YUNPAI GoChek User manual

Document1010-LBL-00021
RevisionV01
Issued2017/10/20
Address: No.9 Haishu Road,Yuhang District
Hangzhou, Zhejiang 311121 China
Phone: +86-571-88566372
Website: http://www.microtechmd.com
EC REP Lotus Global Co., Ltd.
1 Four Seasons Terrace
West Drayton, Middlesex
London, UB7 9GG, United Kingdom
BLOOD GLUCOSE MONITORING SYSTEM
Instruction
Manual

O
Dear User:
Thank you for choosing the GoChek blood glucose meter. Monitoring
your blood glucose level is an integral part of treating diabetes.
MicroTech Medical is committed to help you control your condition so
you can live a healthy and active life.
Please read this user manual to help master the simple functions of
your GoChek meter. If used properly, the blood glucose meter will
give you accurate results.
The GoChek blood glucose monitoring system will give you accurate
readings from fresh capillary or venous whole blood. This product is
for professional use in medical institutes or for self-testing by
diabetics. Test results serve only as helpful information, not as proof
for clinical diagnoses.
Important safety instructions before use:
Do not drop the meter or get it wet.
Only use the meter as directed by this instruction manual.
Only use GoChek blood glucose testing strips (hereafter
referred to as “test strips”).
Do not use the meter if is not working properly or has been
damaged.
Do not place anything on top of the meter.
Unless specifically instructed in this user manual, do not let
anything get inside of the meter openings.
This meter can only be used to determine blood glucose levels
using whole blood samples. Do not use serum or plasma
samples.
Patients who are severely ill, suffer from severe dehydration, or
are in a hyperosmolar state should not use the GoChek system.
Remove the battery if the device will not be used for an
extended period of time.
Do not use test strips or control solution past the expiration date.
Avoid using the meter near devices emitting electromagnetic
radiation such as TV’s, mobile phones, microwaves, and X-ray
machines. Avoid static electricity, especially in very dry
environments.

TSpecificationTIndex
OO
Product Name: GoChek / GoChek Connect / GoChek Connect+
Size: 83.5 mm (L) × 54 mm (W) × 19 mm (Thickness)
Glucose Test Range : 1.1-33.3 mmol/L (20-600 mg/dL)
Results Display: plasma equivalent
0LQLPXP6DPSOH9ROXPHȝ/
Test Time: 5 seconds
Battery: 1x CR 2032 3.0V coin cell battery
Battery life: >1,000 readings (wireless functions disabled)
Glucose Concentration Units: mmol/Lor mg/dLdepending on the
standard of your country
Memory Storage: 500 test results with date and time stamp
Auto Shutdown: Automatic shutdown after 2 minutes
Display Size: 40mm × 42mm
Weight: about 50 grams (including battery)
Operating Temperature: 5-45°C (41-113°F)
Operating Humidity: 20-90% (non-condensing)
Hematocrit Range: 20-70%
Note:
Blood containing high levels of vitamin C or other reducing agents
can lead to inaccurate test results. Triglycerides above 3,000
mg/dL and cholesterol above 500 mg/dL will lead to inaccurate test
results.
Specification
Product Components
Component Descriptions
Blood Glucose Meter
Display
Test Strips
Code Chip
Control Solution
Meter Setup Before Use
Battery Installation
Meter Setup
Quality Control Tests
Testing Your Blood
Blood Sampling
Lancet Removal
Testing Blood Glucose
Understanding Test Results
Suggested Test Times and Target Goals
“HI” and “LO” Messages
“HYPO” and “HYPER” Messages
Meter Memory
Memory Mode
Clearing the Memory
Memory Data Transfer
Comparing Meter and Laboratory Results
Maintenance
Storage
Replacing the Battery
Cleaning
Troubleshooting
Warranty
Symbol Index

T TComponent Descriptions
998
4 5
Code Chip
Battery
Glucose Test Strips
Blood Glucose Meter Control Solution
Lancing Device
Carrying Case
LancetClear Cap
Included components: Glucose meter, lancing device, clear cap, battery,
user guide, carrying case
Optional components (purchased separately): blood glucose test
strips, control solution, lancets, code chip
Product Components
Test Strip Port: Test strips are inserted into this area to perform a test.
Display: Shows test results, settings, and other information.
Code Chip: Code chip insertion area.
“<” Button: When the meter is off, holding the “<” button will turn the
meter on and enter the system settings menu.
“M” Button: When the meter is off, holding the “M” button will enter
memory mode. This button also is used to select/confirm.
“>” Button: When the meter is off, holding the “>” button will enter the test
reminder alarm mode.
Battery Door: Remove the battery door to install one CR2032 coin cell
battery.
Label: Contains product information.
Test Strip Ejector: Slide the ejector forward to discard a used test strip.
Data Port: When used with a data cable, you can transfer the data stored
in the meter to your personal computer, analyze, and print it. (Requires
data cable, please contact customer service)
Data Port
Test Strip Port Test Strip
Ejector
Display
">" Button
"M" Button
"<" Button
Code Chip
Label
Battery Door
The meter reads the test strips and displays the blood glucose
concentration. Use this diagram to become familiar with all of the
parts of your meter.
Blood Glucose Meter

