Wolfgang Warmbier METRISO 2000 User manual

METRISOâ2000 – ESD Test Instrument
High-Resistance Tester for the Measurement of Resistance to Ground and Surface Resistance 3-349-087-53
4/8.01
Operating Instructions

2WOLFGANG WARMBIER
METRISOâ2000 Measuring and Test Instrument METRISOâ2000 Control Panel and Display
LIMIT U>25V
Netz
Mains
START
Keys for the selection of basic
LCD
Potentialdifference³150Vbetween
Mainsorinterferencevoltageatmeasurementinput
Voltageatmeasurementinput>25V
Limitvalueviolation
Left switch contact:
Displays basic functions,
Toggle Switch
Middle switch contact:
Starts measurement
functions and sub-functions,
Right switch contact:
Displays basic functions,
scrolling from right to left scrolling from left to right
Battery level display
fingercontactandmeasurementinput
Indicator lamps
(menu-driven operation)
The desired measuring
functions can be dis-
played with the help of
the and keys.
The scroll bar indicates which
menu is currently open.
Infrared interface Temperature and
humidity measuring adapter
Z541A
METRISO
â
2000
Measurement jack
GUARD
+ pole
Carryingstrap
Closure
Operating and
display unit
Contact surface
Shield jack
Socket for external charger
Part No. 7100.2000.MK.Z501D
Measurement
jack, COM

WOLFGANG WARMBIER 3
Please proceed as follows if the LCD display is not legible, i.e. too bright
or too dark, after switching on the test instrument:
1 Simultaneously press the two right-hand keys to delete the memory
which might be faulty.
2 Wait a few seconds to allow the display to be updated.
Readjust the contrast if necessary, see page 7.
Contents.............................................................................................. Page
1 Applications .................................................................................... 4
2 Safety Features and Precautions .................................................... 4
3 Initial Start-Up ................................................................................ 5
3.1 Battery Test ....................................................................................................5
3.2 Installing and Replacing Batteries .....................................................................5
3.3 Selecting a Menu and Configuring Basic Settings .........................................5
4 General Operation ........................................................................... 8
4.1 Automatic Settings, Monitoring and Shutdown ..................................................8
4.2 Measurement Value Display .............................................................................8
4.3 Online Help .....................................................................................................8
5 ESD Measurement .......................................................................... 9
5.1 Condensed Instructions ...................................................................................9
5.1.1 Setting Limit Values – LIMIT Function ...............................................................9
5.2 Measuring Resistance to Ground REor RGaccording to IEC 61340-4-1 /
DIN IEC 61340-4-1 ...................................................................................... 10
5.3 Measuring Resistance through a Person to Ground RE System or RG System ..........10
5.4 Measuring Surface Resistance ...................................................................... 11
5.4.1 Point to Point Measurement RSwith 2 Measuring Probes according to
IEC 61340-4-1 / DIN IEC 61340-4-1 ............................................................. 11
5.4.2 Measuring Surface Resistance RSof Packaging Materials
with a Ring Probe in compliance with IEC 61340-5-1 / DIN EN 61340-5-1 ...... 11
5.5 Measuring Volume Resistance RVacc. to IEC 61340-4-1 / DIN IEC 61340-4-1 12
5.6 Measurement with the GUARD Terminal ......................................................... 12
6 Database Functions ...................................................................... 13
6.1 Creating a Data Record – Data Function ......................................................... 13
6.2 Saving Measurement Values – STORE Function .............................................. 14
6.3 Querying Data Records – View Function .........................................................14
6.3.1 Deleting a Data Record from within a Memory Address – View Function ..........15
6.3.2 Deleting a Memory Address – Data Function ..................................................15
6.3.3 Delete All Memory Addresses – Data Function ............................................... 16
7 Measuring Insulation Resistance ................................................. 17
7.1 Tests with Variable Test Voltage ....................................................................17
7.2 Tests with Rising Test Voltage ....................................................................... 17
8 Measuring Alternating Voltage ..................................................... 18
9 Measuring Temperature and Humidity with Adapter Z541A ......... 18
10 Characteristic Values .................................................................... 19
11 List of Abbreviations and their Meanings ..................................... 21
12 Maintenance ................................................................................. 21
12.1 Self-Test .................................................................................................21
12.2 Battery Operation .......................................................................................... 22
12.3 Changing the Batteries of the Temperature and Humidity Measuring
Adapter Z541A ............................................................................................. 22
12.4 Housing ....................................................................................................... 22
+

