A&D SJ-3000WP User manual

SJ-3000WP
SJ-6000WP
SJ-15KWP
SJ-30KWP
1WM PD4002705A

© 2013 A&D Company, Limited. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or translated into any
language in any form by any means without the written permission of A&D Company, Limited.
The contents of this manual and the specifications of the instrument covered by this manual are
subject to change for improvement without notice.

1
CONTENTS
1. SAFETY PRECAUTIONS ..........................................................................2
2. PARTS DESCRIPTION..............................................................................3
3. PREPARATION .........................................................................................4
3.1. Installing / Exchanging batteries..............................................................4
3.2. Setting up the scale.................................................................................4
4. DISPLAY AND SYMBOLS.........................................................................5
4.1. Display ....................................................................................................5
4.2. Operations and functions of switches......................................................6
5. OPERATION..............................................................................................7
5.1. Basic weighing operation ........................................................................7
5.2. Notes about operations...........................................................................7
5.3. Weight display resolution ........................................................................8
6. SELECTING A WEIGHING UNIT...............................................................9
6.1. Storing the weighing unit.........................................................................9
6.2. Selecting the weighing unit......................................................................9
7. COUNTING MODE..................................................................................10
8. COMPARATOR .......................................................................................11
8.1. The formula to compare........................................................................12
8.2. Entering the comparator values ............................................................14
9. AUTO-TARE ............................................................................................15
10. CALIBRATION.........................................................................................16
10.1. Calibration mode.................................................................................16
10.2. Gravity acceleration correction............................................................17
10.3. Calibration using a weight...................................................................18
10.4. Restoring the calibration data to the factory set values.......................19
11. FUNCTION SETTINGS............................................................................20
11.1. Setting the parameters........................................................................20
11.2. Restoring the function settings to the factory set values .....................21
11.3. Function list.........................................................................................22
12. MAINTENANCE.......................................................................................24
12.1. Notes on maintenance ........................................................................24
12.2. Error codes..........................................................................................24
13. SPECIFICATIONS...................................................................................25
14. GRAVITY ACCELERATION.....................................................................27

2
1. SAFETY PRECAUTIONS
All safety messages are identified by the following, “
WARNING
” or “
CAUTION
”,
of ANSI Z535.4 (American National Standard Institute: Product Safety Signs
and Labels). The meanings are as follows:
WARNING
A potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
A potentially hazardous situation which, if not avoided,
may result in minor or moderate injury.
This manual is subject to change without notice at any time to improve the
product.
Product specifications are subject to change without any obligation on the part
of the manufacturer.
When using the SJ-WP, the following safety precautions should always be
followed.
WARNING
Internal service or adjustment to this product should be performed
by a qualified person.
CAUTION
Avoid installing the scale in direct sunlight, which may cause discoloration
or malfunctions.
Do not mix battery types, or new and old batteries. Replace with all new
batteries at the same time.
If the scale is not to be used for a long period of time, remove all batteries
from the battery compartment to avoid leakage.
Avoid overloading the scale.
Avoid using the weighing platform to move the scale, as that could cause
damage to the scale.
Avoid chemical solvents. Clean the scale with water.
IP67
Degrees of protection against water: Protected against temporary submergence.
Degrees of protection against solid foreign objects: Dust-tight.
International Protection of IEC60529.

3
2. PARTS DESCRIPTION
Sealing diagrams
Screwdriver
Use this screwdriver to open
or close the battery cover.
Weighing pan
Bubble Spirit level
Adjust the leveling feet so that the bubble is in the
center of circle.
Inner pan (ABS)
Pan support
The pan support can be removed at times such as
when the scale is cleaned.
Battery cove
r
To maintain waterproof performance, close completel
y
this cover and secure by screws.
Weight sensor (Load cell)
Note: To maintain performance, do not touch
this sensor.
Screw to fix the pan support
Screw for
securing
Display and
switch
Leveling foot
Switch cover
When this cover is opened, the CAL and RESET
switches inside are revealed. Keep the cover closed
to maintain waterproof performance.
Weighing pan (SUS304)