Display
Date Time
Apply Blood
Icon AM/PM
Day Average
Pound Sign (#)
Memory
Control Solution
Code
Hypo Icon
Readings Icon
Low Battery Icon
Mute
Test Reminder Alarm
Check Ketone Alert
Hyper Icon
Test Result
Display Area
Blood Glucose
Measurement Units
Before Meal
Icon After Meal
Icon
BlueTooth Icon
GoChek blood glucose test strips contain chemical reagents.After the
test strip is placed into the test strip port and a blood sample is
applied, the blood automatically wicks into the test window. A
transient electrical current is generated, and this current is measured
to determine the correct blood glucose level reading.
Strips Test
Fingerhold Area Use this area to grip the strip
Sample Area
Apply blood or
control solution here
Test Window Check to confirm an
adequate sample has been applied
Contacts
Insert this side into
the test strip port
IMPORTANT:
Make sure that the blood sample is applied at the edge of the
test strip, never on the top surface. Applying blood to the top
surface of the strip will lead to inaccurate results.
Apply blood samples to the edge of the strip until the test window is
full. The blood glucose meter will begin to count down when there is
adequate blood. If the test window is not full, you may add additional
blood within three seconds. If not enough blood has been applied, an
error E-3 will be displayed. Please discard and use a new test strip.
If you see that the test window is not full, but the countdown begins
anyway, please discard and use a new test strip.
Storage and Handling
Please read and follow the storage and handling conditions
below:
Store test strips in a clean, dry environment at 5-30°C (41-86°F).
Do not store test strips in heat or direct sunlight.
Each test strip package is marked with a batch number (LOT),
expiration date ( ) and a control solution range (CTRL1 and CTRL2).
Please use the correct strips recommended for your meter. The use
of non-approved strips will result in the meter not turning on or an
error.
Correct Incorrect

Do not refrigerate or freeze test strips.
Do not store or use strips in a humid environment, such as a bathroom.
Do not store the meter, test strips, or control solution near bleach or
cleaning agents that contain bleach.
Close the cap on the vial immediately after removing a test strip.
Use the test strip immediately after removing it from the package.
Do not use expired test strips. Doing so may lead to inaccurate results.
Tip: The test strip label contains the expiration date in year-month format. For
example, 2017-01 indicates that the test strips are valid until January 2017.
Special Instructions for Test Strips Sold in a Vial
Test strips should be stored in the tightly capped vial that is provided.
Do not store test strips outside of the provided vial. Test strips must be
stored inside the original vial with the lid tightly sealed closed.
Do not transfer test strips from the provided vial into another container.
Close the cap on the vial immediately after removing a test strip.
A new vial of test strips may be used for 6 months after first being
opened. Please take note of the date that the vial was first opened,
and discard after 6 months.
Special Instructions for Test Strips Sold in
Single Use Packets
Tear the packet open carefully beginning from the tear gap. Do not
damage or bend the test strip.
Use the test strip immediately after removing it from the packet.
Test Strip Precautions
For in vitro diagnostic use.
Use the test strip immediately after removing it from the package,
otherwise the test results may not be accurate.
Do not use test strips that are torn, bent, or damaged in any way. Do
not reuse test strips.
Keep the test strip packaging away from children and pets.
Consult your physician or healthcare professional before making any
changes in your treatment plan based on your blood glucose test
results.
Please refer to the test strip instructions for more detailed information.
Code Chip
Insert the code chip into the meter before each test. It is used to
calibrate the meter for use with the associated test strips.
Control Solution
Control solution is a glucose solution of known concentration that is used
to confirm that your meter and test strips are working properly. It is
important to run a quality control test regularly to make sure that you are
getting accurate results.
You should perform a quality control test in the following situations:
Refer to page 12 for instructions on how to perform a quality control test.
Storage and Handling
Please review the following storage and handling instructions:
Tip: The control solution label contains the expiration date in year-month format. For
example, 2017-01 indicates that the test strips are valid until January 2017.
When you suspect that the meter or test strips
are not working properly.
When you suspect that your test results are
inaccurate, or inconsistent with how you feel.
When you suspect that your meter has been
damaged.
After cleaning your meter.
Store the control solution in the temperature range 5-30°C (41-86°F).
Do not refrigerate or freeze the control solution.
If the control solution is cold, do not use until it has warmed to room
temperature.
Do not use expired control solution.
Control solution may be used for 6 months after the bottle is opened
for the first time. Please take note of the date that the bottle was first
opened, and discard after 6 months. Do not use beyond the expiration
date.
Control Solution Precautions
For in vitro diagnostic use. The control solution is for testing only
outside of the body. Do not swallow or inject.
Control solution should be shaken before use.
Quality control tests should be carried out at 15-30°C.
Do not let the control solution bottle touch the test strip.
Use only the control solution that is recommended for your meter.
The control ranges shown on the test strip package are not
recommended ranges for your blood glucose level. Your personal
glucose range should be determined by your healthcare professional.

