Welker HR User manual

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL
WELKER
HEATED INSTRUMENT REGULATOR
MODEL
HR
DRAWING NUMBER
ADHRO02
PATENT
U.S. PATENT 7,471,882
MANUAL NUMBER
IOM-158
REVISION
Rev. 1, 03/08/2023

2
IOM-158 | MODEL: HR | REV: 1 13839 West Bellfort Street, Sugar Land, TX 77498 welker.com Service Department: 281.491.2331
TABLE OF CONTENTS
SAFETY 3
1. PRODUCT INFORMATION 4
1.1 Introduction 4
1.2 Product Description 4
1.3 Important Certification Information 5
1.4 Specific Conditions of Use 5
1.5 Specifications 6
1.6 Equipment Diagram 7
2. INSTALLATION & OPERATION 8
2.1 Before You Begin 8
2.2 Installing the HR 8
3. MAINTENANCE 11
3.1 Before You Begin 11
3.2 Maintenance 12
3.3 Heater Failure 19
APPENDIX 20
A: Referenced or Attached Documents 20
Copyright © 1997–2023 Welker, Inc. All rights reserved. Welker®, W Welker®, W logo, WelkerScope®, Welker Jet®, and OdorEyes® are registered trademarks of Welker, Inc.

3
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
SAFETY
IMPORTANT SAFETY INFORMATION
READ ALL INSTRUCTIONS
Notes emphasize information and/or provide additional information to assist the user.
Caution messages appear before procedures that could result in damage to equipment if not observed.
Warning messages appear before procedures that could result in personal injury if not observed.
This manual is intended to be used as a basic installation and operation guide for the Welker Heated Instrument Regulator, HR. For comprehensive instructions, please refer
to the IOM Manuals for each individual component. A list of relevant component IOM Manuals is provided in Appendix A of this manual.
The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and maintenance of the Welker
equipment described in this manual. Correct installation and operation, however, are the responsibility of the end user. Welker reserves the right to make changes to this
manual and all products in order to improve performance and reliability.
BEFORE YOU BEGIN
Read these instructions completely and carefully.
IMPORTANT
– Save these instructions for local inspector’s use.
IMPORTANT
– Observe all governing codes and ordinances.
Note to Installer – Leave these instructions with the end user.
Note to End User – Keep these instructions for future reference.
Installation of this Heated Instrument Regulator is of a mechanical and electrical nature.
Proper installation is the responsibility of the installer. Product failure due to improper installation is not covered under the warranty.
If you received a damaged Heated Instrument Regulator, please contact a Welker representative immediately.
Phone: 281.491.2331
Address: 13839 West Bellfort Street
Sugar Land, TX 77498

4
IOM-158 | MODEL: HR | REV: 1 13839 West Bellfort Street, Sugar Land, TX 77498 welker.com Service Department: 281.491.2331
SECTION 1: PRODUCT INFORMATION
1.1 Introduction
We appreciate your business and your choice of Welker products. The installation, operation, and maintenance liability for this
equipment becomes that of the purchaser at the time of receipt. Reading the applicable Installation, Operation, and Maintenance
(IOM) Manuals prior to installation and operation of this equipment is required for a full understanding of its application and
performance prior to use.*
If you have any questions, please call Welker at 1-281-491-2331.
*The following procedures have been written for use with standard Welker parts and equipment. Assemblies that have been modified may have additional requirements
and specifications that are not listed in this manual.
1.2 Product Description
The Welker HR Heated Instrument Regulator provides a conditioned gas sample stream at the required temperature and pressure
for the instrumentation.
The HR heats product prior to decreasing the pressure to help compensate for the natural temperature drop that occurs during
regulation. In order to set the desired output pressure, an adjusting screw on the device is turned clockwise. The design of the HR
allows for the sample stream to flow through at a consistent rate of pressure and temperature.
The enclosure for the electrical parts of the equipment consists of one (1) of two (2) certified flameproof enclosures with either
two (2) or three (3) ¾" FNPT entries and a threaded cover. The certified enclosure houses a terminal block and a heater controller
for regulation of the heater cartridge.
Welker may custom design the HR to suit the particular application and specifications of each customer.