4
3. PREPARATION
3.1. Installing / Exchanging batteries
The batteries are not included with the product. Prepare 6 x “D” size (R20P or LR20)
dry-cell batteries.
When the mark is displayed,
exchange the old batteries with new ones.
1. Loosen the screws for securing the
battery cover using the provided
screwdriver, and remove the battery
cover.
2. Remove all the old batteries from the
battery compartment.
3. Install six new batteries properly
according to the + and - indicators of
polarity in the battery compartment.
4. Close the battery cover, and tighten the
screws for securing the battery cover.
CAUTION -
Do not mix used and new batteries. Do not mix the different types
of batteries. That may cause damage to the batteries or the scale.
- Take care of the polarity of batteries. The polarity marks are shown
in the battery compartment.
3.2. Setting up the scale
CAUTION
Avoid installing the scale in direct sunlight, that may cause
discoloration or malfunctions. Place your SJ-WP on a firm weighing
table so that the scale is level. The scale will not perform accurately
when it is not level.
Place the scale on a firm surface and adjust the feet so that the bubble of the
sprit level is in the center of the circle.
(R20P/ LR20
/ D size)
Six new
batteries
Batter
y
cover
Screw fo
r
securing
Old
batteries
OK NG

5
4. DISPLAY AND SYMBOLS
4.1. Display
Symbols
Symbol Description
STABLE Turns on when the weight value is stable.
NET
◄
Turns on when the NET weight is displayed.
(The tare operation is in progress.)
ZERO
◄
Turns on when zero is displayed.
Comparator LEDs Turns on when the comparator results are displayed.
Weighing units “Ib”, “oz”, “ozt” , “Ib-oz” , “tI-s” , “tI-h” , “tI-t” , “t” , “pcs” , “N” , “g”
and “kg” are available.
A selected unit is displayed.
*Turns on while a switch is touched.
Battery indicator
The battery indicator changes as the battery capacity
decreases, as shown below:
New Replace the batteries.

6
4.2. Operations and functions of switches
Switch Description
When the scale is turned off:
Touch the switch to turn ON the scale. The scale will be
automatically set to zero (power-on zero).
When the scale is turned on:
Touch and hold the switch to trun OFF the scale.
Tap the switch to ZERO the scale and display zero.
During measurement:
Touch and hold the switch to proceed to the calibration mode.
In the function setting mode to set the parameters:
Tap the switch to change the parameter of the selected item.
During measurement:
Tap the switch to choose a unit specified in the setting mode.
Touch and hold the switch to enter the comparator value
setting mode.
In the function setting mode to set the parameters:
Tap the switch to select the function item.
During measurement:
Tap the switch to tare the scale and display zero (net weight
display).
In the function setting mode to set the parameters:
Other than at the item "unit":
Tap the switch to store new parameters and return to the
weighing mode.
At the item "unit":
Tap the switch to select active / inactive for the displayed unit.
When the scale is turned off:
Touch the ON/OFF switch while touching and holding the
TARE switch to enter the function setting mode.
Further to the above, continue to touch the TARE switch to
restore the function settings to the factory set values.
When the scale is turned on:
Pressing the switch forcibly turns off the scale.
When the scale is turned on:
By pressing the switch, the scale proceeds to the calibration mode.

7
Object to
be weighed
Weighing pan
Containe
r
Tap
Touch & hold
5. OPERATION
5.1. Basic weighing operation
1. Touch the switch to turn the power ON.
All the symbols except the comparator LEDs are
displayed. When the weight values becomes
stable, the display turns off for a moment and
displays zero (power-on zero) with the weighing
unit used last before turning off.
2. Select a weighing unit using the UNITS switch.
See “6. SELECTING A WEIGHING UNIT” in
detail.
3. When the display doesn’t show zero, tap the
switch to set the display to zero.
4. When using a tare (container), place the
container on the weighing pan, and tap the
TARE switch to set the display to zero.
5. Place the object to be weighed on the weighing
pan or in the container, and wait for the STABLE
indicator to turn on and read the value.
6. Remove the object from the weighing pan.
7. Touch and hold the switch. Then, the
scale displays “0ff” and turns the power off.
5.2. Notes about operations
Power-on zero
If the weight is unstable at power ON, the scale displays “------”. Check anything
touches the weighing pan, or check if there is strong wind or vibration.
The range for power-on zero is within ±50% (±10% for Legal for Trade models) of
the weighing capacity (kg) at the calibrated zero point. If the scale is powered on
with a load beyond this range, the scale displays “------”. Remove the load from
the weighing pan.
ZERO and TARE
The (as ZERO switch) and TARE switches work when the weight value
is stable.
The switch will zero the scale if the weight value is within ±2% of the
weighing capacity (kg) at the power-on zero point. The ZERO
◄
indicator turns on.
(ZERO operation)