4WOLFGANG WARMBIER
1 Applications
ESD protective measures can be tested and documented quickly and
efficiently with the METRISOâ2000 measuring and test instrument within
your EPA (Electrostatic Protected Area).
The test instrument is especially well suited:
• For the selection of ESD materials
• For the installation of selected protective measures
• For periodic testing of protective measures
All relevant measurements in accordance with IEC 61340-5-1 / 5-2,
DIN EN 61340-5-1 and IEC 61340-4-1 can be performed with the
METRISOâ2000.
Included with the METRISOâ2000 Measurement Kit (7100.2000.MK):
• High-resistance ohmmeter with integrated infrared interface
• Plug-on temperature and humidity sensor
• IR interface adapter (RS 232) with data readout software
• 2 measuring probes according to IEC 61340-4-1
• Measuring cables and conductive carrying case with dissipative foam
inserts
Stored measurement values can be uploaded to a PC for printing and
archiving via the integrated infrared interface at the METRISOâ2000. This
is especially important where product liability is concerned.
The following measurements and tests can be performed with the
METRISOâ2000:
• Resistance to ground, surface and volume resistance
(with special measuring probes)
• Insulation resistance
(with variable and rising measurement voltage)
• Temperature and relative humidity
(with included, special sensor)
• Voltage and frequency
2 Safety Features and Precautions
The METRISOâ2000 electronic measuring and test instrument is
manufactured and tested in accordance with safety regulations
IEC/EN 61010-1/VDE 0411-1 and EN 61557.
If used for its intended purpose, the safety of the operator and the
instrument is assured.
Read the operating instructions carefully and thoroughly before using your
instrument, and observe all instructions included therein.
The measuring and test instrument may not be used:
• If the battery compartment cover has been removed
• If external damage is apparent
• With damaged connector cables or measuring adapters
• If it no longer functions flawlessly
• After excessive stress due to transport
• After lengthy periods of storage under unfavorable conditions
(e.g. humidity, dust, extreme temperatures)
• While the batteries are being charged
Meanings of Symbols on the Instrument
Warning concerning a point of danger
(Attention: observe documentation!)
Protection class II device
CAT III Overvoltage category III device
9 V DC charging socket for NA 0100S charging adapter
(part no. 7100.2000.MK.Z501D)
!
–

WOLFGANG WARMBIER 5
3 Initial Start-Up
3.1 Battery Test
Five battery symbol segments ranging from depleted to fully charged
continuously indicate the current battery level at the main menu.
3.2 Installing and Replacing Batteries
New batteries must be installed before initial start-up, or when only one
solid segment remains in the battery symbol.
Attention!
!
The instrument must be disconnected from the measuring circuit
(mains) at all poles before the battery compartment is opened.
Four 1.5 V baby cells in accordance with IEC LR14 are required for
operation of the METRISOâ2000. Use alkaline-manganese batteries only.
Rechargeable NiCd or NiMH batteries can be used as well. Be absolutely
sure to refer to chapter 12.2, page 22, regarding the charging cycle and
the charging adapter.
Always replace the batteries in complete sets.
Dispose of batteries in an environmentally sound fashion.
➭Loosen the two slotted screws at the battery compartment cover on
the housing rear panel and remove the cover.
➭Insert four 1.5 V baby cells making certain they are poled in
accordance with the symbols. Insert the two batteries which are half
covered by the housing first.
➭Replace the cover and retighten the screws.
Attention!
!
The instrument may not be operated if the battery compartment
cover has not been installed and properly tightened!
3.3 Selecting a Menu and Configuring Basic Settings
Press the or the key in order to display the desired measuring
function, the device settings or the database functions.
START
START

