Titan Logix TD80 Installation and operating instructions

Application Note
TPM 060, Rev 1.0
September 2018
Page
1
of
11
TD80 to TD100
Retrofit Procedures
TPM 060

Application Note
TPM 060, Rev 1.0
September 2018
Page
2
of
11
Table of Contents
1. Customer Support ................................................................................................................................................. 3
2. Returns For Repair, Replacement, or Cre it ......................................................................................................... 3
3. Intro uction .......................................................................................................................................................... 4
3.1 Disclaimer ............................................................................................................................................. 4
3.2 Safety .................................................................................................................................................... 4
4. Pre-Installation ...................................................................................................................................................... 4
4.1 Require Materials ............................................................................................................................... 4
4.2 Initial Steps ............................................................................................................................................ 5
5. Retrofit Proce ures with flui in the tank ............................................................................................................ 6
5.1 Retrieve the TD80 strap table ............................................................................................................... 6
5.2 Exchange the Transmitter ..................................................................................................................... 6
5.3 Transferring the TD80 strap table to the TD100 ................................................................................... 6
5.4 Performing Offset Calibration on the TD100 ........................................................................................ 7
5.5 Confirming Offset Calibration on the TD100 ........................................................................................ 7
6. Retrofit Proce ures with empty tank ................................................................................................................... 8
6.1 Retrieve the TD80 strap table ............................................................................................................... 8
6.2 Exchanging the transmitter .................................................................................................................. 8
6.3 Transferring the TD80 strap table to the TD100 ................................................................................... 9
6.4 Performing Offset Calibration on the TD100 ...................................................................................... 10
6.5 Installing the TD100 ............................................................................................................................ 11
6.6 Confirming Level Rea ings of the TD100 ............................................................................................ 11
7. Revision History .................................................................................................................................................. 11

Application Note
TPM 060, Rev 1.0
September 2018
Page
3
of
11
1 Customer Support
24 Hour Technical Support Line
1-877-462-4085
Titan Logix provi es 24-hour technical support for their pro ucts. Call the technical support number:
•to arrange a service call
•if you have imme iate questions regar ing operations an installation
•if you have a request for ocumentation an /or software
•technical assistance
2 Returns For Repair, Replacement, or Credit
Inside Sales (8:00-4:30 Mountain Daylight Time)
1-877-462-4085
NOTE
Returned product for repair, replacement, or credit, must be accompanied by a Titan
Logix generated RMA number
To obtain an RMA number:
1. Call the Insi e Sales number to speak with a representative an obtain an RMA number
2. Provi e the serial number of the item with a escription of issue an expecte resolution (repair,
replacement, return)
3. Describe the issue an in icate whether the item is efective, missing components, not suitable for the
inten e use, or purchase in error
4. Return the item together with the RMA number to the in icate Titan Logix service facility
When speaking with your Titan Logix representative, please indicate whether the item is still under warranty
Shipping cost to an from a Titan Logix service facility is the responsibility of the customer unless the pro uct is
un er warranty. If the pro uct is un er warranty, the freight cost to Titan Logix is incurre by the customer. From
Titan Logix, the cost is incurre by Titan Logix.
All returns for cre it are subject to a restocking fee.

Application Note
TPM 060, Rev 1.0
September 2018
Page
4
of
11
3 Introduction
This ocument escribes the steps that shoul be followe to ensure a successful replacement of the TD80
transmitter with a TD100 transmitter where there may or may not be flui in the tank. This inclu es copying the
strapping table an calibration ata from the TD80 to the TD100.
3 1 Disclaimer
The information in this ocument is subject to change without notice. Titan Logix Corp. makes no representations
or warranties with respect to the contents hereof.
3 2 Safety
This manual will use the following stan ar safety terms an conventions to in icate con itions:
CAUTION
Indicates a hazardous situation which, if not avoided, could result in injury and/or
property damage
NOTE
Indicates an important message not related to personal injury or property damage
The TD100 system must be installe an operate in accor ance with etails escribe in the Titan Logix manuals,
application notes, an all other relevant publications. Only qualifie personnel familiar with the installation an
operation of this equipment shoul install, a just, operate, or service this equipment. Failure to observe this
warning coul result in bo ily injury an or property amage.
Do not install or repair the system where flammable gases an /or fumes may be present.
Follow all fe eral, state/provincial an local safety stan ar s an in ustry recommen e practices for the vehicle.
For example, the Fe eral Motor Vehicle Safety Stan ar s (FMVSS) an the American Trucking Association (ATA)
Technology an Maintenance Council (TMC) Recommen e Practices (RP).
4 Pre-Installation
4 1 Required Materials
•Titan Logix configuration software, which can be ownloa e from the Titan Support portal
(http://support.titanlogix.com/).
oSensorLink software version 2.0 or greater.
TPM 058 SensorLink User Manual V1.1 or greater.
oFinch II firmware up ate utility program.
oFinch II configuration utility program.
•SV Bus Programming Kit that contains the SV-RS232F SV to serial converter

Application Note
TPM 060, Rev 1.0
September 2018
Page
5
of
11
•A 122cm to 152cm (4 to 5 feet) long Dual-Ro , measure from the top of the probe, with shorting
block is require if either the tank is empty, or the level of the flui in the tank cannot be guarantee
to be within the specifie range.
•Han tools an cables.
4 2 Initial Steps
Prior to removing the TD80, etermine if flui exists in the tank, an if so, to what level. If the level of flui in the
tank is between 25cm, approximately 10 inches, above the top of the shorting block an 75cm approximately 30
inches, below the fi ucial procee to section “Retrofit Proce ures with flui in the tank”.
If the level of flui oes not fall within this range or is empty, procee to section “Retrofit Proce ures with empty
tank”.

