ZETLab ZET 7152-N-VER.3 User manual

SEISMOGRAPH
ZET 7152-N-VER.3
USER MANUAL
LLC "ETMS"

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
2
Table of contents
1Description of ZET 7152-N-VER.3....................................................................................4
1.1 Functions......................................................................................................................4
1.2 Technical specifications ...............................................................................................5
1.3 Delivery scope of ZET 7152-N-VER.3 .........................................................................6
1.4 Construction of ZET 7152-N-VER.3 ............................................................................7
2ZET 7152-N-VER.3: setting-up procedures...................................................................10
2.1 Unboxing....................................................................................................................10
2.2 Integrated battery: operating rules............................................................................11
2.3 Installation of ZETLAB software to PC .....................................................................12
2.4 Connecting ZET 7152-N-VER.3 seismograph to PC ................................................13
2.5 Configuring ZET 7176 interface converter................................................................17
2.6 Configuring sensors included into the scope of ZET 7152-N-VER.3.........................19
2.6.1 Configuring digital short-period seismometer ZET 7156..........................................21
2.6.2 Configuring digital accelerometer ZET 7152-N........................................................22
2.6.3 Configuring off-line recorder ZET 7173....................................................................23
2.6.4 Configuring synchronization module ZET 7175........................................................25
3Operation of ZET 7152-N-VER.3 ....................................................................................26
3.1 Mounting of ZET 7152-N-VER.3: off-line use............................................................26
3.2 Mounting of ZET 7152-N-VER.3: stationary use ..........................................................26
3.3 Using ZET 7152-N-VER.3 in off-line mode................................................................27
3.4 Using ZET 7152-N-VER.3 in stationary mode...........................................................28
3.5 Copying the signals from ZET 7152-N-VER.3 memory to PC...................................29
4Possible faults, troubleshooting........................................................................................32
5Technical maintenance......................................................................................................33
6Storage and transportation requirements.......................................................................34
Annex А...................................................................................................................................35
Record of changes...................................................................................................................37

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
3
Introduction
The present User Manual describes structure and operating principles of ZET 7152-N-
VER.3, its operation rules as well as recommendations regarding its setting-up, start, maintenance,
operation, transportation and storage.
Only qualified engineers and technicians can use ZET 7152-N-VER.3.
Unboxing, installation and start of ZET 7152-N-VER.3 can be performed by the user or by
representative of organization responsible for technical maintenance in accordance with the
provisions of the Contract relating to the purchase of ZET 7152-N-VER.3.
The manufacturer hereby reserves the right to introduce changes into design and structure of
ZET 7152-N-VER.3, which do not affect technical characteristics and do not require introducing
any amendments into the technical documentation.
It is necessary to observe the guidelines of the present manual and other documents supplied
together with ZET 7152-N-VER.3 at all stages of its operation.
The present user manual and product certificate are included into ZET 7152-N-VER.3
delivery scope and should be used together with ZET 7152-N-VER.3.
To configure ZET 7152-N-VER.3 as well as to perform registered data analysis, one can use
ZETLAB Software, which is delivered together with ZET 7152-N-VER.3. The software is to be
installed on PC, which is going to be used together with ZET 7152-N-VER.3.
For user’s convenience, ZETLAB Software has a description. It can be activated by “F1”
key when ZET 7152-N-VER.3 is used.
List of abbreviations used
AWS - automated workstation.

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
4
1Description of ZET 7152-N-VER.3
1.1 Functions
ZET 7152-N-VER.3 is used for seismic impact parameters measurements. ZET 7152-N-
VER.3 can be used off-line or within the scope of automated monitoring system for the purpose
of seismic and seismological research by means of reflected and refracted waves method and
converted waves method. ZET 7152-N-VER.3 can also be used for geotechnical studies.