TMeter Setup Before Use
Battery Installation
The meter requires a CR2032 3.0V coin cell battery. You can find one in
the carrying case. Please follow the steps below for installation:
1.
2.
3.
Month Day Year Month Day Time
Turn the meter to the back side, press to open the battery door as
shown in the picture.
Insert the battery. Make sure that the positive side (+) is facing up.
Replace the battery door, making sure that it is closed tightly.
or
2.
3.
The display will now show the month and day in the top left corner.
Change the month and day by pressing the “<” or “>” buttons .
Press “M” to save.
Next, set the correct time. By default, the meter will show the time
in 24 hour display mode. Press “M” to save the correct time. The
meter will now allow you to change to either 24 hour or 12 hour
display mode by pressing “<” or “>”. Press “M” to save and
continue to the next step.
Audio Settings
You can turn your meter’s sound on and off in the audio settings menu.
Press “<” or “>” to turn on or off the sound. Press the “M” button to save
this setting.
When the sound is on, the meter will provide a beep when:
You will hear a one beep warning sound when the meter is powered on,
if the blood sample is adequately applied, and when the test results are
displayed. You will hear 2 short beep warning sounds if an error occurs.
Please refer to the error code table included in this instruction manual.
After you have chosen the audio setting, the meter will enter the high
and low blood glucose alarm settings.
a. The meter is powered on
b. The blood sample is adequate
c. The test results are displayed
d. An error occurs
e. Atest reminder alarm is activated
or
Meter Setup
Follow the steps below to set up your meter:
System Settings
Before using your meter for the first time, make sure all system settings
are correct. For example, setting the correct time is essential for the
history function to work properly.
Press the “<” button for 2 seconds to enter the system settings menu.
Time Settings
1.
The display will show the month, day, and year. In the upper right
portion of the screen, the year will flash. Press the “<” or “>” buttons
to change the year, and then the “M” button to save your choice.

or
or
or
High and Low Blood Glucose Alarm Settings
The meter has high (Hyper) and low (Hypo) blood sugar warning alerts.
You should set these high and low alerts according to your health care
professional’s advice. When your test result is lower than the Hypo alert
value (the meter can be set to a maximum Hypo value of 5.6 mmol/L (100
mg/dL)), the symbol “HYPO” will be displayed. When your test result is
higher than the Hyper alert value (the meter can be set to a minimum
Hyper value of 6.7 mmol/L (120 mg/dL)), the symbol “HYPER” will be
displayed.
The Hyper and Hypo settings are OFF by default.
To change the Hypo limit alert value (or turn off), press the “<” or “>”
buttons and then the “M” button to save. Next, change the Hyper limit alert
value (or turn off) by pressing the “<” or “>” buttons and then the “M” button
to save. The meter will enter the ketone test warning setting.
Ketone Test Warning Settings
To turn on or off the blood ketone test warning, press the “<” or “>” buttons
and then the “M” button to save. When the blood ketone test warning
function is on, the “KETONE?” symbol will show in the upper right section
of the display if your test result is higher than 16.7 mmol/L (300 mg/dL).
After choosing the ketone test warning setting, the system will enter the
meal marker setting.
Meal Marker Settings
The meter has a meal marker function that allows you to mark your test
result as before or after a meal. To turn on or off meal marker function,
press the “<” or “>” buttons and then the “M” button to save. The meter will
power off after this setting is chosen.
Additional Settings
If your meter is has wireless functions (GoChek Connect and GoChek
Connect+), there may be some additional settings in the System Settings
menu. See additional user guide for more information.
Test Reminder Alarm Settings
You can preset up to 10 different test reminder alarms that can remind you
to regularly perform blood glucose tests. You can also turn also turn on or
off any of the reminder alarms.
When the meter is off, long press the “>” button to enter the test reminder
alarm settings menu.
Press the “<” or “>” button to select the number alarm that you would like
to edit, and then press the “M” button to confirm your choice. Now you can
edit the time (hour/minute) that you like the alarm to sound by pressing the
“<” or “>” button, and “M” to confirm/save. After saving the time, the meter
will turn off.
or