6WOLFGANG WARMBIER
Default Settings – Last Used Settings
A selection can be made here as to whether the menus will be displayed
according to the default settings, or if the last opened menus should be
displayed.
➭Activate the Setup key.
➭Press the Default key, if required:
on ✓Settings such as Ton (= 20sec) are reset to the default settings
each time the instrument is switched on.
off ✓The last used settings remain when the instrument is switched
on.
➭Press the key to quit the setup menue.
Setting On-Time, Manual Shutdown
➭Activate the Setup key.
➭Press the Ton key and then the 10sec, 20sec, 30sec or 60sec key depend-
ing upon the desired duration after which the test instrument should
switch off automatically. Additional setting options can be displayed
with the help of the and scroll keys.
The “>>>>>” setting indicates that no automatic shutdown will occur.
The selected setting has a considerable influence on battery service
life.
➭Press the key to quit the setup menue.
The instrument can be switched off manually by simultaneously activating the two
outermost softkeys.

WOLFGANG WARMBIER 7
Background Illumination and Contrast
➭Activate the Display key.
➭In order to extend battery service life, display illumination can be
switched off entirely.
Press the corresponding softkey for this.
If LCD illumination is activated (= ON), it is automatically switched off
several seconds after the last key has been activated in order to extend
battery service life. As soon as a key is activated again, illumination is
switched back on.
➭Contrast can be optimized with the two keys at the far right.
➭Press the START key to quit the setup menue. The data are saved to
memory.
Setting the Clock
➭Activate the Time key.
➭The cursor appears at the first digit in the date. Enter the desired
number with one of the softkeys.
Numbers which do not appear in the window can be displayed with
the help of the or the key. Each time a number is selected,
the cursor moves to the next position to the right.
➭Date and time are saved as soon as the last number has been
entered.
➭Press the START key to quit the setup menue. The data are saved to
memory.
low high
Contrast
off on
LCD Illumination

8WOLFGANG WARMBIER
4 General Operation
4.1 Automatic Settings, Monitoring and Shutdown
The instrument cannot be placed into operation, or it is shut down
automatically, if battery voltage has fallen to below the lower limit value.
Measurement is automatically interrupted and/or the measuring function
is disabled if voltage exceeds the allowable limit in the resistance range
(U > 25 V).
The instrument switches itself off automatically, at the earliest after the
current (automatic) measuring sequence has been completed, and after
the predefined on-time has elapsed (see chapter 3.3). On-time is reset to
the duration selected in the setup menu each time a key is activated.
Lamp Functions
4.2 Measurement Value Display
The following is displayed:
• Measured value with unit and symbol
• The selected function
When automatic measuring sequences are used, the measurement
values are displayed in digital format until the next measuring sequence is
started, or until the instrument switches itself off automatically.
If the measuring range upper value is exceeded, the upper value is dis-
played preceded by the greater than symbol “>” in order to indicate over-
ranging (If the actual value is less than the measuring range lower value,
the value is displayed preceded by the less than symbol “<”, e.g. with
10 V or 100 V test voltage).
4.3 Online Help
Appropriate online help texts can be displayed at the LCD for each of the
basic functions and sub-functions, after the respective function has been
selected in the corresponding menu.
➭Press the key to query online help.
Press any key to exit the help function.
Lamp Status Function
red Potential difference ³150 V
between finger contact and measurement input
Netz
Mains
blinking
red
Line voltage or interference voltage is present at the
measurement inputs (resistance measurements are disabled)
LIMIT red – Measured resistance exceeds the selected limit value.
U>25V red A voltage of greater than 25 V is present at the measurement
inputs.
i

