BrainPort Vision Pro Instructions for Use 9
the device causes discomfort, discontinue use.
•Oral Health. Individuals with high, narrow palatal vaults should
discontinue use of the device if use causes discomfort
•Oral Health. Individuals with maxillary or mandibular tori that
interfere with the IOD placement such that full contact with the
tongue is prevented should seek additional training to gain the
most benefit of the device.
•Dental Appliances (orthodontic appliances, removable partial or
full dentures, lingual amalgam alloy restoration, metal crowns,
etc.) Electrical stimulation of the tongue when metal
appliances/surfaces are present may change results and/or
cause unintended stimulation. If the stimulation causes
discomfort, remove the IOD from your mouth and discontinue
use of the device. If you notice a change in your dental device
or appliance (warmth or looseness), discontinue the use of the
device and contact your dentist.
•Dental Implants. The BrainPort device has not been thoroughly
evaluated in the presence of dental implants. The safety of
dental implants in BrainPort users is unknown. The use of this
device potentially may cause heating of dental implants; chronic
use of this device potentially may result in loosening and failure
of dental implants.
•Condition of device. Before EACH use, tactilely inspect the
device for damage, for example, anything rough or loose on the
IOD, disconnected cables, worn cables, cracked or broken
glasses, cracked or broken headset, etc. If you find these or
similar issues, contact Customer Support and DO NOT use the
device. Using a damaged device could expose sharp edges
and/or the damage could prevent normal operation, thereby
increasing the risk of use.
•Proper environment for use. The BrainPort Vision Pro is
intended for use as a supplemental assistive device.
oDo not use it in environments that could put you in danger.
oDo not operate the device in hot or cold conditions (below
0°C/32°F or above 40°C/104°F). Maintain conditions
between 5% and 95% relative humidity. The device is
intended for operation under normal atmospheric
pressures (700 hPa to 1060 hPa)