998
998
The test reminder alarms sound for 1 minute, and
then repeat at an interval of 5 minutes. The alarm
is cancelled by inserting a test strip or by pressing
any key.
When a test reminder alarm is on, the alarm clock
symbol will be displayed on the screen. An
example is shown in the figure:
When the test reminder alarm sounds, the display
shows:
Note: If you hold the “<” or “>” button, the numbers will decrease or
increase more quickly.
Tip: After you replace the battery, you will only need to reset the
time. Other settings are stored in the meter memory.
Quality Control Tests
Quality control testing is done to verify that the blood glucose meter and
test strips are working correctly, and to ensure the accuracy of the test
results. Please conduct quality control tests in the following situations:
When you suspect that the meter or test strips are not working
properly.
When you suspect that your test results are inaccurate, or if they are
inconsistent with the way you feel.
When you suspect the meter is damaged.
Insert the code chip into glucose meter.
Insert the test strip all of the way into the test strip port, and the
meter will turn on automatically. If the audio is on, you will hear a
“beep” sound, and all of the icons on the display will turn on at the
same time.
After all of the icons appear, the code chip number will display
automatically. Please compare this number with the number on the
test strip packaging. If they do not match, make sure that you are
using the correct code chip, and verify by repeating steps 1 through
3.
The screen will now display the date, time, and code chip number
with the apply blood icon flashing. Test strip icon and flashing blood
drop icon shows that the test strip has been inserted correctly.
Tip: If the test strip is not inserted correctly, the meter will not
power on.
Push the “M” button to mark the test a quality control test, and the
control solution icon will show on the display.

5.2-8.6mmol/L
CTRL 1
5.2-8.6mmol/L
CTRL 1
8.8-13.8mmol/L
CTRL 2
5.2-8.6mmol/L
CTRL 1
8.8-13.8mmol/L
CTRL 2
8.8-13.8mmol/L
CTRL 2
When “ “ and “#” are displayed simultaneously, the result is from a
control solution test, and it will not be included in the 7, 14, 30, 60, or 90
day average. When you access your test history, the control solution
results are not displayed.
If the control solution results are outside of the reference range:
After checking all of the conditions above, repeat the quality control test
with a new test strip. If the quality control test results are still outside of
the range printed on the test vial (or foil pouch), there may be a problem
with your meter. Please seek help and contact your dealer.
Two levels of control solution are available labeled Control Solution 1 and
Control Solution 2. Control Solution 1 is sufficient for most self testing
needs. If you think your meter or strips may not be working correctly, you
may also want to do a level 2 test. The ranges for both (CTRL1 and CTRL
2) are displayed on the test strip vial (or on the foil pouch). Simply repeat
Steps 4 through 6, using Control Solution 2. For confirmation of results,
Control Solution 1 tests should fall within the CTRL1 range, and Control
Solution 2 tests should fall within the CTRL2 range. If the control solution
test results do not fall within the respective ranges, DO NOT use the
system to test blood, as the system may not be working properly. If you
cannot fix the problem, contact your dealer for help.
Please contact your dealer to purchase control solution. Each box of
control solution contains both Control Solution 1 and Control Solution 2.
Confirm you are matching the correct range. Control Solution 1
results should be matched to the CTRL1 range printed on the test
strip vial (or foil pouch).
Check the expiration date of the test strip and control solution. Make
sure that the packages have not been opened for more than 6
months. Discard any expired test strips and control solution.
Confirm that you are testing within the correct temperature range
(15-30 °C).
Make sure that the test strip vial and control solution bottle have
been tightly closed.
Make sure that you are using the correct brand of control solution.
Make sure you are following the user guide instructions properly.
Shake the control solution bottle, gently squeeze out the control
solution, discard the first drop, and drop the second drop onto a clean
nonabsorbent surface. Now touch the second drop to the sample
area of the test strip. Do not let the bottle come into contact with the
test strip. If the audio is on, the meter will beep, telling you that
enough control solution has been applied.
Tips:
After applying enough control solution, the display will count down 5
After the test has been completed, slide the test strip ejector to pop
out the test strip, and the glucose meter will automatically turn off.
If a big bubble forms, wipe with clean cotton paper, and then do the
following steps.
If one drop of control solution does not fill the test window, please
add another drop within 3 seconds. Otherwise discard the test strip,
and try again.
seconds and the control
solution test result will be
displayed on the screen. If the
control solution test result falls
within the control range that is
printed on the package (CTRL
1 on test strip vial or foil
pouch), this indicates that the
meter is working normally, and
the system is functioning
properly.