5
IOM-158 | MODEL: HR | REV: 1 13839 West Bellfort Street, Sugar Land, TX 77498 welker.com Service Department: 281.491.2331
1.3 Important Certification Information
1. Ensure that the field technician installing the HR is trained in the protection practices required to maintain the area
classification for electrical equipment in ATEX and IECEx hazardous locations.
2. Before connecting power to the HR, the technician shall ensure that cabling components supplied by the end-user are
certified for explosion protection flameproof enclosure "d".
3. The cable entry points may reach 179.24 °F (81.8 °C). Appropriate selection of cabling components shall be made by the
end-user to ensure that the installation of the HR maintains its certifications.
4. The phase conductors should have a minimum cross-sectional area no less than 0.8325 mm² (18 AWG). The protective
earth conductors should have the same or larger cross-sectional area as the phase conductors. All external equipotential
bonding connections shall provide an effective connection of a conductor with a minimum cross-sectional area of 4 mm².
5. Do not open the certified enclosure until the surface temperature and any stored electrical energy have decayed to the
point that the certified enclosure can no longer be a potential ignition source.
6. Ensure that the HR is isolated from all incoming and outgoing connections (including neutral conductors) prior to
opening the certified enclosure.
7. When removing the HR to perform maintenance, ensure that the exposed conductors are correctly terminated in an
appropriate enclosure, are insulated and isolated from all power sources, or are properly grounded and isolated from all
power sources.
8. When reassembling the certified enclosure, all joints should be thoroughly cleaned and may be lightly lubricated with
suitable grease, such as Crouse-Hinds STL or HTL Thread Lubricant, to prevent corrosion and assist in weatherproofing.
Use only non-metallic scrapers and non-corrosive cleaning fluids to clean flanges. Keep blind bolt holes clear of grease.
1.4 Specific Conditions of Use
1. Some external parts of the equipment are non-conducting and may generate an ignition-capable level of electrostatic
charge under certain extreme conditions. The end-user should ensure that the equipment is not installed in a location
where it may be subjected to external conditions (such as high pressure steam), which might cause a build-up of
electrostatic charges on non-conducting surfaces. Additionally, cleaning of the equipment should be done with a damp
cloth.
2. The equipment shall be operated within an inlet process temperature range of -76 °F to 140 °F (-60 °C to +60 °C).

6
IOM-158 | MODEL: HR | REV: 1 13839 West Bellfort Street, Sugar Land, TX 77498 welker.com Service Department: 281.491.2331
1.5 Specifications
The specifications listed in this section are generalized for this equipment. Welker can modify the equipment according to your
company's needs. Please note that the specifications may vary depending on the customization of your equipment.
Table 1: HR Specifications
Products Natural Gas and Natural Gas Liquids Compatible With the Materials of
Construction
Materials of Construction
Regulator Base & Body: 6 Moly Stainless Steel, 17-4 PH Stainless Steel,
304/304L Stainless Steel, 316/316L Stainless Steel, 317/317L Stainless
Steel, 321/321H Stainless Steel, 347/347H Stainless Steel, Duplex
Stainless Steel, HASTELLOY® C276 Alloy, Inconel Alloy 625, Monel®
Alloy 400, and Super Duplex Stainless Steel
Seals: Varies Based on Customer Specifications and the MAOP and
MAOT of the Unit
Maximum Allowable Operating Pressure Bolted: 6000 psig @ -76 °F to 140 °F (413 barg @ -60 °C to +60 °C)
Threaded: 3600 psig @ -76 °F to 140 °F (248 barg @ -60 °C to +60 °C)
Process Temperature Range -76 °F to 140 °F (-60 °C to +60 °C)
Connections Product Inlet: ¹⁄" FNPT (Standard) or ¼" FNPT
Product Outlet: ¼" FNPT
Electrical Connections ¾" MNPT
Output Range
0–25 psig (0–1.7 barg)
0–50 psig (0–3.4 barg)
20–100 psig (1.3–6.8 barg)
75–200 psig (5.1–13.7 barg)
Operation Diaphragm- or Piston-Operated Regulator
Features
Explosion-Proof Certified Enclosure
Heating Element With Adjustable Temperature Range
(See Table 2 for Options)
Product Certifications
Patent U.S. Patent 7,471,882
Options
Bracket for Mounting
Regulator Subassembly
Relief Valve With Pressure Gauge
Table 2: Heating Element Options
Electrical Connection Power Temperature Range Current Resistance
AC 110/220 V 150 W 68 °F to 210 °F (20 °C to 98 °C) 1.25 A 196
200 W 180 °F to 380 °F (82 °C to 193 °C) 1.67 A 75
AC 220/240 V 100 W 68 °F to 210 °F (20 °C to 98 °C) 0.417 A 576
200 W 180 °F to 380 °F (82 °C to 193 °C) 0.833 A 288

