Zodiac eXO Series User manual

eXO®
(all models)
H0642500_REVA
INSTALLATION MANUAL
FOR YOUR SAFETY - This product must be installed and serviced by a licensed electrician in
accordance with SANS 10142-1 and local applicable regulations. Before installing this product, read
and follow all warning notices and instructions that accompany this product. Failure to follow warning
notices and instructions may result in property damage, personal injury, or death. Improper installation
and/or operation will void the warranty.
Improper installation and/or operation can create unwanted electrical hazard which can cause serious
injury, property damage, or death.
WARNING

Page 3
eXO®| Installation Manual
Page 2 eXO®| Installation Manual
For information on warranty and service in South Africa go to www.zodiac.co.za
Table of Contents
DATE OF INSTALLATION
INSTALLER INFORMATION
INITIAL PRESSURE GAUGE READING (WITH CLEAN FILTER)
PUMP MODEL HORSEPOWER
FILTER MODEL
CONTROL PANEL MODEL SERIAL NUMBER
NOTES:
EQUIPMENT INFORMATION RECORD
Section 1. Important Safety Instructions....... 4
Section 2. System Overview ........................... 6
2.1 Contents........................................................................6
2.2 Specifications ................................................................6
2.3 Dimensions....................................................................7
2.4 Materials and Tools ....................................................... 7
Section 3. Plumbing......................................... 8
Section 4. Install Controller ............................ 9
4.1 Mount the Bracket .........................................................9
4.2 Install Temperature Sensor ........................................... 9
4.3 Electrical Connections...................................................9
Section 5. Install the pH Link or Dual Link
Module on the Controller ............ 11
5.1 Install the POD Kit....................................................... 11
5.2 Drill the Pipe................................................................ 11
5.3 Install the POD ............................................................ 11
5.4 Install the POD Cap.....................................................12
5.5 Install and Calibrate the Sensor ..................................12
5.6 Install the pH Down Intake and Injection lines.............13
5.7 Calibration Procedure..................................................14
Section 6. Prepare and Balance the Pool .... 15
6.1 pH Regulation Principle...............................................15
6.2 Water Chemistry Table ................................................15
6.3 Stabilizer (Cyanuric Acid) ............................................15
Section 7. Operation...................................... 16
7.1 User Interface Overview..............................................16
7.2 Initial start-Up ..............................................................16
Section 8. Menu and Settings....................... 18
8.1 Filter Pump..................................................................18
8.2 Aux/Lights....................................................................19
8.3 pH Sensor Calibration .................................................20
8.4 Setpoint(s)...................................................................21
8.5 Prime and Test the pH Dosing Pump ..........................21
8.6 Programming/Timers...................................................22
8.7 Reversing ....................................................................24
8.8 Slave Mode .................................................................24
8.9 Help Menu...................................................................24
Section 9. Inspecting and
Cleaning The Electrode............... 25
Section 10. Winterizing.................................... 25
10.1 Winterize the pH Sensor ............................................. 25
10.2 Reopening the Pool.....................................................25
Section 11. Troubleshooting........................... 26

Page 3
eXO®| Installation Manual
Page 2 eXO®| Installation Manual
Section 1. Important Safety Instructions
READ AND FOLLOW ALL INSTRUCTIONS
All electrical work must be performed by a licensed electrician and conform to all national, state, and local
codes. When installing and using this electrical equipment, basic safety precautions should always be
followed, including the following:
WARNING
Prolonged immersion in hot water may induce hyperthermia. Hyperthermia occurs when the internal temperature
of the body reaches a level several degrees above the normal body temperature of 37°C. The symptoms of
hyperthermia include dizziness, fainting, drowsiness, lethargy, and an increase in the internal temperature of the
body. The effects of hyperthermia include: 1) unawareness of impending danger; 2) failure to perceive heat; 3)
failure to recognize the need to exit spa; 4) physical inability to exit spa; 5) fetal damage in pregnant women; 6)
unconsciousness resulting in a danger of drowning.
DANGER
To reduce the risk of severe injury or death, do not remove the suction fittings of your spa or hot tub. Never operate
a spa or hot tub if the suction fittings are broken or missing. Never replace a suction fitting with one rated less than
the flow rate marked on the equipment assembly.
WARNING
To Reduce the Risk of Injury -
a) The water in a spa should never exceed 40°C. Water temperatures should remain between 38°C and 40°C.
The water in a spa should never exceed 40°C. Water temperatures between 38°C and 40°C are considered
safe for a healthy adult. Lower water temperatures are recommended for young children and when spa use
exceeds 10 minutes.
b) Since excessive water temperatures have a high potential for causing fetal damage during the early months of
pregnancy, pregnant or possibly pregnant women should limit spa water temperatures to 38°C. Before entering
a spa or hot tub, the user should measure the water temperature with an accurate thermometer since the
tolerance of water temperature-regulating devices varies.
d) The use of alcohol, drugs, or medication before or during spa or hot tub use may lead to unconsciousness with
the possibility of drowning.
e) Obese persons and persons with a history of heart disease, low or high blood pressure, circulatory system
problems, or diabetes should consult a physician before using a spa.
f) Persons using medication should consult a physician before using a spa or hot tub since some medication
may induce drowsiness while other medication may affect heart rate, blood pressure, and circulation.
WARNING
Risk of electric shock - Install the controller at least 3.5 metres from the inside wall of the pool and/or hot tub
using non-metallic plumbing.
This appliance is not intended for use by persons (including children) with reduced physical, sensory or mental
capabilities, or lack of experience and knowledge, unless they have been given supervision or instruction
concerning use of the appliance by a person responsible for their safety.
Children should be supervised to ensure that they do not play with the appliance.
Children should not use spas or hot tubs without adult supervision.
Do not use spas or hot tubs unless all suction guards are installed to prevent body and hair entrapment.
WARNING
To avoid injury ensure that you use this control system to control only packaged pool/spa heaters which have built-
in operating and high limit controls to limit water temperature for pool/spa applications. This device should not be
relied upon as a safety limit control.