The following steps show how to use the meter, test strips, code chip,
lancing device and sterile lancets together to measure your blood glucose
concentration. The main steps are as follows:
Step 1:
Step 2:
Step 3:
Step 4:
Blood Sampling
Before testing, first become familiar with how to collect blood and then
choose a clean and dry place to conduct the test.
IMPORTANT:
Prior to testing, use with either alcohol or soapy water to disinfect
the sampling site. Use warm water to increase blood flow if
necessary. Dry your hands and the sampling site, ensuring that
there is no soap residue remaining.
Insert the code chip into the glucose meter.
Insert the test strip (that corresponds with the code chip) face up
into the test strip port. The meter will automatically turn on and
display the code chip number and flashing “Apply Blood” icon.
You can choose the blood collection site. Usually a tiny amount
of blood is collected from the fingers, hands, or forearms. Touch
the blood drop lightly to the edge of the sample area. Complete
the test within two minutes, or the meter will automatically power
off.
After the meter detects that there is an adequate amount of
blood, the meter will count down 5 seconds and display the
glucose test results. The test result will be stored in the history
automatically. Slide the test strip ejector to eject the test strip,
and the meter shuts off automatically.
Clear Cap:
Use for sampling on arm or hands
Lancet Release Button
Safety Cap
Lancing Device Cap
Depth Adjustor Lancet
Holder
Cocking Barrel
Fingertip Testing:
Adjust the depth penetration to reduce the discomfort. You do not
need the clear cap for fingertip sampling.
Remove the lancing device
cap. Insert the lancet into the
lancet holder until it comes to a
complete stop.
Twist off the safety cap from
the lancet, save the safety cap
for lancet disposal.
Carefully install the lancing
device cap onto the lancing
device, avoid touching the
lancet needle tip.
Adjust the puncture depth by
rotating the depth adjustor (the
lancing device has 5 puncture
depth settings). To reduce
discomfort, choose the lowest
setting that still produces an
adequate blood sample.
Depth Adjustment:
1 and 2:
3:
4 and 5:
Note: Greater pressure
between the lancing device
against the finger will also
increase the puncture depth.
for delicate skin
for normal skin
for thick or calloused
skin
TTesting Your Blood

Pull back the cocking barrel until
you hear a click. Now the lancing
device is loaded and ready to
draw blood.
Before taking a blood sample,
wash your hands or use an
alcohol swab to clean the area.
Washing your hands in hot water
increases blood circulation. You
can also massage from wrist to
finger to promote better blood
circulation.
Holding the lancing device
against the side of the finger to
be lanced, press the release
button and then put down the
lancing device. Massage
forward slowly from the base
of your finger to the tip to
increase the sample size.
Wipe away the first drop of
blood and use the second drop
for the test strip.
Tip: To reduce pain, lance on
the sides of the fingertips,
where there are less nerve
endings. Rotate finger
locations as much as
possible to accelerate
wound healing and decrease
callouses.
Forearm or Palm Testing
Collecting blood samples from your palm or forearm may be less painful
than your fingertips because there are less nerve endings. The procedure
for collecting samples from your palm or forearm is slightly different. You
will need the clear cap for this type of sampling. You cannot adjust the
puncture depth with the clear cap.
IMPORTANT: There are important differences among forearm, palm
and fingertip samples that you should know. Important information
about forearm, palm glucose testing:
Please refer to "Fingertip Testing" sections 1-3 and 5 to learn how to install
lancets into lancets holder.
When blood levels are changing rapidly such as after a meal,
insulin dose or exercise, blood from the fingertips may show
these changes more rapidly than blood from other areas.
Fingertips should be used if testing is within 2 hours of a meal,
insulin dose or exercise and any time you feel glucose levels are
changing rapidly.
You should test with the fingertips anytime there is a concern for
hypoglycemia or you suffer from hypoglycemia unawareness.
Screw the clear cap onto the
lancing device.
Choose a sampling site on
your forearm or palm that is
dry and clean. Do not choose
an area near bone, visible
veins, or hair. To bring fresh
blood to the surface for
sampling, massage the site
vigorously until it feels warm.
Place the lancing device
against the puncture site. Press
and hold the clear cap against
the puncture site for a few
seconds. Press the release
button of the lancing device,
but do not immediately lift the
lancing device from the
puncture site. Continue to hold
the lancing device against the
puncture site until you can
confirm a sufficient blood
sample has formed.

