output = 12 Volt when the fan is running
13 FS-24 V Remote control input (with 15 –24 Volt)
14 FS-GND Remote control input (to GND)
15 GND Shared reference point = Ground
IMPORTANT: A shielded control cable must always be used for the interface.

- 9 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
INTERFACE DESCRIPTION
1.) The "+15 Volt Out" signal DSUB pin 1
A voltage of 15V is available at this output. This voltage can be loaded with max. 100
mA and can, for example, be used to output a voltage for the "HF-DA" and/or "Error" functions.
This voltage can also be used to switch on the generator on the "FS-24 V" input.
2.) The "POUT" signal DSUB PIN 2
A voltage proportional to the power output of between 0 and 10 V
(= 0 – 100 % power output) is available at this output.
Reference point = "GND"
3.) The "P-EXT.-IN" signal DSUB PIN 3
By applying a voltage of between 5 V and 10 V, the power output of the generator can be set to between 50% and 100% of its
nominal amplitude or power output. To activate this function, open the Settings menu (password 6354)
and select the "external voltage" setting under "Ctrl Assignment".
Reference point = "GND"
4.) The "HF-DA ERROR" signal DSUB PIN 5/9
Shared output/input or root for the internal relay "HF-DA" and "ERROR"
(these pins are internally connected).
5.) The "HF-DA" signal DSUB PIN 6
If the ultrasonic generator has been switched on by one of the signals "FS-24 V", "FS-GND" or by using the test button
on the front panel or the external control console and is emitting HF voltage (i.e. there is no malfunction), an internal
floating relay contact is closed (between pin 6 and pin 5/9).
The "root" of this relay contact leads through to pin 5/9 (these pins are connected internally) of the DSUB socket.
This relay contact can now be queried by an external control system. It is, of course, possible to switch through a
voltage connected to pin 5/9 (max. load 24 VDC/100 mA).
It is particularly useful to query this contact if a timer operation is being executed and the generator switches off automatically
when the time has elapsed. The contact can also be set to the opposite polarity in the "I/O polarities" item of the Settings menu.
Factory default settings: Contact closed when generator emits HF voltage.
6.) The "Error" signal DSUB PIN 7
This is the output of an internal relay (root to pin 5/9).
This relay reports generator malfunctions. This means that if the generator is switched on and the power output does not
correspond to the set level for some reason, this relay is activated.
Factory default settings: Closed in the event of a malfunction. The polarity can be changed in the Settings menu under
"I/O Polarities“ ("Error detect"). An external voltage connected to "HF-DA-Error" can, of course,
also be switched through here (max. load 24 VDC / 100 mA).
7.) The "FAN – ON" signal DSUB PIN 12
At this output, a control voltage of 12V is available for monitoring the fan function when the internal fan is running.

- 10 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
8.) The "FS-24 Volt" signal DSUB PIN 13
Activation of the generator (ultrasound on) by the supply of 15 - 24 V between PIN 13 and GND (PIN 4/15)
9.) The "FS-GND" signal DSUB PIN 14
Activation of the generator (ultrasound on) with a relay contact or switch by connecting PIN 14 on the DSUB socket
to GND (PIN 4/15).
10.) The "<> Nominal" signal DSUB PIN 11
Open Collector output: high signal if a window function is activated and the generator is operated outside the
set window, low signal if a window function is activated and the generator is operated inside the set window.
11.) The "GND" signal DSUB PIN 4, 15
The GND signal is available on multiple pins of the DSUB socket.
It is the common reference point for all input and output signals.