Application Note
TPM 060, Rev 1.0
September 2018
Page
6
of
11
5 Retrofit Procedures with fluid in the tank
5 1 Retrieve the TD80 strap table
1. Ensure power is applie to the Finch II/TD80 system.
2. Recor the volume as shown on the Finch II isplay.
3. Take note if there is a sump an /or riser on the tank. Recor their correspon ing values.
4. Rea the strap table off the TD80 using SensorLink utility program.
a. Information relating to the use of the SensorLink utility program can be foun in TPM 058
SensorLink User Manual.
5. Save the TD80 strap table as a file for future reference.
5 2 Exchange the Transmitter
1. Power own the TD80 an Finch II an procee to replace the TD80 with TD100.
2. Ensure proper electrical connections to the TD100 have been ma e.
5 3 Transferring the TD80 strap table to the TD100
1. Run the SensorLink utility program an follow the steps below.
a. Open the previously save TD80 strap table file.
b. Un er Step 2:
i. Press the “E it” button (on the left-han pane).
c. Un er Step 1:
i. Change the “Transmitter Type” from TD80 to TD100.
ii. Set the “Probe Type” to Dual-Ro probe.
iii. The “Depth Units” an the “Volume Units” will always show inches an Gallon (US)
respectively when rea ing back the strap table from a TD80 regar less. Set the correct
epth an volume units for the TD100.
. Un er Step 3:
i. A just the “HH level” threshol accor ingly bearing in min that the TD80 an TD100
have ifferent minimum istance requirements between HH an Spill threshol .
ii. A just the “Display Precision” as necessary.
e. Un er Step 4:
i. Set the sump an /or riser as previously recor e , if present.
f. From the top menu choose “File” then “Save As” an save the strap table to a new file as a
TD100 strap table.

Application Note
TPM 060, Rev 1.0
September 2018
Page
7
of
11
g. From the top menu choose “Tools” then “Program Table” or un er Step 5, click on “Program”
to program the TD100 with the newly create TD100 strap table.
h. After successfully programming the strap table, power own the TD100 an Finch II an
procee to section 5.4.
5 4 Performing Offset Calibration on the TD100
1. At this point it is assume the TD80 transmitter has been remove an replace with a TD100
transmitter.
2. Apply power to the Finch II / TD100 system. DO NOT INITIATE OFFSET CALIBRATION YET.
3. At the Finch II, select the compartment to be calibrate by pressing the “COMP SELECT” button. Verify
that the Finch II isplays a volume (shoul not show any errors); procce to either step 4-5 or step 6.
4. Press an hol the following four buttons at the same time:
a. COMP SELECT
b. HORN BYPASS
c. ACK
. DIM SLEEP
5. Continue to hol own the four buttons until the isplay flashes “CAL”, then release the buttons.
6. Alternately, the Finch II can enter Calibration mo e by hol ing either ▲ or ▼ arrow while powering the
isplay.
7. The Finch II will then show the current volume of liqui in the tank.
8. Change the Offset Calibration to match the volume base on the previously recor e volume from the
TD80 transmitter by either pressing the ▲ or ▼ arrow on the Finch II.
9. Power cycle the Finch II / TD100 system.
10. The TD100 shoul now contain the same strapping table as store in the TD80 an an Offset
Calibration has been applie .
5 5 Confirming Offset Calibration on the TD100
1. Power cycle the Finch II/TD100 system, wait until a level is isplaye .
2. Confirm that the offset calibration has been applie . The level isplaye shoul match the previously
recor e volume from the TD80 transmitter.
3. Power cycle the Finch II/TD100 system one more time. Observe the level isplaye on the Finch II. It
shoul be consistent an shoul match the previously recor e volume from the TD80 transmitter.
4. Repeat Offset Calibration if the isplay is inconsistent each time a power cycle occurs or is not accurate.

Application Note
TPM 060, Rev 1.0
September 2018
Page
8
of
11
5. As an a itional test, if possible, a flui or rain the flui from the tank an observe the isplay on
the Finch II. The isplay shoul change correspon ingly – ecreasing when raining an increasing
when a ing flui to the tank.
6 Retrofit Procedures with empty tank
6 1 Retrieve the TD80 strap table
1. Ensure power is applie to the Finch II/TD80 system.
2. Take note if there is a sump an /or riser on the tank. Recor their correspon ing values.
3. Rea the strap table off the TD80 using SensorLink utility program.
a. Information relating to the use of the SensorLink utility program can be foun in TPM 058
SensorLink User Manual.
4. Save the TD80 strap table as a file for future reference.
5. Remove power from the Finch II/TD80 system.
6 2 Exchanging the transmitter
1. An external Dual-Ro probe is require for this proce ure. Place an secure the shorting block of the
Dual-Ro probe at the location about 32 inches or 81cm from the fi ucial.