998
998
Lancet Removal
Unscrew the lancing device
cap. Firmly push the needle
into the safety cap.
Pull out the lancet from the
lancet holder. Please dispose
of the used lancet properly.
Lancet Precautions:
Do not use a lancet if the safety cap is loose or missing.
Do not use a lancet if the needle is bent.
Use caution whenever a lancet needle is exposed.
Do not share lancets with other people.
To avoid cross contamination, always use a new sterilized
lancet. Do not reuse lancets.
Avoid contaminating lancets with hand lotion, detergents, oil,
and other debris.
Lancing devices and lancets should not be shared. Each person
should have his own lancing device and lancets.
Clean your lancing device before and after use with alcohol or a
disinfectant wipe. Be sure to clean the part of the device that
touches the finger. Do not immerse the lancing device in water.
Control excess bleeding and disinfect your wound after use.
Reminder:
Testing Blood Glucose
When you insert the code chip and test strip, the meter will automatically
turn on (except in data transfer mode).
Insert the code chip into glucose meter.
Insert the test strip all of the way into the test strip port, and the meter
will turn on automatically. If the audio is on, you will hear a “beep”
sound, and all of the icons on the display will turn on at the same time.
If the strip is inserted upside down, the meter will not turn on.
The test strip has been inserted correctly when you see the “Apply
Blood” icon and drop icon flashing. If the test strip is not correctly
inserted, the meter will not turn on – repeat the steps above.
After all of the icons appear, the code chip number will be displayed
automatically. Please compare this number with the number on the
test strip packaging. If they do not match, make sure that you are
using the correct code chip, and verify by repeating steps 1 - 3.

Apply blood to the sample area of the
test strip. If the audio is on, the meter will beep, telling you that enough
blood has been applied and the meter will start to to measure.
Apply blood to the front or back of the test strip.
Smear the blood drop onto the test strip.
Press your finger against the test strip.
If you press the “>” button, the test results will be marked with the “#”
icon. Results marked with “#” symbol will not be used in the 7, 14, 30,
60 or 90 day average calculations. If you accidentally mark your result
with the “#” icon, simply press the “>” button again to cancel the mark.
After marking a result with “#”, please test again with a new strip.
If the before/after meal marker setting is on, you can press the “<”
button to tag the reading before or after meal, as shown in the figure:
Press the “M” button to confirm.
DO NOT:
The display will count down 5 seconds during
the measurement process. If the audio is on, the
test will end with a beep. If you applied a blood
sample but the meter does not begin a
countdown, you may reapply a second drop of
blood within 3 seconds.
If an error code is displayed, please check the troubleshooting
instructions (page 32). If “HI” or “LO” symbol is displayed, refer to the
“HI” and “LO” Messages section (page 25).
Record valid results in your logbook with the date and time, and
compare them to the target goals set by your healthcare professional.
(Refer to Suggested Testing Times and Target Goals on page 24)
After the test has been completed, slide the test strip ejector to pop
out the test strip, and the glucose meter will automatically turn off.

TUnderstanding Test Results
Suggested Test Times and Target Goals
Tracking your glucose concentrations with regular blood glucose testing is
an important part of managing diabetes. Your healthcare professional will
help you decide the normal target range for you glucose levels and how
often you should test according to your situation. Suggested times to test
include:
More frequent tests may be required when:
When you wake up (before eating)
1-2 hours after breakfast
1-2 hours after lunch
Before dinner
Before bedtime
2-3 o'clock in the morning (if taking insulin)
You adjust your medication dosage.
You think your glucose levels may be too high or too low.
You feel ill.
Before breakfast
Before lunch
Before and after exercises
1-2 hours after dinner
After a snack
Normal blood glucose target ranges (from ADA Clinical Practice
Recommendations, 2011).
Time Target Range (mg/dL) Target Range (mmol/L)
Empty stomach 70-100 3.9-5.6
2 hours after a meal <140 <7.8
(Note: 1 mmol/L = 18 mg/dL)
How many times per day should I test?
Diabetes Type Number of Tests With Empty Stomach
Type 1 At least 3 times per day
Type 2 Test frequently to achieve your target range
Prenatal Diabetes At least twice per day
“HI” and “LO” Messages
Your meter can accurately measure blood glucose concentrations
between 1.1-33.3 mmol/L (20-600 mg/dL). “HI” and “LO” messages
indicate that the test measurement was outside of this range.
If “HI” is displayed, it indicates the test result is above 33.3 mmol/L (600
mg/dL). You should test again to ensure that no mistake was made in the
procedure. If you are sure your glucose meter is functioning properly, no
mistakes were made, and the test results still show “HI”, then you may be
in a state of severe hyperglycemia, please contact your healthcare
professional immediately.
If “LO” is displayed, it indicates the test result is below 1.1 mmol/L (20
mg/dL). You should test again to ensure that no mistake was made in the
procedure. If you are sure your glucose meter is functioning properly, no
mistakes were made, and the test results still show “LO”, then you may be
in a state of severe hypoglycemia, please contact your healthcare
professional immediately.
“ “HYPO” and “HYPER” Messages
If “HYPO” is displayed, it indicates the test result is below the
hypoglycemia limit that you entered in the settings.
If “HYPER” is displayed, it indicates the test result is above the
hyperglycemia limit that you entered in the settings.
Research indicates that measuring your blood glucose at least three times
per day will help you control you glucose level. Please discuss the number
of tests and test times with your healthcare professional.
Record your blood sugar levels and other relevant information in your log
book. Bring your records when you visit your healthcare professional. It
will help him understand your condition, and help him modify your
treatment plan.