WOLFGANG WARMBIER 9
5 ESD Measurement
5.1 Condensed Instructions
Open the RESD menu. Select the test voltage menu.
Select the test voltage. Start measurement.
Press the START key briefly to perform a single, one-time measurement.
If measurement needs to be performed over a longer period of time, the
START key must be pressed and held for the desired test duration.
5.1.1 Setting Limit Values – LIMIT Function
A limit value can be selected for resistance RESD with the LIMIT key. If
measurement values occur which exceed the selected limit value, the red
LIMIT LED lights up.
Select the limit value menu. Set the limit value.
Setting the Limit Value:
Display the desired numbers, and if required the decimal point, with the
help of the and keys. Select the desired numbers with the appro-
priate softkeys. Each time an entry is made, the cursor moves one place
to the right. After a maximum of three numbers and the unit of
measure (k, M or G) have been entered, the entry window is exited
automatically. The cursor can be moved one place to the right during data
entry by pressing the ® softkey, or the entry window can be exited. The
selected limit value is automatically saved to memory when the data entry
window is exited.
START

10 WOLFGANG WARMBIER
5.2 Measuring Resistance to Ground REor RGaccording to IEC 61340-4-1 /
DIN IEC 61340-4-1
(e.g. for measurements at flooring systems, table mats, chairs etc.)
➭Connect the red measuring cable from the “+ pole” to the measuring
probe.
➭The black, shielded measuring cable with the double
“COM / SHIELD” connection is connected to PE (protective earth) in
order to measure REor to a groundable point to measure RG.
➭Select the RESD menu and a test voltage of UM= 10 V. If a value of
less than 100 kWis displayed, the measured value is the result.
➭If the value is displayed with RX> 100 kW, the test voltage must be
changed to 100 volts. The value then displayed is the result.
Test Voltage according to IEC 61340-5-1 / DIN EN 61340-5-1:
5.3 Measuring Resistance through a Person to Ground RE System or RG System
(e.g. for the measurement of a resistance through the combination of a person / shoes /
flooring system according to ESD STM 97.1-1999 and IEC 61340-4-5/NP)
➭Connect the red measuring cable from the “+ pole” to the measuring
probe.
➭The black, shielded measuring cable with the double
“COM / SHIELD” connection is connected to PE (protective earth) in
order to measure RE System or to a groundable point to measure
RG System.
➭Select the RESD menu and a test voltage of UM= 10 volts. If a value of
less than 100 kWis displayed, the measured value is the result.
➭If the value is displayed with RX> 100 kW, the test voltage must be
changed to 100 volts. The value then displayed is the result.
Test Voltage according to IEC 61340-5-1 / DIN EN 61340-5-1:
Rxin WTest Voltage
(open-circuit voltage)
RX£105£10 V
105< RX£1012 100 V
+ Pole COM / SHIELD
Test Object
Rxin WTest Volt age
(open-circuit voltage)
RX£105£10 V
105< RX£1012 100 V
+ Pole COM / SHIELD
Test Object

WOLFGANG WARMBIER 11
5.4 Measuring Surface Resistance
5.4.1 Point to Point Measurement RSwith 2 Measuring Probes according to
IEC 61340-4-1 / DIN IEC 61340-4-1
(e.g. for measurements of working surfaces, storage racks, transport boxes, etc.)
➭Connect the red measuring cable from the “+ pole” to one measuring
probe and the black, shielded measuring cable with the double
“COM / SHIELD” connection to the second measuring probe in order
to measure RS.
➭Select the RESD menu and a test voltage UM= 10 volts. If a value of
less than 100 kWis displayed, the measured value is the result.
➭If the value is displayed with RX> 100 kW, the test voltage must be
changed to 100 volts. The value then displayed is the result.
Test Voltage according to IEC 61340-5-1 / DIN EN 61340-5-1:
5.4.2 Measuring Surface Resistance RSof Packaging Materials
with a Ring Probe in compliance with IEC 61340-5-1 / DIN EN 61340-5-1
(e.g. for the measurement of conductive (dissipative) packaging materials)
➭Connect the red measuring cable from the “+ pole” to one of the
measuring jacks and the black, shielded measuring cable with the
double “COM / SHIELD” connection to the second measuring jack.
➭Connect the “GUARD terminal” at the ring probe to the GUARD jack
at the METRISOâ2000.
➭Select the RESD menu and a test voltage of UM= 10 V. If a value of
less than 100 kWis displayed, the measured value is the result.
Rxin WTest Voltage
(open-circuit voltage)
RX£105£10 V
105< RX£1012 100 V
+ Pole COM / SHIELD
Test Object
+ Pole COM / SHIELD
Test Object
+ Pole
GUARD
COM / SHIELD