Page 5
eXO®| Installation Manual
Page 4 eXO®| Installation Manual
CAUTION
Power to the Aqualink TRI should be supplied by an isolating transformer or through a residual current device
(RCD) with a rated residual operating current not exceeding 30mA.
SAVE THESE INSTRUCTIONS
WARNING
People with infectious diseases should not use a spa or hot tub.
To avoid injury, exercise care when entering or exiting the spa or hot tub.
Do not use drugs or alcohol before or during the use of a spa or hot tub to avoid unconsciousness and possible
drowning.
Before entering a spa or hot tub, measure the water temperature with an accurate thermometer.
Do not use a spa or hot tub immediately following strenuous exercise.
Prolonged immersion in a spa or hot tub may be injurious to your health.
Do not permit any electric appliance (such as a light, telephone, radio, or television) within 3.5 metres of a spa or
hot tub.
The use of alcohol, drugs or medication can greatly increase the risk of fatal hyperthermia in hot tubs and spas.
Water temperature in excess of 38°C may be hazardous to your health.
Attention Installer: Install to provide drainage of compartment for electrical components.
WARNING
A terminal bar marked "GROUND" is provided within the controller. To reduce the risk of electrical shock which can
cause serious injury or death, connect this terminal bar to the grounding terminal of your electric service or supply
panel with a continuous copper conductor having green insulation and one that is equivalent in size to the circuit
conductors supplying this equipment in accordance with SANS 10142-1. In addition, where required, bonding
should be extended in accordance with SANS 10142-1 to any metal ladders, water pipes, or other metal within 3.5
m of the pool/spa.

Page 5
eXO®| Installation Manual
Page 4 eXO®| Installation Manual
Section 2. System Overview
2.1 Contents
Before starting, check that you have the correct
parts as indicated in Table 1. If any parts are
missing or incorrect, please call your local distributor
or technical support at 0860 887 665 for assistance.
2.1.1 Salt Chlorinator System
Clearwater, eXO®, Hydroxinator (all models)
a. Controller
b. Temperature Sensor
c. Wall Mount Bracket
d. Wall Mount Screws Ø7 mm (X2)
e. Wall Mount Anchor (X2)
f. Electrolytic Cell
g. Unions (X2)
h. Glue on Adapter (X2)
i. Adapter O-Rings (X2)
j. Pipe Reducer - 50mm, 60mm (X2),
k. Zip Ties (X7) - (not shown)
Table 1. Salt Chlorinator System Contents
2.2 Specifications
2.2.1 Salt Chlorinator System
Compact Mid Large
Nominal chlorine production 18 g/h 25 g/h 35 g/h
Nominal output amps 3.6 A 5.0 A 7.2 A
Required salt level 4 g/L - 33 g/L
Power supply voltage 240 VAC - 50 Hz
Electric power 200 W
Protection index IP43
Flow through the cell 5m3ph - 18m3ph
Max pressure in the cell 2.75 bar
Operating water temperature 10°C - 40°C
Table 3. Salt Chlorinator System Specifications
2.2.2 Water Care Module
Power Supply Extra low voltage
(connected to controller)
Peristaltic pump flow rate 1.2 L/h
Maximum back pressure (injection) 1.5 bar
pH correct Acidic pH only
(sulfuric/hydrochloric acid)
pH sensor calibration 1 or 2 point (pH 4 and pH 7)
Sensor Cable Length 3 Meters
Table 4. Water Care Module Specifications
NOTE: Water care
modules are optional
accessories and will
not be included with
all models.
2.1.2 Water Care Module Contents
pH Link
a. Module (pH Link)
b. POD Housing Upper
c. POD Housing Lower
d. 50 mm Pipe Adapter
e. POD Cap
f. Cap Plug (X2)
g. pH Injector Valve
h. Hole Saw and Hex Key
i. Compression Fitting
j. Ceramic Weight
k. 6 mm Tube (5 m)
l. -
m. -
n. Locking Collar
o. Locking Collar Tool
p. pH Sensor
q. -
r. Sensor Holder (X2*)
pH Buffer Solutions: pH 7 (X6), pH 4 (X6) - (not shown)
Table 2. Water Care Module System Contents