- 11 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Operation
Operating and display elements on the front panel 11
Operating and display elements on the external control console 12
The LCD display 13
Functions in the Settings menu 18

- 12 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Operator elements and display of the generator
"TEST" button: This button is used to turn on the generator
for testing
LED-Power: Lights up when the device is correctly supplied
with mains voltage.
HF LED: Lights up when the generator emits HF voltage
Error LED: Lights up in the event of an error
Connection for an external control console.
Operating and display elements of the external controller
HF LED: Lights up when the generator emits HF voltage
Aux1 LED: Lights up when a window function is activated and
the generator is operated outside the set window
Aux2 LED: No function
Error LED: Lights up in the event of an error
Power indicator LED
10 – 100%:
Displays the power output in 10% steps.
Encoder / Select: For setting and entering values
Test button: The generator can be switched on by pressing this
button

- 13 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
The LCD display
Various screens and functions are available in the LCD display:
Start screen
On the top left is the frequency in Hertz, in the centre the amplitude set in percent (adjustable
between 50% and 100%), and on the right the last power output in watts. In the bar graph, the
last power output in percent is also shown in diagram form.
If "external voltage" is set in "Ctrl Assignment" in the Settings menu, the following text appears
in the centre of the display instead of the set amplitude: A EXT.
Power output display depending on the welding mode set
There are five different welding modes (the welding modes can be deactivated in the
Administrator Settings menu and are then no longer displayed in the menu).
If "Weld by Remote", "Weld by Time" or "Weld by Pulse" are set, the current power output is
shown on the display in watts. When the device is switched off, the maximum power output
reached is saved until the next time the device is switched on.
If "Weld by Energy" is set, the power output is shown on the display in watt-seconds together
with the required time. When the device is switched off, the peak power output is saved until
the next time the device is switched on.
If "Weld by Peak Power" is set, the maximum power output is shown in watts on the display
together with the required time. When the device is switched off, the maximum power output
reached is saved until the next time the device is switched on.
Frequency display
The display shows the frequency of the generator, in this example of a 40 kHz generator.
Trigger
A delay time of up to 9.99 seconds can be entered here before the actual welding operation
starts. Usage: If the welding sonotrode is installed in a press, for example, the time which the
sonotrode requires to reach the workpiece can be set here.
The delay time is set by pressing the encoder. The delay time can be set by turning the encoder.
The selected delay time is confirmed by pressing the encoder again. In the "off" position this
setting option is deactivated.
This menu item can be deactivated in the Administrator Settings menu and then no longer
appears in the menu.

- 14 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
5 different welding modes
Five different welding modes can be selected via the "Welding" menu item (the welding modes can
be deactivated in the Administrator Settings menu and then no longer appear in the menu). If only the
Remote Mode is enabled and P-Window and T-Window are not, the "Welding Screen" is omitted
completely, as no settings can be made.
Weld by Remote
If the generator is set to "Weld by Remote", welding is carried out for as long as the test button
is pressed or if one of the remote control inputs is active. If a delay time was set in the Trigger
menu, the welding operation does not start until the set time has elapsed. If you release the test
button or deactivate the remote control input before the set trigger time has elapsed, the error
message "Cycle not completed" appears on the display and the red Error LED lights up. If
Remote Mode is selected, the generator ends the welding mode with an inactive remote signal,
and then switches to Holding and After Burst, if enabled. If the remote input or the test button
are activated during After Burst, the generator switches once more to welding mode without
deactivating the output signal.
Welding mode is set by pressing the encoder. The required values can be set by turning.
Pressing the encoder again confirms the selected values. If you then turn the encoder to the
right, the "T-Window" and "P-Window" sub-menus are opened (provided they have not been
deactivated in the Administrator Settings menu).
If you turn the encoder to the left, you exit the sub-menus again.
This welding mode can be deactivated in the Administrator Settings menu and then no longer
appears in the menu. If only one welding mode is enabled, the "Weld by …" screen is omitted.
T-Window function
A time window ranging from 0.01 to 9.99 seconds can be set using the "Select" encoder. If the
operating time is outside the window, the Nominal LED lights up on the control panel or the LED
AUX1 on the hand-held operating device, and there is a high signal at the Nominal output (DSUB
PIN 11). If the operating time is shown within this window, there is a low signal at the Nominal
output (DSUB PIN 11) on the interface (also refer to interface description).
The selected window is set by pressing the encoder. The minimum and maximum values of the
window can be set by turning the encoder. Pressing the encoder again confirms the selected
values. In the "off" position this function is deactivated.
This function can be deactivated in the Administrator Settings menu and then no longer appears
in the menu
P-Window function
The "Select" encoder can be used here to set a power window. If the power output is outside the
window, the Nominal LED lights up on the control panel or the AUX1 LED on the hand-held opera-
ting device.
The selected window is set by pressing the encoder. The minimum and maximum values of the
window can be set by turning the encoder. Pressing the encoder again confirms the selected
values. In the "off" position this function is deactivated.
This function can be deactivated in the Administrator Settings menu and is then no longer
included in the menu.