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
5
1.2 Technical specifications
Technical specifications of ZET 7152-N-VER.3 are shown in Table 1.1.
Table 1.1 Technical specifications of the
Characteristics
Values
Measuring channels specifications
Amount of measuring channels
6
Measuring axes direction
Measuring axis Z is orthogonal to the base and
is forwarded towards the lid of the
seismograph, while axes X and Y –are
parallel to the base
Nominal range of operating frequencies by
measuring axes X, Y, Z:
from 0,3 up to 400 Hz
Relative measurement error for axes X, Y, Z,
max
6 % (ZET 7152-N)
10 % (ZET 7156)
Max. measured value
±8 g (ZET 7152-N)
20 mm/s (ZET 7156)
Noise factor in operating frequencies range,
max
0,2 mg (ZET 7152-N)
0,0001 mm/s (ZET 7156)
Sampling rate by channels
30, 125, 250, 500, 1000, 2000 Hz (ZET 7152-
N)
1, 8, 16, 32, 64, 128 Hz (ZET 7156)
Primary transducer control
Integrated electrical excitation system of
sensing elements with actuator
yes
Primary transducer power control
yes
Off-line mode
Off-line mode operation time, max
18 h
Battery type
lead
Integrated non-volatile memory volume
32 Gb
Time of data recording to integrated SD-card at
500 Hz sampling frequency
Over 1000 h
Synchronization parameters
Synchronization types
GPS or PTP (IEEE1588)
Clock oscillator accuracy
0,1 ррm
General specifications
Power voltage
16-27 V
Consumed power
4 W
Data interface
Ethernet and USB 2.0
Dust and water resistance rating
IP65
Average lifetime
10 years
Weight (without control panel), max
6,5 kg
Dimensions (without control panel), max
base diameter: 160 mm
height: 260 mm
Operating temperature range
-10 … +50 °С

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
6
Guaranteed service life1
10 years
1.3 Delivery scope of ZET 7152-N-VER.3
Delivery scope of ZET 7152-N-VER.3 is shown in Table 1.2.
Table 1.2 Delivery scope
Item
Title
Amount
ZET 7152-N-VER.3
1 pc
HighSpeed USB 2.0 Cable
1 pc
Patch Cord UTP Cable (Cat. 5е)
1 pc
GPS/GLONASS receiver Trimble
1 pc
Battery charger SADP-65KB (220V → 19V)
1 pc
Pins for ground mounting (for off-line product version)
3 pcs
Mounting kit (for stationary use): mounting plates #1,2,
set of mounting screws, junctions box
1 kit
ZETKEY
1 pc.
CD with ZETLAB software
1 pc
Product certificate
1 pc
User manual*
1 pc
* - it is allowed to provide a single copy of User Manual for products lot up to 10 pcs.
1
The guaranteed service life is only valid in the case of ZETLAB software updates at least once in two months. The
warranty does not apply to the integrated battery of ZET 7152-N-VER.3

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
8
Figure 1.1 Overall view of ZET 7152-N-VER.3
ZET 7152-N-VER.3 is a complex multi-functional measuring complex, which includes:
1. ZET 7156 digital short-period seismometer
Digital short-period seismometer ZET 7156 is used for vibration velocity measurements of
high structures (bearing and enclosure structures) and underground structures (foundations, piles,
etc.). The seismograph consists of three identical sensing elements and a measuring module. The
sensing elements detect low-frequency mechanical oscillations by three mutually transverse axes
(X, Y and Z) and convert them into analog signal. The measuring module, in its turn, performs
digital processing of the analog signals and transfers digital data via CAN 2.0 interface.
2. Digital accelerometer ZET 7152-N
Digital accelerometer ZET 7152-N is used for measurements and conversion of acceleration
into a digital signal and further transfer of the data in digital format. The accelerometer consists of
a vibration sensor with integrated triaxial sensing element, which converts acceleration value into
a digital code by three mutually transverse axes X, Y and Z.
3. Off-line recorder ZET 7173
Off-line recorder ZET 7173 is used for recording the data obtained from digital sensors of
ZET 7152-N-VER.3. Data recording starts automatically as ZET 7152-N-VER.3 is switched on.
The data is saved to a microSD card as files in special format.
4. Synchronization module ZET 7175

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
9
Synchronization module ZET 7175 is used for synchronization of sensors time with global
time.
5. Interface converter ZET 7176
Interface converter ZET 7176 is used to connect the digital sensors of ZET 7152-N-VER.3
to PC via Ethernet interface as well as to configure digital sensors parameters.
6. Integrated battery
ZET 7152-N-VER.3 has an integrated battery allowing to record the registered signals to
non-volatile memory for further analysis by means of PC. ZET 7152-N-VER.3 data is transmitted
via Ethernet in real-time mode. Battery capacity enables measurements during 18 hours.
Meanings of indicators located at the lid of ZET 7152-N-VER.3 are shown in Table 1.3.
Table 1.3 Meanings of indicators
Indicator name
Indicator
color
Indicator state
Event
Battery
Red
On
It is necessary to charge the battery.
Blue
On
Battery is charging.
Green
On
Battery is charged (The indicator is
active only when ZET 7152-N-
VER.3 is switched off).
REC
Orange
On
Data recording to a memory card or
data transfer to industrial PC via
USB / Ethernet.
LAN
Green
On
ZET 7152-N-VER.3 is connected to
PC via Ethernet interface.
Sync
Blue
Flashes
ZET 7152-N-VER.3
synchronization by GPS or PTP.
Error
Red
On
Error.