12 WOLFGANG WARMBIER
➭If the value is displayed with RX> 100 kW, the test voltage must be
changed to 100 volts. The value then displayed is the result.
Test Voltage according to IEC 61340-5-1 / DIN EN 61340-5-1:
5.5 Measuring Volume Resistance RVaccording to IEC 61340-4-1 /
DIN IEC 61340-4-1
(e.g. for measuring the volume resistance of materials)
➭Connect the red measuring cable from the “+ pole” to the measuring
probe and the black, shielded measuring cable with the double
“COM / SHIELD” connection to the counter-electrode.
➭Select the RESD menu and a test voltage of UM= 10 volts. If a value of
less than 100 kWis displayed, the measured value is the result.
➭If the value is displayed with RX> 100 kW, the test voltage must be
changed to 100 volts. The value then displayed is the result.
Test Voltage according to IEC 61340-5-1 / DIN EN 61340-5-1:
5.6 Measurement with the GUARD Terminal
The measurement of high-value resistance necessitates extremely small
measuring current and may be rendered problematic by influences such
as electromagnetic fields, humidity or surface current. A precise test setup
is therefore absolutely essential.
For measurements within a range of 10 GWto 1 TW, a guard cable must
be used to prevent surface current from distorting measurement results.
The GUARD terminal at the measuring probe prevents current at the sur-
face of a highly resistive test object from flowing from the + measurement
cable to the - measurement cable, instead of through the highly resistive
material itself.
Ring Probe in compliance with IEC 61340-5-1 / DIN EN 61340-5-1
➭Connect the GUARD terminal at the measuring probe to the GUARD
jack at the METRISOâ2000 for the measurement of highly
resistive materials.
Measurement is otherwise performed as described in the previous
chapters.
Rxin WTest Voltage
(open-circuit voltage)
RX£105£10 V
105< RX£1012 100 V
Rxin WTest Voltage
(open-circuit voltage)
RX£105£10 V
105< RX£1012 100 V
+ Pole
COM / SHIELD
Test Object
+
COM / SHIELD
Counter-Electrode
+ Pole
GUARD
COM / SHIELD

WOLFGANG WARMBIER 13
6 Database Functions
The displayed measurement data for each measurement can be saved to
an internal database, with or without a comment. A data record must be
created and allocated to a specific memory address in order to be able to
assign the individual measurement values to different locations, depart-
ments and systems.
6.1 Creating a Data Record – Data Function
➭Press the Data key.
➭First create the desired memory address with the help of the softkeys.
After acknowledging with the START key (press at center), the cursor
appears at the first data entry position (LOCATION)
➭With the help of the softkeys, entries can be made to the LOCATION,
DEPARTMENT, Nr. and INDEX fields one after the other, and a designation
can be entered for the test object.
Entering Data:
Display the desired alphanumeric character by pressing the or
the key, and then select the character with the corresponding
softkey.
Control characters can be entered in the same way and have the fol-
lowing meanings:
¬: Move cursor to the left (without deleting data)
®: Move cursor to the right (without deleting data)
¿: Same function as the START key
After each character is selected, the cursor moves one position to the
right. If ¿or the START key is activated (press at center), the cursor
moves to the next entry field. After the fields LOCATION, DEPARTMENT, Nr.
and INDEX have been completed, and after acknowledgement has
been confirmed with the ¿softkey, the data fields are displayed as in-
verse images. After the ¿softkey has been activated once again, a
designation for the selected test object can be entered.
Note
☞These entries are required by the PC software in order to enter
measurement values into the database, and to generate reports
with this information automatically.