8
The TARE switch will tare the scale and subtract the weight to zero as a tare
weight when the weight is a plus value. In this case the ZERO
◄
and NET
◄
indicators turn on. (TARE operation)
At the zero point, the net weight display shows the tare weight in negative, and the
ZERO
◄
and NET
◄
indicators turn on.
(Note: In some countries or areas, the ZERO
◄
indicator will not turn on
while the scale is tared.)
When the scale is tared, weighing range for net loads is reduced by the amount of
the tare weight.
When the ZERO operation is performed in the net weight display, the tare operation
previously done is cleared and the NET
◄
indicator turns off.
(Note: In some countries or areas, the ZERO operation will not clear the TARE
operation. Tap the TARE switch after zeroing the scale with nothing
on the weighing pan.)
Auto power-off function
If no switch is touched or tapped and the stable indicator is displayed for a certain
period of time, the scale will automatically turn off. See the function setting “poff”
to set the elapsed time to turn off.
When “ e” or “-e” is displayed (refer to “12.2. Error codes”), the auto power-off
function is enabled.
LCD backlight
The LCD backlight is controlled by the functions “l-it” and “l-i”.
If no switch is touched or tapped and the weight display continues to be stable for a
certain period of time, the LCD backlight will automatically turn off. The elapsed time
to turn off is set by the function setting “l-it”. The backlight always on or off is
also selectable.
The function setting “l-i” adjusts the brightness of the backlight.
5.3. Weight display resolution
The weight display resolution is a ratio of the minimum display to the weighing
capacity.
The SJ-WP series has four types of weight display resolution, as shown below.
Low: 1/3,000
Normal: 1/6,000 or 1/7,500 (depending on the weighing capacity)
High: 1/12,000 or 1/15,000 (depending on the weighing capacity)
Maximum: 1/30,000
The factory setting is the normal resolution. Select the resolution according to your
own application in the function setting “re5o”.
For details about the minimum display and the weighing capacity, refer to “13.
SPECIFICATIONS”.
The weight display resolution of the Legal for Trade models is fixed. The selection in
the function setting “re5o” is not available.
In the counting mode, the scale works with the maximum resolution regardless of
the weight display resolution selected in the function setting “re5o”.

9
6. SELECTING A WEIGHING UNIT
6.1. Storing the weighing unit
1. Touch the ON/OFF switch while touching and holding the TARE switch in
order to display in the function setting mode.
2. Tap the UNITS switch to display .
3. Tap the SAMPLE switch to display a unit. Tap the TARE switch to activate or
deactivate the unit. The indicator is displayed for each active unit.
4. Repeat step 3 for other units.
5. Tap the UNITS switch to move to the next function item when finishing the selection.
6. Tap the TARE switch to store new units. The scale returns to the weighing mode.
Unit Symbol Conversion to gram
Gram 1.00000 g
Kilogram 1000.00 g
Pound (UK) 453.59237 g
Ounce (avoir) 28.349523125 g
Troy ounce 31.1034768 g
Pound - Ounce
-
Tael (Hong Kong general,
Singapore) 37.7994 g
Tael (Hong Kong jewelry) 37.4290 g
Tael (Taiwan) 37.5 g
Tola 11.6638038 g
Counting unit
-
Newton See below
Newtons is a value calculated as follows:
Newtons =(value in grams)x (9.80665 m/s2)/1000
6.2. Selecting the weighing unit
In the weighing mode, tap the UNITS switch to select a weighing unit.
Each time the UNITS switch is pressed, the unit changes as shown below.
Ib kg
g
N
pcs
t
tl
Ib-oz
ozt
oz
UNITS