- 15 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Weld by Time
If the generator is set to "Weld by Time", you can set a time of 0.01 seconds to 9.99 seconds.
The welding process begins when you press the test button or activate the generator via the
remote control input. If you release the test button or deactivate the remote control input before
the set time has elapsed, the error message "Cycle not completed" appears on the display and
the red Error LED lights up. If a time has been set in the Trigger menu and the Holding menu, the
operation starts with the Trigger delay time and only ends when the set Holding cooling time has
elapsed.
The welding mode and time are set by pressing the encoder. The required values can be set by
turning. Pressing the encoder again confirms the selected values. If you then turn it to the right,
the "T-Window" and "P-Window" sub-menus are opened (provided they have not been deacti-
vated in the Administrator Settings menu). If you turn the encoder to the left, you exit the sub-
menus again. Refer to "Weld by Remote" for a description of the window functions.
This welding mode can be deactivated in the Administrator Settings menu and then no longer
appears in the menu. If only one welding mode is enabled, the "Weld by …" screen is omitted.
Weld by Energy
In the "Weld by Energy" menu option, you can set a time of up to 9.99 seconds and an energy
(Watt second) that is to be output within the set time. If the set energy is not reached within the
defined time, the "Timeout" error appears on the display and the red Error LED lights up.
The welding operation starts when the test button is pressed or the generator is activated via the
remote control input. If you release the test button or deactivate the remote control input before
the set time has elapsed, the error message "Cycle not completed" appears on the display and
the red Error LED lights up. If a time was set in the Trigger menu and the Holding menu, the
operation starts with the Trigger delay time and only ends when the set Holding cooling time
has elapsed.
The welding mode, time and energy are set by pressing the encoder. The required values can be
set by turning. Pressing the encoder again confirms the selected values. If you then turn it to the
right, the "T-Window" and "P-Window" sub-menus are opened (provided they have not been
deactivated in the Administrator Settings menu). If you turn the encoder to the left, you exit the
sub-menus again. Refer to "Weld by Remote" for a description of the window functions.
This welding mode can be deactivated in the Administrator Settings menu and then no longer
appears in the menu. If only one welding mode is enabled, the "Weld by …" screen is omitted.

- 16 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Weld by Peak Power
In the "Weld by Peak Power" menu option, you can set a time of up to 9.99 seconds and a
power (Watt), that is to be emitted within the set time. If the set power output is not reached
within the defined time, the "Timeout" error appears on the display and the red Error LED
lights up.
The welding process begins when you press the test button or activate the generator via the
remote control input. If you release the test button or deactivate the remote control input before
the set time has elapsed, the error message "Cycle not completed" appears on the display and
the red Error LED lights up. If a time was set in the Trigger menu and the Holding menu, the
operation starts with the Trigger delay time and only ends when the set Holding cooling time
has elapsed.
The welding mode, time and output are set by pressing the encoder. The required values can be
set by turning. Pressing the encoder again confirms the selected values. If you then turn it to the
right, the "T-Window" and "P-Window" sub-menus are opened (provided they have not been
deactivated in the Administrator Settings menu). If you turn the encoder to the left, you exit the
sub-menus again. Refer to "Weld by Remote" for a description of the window functions.
This welding mode can be deactivated in the Administrator Settings menu and then no longer
appears in the menu. If only one welding mode is enabled, the "Weld by …" screen is omitted.
Weld by Pulse
If the generator is set to "Weld by Pulse", you can set a welding time and an "Off time" of 0.01
seconds to 9.99 seconds. The welding process begins when you press the test button or activate
the generator via the remote control input. If the welding time (Time) has elapsed, the generator
switches off for the set "Time Off" and then restarts. If you release the test button or deactivate
the remote control input, the welding process ends. If a time was set in the Trigger menu and
the Holding menu, the operation starts with the Trigger delay time and only ends when the set
Holding cooling time has elapsed.
The welding mode and time are set by pressing the encoder. The required values can be set by
turning. Pressing the encoder again confirms the selected values. If you then turn it to the right,
the "T-Window" and "P-Window" sub-menus are opened (provided they have not been deacti-
vated in the Administrator Settings menu). If you turn the encoder to the left, you exit the sub-
menus again. Refer to "Weld by Remote" for a description of the window functions.
This welding mode can be deactivated in the Administrator Settings menu and then no longer
appears in the menu. If only one welding mode is enabled, the "Weld by …" screen is omitted.
Holding
A cooling time of up to 9.99 seconds can be entered here before the actual welding operation
starts. Usage: during this time the press remains with the sonotrode on the workpiece until the
welding spot has cooled and the welded material has bonded. The cooling time is set by press-
ing the encoder. The cooling time can be set by turning the encoder. The selected cooling time
is confirmed by pressing the encoder again. In the "off" position this setting option is
deactivated.
This menu item can be deactivated in the Administrator Settings menu and then no longer
appears in the menu.