14 WOLFGANG WARMBIER
6.2 Saving Measurement Values – STORE Function
➭Start the respective measurement. The STORE key is displayed after
the measurement instead of the INFO key.
The STORE key is not displayed until after a given amount of time has
elapsed for measurements which are performed without the START
key. For example, the store key is not displayed after voltage
measurements until a given amount of time has elapsed, so that the
operator can first query the help text with the INFO key.
➭The displayed measurement values are stored to the currently
selected database memory address by briefly acknowledging with
the STORE key. The key is briefly displayed as an inverse image during
storage to memory.
➭Pressing and holding the STORE key allows for the entry of a comment,
and storage of the current measurement.
Entering a comment:
Display the desired alphanumeric character with the or the
key and select the desired character with the corresponding softkey.
Control characters are entered in the same way and have the
following meanings:
¬: reverse and delete,
¿: same as the START key
After each character has been selected, the cursor moves one posi-
tion to the right. Already entered characters can be deleted in reverse
by pressing and holding any softkey (except for ¿).
After entry of up to 15 characters, save the measurement values and
the comment by acknowledging with the START key (press at center) .
The following message appears: “Data are stored”.
6.3 Querying Data Records – View Function
➭Select View.
➭You can scroll forward through the memory addresses with the
key, or backwards with the key.
➭After selecting a memory address, the individual data records can be
queried with the Prev. and Next softkeys, which have been stored to
memory under consecutive numbers.
In case a measurement value is missing for the currently selected location,
perform the measurement and the value will be immediately stored.
START
START

WOLFGANG WARMBIER 15
6.3.1 Deleting a Data Record from within a Memory Address – View Function
➭Activate the Del key. No security request appears.
Data record numbering is changed as soon as an individual data
record is deleted.
6.3.2 Deleting a Memory Address – Data Function
➭Select Data.
➭Enter blanks to the data fields LOCATION, Department, Nr. and INDEX.
After all of these fields have been entirely overwritten with blanks,they
are displayed as inverse images.
➭Acknowledge with the START key (press at center). All data from the
selected memory address are deleted.
START

16 WOLFGANG WARMBIER
6.3.3 Delete All Memory Addresses – Data Function
Approximately 2,500 data records can be stored to memory. The memory
is full when the bar graph to the right of “MEMORY:” is entirely filled in. The
entire memory, i.e. all data records from all memory addresses, can be
deleted at once. We recommend uploading and saving your data to a PC
before deletion.
➭Select Data.
➭Enter memory address “000”. After acknowledgement by pressing
button START (press at center) a safety request appears.
➭By simultaneously activating O and K all data that have been saved
are deleted from memory. The bar graph to the right of the parameter
„MEMORY:“ is empty. To the left, memory address „001“ is displayed.
New data can now be entered for the first address, or the
database can be exited (press ¿or START 9 times).
If the above message appears upon switching on the test instrument, all
data can be saved to a PC before finally deleting the database in order to
rectify the error.
3x
+
+

WOLFGANG WARMBIER 17
7 Measuring Insulation Resistance
7.1 Tests with Variable Test Voltage
DC test voltages within a range of 50.0 to 600 V can be selected under
UVAR for measurements at sensitive components, as well as within sys-
tems equipped with voltage limiting devices.
Test Voltage Menu: Display and select variable test voltage.
Select test voltage. Start measurement.
Selecting the Test Voltage:
Display the desired numbers, and if required the decimal point, with the
help of the and keys. Select the desired numbers with the appro-
priate softkeys. Each time an entry is made, the cursor moves one place
to the right. After a maximum of three numbers have been entered, exit
the data entry window by pressing the ® softkey. If the selected
voltage is not within the valid range, the cursor jumps back to the entry
position for the first digit.
7.2 Tests with Rising Test Voltage
The “U ” function is used for the detection of insulation weak points, and
for the determination of response voltage for voltage limiting devices.
Select rising test voltage. Start measurement.
As long as the START key is depressed, test voltage rises
continuously. Insulation measurement is started:
• As soon as maximum voltage has been reached
(variable test voltage),
• As soon as the START key is released
(when the desired voltage is displayed) or
• As soon as a measurable test current is detected
(e.g. after a discharge voltage sparkover).
Test voltage, any response and discharge voltage values and
insulation resistance are all displayed.
START
START
START