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
10
At the lid of ZET 7152-N-VER.3, there are ports for connecting external devices.
Information on ports functions is shown in Table 1.4.
Table 1.4 Functions of ports
Icon
Title
Function
USB
Transfer of the recorded data to PC via USB interface.
Ethernet
Connection of ZET 7152-N-VER.3 to industrial PC via local
Ethernet network.
Charge
Connection of charging device SADP-65KB to ZET 7152-N-
VER.3.
Sync
Connection of external aerial GPS/GLONASS Trimble to the
ZET 7152-N-VER.3.
«Power»key is used to switch ZET 7152-N-VER.3 on/off. In order to switch ZET 7152-N-
VER.3 on, press and hold “Power” key for at least 3 seconds. To switch ZET 7152-N-VER.3 off,
click the “Power” key.

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
11
2ZET 7152-N-VER.3: setting-up procedures
2.1 Unboxing
Unboxing should be performed on a horizontal stable surface, after that one should:
Check the completeness of ZET 7152-N-VER.3 in accordance with the items specified in
Table 1.2 (see Clause Ошибка! Источник ссылки не найден.);
Perform visual examination, check ZET 7152-N-VER.3 for any mechanical damage.
Recommendation: In the case if there is a room for storing of ZET 7152-N-VER.3, it is
recommended to save the package. If ZET 7152-N-VER.3 is transported, it should be placed into
the package, so that to avoid damage.

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
12
2.2 Integrated battery: operating rules
ZET 7152-N-VER.3 has lead battery (5 Ah capacity). The battery has leak-proof package
and gas recombination system. Table 2.1 shows battery operational requirements.
Table 2.1 Battery operational requirements
Parameter
Value
Storage temperature range, ºС
-35…50
Operating temperature range, ºС
-10…50
Battery charging should be performed in the following sequence:
1. Connect the charging device (included into delivery scope) to the “Charge”port of
ZET 7152-N-VER.3 panel. Connect the charging device to 220 V AC network.
2. During battery charging the “Battery” indicator should have blue color.
3. Green color of “Battery” indicator means that the battery charging process is complete.
Upon charging process completion, disconnect charging device from ZET 7152-N-VER.3 and
220V AC.
Note! Battery replacement is not covered by warranty provisions and is to be performed
by the Manufacturer.
Attention! In order to extend battery service life in the case of long-term storage, it is
necessary to charge the battery at least once in 6 months.
Attention! It is allowed to use the battery at negative temperatures; however, it may lead to
a shorter time of off-line operation.

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
13
2.3 Installation of ZETLAB software to PC
In order to install ZETLAB software to PC, one should:
1. Put the Software CD (included into delivery scope) into the PC;
2. Run «ZetLab.msi», follow the instructions and complete installing ZETLAB Software
to the PC;
Attention! You should be logged-in as an administrator to install the software.
Note: the following system requirements should be met for proper operation of the software:
Dual-core processor or more;
Core speed - over 1,6 GHz;
RAM - over 2 Gb;
HD free space –over 20 Gb;
Video-controller: 3D-graphics acceleration, support of ОреnGL, DirectХ,
memory: over 128 Mb;
Resolution: min. 1280×1024;
Mouse or any other pointing device;
Standard keyboard or any other entry device;
CD-ROM for programs installation;
Supported OS versions:
o Microsoft® Windows® 7 32/64 with SP1;
o Microsoft® Windows® 8 32/64;
o Microsoft® Windows® 8.1 32/64;
o Microsoft® Windows® 10 32/64.