Page 7
Page 6 eXO®| Installation Manual
2.3 Dimensions
2.3.1 Controller
31.5 cm 9.6 cm
36.4 cm
Figure 1. Controller Dimensions
2.3.2 Cell
32 cm
11.5 cm
21.5 cm
11 cm
21.5 cm
15.5 cm
Figure 2. Electrolytic Cell Dimensions
2.4 Materials and Tools
2.4.1 Controller
Tools Needed for Installation
• Power Drill
• 7 mm Drill Bit - Hammer Drill Bit
(only necessary to drill into brick or concrete)
• Pencil or Marking Pen
• Flat Head Screwdriver
• Pozidiv Screwdriver
• Small Flathead or Slotted Screwdriver
2.4.2 Cell
Tools Needed for Installation
• PVC Cutter
• PVC Cement
• Pencil or Marking Pen
• Check Valve
• 50 mm or 63 mm PVC pipe
2.4.3 pH POD
Tools Needed for Installation
• Power Drill
• Pencil or Marking Pen
• Flat Head Screwdriver
• Phillips Head Screwdriver
• Small Flathead or Slotted Screwdriver

Page 7
Page 6 eXO®| Installation Manual
Section 3. Plumbing
The cell must be plumbed in a position that is
accessible for maintenance and within 2 meters of
the controller. The cell should be the last piece of
equipment in the circulation system. The cell must
be installed horizontally, level and with correct ow
orientation, see Figure 3 and Figure 4.
1. If you intend to plumb the cell on a bypass,
the bypass must be equipped with isolation
valves, see Figure 3 and Figure 4.
2. Plumb the cell inlet and outlet on vertical
lengths of 50 mm or 63 mm PVC pipe. The
cell inlet and outlet ports are 11.5 cm apart,
see Figure 2. The inlet of the cell is on the side
closest to the electrical lead, see Figure 5.
3. Put the unions onto the inlet and outlet pipes,
see Figure 5.
4. For 63 mm plumbing, glue the adapters
directly onto the pipes.
5. For 50 mm plumbing, glue the reducers into the
adapters then glue the adapters onto the pipes.
6. Ensure the o-rings are seated properly on the
adapters.
7. Secure the cell to the plumbing by
tightening the unions hand tight. DO NOT
OVERTIGHTEN.
8. Double check cell orientation. The cell inlet is
closest to the cell electrical lead.
Figure 3. Plumbing Cell Only
pH Minus
CAUTION: Never store acid directly underneath any
equipment as fumes will cause corrosion and resulting
damage can void your warranty. It is recommended to store
your acid container in a safe, protected well ventilated area.
NOTE: The sensor POD must be positioned after
filtration and heating systems and before the cell.
Figure 4. Plumbing Cell and POD assembly
Figure 5. Cell Plumbing Assembly
WARNING
• The cell must be installed horizontally and level.
Improper installation can lead to gas build up which
could result in equipment damage or serious injury.
• The cell must be the last piece of equipment on
the return line, see Figure 3 and Figure 4.
• It is recommended in all installations that the cell is
installed on a bypass equipped with isolation valves,
see Figure 3 and Figure 4.
• In order to avoid load loss, installing the cell on a
bypass is MANDATORY if system flow rate exceeds
18m3/hr, see Figure 3 and Figure 4.
• If installing on a bypass, use a downstream check
valve instead of a manual valve to prevent improper
back flow into the cell.
Cell
Adapter
Union
Reducer
Plumbing
Adapter
O-ring
CELL INLET

Page 9
Page 8 eXO®| Installation Manual
Section 4. Install Controller
The controller should be located at or near the
equipment pad, at least 3.5 metres or more away
from the inside wall of the pool/spa, 1.5 metres
off the ground, and within 2 meters of the cell. All
national, state, and local codes are applicable.
4.1 Mount the Bracket
1. Determine the controller location.
2. Use the enclosed bracket to mark the
mounting surface through the screw holes.
3. Drill surface with 7 mm bit.
4. Install the included wall anchors.
5. Hang bracket using the included screws.
6. Hang the controller on the wall bracket and
slide to the right to secure in place.
Figure 6. Controller Mount
4.2 Install Temperature Sensor
The temperature sensor must be installed
upstream from any heating equipment, see Figure 3
and Figure 4. The sensor will need to be placed into
the plumbing system at any suitable position after
the filter pump and before any heating equipment.
1. Drill a Ø9 mm hole in the top of the pipe.
2. Ensure the o-ring is properly seated on the
sensor body.
3. Secure the sensor and o-ring with the screw
clamp provided. DO NOT OVERTIGHTEN.
Figure 7. Install Temperature Sensor
4.3 Electrical Connections
WARNING
FOR YOUR SAFETY: This product must be
serviced by a professional pool/spa service
technician as described on the front cover of this
manual. The procedures in this manual must be
followed exactly. Failure to follow warning notices
and instructions may result in property damage,
serious injury, or death. Improper installation and/or
operation will void the warranty.
The controller must be permanently powered by
connection to a 30 mA residential circuit breaker.
The controller can be used to connect and control
multiple devices see Table 5 for a summary of functions.
WARNING
The controller is not to be considered as suitable for
use as service equipment. Therefore it is required to
have the appropriate means of disconnection, circuit
isolation, and/or branch circuit protection upstream
of the controller/device, in according with local wiring
rules.
4.3.1 Remove Lower Cover
1. Unplug power supply from mains.
2. Remove the external cover.
3. Remove the 2 screws from the lower housing.
4. Remove the lower housing by pulling down
then out see Figure 8.
Ø9 MM