18 WOLFGANG WARMBIER
8 Measuring Alternating Voltage
Sinusoidal alternating voltage within a frequency range of 15 to 400 Hz
can be measured with the test instrument.
➭Press the U~/f key.
➭Contact the measuring points with the measuring probes.
Notes
• The voltage measuring range has an overload capacity of 1200 V.
• The input impedance for the voltage measuring range is 5 MW.
9 Measuring Temperature and Humidity with Adapter Z541A
Temperatures within a range of –10.0 °C to +50.0 °C, as well as
relative humidity from 10.0% to 90.0% can be measured with the temper-
ature-humidity combination adapter.
➭Connect the temperature-humidity adapter to the IR interface of the
test instrument (see drawing on page 2) by inserting the guide pin of
the adapter in the opening provided at the top of the test instrument
such that the adapter is located on the two rubber pads in central
position to the housing. Then push the adapter down to fix it tightly.
➭Press the T/R.H. key. The adapter is activated via the interface.
➭Select the desired temperature unit of measure, °C or °F, with the
corresponding key. Temperature and humidity are displayed directly.
➭We recommend moving the test instrument with the adapter for a few
seconds to enable the adapter to adjust to the prevailing room climate
more quickly. The air current can thus penetrate through the opening
into the adapter housing more quickly. It would otherwise take several
minutes to achieve this condition.
➭When changing to another function, the adapter is deactivated.
➭The measured values are only updated every 5 seconds to extend the
service life of the batteries in the adapter.
Display at the Test Instrument
The „———“ symbol shown on the display can be attributed to the fol-
lowing causes:
– Exposure to intense direct sunlight
– depleted batteries in adapter Z541A
– adapter defective or not properly positioned
Adapter Z541A should not be exposed to intensive sunlight to prevent
inadvertent activation via infrared (battery life!).

WOLFGANG WARMBIER 19
10 Characteristic Values
1) With external adapter
2) DC test voltage = UM±5%
3) At nominal resistance RN= 1000 W/V
4) No-load voltage at R ³ 10 MW
Reference Conditions
Ambient Temperature + 23 °C±2 K
Relative Humidity 45 ... 55%
Battery Voltage 5.5 V ±1%
Meas. Qty. Frequency 50 Hz ±0.2 Hz
Live Voltage Waveshape sine, deviation between RMS and
rectified values < 1 %
Power Supply
Batteries 4 ea. 1.5 V baby cells (4 x C-Size)
(alkaline-manganese per IEC LR14)
or 4 ea. NiCd rechargeable batteries
Nominal Range of Use 4.6 ... 6.5 V
Battery Test symbolic display
Battery Saver Circuit Display illumination can be switched off. The
test instrument switches itself off automatically
10 to 60 seconds after the last key operation.
On-time can be selected by the user.
Service Life For RESD:3000 measurements with one set of
batteries (alkaline-manganese) with 10 s
on-time and performance of one measurement
before each automatic instrument shutdown
For U~/f and T/FREL:
5 hours (same conditions as above)
Measured
Quantity Measuring Range Test Current Nominal Range of Use Nominal Values /
Impedance Intrinsic Error Measuring Error
RESD 01.0 kW ... 99.9 kW—1 kW ... 100 kWUM= 10 V 4) ±(5% rdg. + 3 d) ±(7% rdg. + 3 d)
RESD > 100 kW ... 10.0 GW1 mA 3) > 100 kW ... 10 GWUM= 100 V 2) ±(5% rdg. + 3 d) ±(7% rdg. + 3 d)
RESD > 10 GW ... 100 GW1 mA 3) > 10 GW ... 100 GWUM= 100 V 2) ±(8% rdg. + 3 d) ±(10% rdg. + 3 d)
RESD 0.00 MW ... 10.0 GW1 mA 3) > 100 kW ... 10 GWUM= 250/500 V 2) ±(5% rdg. + 3 d) ±(7% rdg. + 3 d)
RESD 0.00 MW ... 100 GW1 mA 3) > 10 GW ... 100 GWUM=250/500 V 2) ±(8% rdg. + 3 d) ±(10% rdg. + 3 d)
RESD > 100 GW ... 999 GW1 mA 3) > 100 GW ... 1 TWUM=100/250/500 V 2) ±(25% rdg. + 5 d) ±(50% rdg. + 20 d)
UM000 V ... 720 V 50 ¼ 600 V 5 MW±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
U~ 00.0 V ... 720 V 10 ¼600 V 5 MW±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
f15.0 ¼400 Hz 45 ¼200 Hz 5 MW±(0.5% rdg. + 2 d) ±(1% rdg. + 2 d)
T 1) –10.0 ¼+50.0 °C0¼+40 °C±2 °C
Frel 1) 10.0 ¼90.0% 20 ¼80% ±5%
Phase
Test LED PE > 100 V 100 ¼ 1000 V > 100 MW/50 Hz