7
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
1.4 Equipment Diagram
Figure 1: Welker Heated Instrument Regulator Diagram

8
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
SECTION 2: INSTALLATION & OPERATION
2.1 Before You Begin
After unpacking the unit, check the equipment for compliance and any damage that may have occurred during shipment.
Immediately contact a Welker representative if you received damaged equipment.
When sealing fittings with PTFE tape, refer to the proper sealing instructions for the brand used.
2.2 Installing the HR
Ensure that the field technician installing the HR is trained in performing installations as well as the protection practices required to
maintain the area classification of the electrical equipment in ATEX and IECEx hazardous locations.
1. As necessary, install a pressure gauge to the pressure gauge port (Figure 2).
Figure 2: Port Diagram
Welker can install a relief valve and pressure gauge if requested at the time of order.
2. As necessary, use a safe auxiliary gas supply to set the relief valve to the proper pressure. Refer to the Installation,
Operation, and Maintenance (IOM) Manual for the relief valve for instructions on setting the relief.
If a Welker relief valve is used, Welker can set the relief valve prior to shipment if requested at the time of order.
3. As necessary, install the set relief valve to the relief valve port (Figure 2).

9
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
4. With the customer-supplied electrical power turned OFF, connect the appropriate leads to terminals 2 and 4 in the
provided certified enclosure (Figure 3).
See Section 1.3, Important Certification Information, for information on selecting appropriate cabling components for the HR to
ensure it maintains its certifications.
Figure 3: Wiring Diagram
5. Conductors used to supply power to the certified enclosure are required to use adequate barrier glands or conduit seals
and an appropriate sealing compound to maintain flameproof "d" rating.
For systems used in hazardous locations, sealing compound is required to seal all fittings to restrict the passage of gases, vapors, or
flames.
When sealing fittings with sealing compound, refer to the proper sealing instructions for the brand used.
6. Set the heater controller dial to the desired temperature.
Note that the temperature scale on the heater controller is for reference only.
7. Secure the cover to the certified enclosure and fasten the set screw.
8. Using customer-supplied ¼" tubing, connect from the product supply outlet to the product inlet of the HR (Figure 1).
9. Install customer-supplied tubing or other fittings to the product outlet port of the HR (Figure 1 or Figure 2).

10
IOM-158 | MODEL: HR | REV: 1 13839 West Bellfort Street, Sugar Land, TX 77498 welker.com Service Department: 281.491.2331
Welker recommends that this tubing be insulated and heat-traced so that the sample maintains its temperature increase prior to
reaching the analyzer.
10. Loosen the jam nut on the adjusting screw (Figure 1).
11. Screw the adjusting screw clockwise to adjust the outlet pressure. Tighten the jam nut to secure the adjusting screw at
the desired outlet pressure (Figure 1).
12. Turn ON the electrical power to begin operation. Allow at least thirty (30) minutes for the HR to warm up.
Welker recommends that the unit be enclosed or insulated to maximize the heating effect of the HR.

11
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
SECTION 3: MAINTENANCE
3.1 Before You Begin
1. Welker recommends that the unit have standard yearly maintenance. Based on the operating conditions and/or site
requirements, adjustments to the maintenance schedule may be necessary.
2. Prior to maintenance or disassembly of the unit, it is advisable to have a repair kit available for repairs of the system in
case of unexpected wear or faulty seals.
New seals supplied in spare parts kits should be lightly lubricated before being installed to ease the installation of the seals and
reduce the risk of damage when positioning them on parts. Wipe excess lubricant from the seals, as it may adversely affect analytical
instrument results.
For sample-exposed seals, Welker recommends non-hydrocarbon-based lubricants, such as Krytox®.
For non-sample-exposed seals, Welker recommends either non-hydrocarbon-based lubricants or silicone-based lubricants, such as
Molykote® 111.
After the seals are installed, the outer diameter of shafts and inner diameter of cylinders may be lubricated to allow smooth
transition of parts.
3. All maintenance and cleaning of the unit should be performed on a smooth, clean surface.
4. Welker recommends having the following tools available for maintenance. Please note that the exact tools required may
vary by model.
a. 6" and 10" Channel Lock Pliers
b. 10" Adjustable Wrench
c. Gloves
d. Hex Key Set
e. Non-autoranging Multimeter
f. Seal Pick
g. Tubing Crimp and Accessories

