Hanna Instruments HI702 User manual

HI702
Copper High Range
INSTRUCTION MANUAL
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Tips for an Accurate Measurement
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Battery Management
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Operation of
these instruments may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct such interferences. Any variation
introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance. To avoid damages or burns, do not put the instrument in microwave oven. For
yours and the instrument safety do not use or store the instrument in hazardous environments.
Recommendations for Users
Accessories
Reagent Sets
HI702-25 Reagents for 25 Copper High Range tests
Other Accessories
HI702-11 Copper HR Certified Standard Kit
HI731225 Cuvette Black Cap for Checker®HC Colorimeters (4 pcs.)
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731321 Glass cuvettes and Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI731353 Cuvette Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI93703-50 Cuvette cleaning solution (230 mL)
IST702 05/18
HI702 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Warranty
• Turn the meter on by pressing the button. All
segments are displayed. When the display
shows “Add”, “C.1” with “Press” blinking,
the meter is ready.
• Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
• Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the meter
is zeroed.
• Remove the cuvette from the meter and
unscrew the cap. Add the content of one
packet of HI702-0 reagent. Replace the cap
and shake gently for about 15 seconds. Place
the cuvette back into the meter.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 45 seconds and press
the button.
• The instrument displays the concentration of
copper in ppm. The meter automatically turns
off after 10 minutes.
Measurement Procedure
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.00” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you use
the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the meter.
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Functional Description
Errors and Warnings
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction
manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit
www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected].
Find your local Hanna Instruments Office on www.hannainst.com.
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI702 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 6 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Preliminary Examination
Range 0.00 to 5.00 ppm
Resolution 0.01 ppm
Accuracy ±0.05 ppm ±5% of reading @25 °C/77 °F
Light Source Light Emitting Diode @575 nm
Light Detector Silicon Photocell
Method
Adaptation of the EPA approved method. The reaction between
copper and the bicinchoninate reagent causes a purple tint in the
sample.
Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condesing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 10 minutes of non-use
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Specifications

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 4105 User manual

Hanna Instruments

Hanna Instruments HI 4105 User manual

Hanna Instruments pH 21 User manual

Hanna Instruments

Hanna Instruments pH 21 User manual

Hanna Instruments HI 721 User manual

Hanna Instruments

Hanna Instruments HI 721 User manual

Hanna Instruments HI98165 User manual

Hanna Instruments

Hanna Instruments HI98165 User manual

Hanna Instruments HI97745 User manual

Hanna Instruments

Hanna Instruments HI97745 User manual

Hanna Instruments TDS 1 User manual

Hanna Instruments

Hanna Instruments TDS 1 User manual

Hanna Instruments HI98163 User manual

Hanna Instruments

Hanna Instruments HI98163 User manual

Hanna Instruments HI 93749 User manual

Hanna Instruments

Hanna Instruments HI 93749 User manual

Hanna Instruments HI 9023C User manual

Hanna Instruments

Hanna Instruments HI 9023C User manual

Hanna Instruments HALO 2 HI9810402 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810402 User manual

Hanna Instruments HI 8424NEW User manual

Hanna Instruments

Hanna Instruments HI 8424NEW User manual

Hanna Instruments HI 84181 User manual

Hanna Instruments

Hanna Instruments HI 84181 User manual

Hanna Instruments HI769828 User manual

Hanna Instruments

Hanna Instruments HI769828 User manual

Hanna Instruments HI97700 User manual

Hanna Instruments

Hanna Instruments HI97700 User manual

Hanna Instruments edge EC HI2003 User manual

Hanna Instruments

Hanna Instruments edge EC HI2003 User manual

Hanna Instruments HI 931100 User manual

Hanna Instruments

Hanna Instruments HI 931100 User manual

Hanna Instruments HI 701 User manual

Hanna Instruments

Hanna Instruments HI 701 User manual

Hanna Instruments HI97739 User manual

Hanna Instruments

Hanna Instruments HI97739 User manual

Hanna Instruments HI 98703 User manual

Hanna Instruments

Hanna Instruments HI 98703 User manual

Hanna Instruments HI83326 User manual

Hanna Instruments

Hanna Instruments HI83326 User manual

Hanna Instruments HI97101 User manual

Hanna Instruments

Hanna Instruments HI97101 User manual

Hanna Instruments HI 96762C User manual

Hanna Instruments

Hanna Instruments HI 96762C User manual

Hanna Instruments HI 99181 User manual

Hanna Instruments

Hanna Instruments HI 99181 User manual

Hanna Instruments HI96708C User manual

Hanna Instruments

Hanna Instruments HI96708C User manual

Popular Measuring Instrument manuals by other brands

Vutlan HAT-100Q1 Quick installation manual

Vutlan

Vutlan HAT-100Q1 Quick installation manual

ITRON Intelis Setup guide

ITRON

ITRON Intelis Setup guide

BRUEL & KJAER 2203 Instructions and applications

BRUEL & KJAER

BRUEL & KJAER 2203 Instructions and applications

Sequoia LISST-VSF user manual

Sequoia

Sequoia LISST-VSF user manual

TTI 1604 Service manual

TTI

TTI 1604 Service manual

Goobay 67000 instruction manual

Goobay

Goobay 67000 instruction manual

Mitchell Instrument WENS 900C user manual

Mitchell Instrument

Mitchell Instrument WENS 900C user manual

Syco SWR quick start guide

Syco

Syco SWR quick start guide

Pacific Data Systems DPM-RT-2 operating manual

Pacific Data Systems

Pacific Data Systems DPM-RT-2 operating manual

YOKOGAWA Power Series Plus IM 2493WVP-1 user manual

YOKOGAWA

YOKOGAWA Power Series Plus IM 2493WVP-1 user manual

Lutron Electronics MHT-381SD Operation manual

Lutron Electronics

Lutron Electronics MHT-381SD Operation manual

Myzone Inbody 230 Setup instructions

Myzone

Myzone Inbody 230 Setup instructions

Endress+Hauser Prosonic T FMU 130 operating manual

Endress+Hauser

Endress+Hauser Prosonic T FMU 130 operating manual

XS Instruments Temp70 PT100 vio instruction manual

XS Instruments

XS Instruments Temp70 PT100 vio instruction manual

RadComm RC2PLUS user manual

RadComm

RadComm RC2PLUS user manual

ZBM MINI OTDR user manual

ZBM

ZBM MINI OTDR user manual

Dostmann Electronic P705 Operation manuals

Dostmann Electronic

Dostmann Electronic P705 Operation manuals

Simrad PI50 -  REV A installation manual

Simrad

Simrad PI50 - REV A installation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.