- 17 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
After Burst
In order to clean the sonotrode of residue from the previous welding process, you can again emit
energy to the sonotrode here. You can set a delay time before the After Burst of up to 9.99
seconds. In addition, you can set a maximum after burst time (sonotrode emits ultra-sound) of
up to 30 seconds. The times are set by pressing the encoder. The required time can be set by
turning. The selected time is confirmed by pressing the encoder again. In the "off" position this
setting option is deactivated.
This menu item can be deactivated in the Administrator Settings menu and then no longer
appears in the menu.
Setting the amplitude
The required output amplitude can be set within a range of 50 - 100% using the "Select"
encoder.
The required amplitude is set by pressing the encoder. The required amplitude can be set by
turning the encoder. Pressing the encoder again confirms the selected amplitude. You can also
adjust the amplitude while the generator is emitting ultrasound.
If "external voltage" is set in "Ctrl Assignment" in the Settings menu, the following text appears
in the centre of the display instead of the set amplitude: analog source.
The amplitude is then set to 50 to 100% of the nominal amplitude by applying an external
control voltage of 5V to 10V to the interface. Also refer to the description of the interface
assignment.
Counter
This counts how often the welding procedure has been started. This counter can be reset by
the operator. It is always set to zero after "Power on".
Contrast setting
The contrast of the LCD display can be adjusted by pressing and turning the encoder. Pressing
the encoder again applies the set value. Set the contrast so that the LCD display is easily read-
able.
Display backlight
The backlight of the LED display can be switched on and off by pressing and turning the
encoder. For optimum readability, we recommend turning the backlight on.
Settings
The Settings menu can be accessed by entering the password "6354":
This is also accessed by turning the encoder.
If the correct number is selected in each case, confirm by pressing the encoder.
If you enter the incorrect number, "you loose" appears on the display and the start screen
appears.

- 18 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Functions in the Settings menu
Control Assignment
Here you can set how the generator receives the reference value specifications.
Internal Front means with the front control panel (or with the external control console on the
Basic version), ext. Voltage means that the generator can be operated by means of an external
voltage (refer to Interface assignment).
I/O Polarities
Remote IN: By changing this value, you can change the polarity of the remote control input.
Factory default settings: A closed contact switches the generator on.
Symbol "H": When the contact is closed, the generator is switched on
Symbol "L": When the contact is open, the generator is switched on
Error out: By changing this value, you can switch the polarity of the internal error relay.
Factory default settings: Open contact is an error, which safeguards the
system in the event of a broken cable. By changing this value, you can switch
the polarity of the internal error relay.
Symbol "L": Closed contact is an error
Symbol "H": Open contact is an error
If the polarity is changed in such a way that a closed contact generates an error message, it should be noted
that when the generator is disconnected from the mains or there is a general failure, the contact is open.
Recommendation: Leave the factory default setting, as this will safely detect the
above-mentioned errors.
RF detect: By changing this value, you can change the HF-DA status signal
(generator is emitting HF voltage).
Factory default settings: Closed contact when generator emits HF voltage.
Symbol "H": Closed contact – generator emits HF voltage
Symbol "L": Open contact – generator emits HF voltage
Please note that when the generator is not connected to the mains or there is a general failure, this contact is open.
Recommendation: Do not change the factory settings.