Page 9
Page 8 eXO®| Installation Manual
Figure 8. Remove Controller Lower Cover
4.3.2 Controller Enclosure Wiring
1. Wire low voltage connections on the low
voltage side of the enclosure, see Figure 10.
2. Wire high voltage connections on the high
voltage side of the enclosure, see Figure 10.
3. Use the internal labels (low voltage - yellow, high
voltage - orange) to identify connection points.
4. Use the cable ties provided to secure internal
wiring to the controller frame. See Figure
10 for recommended cable tie locations and
Table 5 for connection point functions.
5. Pass cables through the appropriately labeled
strain relief ports. It may be necessary to
pierce the strain relief grommet prior to
passing the cables through it.
Figure 9. Controller Connection Ports
NAME TYPE ROUTE FUNCTION
SENS Input - Connection of the pH sensing PCBA
PUMP Input - Connection of the pH pump for pH Link module
CELL Output 1 Connection of the electrolysis cell
UI Output - Connection of the user interface
TEMP Input 3 Connection of the water temperature probe
Flow Input 2 Not used in AUS / South Africa
iAL RS485 Input 4 Connectivity dedicated connector
VSP RS485 Output 5 RS485 variable speed pump connection
DRED Input 6 DRED Function (Demand Response Enabled Device) – for Australia only
COVER Input 7 Pool cover contact – lower the chlorine production
(dry contact: cover open = open contact/cover closed = closed contact)
SLAVE Input 8 External connection of a device controlling the chlorinator
AUX 2 12-24V Output 9 ON/OFF control of a LOW VOLTAGE external device (12 – 24V)
AUX 1 230V Output 10 ON/OFF control of a HIGH VOLTAGE device. - Not available in Australia
PUMP 230V Output 11 230 V Power supply for the filtration pump Maximum 8A
MAINS Input 12 Device power supply 230 VAC – 50Hz
Table 5. Controller Wiring Positions

Page 11
Page 10 eXO®| Installation Manual
H0632100 Rev A
CABLE IDENTIFICATION TNG EU
SENS PUMP CELL
UI
TEMP FLOW
MAINSPUMP 230V
iAL RS485 VSP RS485 COVER SLAVE AUX 2 12-24V
AUX 230V
L N N L
H0632100 Rev A
CABLE IDENTIFICATION TNG EU
SENS PUMP CELL UI TEMP FLOW
MAINSPUMP 230V
iAL RS485 VSP RS485 COVER SLAVE AUX 2 12-24V
AUX 230V
L N N L
Figure 10. Controller Wiring

Page 11
Page 10 eXO®| Installation Manual
Section 5. Install the pH Link Module
on the Controller
NOTE: The water care module is an optional accessory
and will not be included with all models.
If your controller model does not have a pH Link
module, skip to Section 6.
The controller must be installed and operational
before you install the module, refer to Section 4.
pH balance and suitable alkalinity levels in the
water must be maintained. Be sure to prepare and
balance the pool before installing the pH Link or
Dual Link module, for details see Section 6.
1. Turn OFF power to the controller.
2. Remove the front cover from the controller,
see section 4.3.1 for details.
3. Connect Sens and Pump wires from the
Module to the controller board see section 4.3
for wiring details.
4. Seat the Module housing on the controller.
5. Fasten two screws to secure the module
to the controller.
6. Replace the external cover.
WARNING
Do not reconnect power supply until all components
are installed: (1) pH Link/Dual Link module, (2) POD
kit, and (3) pH down injection pipe.
5.1 Install the POD Kit
1. Make sure power to the controller is OFF.
2. For optimal performance, the POD must be installed:
• On a 30 cm length of rigid 50 mm or 63 mm PVC pipe.
• As far as possible from any angles.
• Horizontally.
• Vertically within 30°.
• After the ltering equipment.
• After any heating device but before the electrolytic
cell or other water treatment equipment.
5.2 Drill the Pipe
1. Inspect the POD O-ring to ensure it is in good
condition and properly seated.
2. Press tabs to remove the upper POD housing
from the lower POD housing.
O-ring
Upper Housing
Lower Housing
Figure 11. POD Assembly
3. Remove the 50 mm adapter from the lower
POD housing.
4. Place the 63 mm guide on the top of the pipe.
NOTE: For 50 mm installations use the 50 mm adapter
marked “EU” as the guide for marking the pipe,
for 63 mm installations use just the bottom of
the POD assembly.
5. Mark the pipe through the holes.
6. Drill holes in the PVC pipe using the hole saw.
Make sure the edges are smooth without burrs.
Hole Saw
50 mm Adapter
Figure 12. Mark and Drill Pipe
5.3 Install the POD
1. For 50 mm pipe installations, insert the
adapter into the holes in lower POD housing.
* with 50 mm pipe
Figure 13. Assemble the POD