12
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
3.2 Maintenance
When removing the HR to perform maintenance, ensure that the exposed conductors are correctly terminated in an appropriate
enclosure and are properly grounded and isolated from all energy sources.
1. Turn OFF all electrical power to the unit and isolate it from all applicable energy sources.
The HR will be HOT after use. Allow an appropriate amount of time for the regulator to cool down prior to performing maintenance.
2. Disconnect the electrical wiring and ensure that it cannot become live while maintenance is performed (Figure 3).
3. Turn OFF the product supply to the unit to depressurize the HR.
4. Disconnect the customer-supplied tubing, fittings, or instrument attached to the HR.
5. Remove the tie bolts from the bottom of the base and separate the regulator subassembly from the base (Figure 4).
Base Maintenance
Figure 4: Base Maintenance Diagram
6. Replace the O-ring in the top of the base (Figure 4).
7. Remove the chamber insert and inspect it for any dents that would impede the flow of gas (Figure 4). Replace as
necessary.
8. Return the chamber insert to the base, and then set the base aside.

13
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
Regulator Upper Housing Maintenance
Maintenance should not be performed on the regulator until the regulator has been isolated from all pressure.
9. Loosen the jam nut on the adjusting screw (Figure 1).
10. In a counterclockwise direction, back off the adjusting screw to relieve tension on the spring (Figure 1).
11. As necessary, perform maintenance on the relief valve. Refer to the Installation, Operation, and Maintenance (IOM) Manual
for the relief valve for maintenance instructions.
12. Separate the spring housing from the regulator body (Figure 1).
To perform maintenance on a cartridge-style regulator, continue to step 13.
To perform maintenance on a diaphragm-style regulator, proceed to step 20.
To perform maintenance on a piston-style regulator, proceed to step 28.

14
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
Maintenance: Cartridge-Style Regulator
Figure 5: Cartridge-Style Regulator
13. Remove the diaphragm assembly, spring, and spring guide (Figure 5).
14. Inspect the diaphragm for wear. If the diaphragm shows signs of wear, the diaphragm assembly must be replaced.
The diaphragm assembly must be replaced at least every five (5) years.
15. Place the diaphragm assembly on the top of the regulator body with the bottom spring guide facing up (Figure 5).
16. Place the spring on the diaphragm assembly (Figure 5). Ensure that the spring is sitting on the bottom spring guide.
17. Return the spring guide to the top of the spring (Figure 5).
18. Install the spring housing to the regulator body (Figure 5).
When reassembling the upper housing, HAND TIGHTEN ONLY.
19. Proceed to step 38 for instructions on maintaining the regulator body.

15
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
Maintenance: Diaphragm-Style Regulator
Figure 6: Diaphragm-Style Regulator
20. Remove the spring guides and spring (Figure 6).
21. Remove the diaphragm assembly (Figure 6).
22. Inspect the diaphragm for wear. Replace as necessary.
The diaphragm assembly must be replaced at least every five (5) years.
23. Place the diaphragm assembly on top of the regulator body with the bottom spring guide facing up (Figure 6).
24. Place the spring on the diaphragm assembly (Figure 6). Ensure that the spring is sitting on the bottom spring guide.
25. Return the spring guide to the top of the spring (Figure 6).
26. Install the spring housing to the regulator body (Figure 6).
When reassembling the upper housing, HAND TIGHTEN ONLY.
27. Proceed to step 38 for instructions on maintaining the regulator body.

16
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
Maintenance: Piston-Style Regulator
Figure 7: Piston-Style Regulator
28. Remove the piston, adapter, spring, and spring guide (Figure 7).
29. Replace the O-ring in the regulator body and in the adapter (Figure 7).
30. As necessary, remove the retaining ring from the spring housing and inspect the retaining ring for scratches or damage
(Figure 7). Replace as necessary.
31. Inspect the sealing surfaces of the piston for scratches (Figure 7). Replace as necessary.
32. Install the piston to the adapter, and then install this subassembly to the regulator body with the piston facing up.
33. Place the spring on the piston assembly. Ensure that the spring is sitting on the piston.
34. Return the spring guide to the top of the spring.
35. As necessary, return the retaining ring to the spring housing.
36. Following a cross-bolting sequence, install the cap screws to the spring housing, and then tighten to secure the spring
housing to the regulator body (Figure 8).
Figure 8: Cross-Bolting Sequence
37. Continue to step 38 for instructions on maintaining the regulator body.