10
7. COUNTING MODE
Determines a unit weight (the weight of one piece) from a known sample quantity, and
calculates how many pieces are on the weighing pan using the unit weight.
The unit weight is maintained even if the power is turned OFF.
1. Tap the UNITS switch to se lect “pcs”.
(“pcs” = pieces)
2. Tap the SAMPLE switch to enter the sample
unit weight storing mode.
The numerical value on the left indicates the
number of samples.
3. Tapping the SAMPLE switch allows you to
change the number of samples in the order 5
10 20 50 100 ESC 5.
•By tapping the UNITS switch when “e5C” is
displayed, the scale exits the sample unit
weight storing mode to the count display.
4. When “-” appears at the right side of the number of
samples, tap the switch to zero the
scale.
If necessary, place a container on the weighing
pan, and tap the TARE switch. Confirm that
the right side of the number of samples shows
zero.
5. Place the correct number of samples on the
weighing pan or in the container.
6. Confirm that the STABLE indicator is turned on.
Tap the UNITS switch to calculate and store
the unit weight. Remove the samples. The
scale is set to count objects with this unit
weight.
The total weight of samples should be more than
shown below, regardless of the number of samples.
SJ-3000WP: 2.5 g
SJ-6000WP: 5 g
SJ-15KWP: 12.5 g
SJ-30KWP: 25 g
If not, the display shows “lo ut” and returns to
the display of step 5. Increase the number of
samples (step 3) and try again.
7. Place the objects to be counted on the weighing pan.
Confirm
zero
Number of
samples
Samples
Weighing
pan
Containe
r
Tap
Tap
Tap
Tap
Tap

11
8. COMPARATOR
The scale has three-, five- and seven-level comparators.
Each comparator mode compares the weight value against the preset limit values and
outputs the results using LEDs (yellow / green / red).
Note: When the unit is “lb-oz” or “tl”, this function can not be used.
•Five-level comparator mode:
Uses four comparator values
to compare the weight value
and outputs results in five
levels of LOLO, LO, OK, HI
and HIHI.
•Three-level comparator mode:
Uses two comparator values
(upper and lower limit values)
to compare the weight value
and outputs results in three
levels of LO, OK and HI.
•Seven-level comparator mode (portion weighing mode):
Uses six comparator
values to compare the
weight value and outputs
results in seven levels of
outside the lowest limit,
level 1 (LOLO), level 2
(LO), Level 3 (OK), level
4 (HI), level 5 (HIHI) and
outside the highest limit.
To use the comparator modes, the function settings “Cp-l” and “Cp” must be
specified and the comparator values must be set.
Using the function setting “Cp-l”, select a comparator mode.
0: five-level comparator mode (Result LED blinks)
1: five-level comparator mode (Result LED lights)
2: three-level comparator mode (Result LED blinks)
3: three-level comparator mode (Result LED lights)
4: Seven-level comparator mode (Result LED blinks)
5: Seven-level comparator mode (Result LED lights)
Using the function setting “Cp”, select comparison conditions.
0: No comparison (comparator disabled).
1: To compare all data.
2: To compare all stable data.
3: To compare all data which are ≥+5d or ≤-5d.
4: To compare stable data which are ≥+5d or ≤-5d.
5: To compare all data which are ≥+5d.

12
6: To compare stable data which are ≥+5d.
d = minimum display in kg (Refer to “13. SPECIFICATIONS”.)
In the counting mode, “d” is equal to the minimum weight display of kg mode.
8.1. The formula to compare
Comparison is performed using the formula listed below and the results are output.
Five-level comparator mode
Results Comparison formula LED display
LOLO Displayed value < LOLO limit, or “-e”
( Red LED on)
LO LOLO limit ≤Displayed value < LO limit
( Yellow LED on)
OK LO limit ≤Displayed value ≤HI limit
( Green LED on)
HI HI limit < Displayed value ≤HIHI limit
( Yellow LED on)
HIHI HIHI limit < Displayed value, or “ e”
( Red LED on)
Three-level comparator mode
Results Comparison formula LED display
LO Displayed value < LO limit, or “-e”
( Red LED on)
OK LO limit ≤Displayed value ≤HI limit
( Green LED on)
HI HI limit < Displayed value, or “ e”
( Red LED on)