Page 13
Page 12 eXO®| Installation Manual
2. Note the arrows on the upper housing
indicating ow direction. Make sure the pod is
in proper orientation.
Flow Flow
InletOutlet
Figure 14. POD Orientation
3. Assemble the POD around the pipe, the tabs
will click into place.
5.4 Install the POD Cap
1. Place the POD cap on the POD assembly.
The cover and POD assembly are keyed for
proper orientation. Use caution to ensure the
POD O-ring is clean and properly seated.
2. Align the arrow on the POD with the locking ring.
Figure 15. POD Locking Ring Orientation
3. Tighten the locking ring with the locking collar
tool until snug. Do not overtighten.
5.5 Install and Calibrate
the Sensor
1. Install the sensor holder into the correct port
on the POD cap.
Sensor
Holder
Cap
Plug
Figure 16. Sensor Holder Installation
2. Carefully remove the protective tube
from the sensor.
3. Rinse the end of the sensor with tap water.
Shake o excess water.
4. Carefully screw the sensor into sensor holder.
Figure 17. pH Sensor Installation
5. Calibrate the sensor.(See Section 5.7.
Calibration Procedure)
To maintain accurate and reliable operation,
the sensors must be calibrated before the initial
use of the controller with pH Link module. After
initial calibration, it is recommended to recalibrate
sensors every few months.
6. Connect the sensor lead to the module at the
base of the controller.
pH
Figure 18. pH Connection Port

Page 13
Page 12 eXO®| Installation Manual
5.6 Install the pH Down Intake and
Injection lines.
Intake line
(Pump to
pH down)
Injection line
(Pump to POD)
Figure 19. pH- Pump Conguration
The pump rotates in a clockwise direction. Acid is
fed into the left side of the pump (via the Intake
line) and injected into the POD from the right side
(via the Injection line).
CAUTION: For safety reasons, install the Injection line
before the Intake line to ensure pH is not
accidentally dosed into the system before all
connections are made.
Never store acid directly underneath any
equipment as fumes will cause corrosion
and resulting damage can void your
warranty. It is recommended to store your
acid container in a safe, protected well
ventilated area.
ow direction ( pump inlet, Injection)
Figure 20. pH- Pump Port
5.6.1 Install the Injection Line
1. Remove the protective cover from the pump.
2. Cut a section of the plastic tubing that has
adequate length to reach from the injection valve
at the pump to the non-return valve on the POD.
3. Remove the threaded cap from the pump
outlet (right).
4. Thread tubing through the hole in the cap,
connect the tubing to the nipple on the module
and reattach the threaded cap.
NOTE: For a better t, soften both ends of the tubing by
immersing in warm water.
5. Remove the threaded cap from the POD
intake valve.
6. Thread tubing through the hole in the cap,
connect the tubing to the nipple on the POD
and reattach the threaded cap.
5.6.2 Install the pH Down Intake Line
1. Cut a section of the plastic tubing that has
adequate length to reach from the pH down
reservoir to the pump.
2. Remove the threaded cap from
the pump inlet (left side).
3. Pass one end of the tubing through the
threaded cap and attach tubing to the nipple
on the pump inlet.
4. Remove the cap from the pH down container.
5. Thread tubing through the hole in the cap
through the ceramic weight.
6. Remove the nut from the blue
compression tting.
7. Pass the tubing through the nut and attach it
to the nipple on the compression tting.
8. Reattach the threaded nut.
9. Put the weighted end of the hose into the pH
down reservoir.
10. Secure the pH down cap tightly on the container.
11. Replace the protective cover on the pump.

Page 15
Page 14 eXO®| Installation Manual
5.7 Calibration Procedure
To maintain accurate and reliable operation,
the sensors must be calibrated before using
the controller with pH Link module. After initial
calibration, it is recommended to recalibrate the
sensor every few months.
5.7.1 Clean the Sensor
The sensor must be cleaned before calibration,
including prior to initial use after installation.
1. Turn o the lter pump.
2. Close all valves.
3. Remove the sensor and the sensor holder
from the POD.
4. Rinse the sensor with tap water for 1 minute.
Shake o excess water.
NOTE: To avoid damaging the active sensor,
do not rub or dry with a cloth.
5. Brush the junctions and metal strip using a
toothbrush for 1 minute.
Figure 21. Cleaning Probe
6. Prepare a diluted hydrochloric acid solution by
pouring 50 mL tap water into 1 mL (10 drops)
of commercially-available hydrochloric acid
(HCl 37%).
WARNING
• Hydrochloric acid is a hazardous chemical that
may cause burns, lesions, and irritations. Handle
with extreme care using protective equipment
(gloves, goggles, coveralls). Refer to the
substance SDS sheet for more information.
• Always pour acid into water.
• Once cleaning is complete, dispose of the solution
according to standard in effect in the country of
use.
7. Wash the sensor in the diluted hydrochloric
acid solution for 2 minutes.
8. Rinse sensor in clean tap water for 1 minute.
Shake o excess water.
5.7.2 Calibrate the pH Sensor
The pH sensor (blue) can be calibrated in 1 or 2
points. We recommend 2-point calibration for a
more reliable measurement throughout the season.
1. Turn on power to the controller.
2. Turn o the pool pump and close valves as
necessary to isolate the cell and sensors.
3. Unscrew the pH sensor from the POD.
4. Rinse the end of the sensor with tap water.
Shake o excess water. Do not touch the
glass bulb at the end of the pH sensor.
5. Start 1 or 2 point calibration via the controller
User Interface, see Section 8.3 For details.
6. If calibration is unsuccessful, see Section 11.
Troubleshooting.