Application Note
TPM 060, Rev 1.0
September 2018
Page
9
of
11
2. Remove the TD80 transmitter from existing installation an connect it to the external Dual-Ro probe
with all electrical cabling in place.
3. Power up the TD80 transmitter an recor the volume as shown on the Finch II. Note that any objects
close to the probe may affect the rea ing.
4. Power own the system.
5. Remove the TD80 transmitter an connect the TD100 transmitter to the external Dual-Ro probe as
well as move all TD80 cabling to the TD100.
6 3 Transferring the TD80 strap table to the TD100
1. Run the SensorLink utility program an follow the steps below.
a. Open the previously save TD80 strap table file.
b. Un er Step 2:
i. Press the “E it” button (on the left-han pane).
c. Un er Step 1:
i. Change the “Transmitter Type” from TD80 to TD100.

Application Note
TPM 060, Rev 1.0
September 2018
Page
10
of
11
ii. Set the “Probe Type” to Dual-Ro probe.
iii. The “Depth Units” an the “Volume Units” will always show inches an Gallon (US)
respectively when rea ing back the strap table from a TD80 regar less. Set the correct
epth an volume units for the TD100.
. Un er Step 3:
i. A just the “HH level” threshol accor ingly bearing in min that the TD80 an TD100
have ifferent minimum istance requirements between HH an Spill threshol .
ii. A just the “Display Precision” as necessary.
e. Un er Step 4:
i. Set the sump an /or riser as previously recor e , if present.
f. From the top menu choose “File” then “Save As” an save the strap table to a new file as a
TD100 strap table.
g. From the top menu choose “Tools” then “Program Table” or un er Step 5, click on “Program”
to program the TD100 with the newly create TD100 strap table.
h. After successfully programming the strap table, power own the TD100 an Finch II an
procee to section 6.4.
6 4 Performing Offset Calibration on the TD100
1. At this point it is assume the TD100 transmitter is connecte to the external Dual-Ro probe.
2. Apply power to the Finch II / TD100 system. DO NOT INITIATE OFFSET CALIBRATION YET.
3. At the Finch II, select the compartment to be calibrate by pressing the “COMP SELECT” button. Verify
that the Finch II isplays a volume (shoul not show any errors); procce to either step 4-5 or step 6.
4. Press an hol the following four buttons at the same time:
a. COMP SELECT
b. HORN BYPASS
c. ACK
. DIM SLEEP
5. Continue to hol own the four buttons until the isplay flashes “CAL”, then release the buttons.
6. Alternately, the Finch II can enter Calibration mo e by hol ing either ▲ or ▼ arrow while powering the
isplay.
7. The Finch II will then show the current volume in the tank, as set by the position of the shorting block
on the external probe.
8. Change the Offset Calibration to match the volume base on the previously recor e volume from the
TD80 transmitter by either pressing the ▲ or ▼ arrow on the Finch II.
9. Power cycle the Finch II / TD100 system.

Application Note
TPM 060, Rev 1.0
September 2018
Page
11
of
11
10. The TD100 shoul now contain the same strapping table as store in the TD80 an an Offset
Calibration has been applie .
6 5 Installing the TD100
1. Detach the TD100 from the external Dual-Ro probe an install the TD100 onto the truck.
2. Power up the system an verify system is working. There shoul n’t be any error co es reporte by the
TD100
6 6 Confirming Level Readings of the TD100
1. Power cycle the Finch II/TD100 system, wait until a level is isplaye .
2. There shoul be no error isplaye on the Finch II. Since the tank is empty at this point, the isplay
shoul in icate 2Lo.
3. If it is safe to o so, run a metallic con uctor shorting the ual ro s at various points along the length
of the probe. The isplay shoul change accor ingly. Check the HH an Spill isplays near the top of
the probe.
4. Alternatively, if possible, fill the tank with liqui an confirm that the rea ing increases. Drain the tank
an confirm that the rea ing ecreases.
7 Revision History
Revision
Description of Change
Date
Author
1.0
Initial version for review
14SEP2018
ML
/EM
Other manuals for TD80
1
This manual suits for next models
1
Table of contents
Other Titan Logix Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Shenzhen Linshang Technology
Shenzhen Linshang Technology LS117 user manual

Neoss
Neoss NeoTell & SmartPegs Instructions for use

TS line
TS line TS010 operating instructions

McCrometer
McCrometer Ultra Mag INSTALLATION WIRING GUIDE

ADInstruments
ADInstruments AD3343 Operating instruction

Sokkia
Sokkia SET2CII Operator's manual