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
14
2.4 Connecting ZET 7152-N-VER.3 seismograph to PC
The PC should have Windows OS and ZETLAB software installed. It is also necessary to
connect ZETKEY (which is included into delivery scope) to USB port.
ZET 7152-N-VER.3 is connected to PC via Ethernet interface. In order to connect the
ZET 7152-N-VER.3 to PC via Ethernet interface, connect the cable PatchCord UTP (Cat. 5е) to
Ethernet port of ZET 7152-N-VER.3 and free Ethernet port of the PC. As ZET 7152-N-VER.3 is
connected to the PC, enable “Power” key at the top panel of ZET 7152-N-VER.3. The system will
find and install the corresponding driver.
IP address of the PC should be in the same network with IP address of interface converter
ZET 7176 (included into seismograph). Local network should not block UDP (multicast).
Note: By default, interface converter has IP-address 192.168.1.76 with mask 255.255.255.0.
Configuring of the devices included into the scope of ZET 7152-N-VER.3 is performed in
“Device manager” program from “Service” menu of ZETLAB panel. (Figure 2.1).
Figure 2.1 ZETLAB panel

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
15
The program “ZET Device manager” will search for available devices in the local network
and depict them in the list of devices (Figure 2.2).
Figure 2.2 “ZET Device manager”: list of devices
The devices in the list are shown in grey color, which means that they are available, but not
active. In order to connect to the interface converter, activate it from the menu with right click on
its serial number. (Figure. 2.3).
Figure. 2.3 Connecting to the interface converter
Note: Serial number of interface converter is specified in its product certificate.
During connection process, the device name will be highlighted with bold type. If the device
name is depicted in bold type for a long time, it means that it is impossible to establish connection,
Main reasons for that are as follows:
IP-address of the PC connected to ZET 7176 is not in the same network with IP-
address of the interface converter;

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
16
There is a conflict of IP-addresses: in the local network there is a device with IP-
address, identical with that of the interface converter.
In order to check current IP-address of the interface converter, enter the menu (with right
click on the interface converter), enter “Properties” menu and go to “Device” tab (Figure 2.4). The
parameter «IPv4 Address» will indicate the current IP-address of the interface converter.
Figure 2.4 “Device” tab

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
17
As the connection is established, there will be formed a list of digital sensors connected to
the interface converter (Figure 2.5).
Figure 2.5 Active ZET 7176 and corresponding list of the sensors connected
The left section of the window depicts tree of devices, that are connected to the PC. The first
level depicts interface converters and devices, that are connected to PC. The second level shows
sensors, that are connected to a particular interface converter. To the right from the digital sensor
one can see device address in the measuring network (Figure 2.6).
Figure 2.6 Addresses of the devices in the measuring network
In the case if detailed display mode is enabled, the right section of the window will display
basic parameters of measuring channels.
In order to select a sensor to be configured, double-click its name (more information is
available in “ZETLAB Software. User manual”).

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
18
2.5 Configuring ZET 7176 interface converter
In order to configure interface converter ZET 7176, one should:
Set the network address in accordance with applicable requirements in the fields “Ipv4
address”, “Subnet mask”, “Default gateway” of “Ethernet” tab (Figure 2.7).
Figure 2.7 «Ethernet» tab

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
19
Select the bit rate in accordance with the applicable requirements in the “Bit rate, kbps”
field of “CAN” tab (Figure 2.8).
Figure 2.8 «CAN» tab.
Note: After changing the bit rate of interface converter it is necessary to cycle power ZET
7152-N-VER.3. The system will automatically change the bit rate of digital sensors, which are
connected to the interface converter.
Note: additional information regarding configuration of ZET 7176 interface converter is
available in the document “ZET 7176 User Manual”.

LLC "ETMS"
Tel./Fax.: +7 (495) 739-39-19, www.zetlab.com
20
2.6 Configuring sensors included into the scope of ZET 7152-N-VER.3
Before starting to use ZET 7152-N-VER.3, it is necessary to configure corresponding digital
sensors.
Note: primary configuration of ZET 7152-N-VER.3 is performed by the manufacturer.
Please, note that in the field “Node 2 ÷ 63” of “Information” tab there should be set a unique
address of each device of the measuring network. All the devices within the scope of the
seismograph should have unique address –it is an obligatory condition for normal operation of
ZET 7152-N-VER.3. The nodes are set in the range from 3 to 63. Figure 2.9 shows an example of
“Information” tab of digital accelerometer ZET 7152-N with «3» node set.
Figure 2.9 “Information” tab
Table of contents
Other ZETLab Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

USAG
USAG 960 D manual

janitza
janitza 800-CT8-LP installation manual

Union Instruments
Union Instruments INCA4003 Translation of the original operating instructions

Titan Logix
Titan Logix TD80 Application note

ACR Electronics
ACR Electronics Vecta 3 Product support manual

B-I Inductrial
B-I Inductrial BI-SM2004 manual