Page 15
Page 14 eXO®| Installation Manual
Section 6. Prepare and Balance the Pool
6.1 pH Regulation Principle
The amount of chlorine needing to be produced is
dictated by the pH level in the pool. As the pH level
rises, the amount of chlorine must be increased to
compensate, see Table 6.
6.2 Water Chemistry Table
Test and maintain correct water balance throughout the season, according to the table.
Free
Chlorine
pH Total Alkalinity
(ppm)
Calcium Hard-
ness (ppm)
Cyanuric Acid
(ppm)
Salt
Level (ppm)
Metal
South
African
Standard
1 -3 7.2 - 7.8 80 - 120 175 - 275 30 -50 4000 < 0.10
Ideal range 1 - 3 7.4 80 - 140 90 - 300 up to 50 4000 at 27°C < 0.10
To Increase Add chlorine or
increase equip-
ment output
Add buer or
soda ash (so-
dium carbonate)
Add sodium
bicarbonate
Add calcium
chloride
Add cyanuric
acid
Add salt or
minerals **
-
To Decrease Add muriatic
acid
Add muriatic
acid or dry acid
Partially drain
and rell pool*
Partially drain
and rell pool*
Partially drain
and rell pool*
Use appropriate
metal remover
In Season
Testing
Frequency
Weekly Weekly Weekly Weekly Weekly Monthly
Monthly
NOTE: Test all equipment sensors quarterly.
* Fill pool with water from the mains water supply. Do not use rain water or well water.
** Do not add salt directly into the skimmer. Do not initiate electrolysis until salt has fully dissolved.
*** Reading is True Calcium Hardness, not Total Hardness.
Table 7. Water Chemistry
+
-
10 min 10 min 10 min 10 min
Dosing Cycles
Time
pH pH evolution
7,30
7,35
7,40
7,45
7,50
7,55
7,60
7,65
7,70
7,75
7,80
7,85
7,90
7,95
8,00
8,05
8,10
8,15
8,20
100%
75%
50%
20%
Factory Set Point = 7,4 pH
Table 6. pH Evolution

Page 17
Page 16 eXO®| Installation Manual
Section 7. Operation
7.1 User Interface Overview
CAUTION
Before activating the chlorination function of your system, ensure that all salt added to the pool has completely dissolved.
User Menu
Browse current menu;
Increase or decrease a setting;
Lock/unlock the UI (press and hold both)
On, O
Standby mode
Enter the highlighted selection;
Delete an error message
(press and hold 4 seconds)
Activate reduced
chlorine production mode
Activate 100% chlorination for
24 hours*
LED indicates information or
error message on screen
LED indicates low salt/minerals
or poor water conductivity
*The lter pump will also operate if ltration pump is connected to the controller
Figure 22. Controller User Interface
7.2 Initial start-Up
The following setting must be configured at the
initial start up of the device. Settings can be
changed or updated at a later time using the
general menu, details can be found in Section 8.
From factory, the chlorination is set at 50 %. It can
be set between 0 and 100 % in 10 % increments
from the “main screen” by pressing . The set
value is valid until the next modification.
7.2.1 System Power
1. Press to enter Standby mode.
2. Press again to start chlorination and
ltration.
3. Press and hold for 3 seconds to turn the
controller o.
7.2.2 Language
1. Press .
2. Use to select “Language”
then press .
3. Use to browse language the options.
4. Press to make a selection.
<-LANGUAGE
MAIN MENU
<-ENGLISH
-FRANCAIS
-ESPANOL
-ITALIANO
-DEUTSCH
-NEDERLANDS
-PORTUGUES
-CESTINA
-MAGYAR
-SVENSKA
-SLOVENSKY
SELECT LANGUAGE
7.2.3 Set Clock
1. Press .
2. Use to select “Programming”
then press .
3. Use to select “Clock”
then press .
4. Use to set hour then press .
5. Use to set minutes then press .
NOTE:
Press and hold or to increment time by 10.
6. Press to exit.
<-PROGRAMMING
-LOW MODE
-SLAVE MODE
MAIN MENU
23:04
CLOCK SET UP
7.2.4 Boost Mode
In some cases, the pool may require a higher than
normal chlorination level such as in cases of high
use, poor weather or at the start of the season. Boost
mode is used to increase the chlorine level quickly.
Boost mode lasts for 24 cumulative hours at 100%
production output. If the program is set to perform
chlorination 12 hours per day, Boost mode will be
active for 12 hours on the first day and 12 hours
on the second day. If the filter pump is connected
to the controller, the filter pump will operate in
Boost mode also. The controllers operating times
including filtration schedules are temporarily
Wi-Fi to phone connection