Blood Ketone Test Reminder: if “KETONE?” is
displayed, this indicates a blood ketone test is
necessary because your blood glucose test results
are higher than 16.7mmol/L (300 mg/dL).
Please contact your healthcare provider.
Notes:
This meter sytem is only for in vitro diagnostic use, and only can be
used with GoChek test strips. Use of other test strips can lead to
wrong test results.
This meter can only be used to determine blood glucose levels with
whole blood samples. Do not use serum or plasma samples.
This meter can be used within a hematocrit (HCT) range of 20% ~
70%. Please do not use this product for testing if your hematocrit
value is out of this range.
Abnormally high levels of Vitamin C (ascorbic acid) or other reducing
substances will produce falsely high blood glucose measurements.
Fatty substances, such as triglycerides less than 3,000 mg/dL or
cholesterol less than 500 mg/dL, have no major effect on blood
glucose test results.
Diabetic patients receiving dialysis treatment should do not use the
GoChek glucose meter and its test strips.
Patients who are severely ill, suffer from severe dehydration, or are in
a hyperosmolar state (with or without ketosis) should not use the
GoChek system.
This product is only suitable for clinical screening tests or family
self-monitoring. The test results can’t be referred to as confirmed
cases. In order to ensure the accuracy of the results, the test results
can be further confirmed by other methods, such as biochemical
methods.
As with all diagnostic reagents, test results must be linked with a
professional doctor’s diagnosis from the other clinical symptoms.
Carefully process waste caused by blood glucose tests according to
the relevant local laws and regulations, because blood samples are
considered a biohazard.
Press the “M” button to enter memory mode. The
“ ” icon and latest test result will be displayed.
If the blood glucose meter is used for the first
time, “---” and “ ” will be displayed.The date will
not be shown for the first test.
The stored test result along with the associated
time and date are displayed at the same time.
Test results marked with “#” icon will not be
included in the 7, 14, 30, 60 and 90 day
averages.
Press the “<” or “>” button to review all stored test
results.
Press button “M” again to enter the data average
screen. The number of days (“DAY AVG”) and
number of readings (“READINGS”) used in the
average calculation will be displayed. If no data
has been recorded, then the meter will turn off.
Meter Memory
Memory Mode
The glucose meter can store up to 500 test results with the
corresponding time and date. If 500 test results already exist, a new
result will overwrite the oldest one. The glucose meter also is capable of
calculating 7, 14, 30, 60 and 90 day averages from the stored results.

Memory Data Transfer
GoChek meters are capable of transferring memory data to a personal
computer. Please refer to the software instructions for correct usage and
the proper equipment to connect to your computer.
GoChek Connect and GoChek Connect+ meters can transmit data
wirelessly to other devices such as mobile phones. Please refer to the
additional user guide for more information.
Your blood glucose meter and laboratory equipment both report glucose
concentrations in the serum or plasma component of your blood.
However variations between the two are normal, and your meter results
and laboratory results may be slightly different. Glucose concentration
results can be affected by a number of factors and conditions, but these
factors and conditions will not affect the test results of biochemical
analyzers.
Under normal conditions, the difference between measurements with your
meter and laboratory results are within the range allowed by national
standards.
To ensure a reasonable comparison between your meter and laboratory
results, please follow these guidelines:
1.
2.
3.
4.
5.
6.
Test results may show small deviations, this may due to the following
reasons:
Blood oxygen and red blood cell count vary from person to person, and
even within the same person. GoChek glucose meters test blood glucose
concentrations for the widest range of people possible. If the user’s blood
indexes fall within the middle of the range, the result will be ideal.
Otherwise, there will be some small deviations. (The deviations should be
within the range allowed by local government.)
Comparing Meter and Laboratory
Results
Make sure your meter is working properly.
Comparisons will be more accurate if you do not eat for at least four
hours (preferably eight hours) before testing.
Bring your blood glucose meter, test strips, and control solution to the
lab.
Ensure that the time between tests with your meter and the laboratory
is within 15 minutes.
Wash and dry your hands before obtaining a blood sample.
Make sure you closely follow the instructions in this manual.
Press the “M” button again, and then use the
“<” or “>” buttons to scroll through the 7, 14, 30,
60, and 90 day averages. Press the “M” button
to see the before meal and after meal marker
averages. The meter will calculate the
historical averages according to the
parameters that you choose and the meter will
show how many records are used to calculate
the average.
If the meter has not been used for the amount
of time indicated as the “day average”, the
meter will not show an average.
Finally, press “M” one last time to turn off the
meter.
Note: Test results marked as quality control test
or with the “#” symbol will not be used in the
calculated average.
Clearing the Memory
Please be careful when using the clear memory function because the
action is irreversible. Once the memory is deleted, it cannot be restored.
The clear memory function will erase all test records.
While in the memory mode, press the “<” and “>”
buttons at the same time enter the clear memory
mode.
Press “M” to confirm that you would like all
records to be erased. The “ ” and "---" icons will
display at the same time, and the meter will turn
off automatically after a few seconds.
If you want to abort the memory delete function,
you can cancel by pressing the “<” and “>” button
instead of the “M” button. The memory will not be
erased.