13
Seven-level comparator mode (portion weighing mode)
Results Comparison formula LED display
None Displayed value < Level 1 lower limit, or “-e”
( No LEDs on)
LOLO
(Level 1) Level 1
lower limit ≤Displayed value < Level 2
lower limit
( Red LED on)
LO
(Level 2) Level 2
lower limit ≤Displayed value < Level 3
lower limit
( Yellow LED on)
OK
(Level 3) Level 3
lower limit ≤Displayed value ≤Level 3
upper limit
( Green LED on)
HI
(Level 4) Level 3
upper limit < Displayed value ≤Level 4
upper limit
( Yellow LED on)
HIHI
(Level 5) Level 4
upper limit < Displayed value ≤Level 5
upper limit
( Red LED on)
None Level 5 upper limit < Displayed value, or “ e”
( No LEDs on)
The comparator values are common to the weighing and counting mode.
Ignore the decimal point when setting the comparator values.
Example for SJ-6000WP when the setting value is “001000”:
Display mode Limit value Capacity / Minimum display
Normal resolution kg 1.000 kg 6.000 kg / 0.001 kg
High resolution kg 0.1000 kg 6.0000 kg / 0.0005 kg
Maximum resolution g 100.0 g 6000.0 g / 0.2 g
Low resolution oz 100.0 oz 210.0 oz / 0.1 oz
Normal resolution oz 10.00 oz 210.00 oz / 0.05 oz
High resolution oz 10.00 oz 210.00 oz / 0.02 oz
Counting mode 1000 pcs
The comparator values are maintained even if the power is turned OFF.
The scale does not judge magnitude relation among the comparator values.
Even if the wrong values are set, no error will be shown.

14
8.2. Entering the comparator values
1. In the weighing mode, touch and hold the UNITS switch to enter the comparator
value setting mode.
2. Enter the comparator values using the following switches.
UNITS To shift the blinking digit to the right.
SAMPLE To increase the value of the blinking digit by one.
TARE To store the setting value and to go to the next limit setting or “end”.
The minus sign can be set at the next digit of the least significant digit. The
SAMPLE switch alternates the minus sign on and off. The blinking “-” shows minus
and no sign shows plus.
3. When the setting is complete, the scale memorizes the values and displays “end”,
then returns to the weighing mode. (At this time, power-on-zero is not performed.)
Five-level Three-level Seven-level
comparator mode comparator mode comparator mode
Touch
& hold Touch
& hold Touch
& hold
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Weighing
mode
To set the
LOLO
limit value
To set the LO
limit value
To set the HI
limit value
To set the HIHI
limit value
Weighing
mode
Weighing
mode Weighing
mode
Weighing
mode
Weighing
mode
To set the LO
limit value
To set the HI
limit value
To set level 1
lower limit value
To set level 2
lower limit value
To set level 3
lower limit value
To set level 3
upper limit value
To set level 4
upper limit value
To set level 5
upper limit value

15
9. AUTO-TARE
The SJ-WP series has an auto-tare function to be used with the comparator mode
enabled. If the weight values are in the OK range of comparator limits and stable for a
preset period of time, the scale will automatic 1 ally tare the weight and show zero.
In some countries or areas, the auto-tare function can not be used on the Legal for Trade
models and the selection in the function settings “at”, “at-t” and “at-f” is not available.
To use the auto-tare function, set the function settings below.
Cp 1: Compare all weighing data (other settings may be used depending on
the application).
at 1: Auto-tare function enabled.
at-t 0~ 9:
Select the timing to tare automatically to avoid the wrong tare operation, for
example; too early to tare, to take a longer time to go to the next weighing.
Normal comparion “Cp-p 0”
Start with display zero after tare operation. Place or take away objects until the
comparison result will show OK. When the stable indicator is ON for a the preset
period of time specified in the function setting “at-t”, the scale will automatically
tare the weight, show zero and be ready for next weighing to repeat.
Negative comparison for take-away “Cp-p 1”(example with “Cp-l 0”)
Take-away check weighing (negative comparison) is the way to compare the
negative weight while taking away objects from a container.
Set the function “Cp-p 1” together with the auto-tare function enabled “at 1”. In this
operation mode, the scale operates as “take-away the objects” “OK and stable”
“auto-tare” “take-away the objects” ······.
In this setting, the polarity of LOLO, LO, HI, and HIHI limit values are ignored and
the scale shows the comparator results as below.
Note: To start the take-away check weighing, be sure to use the TARE switch to
tare the weight of the container filled with objects. The ZERO switch may zero
the display, and the scale goes below the zero point by taking out the objects.
Then, the auto-tare function does not work.
When the function “at-f 1Tares the initial (container) weight.” is selected:
To start the auto-tare function, usually the container (filled with objects) will be placed on
the weighing pan and its weight must be tared using the TARE switch. When the
function “at-f 1” is selected, the scale will tare the initial (container) weight automatically.
When all load on the weighing pan is removed, the scale will return to the zero point
and the tare weight will be automatically cleared. If the scale does not return to the
zero point, tap the switch to clear the tare weight.