Page 17
Page 16 eXO®| Installation Manual
ignored for the duration of the Boost mode. Once
Boost mode completes, the controller and filter
pump will resume programmed operations.
NOTE: If the device is equipped with a Dual Link
module, the Boost mode will not take the ORP
value into account. The Boost mode takes
priority over the ORP regulation
1. Press .
23:04 ON
BOOST MODE 23:59
2. If the controller is in Low/Cover mode you will
need to conrm that you want Boost mode to
override Cover or Low mode settings.
7.2.5 Low Mode
Low mode is designed to reduce chlorine
production in situations where the pool is covered
or when pool use is limited. Chlorine production
needs will be lower when pool use is low and/or
when pool water is not exposed to UV rays etc.
Low mode output can be set in the main menu
settings for Low mode. Low can be set from
0%-30% in 10% increments. Programs are still
active while the controller is in Low/Cover mode.
1. To manually enter Low mode press .
2. To exit Low mode press again.
23:04 ON
LOW MODE 10%
7.2.6 Cover Mode
If the pool is equipped with a compatible electric roller
cover, this can be connected to the controller in order
to automatically reduce chlorination upon closing the
cover, this is known as Cover mode, see Table 5 and
Figure 10 for wiring details. Chlorination will resume
at the level determined by the programming upon
opening the compatible electric cover.
Cover mode output can be set in the main menu
settings for Low mode. Low can be set from
0%-30% in 10% increments. Programs are still
active while the controller is in Low/Cover mode.
Check that the cover is compatible and connected
to the chlorinator.
Cover mode will automatically be activated when
the cover is closed. The Cover mode message and
the output percentage are displayed on the screen.
23:04 ON
COVER MODE 10%
Cover mode will stop as soon as the cover is
completely open.
If the controller is equipped with a Dual Link
module we do not recommend connecting the
Cover mode. More specifically, chlorination is
managed via the Dual Link module. In cases where
the Cover mode is connected in the presence of
a Dual Link module: chlorination will take place
upon closing the roller cover, even if the ORP
measurement is greater than the setpoint value.
7.2.7 Cold Water Protection
In addition to recording and displaying the water
temperature, the temperature sensor is used to
protect the cell, which is sensitive to cold water.
The temperature displayed on the top left of the
home screen will begin to flash at 15°C.
23:04 ON15°
When water temperature is under or equal to 15°C,
chlorine production automatically switches to the
level defined in Low/Cover mode.
When water temperature is under or equal to 10°C,
chlorine production is stopped. The absence of
chlorination at this temperature is not hazardous,
as bacterial development is slowed in cold water.
In addition to the flashing temperature, a “LOW
TEMPERATURE” message is alternately
displayed.
When the temperature goes over 10°C again, the
output percentage will be set to Low/Cover.
When the temperature goes over 15°C again,
chlorination resumes its level of operation
configured using programs.
7.2.8 Locking the Controller
The controller can be locked, thus deactivating the
buttons on the user interface panel. Simultaneously
press and hold the and buttons for 3
seconds. This function can be accessed from any
screen/menu.
Locking the controller automatically returns the user to
the home screen. To unlock the controller, press and
hold the and buttons for another 3 seconds.

Page 19
Page 18 eXO®| Installation Manual
Section 8. Menu and Settings
8.1 Filter Pump
Both single and variable speed pumps can be
connected to and operated by the controller.
8.1.1 Single Speed Filter Pump
1. Connect the lter pump to the controller at
the pump connection point in the high voltage
raceway. For Australian installations plug the lter
pump into the socket provided on the underside
of the control unit see Table 5 and Figure 10 for
wiring details.
2. Press .
3. Use to select “Filter Pump”
then press .
4. Use to select “Single Speed”
then press .
5. The user interface will prompt “Single Speed
Plugd to Device”. Use to select “Yes”
then press .
<-FILTER PUMP
-AUX/LIGHTS
-PROGRAMMING
-LOW MODE
-SLAVE MODE
-REVERSING
-HELP MENU
-LANGUAGE
MAIN MENU
<
-BACKWASH
-SINGLE SPEED
-VARIABLE SPEED
FILTER PUMP
<
PLUGD TO DEVICE?
-YES
-NO
SINGLE SPEED
8.1.2 Variable Speed Filter Pump
1. The following pump models are compatible:
FloPro™ VS, Flopro™ VS AUT
2. Connect the lter pump to the controller at
the VSP RS485 connection point in the low
voltage raceway, see Table 5 and Figure 10 for
wiring details.
3. Ensure the RS485 wires are in the proper order.
LOW VOLTAGE HIGH VOLTAGE
RED
BLK
YEL
GRN
On the Controller
RED
BLK
YEL
GRN
On the Pump
4. Ensure the pump DIP switches are in the
correct position as follows:
1 = up
2 = up
3 = down
4 = down
5 = up
5. Press .
6. Use to select “Filter Pump”
then press .
7. Use to select “Variable Speed”
then press .
8. The user interface will prompt “Variable Speed
Plgd to Device”. Use to select “Yes”
then press .
9. Use to select a speed then press .
10. Use to adjust the RPM.
11. Once desired RPM is reached, press .
12. Press to exit.
NOTE: If multiple VSP timers have a time slot in common
: it is the higher RPM that will be operated.
<-FILTER PUMP
-AUX/LIGHTS
-PROGRAMMING
-LOW MODE
-SLAVE MODE
-REVERSING
-HELP MENU
-LANGUAGE
MAIN MENU
<
-BACKWASH
-SINGLE SPEED
-VARIABLE SPEED
FILTER PUMP
<
PLUGD TO DEVICE?
-YES
-NO
VARIABLE SPEED
<
VARIABLE SPEED
-SPEEDS
-MIN/MAX
-PRIMING
-PROGRAMMING
<
VSP SPEEDS
-RPM 1 1000
-RPM 2 1500
-RPM 3 2000
-RPM 4 2850
-QUICK 3450
8.1.3 Quick Clean
Quick Clean is a backwash mode is used to quickly
start/stop the filter pump (single- or variable-speed
pump) in order to backwash the filter.
For variable speed pumps, the backwash RPM is
set under “Quick”.
1. Select YES to activate ltration.
2. Select NO to stop ltration.
For safety reasons, the chlorination is stopped
during the backwash mode. To avoid emptying the
pool, the backwash mode automatically stops after
5 minutes. The variable speed pump RPM is by
default set at 3450 rpm (max speed). This can be
changed in the priming menu.