17
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
Regulator Body Maintenance
38. Ensure that the regulator subassembly is separated from the base.
To perform maintenance on a standard regulator subassembly, continue to step 39.
To perform maintenance on a cartridge-style regulator subassembly, proceed to step 50.
Regulator Subassembly: Standard
Figure 9: Standard Regulator Subassembly
39. Use a socket wrench to loosen the flow ring (Figure 9).
40. Remove the flow ring, spring, and poppet (Figure 9).
41. Examine the seating face of the poppet for scratches or damage. Replace as necessary.
Debris or scratches on the poppet will prevent positive shutoff of the regulator.
42. Use a small, pointed instrument to carefully remove the seat from the regulator body (Figure 9).
43. Inspect the seat for debris or scratches. Replace as necessary.
Debris or scratches on the seat will prevent positive shutoff of the regulator.
44. Replace the O-ring on the flow ring (Figure 9).
45. Install the seat to the regulator body.
46. Guide the poppet into the seat.
47. Return the spring to the poppet.
48. Return the flow ring to the regulator body and tighten firmly using a socket wrench.
49. Proceed to step 54 for reassembly instructions.

18
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
Regulator Subassembly: Cartridge
Figure 10: Cartridge Regulator Subassembly
50. Remove the cartridge assembly from the regulator body (Figure 10).
51. Replace the O-ring in the regulator body (Figure 10).
52. Install a replacement cartridge assembly to the regulator body.
53. Install the spring housing to the regulator body and hand tighten only.
Reassembly
54. Return the regulator subassembly to the top of the base.
55. Following a cross-bolting sequence, insert the tie bolts through the base to the regulator body, and then tighten firmly
(Figure 9).
56. The HR is now ready for installation. See Section 2.2, Installing the HR, for instructions on correctly installing the unit.

19
IOM-158 | MODEL: HR | REV: 1
13839 West Bellfort Street, Sugar Land, TX 77498 welker.com
Service Department: 281.491.2331
3.3 Heater Failure
Neither the controller nor the heating element is prone to fail; however, failure is possible in instances of misuse and extreme
malfunction of the electrical system.
If the thermostat controller or heating element do not produce heat, replacement is required. The heated regulator must be
returned to Welker. This is necessary to ensure the unit maintains its certifications.
Returning Unit to Welker
1. Turn OFF all electrical power to the unit and isolate it from all applicable energy sources.
DO NOT open the certified enclosure until it can no longer be a potential ignition source.
Ensure that the HR is isolated from all incoming and outgoing connections (including neutral conductors) prior to opening the
certified enclosure.
2. Unfasten the set screw, and then remove the cover from the certified enclosure.
3. Disconnect the two (2) power supply leads from the terminal block.
4. Return the cover to the certified enclosure and refasten the set screw.
5. As necessary, disconnect the customer-supplied tubing, fittings, or instrument attached to the unit.
6. Send unit back to Welker.

20
IOM-158 | MODEL: HR | REV: 1 13839 West Bellfort Street, Sugar Land, TX 77498 welker.com Service Department: 281.491.2331
APPENDIX A: REFERENCED OR ATTACHED DOCUMENTS
Welker Installation, Operation, and Maintenance (IOM) Manuals suggested for use with this unit:
IOM-033: Welker RV-1, RV-2, RV-2CP, and RV-3 Relief Valves
IOM-234: Welker RV-110A and RV-110V Relief Valves
Other Installation, Operation, and Maintenance (IOM) Manuals suggested for use with this unit:
Adalet Explosionproof Instrument Housing XIH Series Single Ended Housing (Welker IOM-V236)
International Metal Engineering Pte. Ltd. Model 8080 Explosion Proof Instrument Enclosure (Welker IOM-V237)
Welker drawings and schematics suggested for use with this unit:
Assembly Drawing: ADHRO02
Table of contents
Other Welker Controllers manuals
Popular Controllers manuals by other brands

Murphy
Murphy Autostart 710 AS710S owner's manual

Timewave
Timewave PK-232 MBX operating manual

Novar
Novar NOVAR 300 Operation & maintenance manual

National Instruments
National Instruments IC-3121 user manual

4O3A
4O3A Station Genius v2 user manual

Johnson Controls
Johnson Controls Joventa DMS1.1 installation instructions