WIKA GPD-1000 User manual

EN
Operating instructions
Gas presence detector, model GPD-1000
Gas presence detector, model GPD-1000

2
14314459.01 12/2018 EN
WIKA operating instructions model GPD-1000
EN
Operating instructions model GPD-1000 Page 4 - 7
© 12/2018 WIKA Alexander Wiegand SE & Co. KG
All rights reserved.
WIKA® is a registered trademark in various countries.
Prior to starting any work, read the operating instructions!
Keep for later use!

EN
WIKA operating instructions model GPD-1000
14314459.01 12/2018 EN
3
Declarations of conformity can be found online at www.wika.com.
Contents
Contents
1. General information 4
2. Technical data 5
3. Instructions 6
4. Features 6
5. Battery installation and replacement 7
6. Sensor 7
7. Sensor replacement 7
8. Maintenance 7

EN
14314459.01 12/2018 EN
4
WIKA operating instructions model GPD-1000
1. General information
■
The detection instrument described in the operating instructions has
been designed and manufactured using state-of-the-art technology.
■
All components are subject to stringent quality and environmental
criteria during production. Our management systems are certied to
ISO 9001 and ISO 14001.
■
These operating instructions contain important information on
handling the instrument. Working safely requires that all safety
instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and
general safety regulations for the instrument’s range of use.
■
The operating instructions are part of the product and must be kept
in the immediate vicinity of the instrument and readily accessible to
skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the
operating instructions prior to beginning any work.
■
The manufacturer‘s liability is void in the case of any damage caused
by using the product contrary to its intended use, non-compliance
with these operating instructions, assignment of insuciently
qualied skilled personnel or unauthorised modications to the
instrument.
■
The general terms and conditions contained in the sales
documentation shall apply.
■
Subject to technical modications.
■
Gas presence detector model GPD-1000 should not be used on
systems that contain ammable refrigerants such as Propane,
Isobutane, etc.
■
Factory calibrations/DKD/DAkkS calibrations are carried out in
accordance with international standards.
■
Further information:
- Internet address: www.wika.com/sf6
- Relevant data sheet: SP 62.11
- Application consultant Tel.: +49 9372 132-0
1. General information

EN
WIKA operating instructions model GPD-1000
14314459.01 12/2018 EN
5
2. Technical data
Technical data
Area of application Leakage detection
Detection limit < 3 ppmv
Battery lifetime 35 hours
Sensitivity < 3 g/year (0.1 oz/year)
Sensor lifetime 30 hours
Operating temperature 0 ... 50 °C
Warm up time < 2 seconds
Response time Instantaneous
Reset time Instantaneous
Probe length 300 mm
Replacement sensor Part-No. 14291469
2. Technical data

EN
14314459.01 12/2018 EN
6
WIKA operating instructions model GPD-1000
3. Instructions
1. Press the button to turn the unit on.
2. The LED will ash orange for a short moment to indicate auto-reset,
afterwards the unit will begin beeping and ashing green.
3. To reset the unit to the existing level of ambient SF6 or alternative
gases, press and release the button. The LED will ash orange briey
to indicate the reset. All levels of SF6 or its alternative gas less than
the reset level will be ignored.
4. Unit turns on in the low sensitivity level, which is adequate for most
leaks. Press the button twice (double-click like a computer mouse).
The LED will fade from red to green and a sweeping sound eect of
low to high pitch will be heard. The sensitivity is now set for 3g/year
(0.1oz/yr). This mode should only be used for looking for leaks of less
than 14g/year (0.5 oz/yr).
5. Double click again to change back to low sensitivity.
6. Move the probe towards a suspected leak at a rate of less than 2
inches (50 mm) per second, no more than ¼ inch (5 mm) away from
the suspected source.
7. If a leak exists, the sound will increase in rate and pitch and the LED
will start ashing rapidly.
8. To turn the GPD-1000 o, press and hold the button for 3 seconds.
4. Features
If the sensor is damaged, a sweeping alarm and a red-green ash will
indicate the problem.
A ashing LED during normal operation also reects the battery level:
■
Green: batteries are fresh
■
Orange: batteries should be changed
If the batteries are nearly empty, the LED will turn solid red and a
two-tone alarm will be noted for 5 seconds, before GPD-1000 will power
itself o.
If the unit is left on unattended, it will automatically turn o after about 5
minutes to conserve batteries.
3. Instructions / 4. Features

EN
WIKA operating instructions model GPD-1000
14314459.01 12/2018 EN
7
5. Battery installation and replacement
1. Turn the power o.
2. Remove the battery cover from the back of the instrument by pressing
down while simultaneously sliding the cover back. Be careful not to
lose the extra sensor stored in the battery compartment.
3. Insert 2 size ‘AA’ alkaline batteries.
4. Slide battery cover back into place.
6. Sensor
A GPD-1000 sensor will last between 25 and 30 hours of continuous use.
A failed or failing sensor will be indicated in one of the following ways:
1. Unstable or erratic operation of the unit and many false alarms, even
in pure air.
2. A continuous “siren” sound.
7. Sensor replacement
In either case, replacement of the sensor is necessary:
1. Turn the power o.
2. Locate the replacement sensor in the battery compartment.
3. Unscrew (turn counter clockwise) the old sensor from the end of the
probe.
4. Screw (turn clockwise) the new sensor into place.
8. Maintenance
The GPD-1000 should provide years of service with little maintenance
aside from changing batteries and sensors. The case may be cleaned
with a shop towel moistened with water and a mild detergent. Do not use
solvent of any kind.
5. Battery ... 8. Maintenance

WIKA subsidiaries worldwide can be found online at www.wika.com.
14314459.01 12/2018 EN
8
WIKA operating instructions model GPD-1000
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Strasse 30
63911 Klingenberg • Germany
Tel. +49 9372 132-0
Fax +49 9372 132-406
www.wika.de
Other manuals for GPD-1000
1
Table of contents
Other WIKA Security Sensor manuals