Page 19
Page 18 eXO®| Installation Manual
8.2 Aux/Lights
The chlorinator is capable of controlling devices in
addition to the filter pump. The equipment must be
connected to the chlorinator using the appropriate
auxiliary line:
Aux 1: for equipment connected to the high voltage
supply (230 V) (not applicable for Australia).
Aux 2: for equipment connected to the low voltage
supply (12/24 V).
The following functions are available in the
AUX/LIGHTS menu for auxiliaries 1 and 2:
● Turn the equipment on/off
● Program a timer (See section 8.7.6 for details)
For Australian products, in order to control
the light transformer, an additional switching
box is required. Please contact your local pool
care professional for this relay box, item code
WW000222.
1. Remove the lower cover from the housing.
2. Attach the 2 x core wires from the switch box
to the terminal bar marked ‘AUX 2’.
3. Replace the cover/s.
4. Plug the switch box into any
available 240V GPO.
5. Plug the light transformer directly into the
socket located in the switch box. This will now
control the light directly from the chlorinator.
8.2.1 Assign Lights
1. Press .
2. Use to select “Aux/Light” then press .
3. Use to select “Assign Lights” then press .
4. Use to select “Aux 2” then press .
5. Use to select “Yes” then press .
6. Use to select “White Light” or “Color
Light” then press .
7. Press to exit.
<-AUX/LIGHTS
-PROGRAMMING
-LOW MODE
-SLAVE MODE
-REVERSING
-HELP MENU
-LANGUAGE
MAIN MENU
<
AUX/LIGHTS
-ASSIGN LIGHTS
-AUX 2
-AUX 1
<
ASSIGN LIGHTS
-AUX 2
-AUX 1
<
AUX SETUP
AUX 2 = LIGHT?
YES
NO
<
AUX 2 = LIGHT
-WHITE LIGHT
-COLOR LIGHT
8.2.2 White Light Operation
1. Press .
2. Use to select “Aux/Light” then press .
3. Use to select “Aux 2/Light” then press .
4. Use to select “On”, “O”,
or “Set Timers”, then press .
5. Press to exit.
<
AUX 2
-OFF
-ON
-SET TIMERS
8.2.3 Color Light Operation
1. Press .
2. Use to select “Aux/Light” then press .
3. Use to select “Aux 2/Light” then press .
4. Use to select desired color, “O”,
“On” or “Dene Color”, then press .
5. If “Dene Color” selected, use to select
desired color then press .
6. Press to exit.
<
AUX 2 / LIGHT
-OFF
-ON
-DEFINE COLOR
<
COLOR LIGHT
-ALPINE WHITE
-SKY BLUE
-COBALT BLUE
-CARIBBEAN BLUE
-SPRING GREEN
-EMERALD GREEN
-EMERALD ROSE
-MAGENTA
-VIOLET
-SLOW SPLASH
-FAST SPLASH
-AMERICA
-FAT TUESDAY
-DISCO TECH
Note: colour sequence is relevant for Zodiac colour
lights only. If other manufacturers lights are
installed the colour selection will vary

Page 21
Page 20 eXO®| Installation Manual
8.3 pH Sensor Calibration
The pH sensor may be calibrated in 1 point
calibration or 2 point calibration (pH 4 and pH 7); 2
point calibration is recommended.
1. Turn on power to the controller.
2. Turn o the pool pump and close valves as
necessary to isolate the cell and sensors.
3. Press .
4. Use to select “pH Setup” then press .
5. Use to select “pH Calibration”
then press .
6. Use to select “ 1 Point”, or “ 2 Points”,
then press .
7. Remove the pH sensor from the POD.
8. Insert the pH sensor into the
pH 7 buer solution.
9. Press while “Start” is selected.
10. Press while “Go to Step 2” is selected.
11. Rinse the pH sensor with tap water.
12. Insert the pH sensor into the pH 4 buer
solution.
13. Press while “Start” is selected.
14. Press while “Continue” is selected.
15. Press to exit.
<-PH SETUP
-ORP SETUP
-LOW MODE
MAIN MENU
<
-PH SETPOINT
-PH CALIBRATION
-ACID PUMP
-TEST DOSAGE
PH SETUP
<
-1 POINT
-2 POINTS
PH CALIBRATION
<
PH SENSOR INTO
BUFFER PH 7
START
STEP 1
PH 7 CALIBRATION
PROCEEDING
STEP 1
<
PH 7 CALIBRATION
COMPLETE
GO TO STEP 2
STEP 1
IN TAP WATER
RINSE SENSOR
<
PH SENSOR INTO
BUFFER PH 4
START
STEP 2
PH 4 CALIBRATION
PROCEEDING
STEP 2
<
PH 4 CALIBRATION
COMPLETE
CONTINUE
STEP 2
Table of contents