TMaintenance
Proper maintenance of your meter is recommended for the best results.
Storage
Replacing the Battery
The meter uses 1 CR 2032 3.0V coin cell battery. When the meter
displays the battery symbol ( ), this indicates that the battery is low and
you should replace the battery as soon as possible. If available, turn off
all wireless functions. The “E-6” error code means that the battery is
empty and you cannot use the meter until the battery is replaced.
How to replace the battery:
1.
2.
3.
4.
5.
Keep the test strip port area clean.
Keep the meter dry. Do not allow liquids to enter the housings. Avoid
extreme temperatures and humidity.
Do not leave the meter in your car.
Avoid dropping the meter. If you accidentally drop it, perform a quality
control test (see page 12) to verify the system is working properly.
Keep the meter and components away from children and pets.
Do not disassemble the meter. Disassembly will void the warranty.
Please follow the local regulations to dispose of the meter and battery
properly.
Make sure the meter is off.
Open the battery door on the back of the meter.
Remove the old battery and replace with a new CR 2032 3.0V coin
cell battery, making sure that the positive (+) side is facing up.
Replace the battery door.
After replacing the battery, you may need to re-enter some settings in
the settings menu (for more information, see page 8).
Cleaning
Glucose Meter
Under normal circumstances, no special cleaning is required. If cleaning
is needed, wear gloves before applying a mild detergent and water with
a soft damp cloth. Be careful not to allow water, dirt, blood, or control
solution inside of the meter. We recommend that you store the meter in
the carrying case after use.
Your blood glucose meter is a precision instrument. Please be careful.
Lancing Device
As needed, use a soft damp cloth moistened with soap and hot water to
wipe the surface. Do not immerse the lancing device in water.

TWarranty
TSymbol Index
TTroubleshooting
The meter has built-in messages to alert you of problems. When error
messages appear, note the error number, turn off the meter, and then
follow these instructions.
Display Causes Solution
Meter
does not
turn on
Battery may be damaged or
not have enough charge
Meter is too cold
Code chip error
Test strip is wet,
contaminated, or used
Meter self test error
Sample applied to the test
strip too soon
Insufficient sample
Test strip was removed during
the test
Temperature exceeds the
normal operating range (5°C
~ 45°C)
The battery is low, but may be
used for another 20 tests (if
wireless function is disabled)
The battery has been fully
discharged - no more tests
are possible
The meter has recorded a
level that is higher than the
meter test range
The meter has recorded a
level that is lower than the
meter test range
Replace the battery.
Allow the meter to warm to
room temperature for 30
minutes, then use.
Insert or reinsert the code chip
Make sure that the strip has not
expired, is not damp, or used. If
the test strip is ok, remove the
battery for 30 seconds, and
retest with a new strip. If the
problem persists, contact
customer service.
Retest with a new strip. Make
sure there is enough blood to fill
the test window.
Repeat the test and ensure test
strip remains in place.
Move to a place within the
normal operating temperature
range and repeat the test.
The next 20 results will still be
accurate, but replace the battery
as soon as possible.
Replace the battery and repeat
the test.
Repeat test. If you see HI
again, contact your doctor
immediately.
Repeat test. If you see LO
again, contact your doctor
immediately.
Consult instructions for use
Use by
Manufacturer
Code Number
Serial Number
1 Time Use
For in vitro diagnostic use
Lot Number
DC Power
Control Range
Do not dispose with
household waste
Biohazard
Please complete the warranty card that came with this product and mail
it to your dealer to register your purchase. Please keep your sales
receipt and other related purchase documents. If the meter fails for any
reason other than obvious abuse within the warranty period, we will
repair or replace it free of charge. Please note the date of purchase.
Date of Purchase:
Note: The warranty covers only the blood glucose meter and does not
include the battery
This manual suits for next models
1
Table of contents