20 WOLFGANG WARMBIER
Safety Shutdown The instrument is switched off, or cannot be
switched on, if supply voltage has dropped to
below the critical level.
Charging Socket Installed rechargeable batteries can be
recharged inside the instrument by connecting
the NA0100S battery charger
(7100.2000.MK.Z501D) to the charging socket.
Overload Capacity
RESD 600 V continuous, 720 V for 10 s
U~ 600 V~ continuous, 720 V for 10 s
Electrical Safety
Safety Class II per IEC 61010-1/EN 61010-1/
VDE 0411-1
Nominal Voltage 1000 V
Test Voltage 5.55 kV
Overvoltage
Category 600 V CAT III
Fouling Factor 2
Electromagnetic
Compatibility (EMC) IEC 61326/EN 61326
Ambient Conditions
Nominal Temperature 0 ... +40 °C
Operating Temperature –10 ... +50 °C
Storage Temperature –20 ... +60 °C (without batteries)
Relative Humidity to 75% (max. 85% for storage and transport),
no condensation allowed
Elevation max. 2000 m
Deployment indoors only
Mechanical Design
Display multiple dot matrix display
128 x 64 pixels (65 mm x 38 mm), illuminated
Protection housing: IP 52 per DIN VDE 0470 part 1/
EN 60529
Dimensions 275 mm x 140 mm x 65 mm
Weight approx. 1.2 kg with batteries
Data Interface
Type infrared interface (SIR/IrDa)
bidirectional, half-duplex
Format 9600 baud,
1 start bit, 1 stop bit, 8 data bits,
no parity, no handshake
Range max. 10 cm
recommended: < 4 cm
Table of contents
Other Wolfgang Warmbier Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

PCB Piezotronics
PCB Piezotronics 3503A1160KG/-0002 Installation and operating manual

Triplett
Triplett CM1050 user manual

Zhejiang Uniview Technologies
Zhejiang Uniview Technologies B3223P30 Release notes

LOVATO ELECTRIC
LOVATO ELECTRIC DMED320 instruction manual

Ruskin
Ruskin AIRFLOW-IQ Start-up & Commissioning Guide

X-Rite
X-Rite Ci7860 Series user guide

Garmin
Garmin LIDAR-Lite v1 Silver Label operating manual

Han Seung
Han Seung HTS-CT10 Operation manual

autosen
autosen AS005 operating instructions

Eleaf
Eleaf LCD Meter user manual

PCB Piezotronics
PCB Piezotronics IMI Sensors 625B61 Installation and operating manual

Trace2O
Trace2O ARSENOMETER HMAS100 instruction manual