16
10. CALIBRATION
Adjusts the scale for accurate weighing.
Calibrate the scale in the following cases.
When the scale is first installed.
When the scale has been moved.
When the ambient environment has changed.
For regular calibration.
Note: The Legal for Trade models can not be re-calibrated if they have been
sealed.
10.1. Calibration mode
The calibration mode has the following three functions.
•Gravity acceleration correction
•Calibration using a weight
•Restoring the factory set values
How to enter the calibration mode
Method 1:
1. Make sure that the scale is in the weighing mode.
2. Touch and hold the SAMPLE switch until the gravity acceleration value (in this
example, “9.7985”) appears and release the switch.
Note: The above operation is disabled for the Legal for Trade models.
Method 2:
1. Make sure that the scale is in the
weighing mode.
2. Loosen the two screws on the switch
cover and open the switch cover. The
calibration (CAL) switch is located inside.
3. Press the CAL switch. The scale
displays the gravity acceleration value (in
this example, “9.7985”).
Switch cover
CAL switch

17
Tap the UNITS switch
to select an item
and
Tap the TARE switch
to execute it.
To exit from the calibration mode, touch
the switch or Press the CAL
switch. The scale turns the power OFF.
10.2. Gravity acceleration correction
When the scale is first used or has been moved to another location, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value stored in the scale to the
value of the area where the scale will be used. Refer to the gravity acceleration map at
the end of this manual.
Note: Gravity acceleration correction is not required when the scale will be
calibrated using a calibration weight at the place where it is to be used.
1. Refer to “10.1. Calibration mode” to enter the calibration
mode. The gravity acceleration value is displayed.
2. Tap the TARE switch to enter the gravity
acceleration value setting mode.
3. Change the displayed value using the following switches.
UNITS To shift the blinking digit to the right.
SAMPLE To increase the value of the blinking
digit by one.
4. Tap the TARE switch. The display shows “end”and
returns to the newly stored gravity acceleration value.
5. When calibration using a calibration weight is to be
performed, go to step 3 of “10.3. Calibration using a
weight”.
To finish the setting procedure, touch the switch
or the CAL switch. The scale returns to the weighing mode.
Weighing mode
Power OFF
Gravity acceleration
correction
Calibration using
a weight
Restoring the factory
set values
or
Method 1
Touch & hold
Calibration mode
Method 2
Press
Tap
Tap
Tap

18
10.3. Calibration using a weight
Prepare a weight, preferably a weight with the same value as the weighing capacity of
the scale to be calibrated. Note that the calibration weight value can be changed.
1. Turn the power ON and warm up the scale for at least half an hour.
Change the function setting “poff” or place something on the weighing pan to
disable the auto power-off function.
2. Refer to “10.1. Calibration mode” to enter the
calibration mode. The gravity acceleration value is
displayed.
3. Tap the UNITS switch and the scale displays “Cal”.
4. Tap the TARE switch, then “Cal 0”is displayed.
Confirm that nothing is placed on the weighing pan and
wait for the STABLE indicator to turn on.
5. Tap the TARE switch. The scale calibrates the zero
point and displays the value of the calibration weight
(SPAN calibration).
The calibration weight value is equal to the weighing
capacity. (factory setting)
If SPAN calibration is not to be performed, turn the
power OFF to exit from the calibration procedure.
6. To calibrate with a weight different from the weighing
capacity, change the displayed value using the
following switches.
UNITS To shift the digit that is blinking to the
right.
SAMPLE To increase the value of the blinking
digit by one.
Using a weight with the same value as the weighing
capacity is recommended. If other weights are used,
use one with a value greater than two-thirds of the
capacity.
7. Place the calibration weight with the same value as
displayed on the weighing pan, and wait for the
STABLE indicator to turn on.
8. Tap the TARE switch. The scale calibrates SPAN and
“end” is displayed. Then, the display returns to “Cal”.
To finish the procedure, touch the switch or
press the CAL switch. The scale turns the power
OFF.
Note: If the scale will be moved to another location, set the gravity acceleration
value for the present location first and calibrate the scale using a weight.
Then, change the gravity acceleration value for the new location.
Tap
Tap
Tap
Tap
Tap
This manual suits for next models
4
Table of contents
Other A&D Scale manuals