- 19 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
TEST BUTTON
Here are three configuration options for the test button on the front plate:
PUSH Generator is switched on when the test button is pushed and switched off
when the button is released.
TOGGLE Generator is switched on when the test button is pushed.
Pressing the button again switches the generator off.
DISABLED Test button has no function
Operating hours indicator
T Power: This displays the activation time of the generator.
T Sonic: This displays the time during which the generator emitted HF.
Display format: Days, hours, minutes.
Info
In the Info menu option, you can see the software version number of the front plate and the
software version number of the generator. If you press the encoder again, you can see the serial
number and the date of manufacture of the generator.
Sensorics
By connecting a voltage between 5V and 10V to PIN 3 of the interface, the output of the generator
can be set to between 50% and 100% of the nominal amplitude or power. The voltage connected
can be read here as a percentage. 10 V corresponds to 100%, so in this example we can see that
the voltage is 6.5 V. Reference point is "GND".
EXIT – exit the Settings menu and save any changed parameters

- 20 -
SONIC DIGITAL HS - Ultrasonic Generators
www.weber-ultrasonics.com
Functions in the Administrator Settings menu
This Administrator Settings menu is only for qualified service personnel. Faulty settings in this
menu can irreparably damage the generator or the connected transducer system. The password
is only issued to registered administrators and is available from our service contact number.
Adjusting the starting frequency
The resonant frequency of the connected system must be known. Generally a starting frequency
which is approximately 500 Hz above the resonant frequency of the connected system produces
good results. The capture range of the generator is 2 kHz.
In special cases, however, it can also be necessary to set the starting frequency closer to the
resonance of the transducer system. Under no circumstances should the starting frequency be
set so that it is below the resonant frequency of the connected transducer system! The starting
frequency can be changed by pressing and then turning the encoder. The area which is covered
depends on the device type. The capture range of the generator can be limited with the stop
frequency.
Configure
Various menu items from the operator menu can be activated and deactivated with the Configure
menu item.
If "YES" is selected in "Configure Trigger", the Trigger function is active in the operator menu
(delay before welding). If "NO" is selected, this function is inactive and does not appear.
If "YES" is selected in "Configure Holding", the Holding function is active in the Operator menu
(wait time after welding). If "NO" is selected, this function is inactive and does not appear.
If "YES" is selected in "Configure After Burst", the After Burst function is active in the Operator
menu (energy for cleaning the sonotrode). If "NO" is selected, this function is inactive and does
not appear.
With the "Configure RF Output" function, you can switch the output on the interface between the
RF Output Signal ("YES") and the Control Output Signal ("NO"). When the RF output is set, the
interface output is only active when ultrasound is active. When the control output is set, the
interface output is only active during the Trigger Welding and Holding phases.
Trigger Delay Welding Holding After Burst Delay After Burst
RF
Output
Trigger Delay Welding Holding After Burst Delay After Burst
Control
Output
Other manuals for SONIC DIGITAL HS MD
1
Table of contents
Other Weber Ultrasonics Portable Generator manuals
Popular Portable Generator manuals by other brands

Chicago Pneumatic
Chicago Pneumatic CPG 90-120 JD instruction manual

Fischer Panda
Fischer Panda Panda 15000i PVMV-N manual

Ward-Beck Systems
Ward-Beck Systems POD33 manual

Golden engineering
Golden engineering XRS4 Operator's manual

Lion Energy
Lion Energy SAFARI LT user manual

Toyama
Toyama ATS owner's manual

Champion Global Power Equipment
Champion Global Power Equipment 100592 quick start guide

Casals
Casals GEN7000 Original instructions

MKS
MKS SG 524 manual

NB
NB GEN4065 instruction manual

Champion
Champion 75551i Owner's manual & operating instructions

Rheem Manufacturing
Rheem Manufacturing GEN12AD Operator's manual